K Number
K231720
Date Cleared
2023-09-08

(87 days)

Product Code
Regulation Number
880.5570
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterile Hypodermic Needles for Single Use are intended to be used with a luer slip syringe and injection devices for general purpose fluid injection/aspiration.

Device Description

The Sterile Hypodermic Needles for Single Use is composed of a needle hub, protective cover, needle tube and jointing, connected with a luer lock or luer slip syringe and injection devices for general purpose fluid injection. The Sterile Hypodermic Needles for Single Use is for single use only. After opening the inner packaging and taking out the product, tighten the needle to the syringe and remove the protective cover, extract the drug liquid, exhaust the air and then inject. It is provided sterile. The sterilization method is EO sterilization and the sterilization assurance level is 10-6.

AI/ML Overview

This document outlines the acceptance criteria and supporting study details for a Sterile Hypodermic Needle for Single Use (K231720) according to its 510(k) summary. This is not an AI/ML-based device, so many of the requested fields (especially those related to AI algorithm performance, human reader studies, and ground truth establishment for AI training/test sets) are not applicable.

The "device" in question is a physical medical device (hypodermic needles), not a software or AI-driven system. Therefore, the acceptance criteria and study descriptions are focused on physical and biological performance, rather than algorithm performance.


Acceptance Criteria and Device Performance for Sterile Hypodermic Needles for Single Use (K231720)

The substantial equivalence determination for this device is based on its compliance with established international and national standards for medical devices, rather than a comparative effectiveness study involving human readers or AI.

1. Table of Acceptance Criteria and Reported Device Performance

CategoryAcceptance Criteria (Standard / Test)Reported Device Performance
I. Biocompatibility
CytotoxicityISO 10993-5:2009All evaluation acceptance criteria were met.
Skin SensitizationISO 10993-10:2010All evaluation acceptance criteria were met.
HemolysisISO 10993-4:2017All evaluation acceptance criteria were met.
Intracutaneous ReactivityISO 10993-10:2010All evaluation acceptance criteria were met.
Acute Systemic ToxicityISO 10993-11:2017All evaluation acceptance criteria were met.
PyrogenicityISO 10993-11:2017 (for testing); Device non-pyrogenic (for general claim)All evaluation acceptance criteria were met; Non-pyrogenic.
Particulate MatterUSP Particulate Matter in InjectionsMet USP acceptance criteria.
II. Sterilization & Shelf Life
Sterilization ValidationISO 11135:2014 (EO Sterilization)Sterilization method validated to ISO11135.
EO ResidueISO 10993-7:2008Compliant with ISO 10993-7:2008.
ECH ResidueISO 10993-7:2008Compliant with ISO 10993-7:2008.
Bacterial EndotoxinUSP42-NF37Compliant with USP42-NF37.
Shelf Life (Stability)Ageing test (method not specified but implies meeting performance over time)Five (5) years.
Sterile Barrier PackagingASTM F88/F88M-15 (Seal Strength)Compliant.
ASTM F1929-15 (Dye Penetration)Compliant.
ASTMF1140/F1140M-13 (Creep/Burst Testing)Compliant.
ASTM F2096-11 (Gross Leakage)Compliant.
III. Performance Testing
CleanlinessISO 7864:2016 Clause 4.3; ISO 9626:2016 Clause 5.3Complied (implied by "all evaluation acceptance criteria were met" for biocompatibility tests which can be affected by cleanliness, and by general statement of compliance with these standards).
Limits for acidity or alkalinityISO 7864:2016 Clause 4.4; ISO 9626:2016 Clause 5.4Complied.
Limits for extractable metalsISO 7864:2016 Clause 4.5Complied.
Tubular needle designationISO 7864:2016 Clause 4.6Complied.
Color codingISO 7864:2016 Clause 4.7; ISO 6009Complied (Color-coded per ISO 6009).
Needle hubISO 7864:2016 Clause 4.8; ISO 80369-7Complied.
Needle capISO 7864:2016 Clause 4.9Complied.
Needle tube (Tolerance on length, Freedom from defects, Lubricant)ISO 7864:2016 Clause 4.10Complied.
Needle PointISO 7864:2016 Clause 4.11Complied.
Bond between Tube and HubISO 7864:2016 Clause 4.12Complied.
Patency of LumenISO 7864:2016 Clause 4.13Complied.
Surface finish & visual appearanceISO 9626:2016 Clause 5.2Complied.
Size designationISO 9626:2016 Clause 5.5Complied.
DimensionsISO 9626:2016 Clause 5.6Complied.
StiffnessISO 9626:2016 Clause 5.8Complied.
Resistance to breakageISO 9626:2016 Clause 5.9Complied.
Resistance to corrosionISO 9626:2016 Clause 5.10Complied.
Dimensional requirements for luer connectorsISO 80369-7:2021 Clause 5Complied.
Fluid leakage (Positive pressure liquid leakage)ISO 80369-7:2021 Clause 6.1.3Complied.
Sub-atmospheric pressure air leakageISO 80369-7:2021 Clause 6.2Complied.
Stress crackingISO 80369-7:2021 Clause 6.3Complied.
Resistance to separation from axial loadISO 80369-7:2021 Clause 6.4Complied.
Resistance to separation from unscrewingISO 80369-7:2021 Clause 6.5Complied.
Resistance to overridingISO 80369-7:2021 Clause 6.6Complied.
IV. Labeling21 CFR Part 801Met the requirements of 21 CFR Part 801.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify exact sample sizes for each test. Testing of physical medical devices typically involves testing a statistically significant sample size from a production lot to demonstrate compliance with standards. The data provenance is implied to be from laboratory testing performed on the manufacturer's device, likely in China (where the manufacturer is located). The testing is prospective for the purpose of demonstrating compliance for this specific device model.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the device is a physical medical device (hypodermic needle), not an AI/ML diagnostic or image analysis system. Ground truth is established through validated laboratory test methods and established international/national standards, not through expert consensus in the medical imaging context.

4. Adjudication Method for the Test Set

This is not applicable as the ground truth is established through objective laboratory testing against defined standards (e.g., ISO, ASTM, USP), not through the adjudication of medical opinion.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML-driven diagnostic aids that assist human readers in interpreting medical images. This device is a sterile hypodermic needle, a physical disposable medical tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone performance evaluation in the context of an AI algorithm was not done. This is not an algorithmic device.

7. The Type of Ground Truth Used

The "ground truth" for this physical device is defined by the acceptance criteria specified in the referenced international standards (e.g., ISO 10993, ISO 7864, ISO 9626, ISO 80369, ASTM, USP). These standards define the physical, chemical, and biological properties a device must meet to be considered safe and effective for its intended use. Testing protocols outlined within these standards serve as the objective 'ground truth' against which the device's performance is measured. Compliance is determined by the device meeting all the specified numerical and qualitative requirements of these standards.

8. The Sample Size for the Training Set

This information is not applicable as there is no "training set" in the context of a physical medical device. The manufacturing process is controlled by Good Manufacturing Practices (GMP) and validated testing, not machine learning training.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no "training set" involved in the development or evaluation of this physical medical device.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).