K Number
K231785

Validate with FDA (Live)

Device Name
Perifit Pump
Manufacturer
Date Cleared
2024-01-26

(220 days)

Product Code
Regulation Number
884.5160
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Perifit Pump is a wearable electric breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

Device Description

The Perifit Pump is an electrically powered wearable single breast pump consisting of the following key components: a housing ("hub") containing pump electronics (motor, valve, PCB, rechargeable battery and associated components), and detachable milk-contacting components: a silicone flange (breast shield), a plastic (polypropylene) connector, silicone diaphragm, a silicone duckbill valve, a plastic bottle. Accessories provided with the Perifit Pump include a screw-on container lid, a USB charging cable, and bra adjustment buckle. The Perifit pump is a closed-system breast pump.

The hub user interface allows the user to power the breast pump on and off, and to operate the breast pump in two modes (stimulation, expression) and to control the vacuum settings within these modes. Each mode consists of 8 vacuum levels, capable of providing suction from 30 to 285 mmHg, with cycle speeds ranging from 68 to 120 cycles per minute in stimulation mode and 28 to 80 cycles per minute in expression mode.

The Perifit Pump is powered by an internal rechargeable lithium-ion polymer battery (3.7V 1500mAh). The breast pump cannot be operated while it is being charged. The Perifit Pump operates on embedded software, and software updates by end-users are supported. The subject device is intended for use in a home healthcare environment.

An optional smartphone app may be used with the Perifit Pump. The app may be used to control the operation of the breast pump (select mode and vacuum level, start and stop pumping), save pumping session history and display the quantity (in ml) of milk in the bottle on the user's smartphone. Use of the app is not mandatory to operate the Perifit Pump.

The Perifit Pump is a closed-system breast pump with an anti-backflow diaphragm which physically separates the air-pumping system from the milk-flow system.

AI/ML Overview

The Perifit Pump, a wearable electric breast pump, underwent various non-clinical performance tests to demonstrate its substantial equivalence to the predicate device, the Elvie Stride (K210936). The acceptance criteria for these tests were based on relevant international and national standards, as well as FDA guidance documents.

Here's a breakdown of the acceptance criteria and the study details:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test/CriteriaDevice Performance (Result)
BiocompatibilityCytotoxicity (ISO 10993-5:2009)Non-cytotoxic
Skin Sensitization (ISO 10993-10:2021)Non-sensitizing
Skin Irritation (ISO 10993-23:2021)Non-irritating
Electrical Safety & EMCElectrical Safety (AAMI ES60601-1:2005 + A1:2012 + A2:2021)Compliant with relevant standards
EMC (IEC 60601-1-2:2014 + A1:2020)Compliant with relevant standards
Home Healthcare Environment (ANSI/AAMI HA60601-1-11:2015+A1:2021)Compliant with relevant standards
Battery Safety (IEC 62133-2:2017/AMD1:2021)Compliant with relevant standards
Battery Altitude Testing (UN 38.3)Compliant with UN 38.3
SoftwareSoftware Verification and Validation (FDA guidance, IEC 62304:2006)Compliant with FDA guidance and IEC 62304:2006
Material ComplianceMilk-contacting components (21 CFR 177.1520, 21 CFR 177.2420, 21 CFR 177.2600)Compliant with material-specific requirements
Performance TestingVacuum Level and Cycle Speed (all intensity levels, both modes)Demonstrated meeting cycle speed and suction specifications
Backflow ProtectionEnsured liquid did not contact the air-pumping system
Device Use LifeMaintained vacuum level and cycle speed specifications over proposed use life
Battery Use LifeMaintained capacity over the number of charging cycles associated with device use life
Battery Charge and Discharge TimeOperating time and charge time fell within specifications
Battery Status IndicatorSatisfied functional requirements for battery status indication (implied by "testing")
Milk Level ValidationSatisfied functional requirements for milk level validation (implied by "testing")
Ingress Protection (IP22)Achieved IP22 rating
Packaging Drop and Vibration Testing (ISTA-3A)Compliant with ISTA-3A (implied by "testing")

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify the sample size for any of the non-clinical tests conducted. It generally states that "Performance testing... included" these tests, implying empirical evaluation. The data provenance is also not explicitly stated regarding whether it was retrospective or prospective, or country of origin. However, given the nature of the tests (biocompatibility, electrical safety, mechanical performance), they are typically conducted prospectively in a laboratory setting.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This information is not applicable to the described non-clinical performance and safety testing. These tests rely on established scientific and engineering standards and measurements, not expert human interpretation for ground truth.

