K Number
K231478
Date Cleared
2023-12-15

(207 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adults via non-invasive oscillometric technique in which an inflatable CUFF (22-32cm) is wrapped around the upper arm at medical facilities or at home, it's supplied for OTC use.

Device Description

The arm-type electronic blood pressure monitor uses fuzzy logic intelligence to detect both upper and lower pressure value simultaneously. The personalized optimal inflation level determines the result of each measurement. This equipment is composed of a main body and a cuff. The main body is composed of a central processing unit, a pressure sensor, an air pump, a solenoid valve, a uniform speed vent valve, a PCB board, and an LCD liquid crystal display.

AI/ML Overview

The provided text describes the regulatory clearance of a Digital Blood Pressure Monitor (Model: YB-800) and includes information about its performance testing. I will extract the requested information based on the provided document.

1. Table of Acceptance Criteria and Reported Device Performance

The document refers to the ISO 81060-2:2018 standard for clinical validation of automated non-invasive sphygmomanometers. For this standard, the acceptance criteria for accuracy are typically:

  • Mean difference: ≤ ±5 mmHg for both systolic and diastolic blood pressure.
  • Standard deviation of differences: ≤ 8 mmHg for both systolic and diastolic blood pressure.

The document states: "All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018." This confirms the device met these criteria.

Metric (Standard: ISO 81060-2: 2018)Acceptance CriteriaReported Device Performance
Blood Pressure Accuracy
Mean Difference (Systolic)≤ ±5 mmHgMet (Not over the limits)
Standard Deviation (Systolic)≤ 8 mmHgMet (Not over the limits)
Mean Difference (Diastolic)≤ ±5 mmHgMet (Not over the limits)
Standard Deviation (Diastolic)≤ 8 mmHgMet (Not over the limits)
Pulse Rate Accuracy
Static Pressure± 3 mmHg (Bench test reference)± 3 mmHg
Pulse Rate± 5% (Bench test reference)± 5%

Note: The "Static Pressure" and "Pulse Rate" accuracy in the comparison table of the 510(k) summary (Table 1, page 5) refers to the technical specification of the device, not necessarily the clinical accuracy determined by ISO 81060-2 which focuses on differences compared to a reference measurement.

2. Sample Size for the Test Set and Data Provenance

  • Sample Size: 85 adult subjects (41 females, 44 males).
  • Data Provenance: Not explicitly stated (e.g., country of origin). The study is described as "clinical data" testing the device, implying a prospective study specifically for this clearance.

3. Number of Experts and Qualifications for Ground Truth

The document does not specify the number of experts used to establish ground truth or their qualifications. ISO 81060-2, which was used for validation, requires comparison to reference blood pressure measurements (e.g., auscultation by trained observers), but the specific details of these observers are not provided in this summary.

4. Adjudication Method for the Test Set

The document does not explicitly state the adjudication method used for the clinical validation study. For ISO 81060-2, it typically involves comparing the device's readings to those obtained by multiple trained observers (e.g., two observers) using a standardized method, with an agreed-upon difference or resolution process if discrepancies occur. However, the specific method (e.g., 2+1, 3+1) is not detailed here.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or described. This type of study typically assesses human readers' performance with and without AI assistance, which is not relevant for a standalone blood pressure monitor.

6. Standalone Performance Study

Yes, a standalone performance study was done. The "Clinical data" section explicitly states: "YB-800 was tested to ISO 81060-2: 2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type." This standard evaluates the accuracy of automated non-invasive blood pressure monitors on their own, without human intervention in the measurement process (beyond positioning the cuff and starting the device).

7. Type of Ground Truth Used

The ground truth used for the clinical validation was established by reference blood pressure measurements as mandated by the ISO 81060-2:2018 standard. This standard requires "simultaneous determination of blood pressure by a reference method (e.g. auscultation by trained observers) and by the automated measurement type under test."

8. Sample Size for the Training Set

The document does not provide information about a separate "training set" or its sample size. This type of information is typically relevant for machine learning algorithms where data is split into training, validation, and test sets. For a traditional medical device like a blood pressure monitor, the "clinical data" described generally serves as the primary validation/test set for regulatory submission.

9. How Ground Truth for the Training Set Was Established

As no training set is described, this question is not applicable. The clinical validation focused on the performance against reference measurements in a test population.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

December 15, 2023

Wenzhou Yosun Medical Technology Co.,Ltd Xiao Yuan Lan, Regulatory Affair Manager No.17 Shahong Road, Lingmen Beibaixiang Town Yueaing Wenzhou, Zhejiang 325603 China

Re: K231478

Trade/Device Name: Digital Blood Pressure Monitor (Model: YB-800) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 21, 2023 Received: November 21, 2023

Dear Xiao Yuan Lan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K231478

Device Name

Digital Blood Pressure Monitor (Model: YB-800)

Indications for Use (Describe)

The Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adults via non-invasive oscillometric technique in which an inflatable CUFF (22-32cm) is wrapped around the upper arm at medical facilities or at home, it's supplied for OTC use.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{4}------------------------------------------------