4. Adjudication Method for the Test Set

This is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies where expert consensus is needed to establish ground truth for human interpretation of medical images or data. The described tests are based on objective measurements against predefined specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not explicitly stated or described. The document focuses on non-clinical performance and safety testing to demonstrate substantial equivalence to a predicate device, rather than comparing human reader performance with and without AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is largely not applicable in the context of the Perifit Pump as a physical medical device. While the device does use embedded software and has an optional smartphone app, the "standalone performance" typically refers to the diagnostic accuracy of AI algorithms without human intervention. The software in the Perifit Pump is for controlling pump operations, saving history, and estimating milk volume, not for diagnostic purposes. The software verification and validation would confirm its functionality and adherence to specifications, which is a form of standalone testing for the software component.

7. The type of ground truth used

The ground truth for the non-clinical tests was based on established physical, chemical, and electrical standards and specifications. For instance:

  • Biocompatibility: In vitro and in vivo testing against established criteria for cytotoxicity, sensitization, and irritation.
  • Electrical Safety & EMC: Compliance with limits and requirements defined in IEC and AAMI standards.
  • Material Compliance: Adherence to specific chemical composition and safety requirements outlined in 21 CFR.
  • Performance Testing: Measured vacuum levels, cycle speeds, and other operational parameters against predefined engineering specifications. For backflow protection, the ground truth is the absence of liquid in the air-pumping system. For use life, it's maintaining performance over a specified duration.

8. The Sample Size for the Training Set

This information is not applicable as the document describes non-clinical performance and safety testing, not the development or evaluation of a machine learning algorithm.

9. How the Ground Truth for the Training Set was Established

This information is not applicable as the document describes non-clinical performance and safety testing, not the development or evaluation of a machine learning algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 26, 2024

X6 Innovations % Lina Kontos Partner Hogan Lovells US LLP 555 13th Street NW Washington, DC 20004

Re: K231785

Trade/Device Name: Perifit Pump Regulation Number: 21 CFR$ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: December 27, 2023 Received: December 27, 2023

Dear Lina Kontos:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K231785

Device Name Perifit Pump

Indications for Use (Describe)

The Perifit Pump is a wearable electric breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY - K231785

X6 Innovation's Perifit Pump

l. Submitter:

X6 Innovations 128 rue la Boétie, 75008 Paris, France Phone: +33 6 71 26 46 45 Contact Person: Erica Perrier

Date Prepared: January 22, 2024

II. Device

Name of Device: Perifit Pump Common or Usual Name: Powered breast pump Regulation Name: Powered breast pump Regulation Number: 21 CFR 884.5160 Regulatory Class: Class II Product Code: HGX (Pump, Breast, Powered)

III. Predicate Device:

Elvie Stride (K210936), Chiaro Technology Limited, 2nd Floor 63-66 Hatton Garden, London, GB EC1M 8LE

The predicate device has not been subject to a design-related recall.

IV. Device Description

The Perifit Pump is an electrically powered wearable single breast pump consisting of the following key components: a housing ("hub") containing pump electronics (motor, valve, PCB, rechargeable battery and associated components), and detachable milk-contacting components: a silicone flange (breast shield), a plastic (polypropylene) connector, silicone diaphragm, a silicone duckbill valve, a plastic bottle. Accessories provided with the Perifit Pump include a screw-on container lid, a USB charging cable, and bra adjustment buckle. The Perifit pump is a closed-system breast pump.

The hub user interface allows the user to power the breast pump on and off, and to operate the breast pump in two modes (stimulation, expression) and to control the vacuum settings within these modes. Each mode consists of 8 vacuum levels, capable of providing suction from 30 to 285 mmHg, with cycle speeds ranging from 68 to 120 cycles per minute in stimulation mode and 28 to 80 cycles per minute in expression mode.

The Perifit Pump is powered by an internal rechargeable lithium-ion polymer battery (3.7V 1500mAh). The breast pump cannot be operated while it is being charged. The Perifit Pump

{5}------------------------------------------------

operates on embedded software, and software updates by end-users are supported. The subject device is intended for use in a home healthcare environment.

An optional smartphone app may be used with the Perifit Pump. The app may be used to control the operation of the breast pump (select mode and vacuum level, start and stop pumping), save pumping session history and display the quantity (in ml) of milk in the bottle on the user's smartphone. Use of the app is not mandatory to operate the Perifit Pump.

The Perifit Pump is a closed-system breast pump with an anti-backflow diaphragm which physically separates the air-pumping system from the milk-flow system.

V. Intended Use / Indications for Use

The Perifit Pump is a wearable electric breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

VI. Comparison of Intended Use and Technological Characteristics

The table below compares the intended use and technological characteristics of the subject and predicate device.

Perifit PumpSubject DeviceElvie Stride (K210936)Comparison
Product NamePerifit PumpElvie StrideN/A
Product CodeHGXHGXIdentical
Regulation Number21 CFR 884.516021 CFR 884.5160Identical
Regulatory ClassClass IIClass IIIdentical
Patient PopulationLactating WomenLactating WomenIdentical
Indication for UseThe Perifit Pump is a wearableelectric breast pump intendedto express milk from lactatingwomen in order to collect milkfrom their breasts. The deviceis intended for a single user.The Elvie Stride is a poweredbreast pump to be used bylactating women to expressand collect milk from theirbreasts. The Elvie Stride isintended for a single user.Identical
Single/double pumpSingle (or Double by using twounits simultaneously)Single or DoubleIdentical
Cycling controlmechanismMicrocontrollerMicrocontrollerIdentical
Backflow ProtectionYesYesIdentical
Suction ModesStimulation Mode andExpression ModeStimulation mode andExpression modeIdentical
Suction levels810Similar: the difference in thenumber of suction levels doesnot raise different questions of
Perifit PumpSubject DeviceElvie Stride (K210936)Comparison
Adjustable suctionlevelsYesYesIdenticalsafety or effectiveness as thevacuum range is comparable.
Flange Size18mm, 21 mm, 24 mm, and 27mm15, 17, and 19mm (siliconeinserts), 21, 24, 28mm (plasticflanges)Similar: the difference inflange sizes does not raisedifferent questions of safety oreffectiveness as the sizerange they cover iscomparable. The user willselect the specific size thatworks best and is comfortable.
Vacuum range-30 to -300 mmHg-55 to -300mmHg in singlepumping modeSimilar: the difference inspecification does not raisedifferent questions of safety oreffectiveness.
(-35 to -260mmHg in doublepumping mode)
Cycle speed:Stimulation68 to 120 cycles per minute31 to 83 cycles per minuteSimilar: the difference inspecification does not raisedifferent questions of safety oreffectiveness.
Cycle speed:Expression28 to 80 cycles per minute18 to 51 cycles per minuteSimilar: the difference inspecification does not raisedifferent questions of safety oreffectiveness.
ControlsOn/Off button (which alsocontrols Play/Pause pumpfunction); Increase/decreasevacuum buttons; Modeselection buttonOn/Off button;Play/pause button;Increase/decrease vacuumbuttons, Mode selection buttonIdentical
Power SupplyLi-Ion BatteryLi-ion BatteryIdentical
IndicatorsYes, LEDYes, LEDIdentical
MaterialsMilk-contacting parts are madeof silicone, polypropylene, andTritan copolyester (bottle only)Pump Outer Housing: PC/ABSplasticMilk-contacting parts are madeof silicone and polypropylenePump Outer Housing: ABSplasticSimilar: the difference inbottle/container material doesnot raise different questions ofsafety or effectiveness.
SoftwarePump operates on embeddedsoftware.Pump operates on embeddedsoftware.Identical
A smartphone app (optional)may also be used to controlthe pumpA smartphone app (optional)may also be used to controlthe pump.Similar: the smartphone appsfor both devices providesimilar functionalities and anydifferences do not raisedifferent questions of safety oreffectiveness.

Table 1: Substantial Equivalence Table

{6}------------------------------------------------

{7}------------------------------------------------

The subject device and predicate device have the same intended use. Both the subject and predicate devices work according to the same technological principle. Suction is applied, then released, to the user's nipple in order to stimulate milk expression from a lactating woman's breasts. The user can control the vacuum intensity and select between two suction modes offering faster and slower suction cycles. The milk expressed from the breasts is collected in a collection container that is self-contained within the user's bra. Both systems are closed-system breast pumps with an anti back-flow diaphraqm physically separating the air-pumping system from the milk-collecting system. The subject device (hub containing the pump electronics plus milk-contacting elements) is fully contained within the user's bra, while the predicate's milkcontacting elements are fully contained within the nursing bra and connected to the pump using external tubing. Despite the difference in configuration, both pumps operate in the same manner, using an anti-backflow diaphragm to physically separate the pumping system from the milk collection system.

Both the subject and predicate device have the option of being used in conjunction with a smartphone app. Use of this app is not mandatory for either device, and both the subject and predicate breast pumps can be operated using only the user control interface on the hub.

The Perifit Pump app also includes some additional features that are not available with the predicate (e.g., estimate of milk volume during pumping session, selecting between several suction patterns).

The technological differences between the Perifit Pump and the predicate device do not raise different questions of safety or effectiveness.

VII. Summary of Non-Clinical Performance Testing

Biocompatibility

  • Cytotoxicity (ISO 10993-5:2009) .
  • Skin Sensitization (ISO 10993-10:2021) ●
  • Skin Irritation (ISO 10993-23:2021). ●

Results of this testing support the biocompatibility of the device. The user-contacting materials were found to be non-cytotoxic, non-sensitizing, and non-irritating.

Electrical Safety and Electromagnetic Compatibility

Electrical safety and electromagnetic compatibility testing were conducted in accordance with the following standards:

  • . IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601- 1:2005/AMD2:2020 plus US National Differences: National standard AAMI ES60601-1:2005, ES60601-1:2005/AMD1 1:2012, ES60601-1:2005/AMD2:2021
  • IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020 for use in conjunction with IEC 60601-1:2005, IEC 60601- 1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 plus US NATIONAL DIFFERENCES - Differences according to US National standard ANSI/AAMI HA60601-1-11:2015+A1: 2021
  • . IEC 60601- 1-2:2014 + A1:2020, EN 60601- 1-2:2015 + A1:2021, IEC 60601-1 -11:2015 + A1:2020 Clause 12, EN 60601-1 -11:2015 + A1:2021 Clause 12
  • IEC 62133-2:2017/AMD1:2021
  • Altitude testing of the device battery per UN 38.3

{8}------------------------------------------------

Software verification and validation testing was conducted in accordance with FDA guidance ("Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", 2005 and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", 2023) and IEC 62304:2006.

All milk-contacting components were tested for compliance with material-specific requirements as outlined in 21 CFR 177.1520, 21 CFR 177.2420 and 21 CFR 177.2600.

Performance testing to substantiate the design and performance of the Perifit Pump included:

  • Vacuum level and cycle speed verification testing to demonstrate that the device met ● cycle speed and suction specifications at all intensity levels in both modes of operation.
  • Backflow protection testing to ensure that liquid was not able to come into contact with ● the air-pumping system.
  • Device use life testing to demonstrate that the device maintained its vacuum level and cycle speed specifications over its proposed use life.
  • . Battery use life testing to demonstrate that the battery maintained its capacity over the number of charging cycles associated with the device use life.
  • . Battery charge and discharge time to demonstrate that the operating time and charge time of the breast pump device fall within specifications communicated to the user.
  • . Battery status indicator testing
  • Milk level validation testing
  • Ingress protection (IP22)
  • Packaging drop and vibration testing according to ISTA-3A ●

Conclusion

The results of the performance testing described above demonstrate that the Perifit Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).