510(k) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

1. ApplicantWenzhou Yosun Medical Technology Co.,LtdNo.17, Shahong Road, Lingmen Beibaixiang Town, Yueqing Wenzhou 325603Zhejiang P.R. China086-0577-61896987
Contact PersonXiaoYuan Lan, 610826399@qq.com
Prepare date2023-04-15
2. Device nameandclassificationDevice Name: Digital Blood Pressure MonitorModels: YB-800Classification Name:21 CFR 870.1130Noninvasive Blood Pressure Measurement SystemProduct code: DXNRegulatory Class: Class II
3.PredicateDevice(s)Shenzhen Jamr Technology Co., Ltd. Upper Arm Type Blood PressureMonitor, Model F1701T, K220886
4. DeviceDescriptionThe arm-type electronic blood pressure monitor uses fuzzy logic intelligence todetect both upper and lower pressure value simultaneously. The personalizedoptimal inflation level determines the result of each measurement.This equipment is composed of a main body and a cuff. The main body iscomposed of a central processing unit, a pressure sensor, an air pump, asolenoid valve, a uniform speed vent valve, a PCB board, and an LCD liquidcrystal display.
5. Indications forUseThe Digital Blood Pressure Monitor is intended to measure the systolic anddiastolic blood pressure, as well as the pulse rate of adults via non-invasiveoscillometric technique in which an inflatable CUFF (22-32cm) is wrappedaround the upper arm at medical facilities or at home, it's supplied for OTCuse.

{5}------------------------------------------------

6. Predicate Device Comparison

Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device.

Please refer to following table to find differences between the subject device and predicate device.

Table 1 Comparison between the predicate PG-800B36 and the subject device

ITEMProposed DeviceYB-800Predicate DeviceK22086/ F1701TComparisonResult
Device NameDigital Blood Pressure MonitorUpper Arm Type Blood PressureMonitor--
AppearanceImage: Digital Blood Pressure MonitorImage: Upper Arm Type Blood Pressure Monitor--
ManufacturerWenzhou Yosun MedicalTechnology Co.,LtdShenzhen Jamr Technology Co.,Ltd--
Indications for UseThe Digital Blood Pressure Monitoris intended to measure the systolicand diastolic blood pressure, aswell as the pulse rate of adults vianon-invasive oscillometrictechnique in which an inflatableCUFF (22-32cm) is wrappedaround the upper arm at medicalfacilities or at home, it's suppliedfor OTC use.The Upper Arm Type Blood PressureMonitor is intended to measure thesystolic and diastolic blood pressureas well as the pulse rate of adult byusing the arm cuff (22-42cm), it can beused in medical facilities or at home. Itis supplied for OTC use.Same
Prescription & OTCOTCOTCOTCSame
ContraindicationNot KnownUse of this instrument on patients underdialysis therapy or on anticoagulant,antiplatelets, or steroids could causeinternal bleeding.Different¹(Note 1)
Clinical UseMedical Facilities and Home UseMedical Facilities and Home UseSame
Patient PopulationAdultsAdultsSame
Measurement siteUpper armUpper armSame
MeasurementPrincipleOscillometricOscillometricSame
ComponentsLCD / Key / Cuff / MCU / Pump / BatteriesLCD / Key / Cuff / MCU / Pump / BatteriesSame
Memory120 memories for 2 users (SYS, DIA, Pulse)Automatically stores the last 120 measurements for measureSame
Mode of operationContinuous operationContinuous operationSame
Power Source4x1.5V AA Alkaline Batteries3x1.5V AA Alkaline BatteriesDifferent1(Note 1)
Mode of operationContinuous operationContinuous operationSame
Safety classificationsType BF applied partType BF applied partSame
Measurement RangeBlood Pressure0 ~ 300 mmHgBlood Pressure 0 ~ 295 mmHgDifferent2(Note 2)
Pulse Rate40 -199bpmPulse Rate 40-180 bpm
AccuracyStatic Pressure$\pm$ 3 mmHgStatic Pressure $\pm$ 3 mmHgSame
Pulse Rate$\pm$ 5%Pulse Rate $\pm$ 5%
Arm Circumference22 cm~32cm22 cm~42 cmDifferent2(Note 2)
Patient Contact MaterialCuff -Flannelette/Sticky fabricCuff -NylonDifferent3(Note 3)
Physical Dimensions (LengthWidthHeight)Approx:131mm111mm79mm109*121.5 * 52.5mmDifferent4(Note 4)
WeightAbout 265g (including battery and cuff)290g (batteries and AC adapter is not included)
Safety and essential performanceIEC 60601-1IEC 60601-1-2IEC 60601-1-11ISO 80601-2-30ISO 81060-2IEC 60601-1IEC 60601-1-2IEC 60601-1-11ISO 80601-2-30ISO 81060-2Same
BiocompatibilityISO 10993-1;ISO 10993-5;ISO 10993-23;ISO 10993-1;ISO 10993-5;ISO 10993-10Same
Operation Environments+ 5℃~ + 40℃,15%RH80%RH,80 kPa105 kPa+ 5℃~ + 40℃, 15%RH93%RH,70 kPa106 kPaDifferent5(Note 5)
Storage Environments- 20℃~ + 55℃,10%RH90%RH,80 kPa106 kPa- 25℃~ + 70℃, ≤93%RH, 70 kPa~106 kPa

{6}------------------------------------------------

{7}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]

The proposed device and the predicate device have similar technological characteristics. Both devices are battery-operated arm blood pressure monitors intended for adult use and employ the oscillometric method for measuring blood pressure and pulse. The devices share the same accuracy range of pressure reading of ±3 mmHg and the same accuracy range for pulse rate reading of ± 5%.

There are minor differences in technical specifications and features between the YB-800 proposed device and predicate device would be discussed below:

SUBSTANTIAL EQUIVALENCE

The contraindication of the proposed device is different form the predicate device. And they both work with alkaline batteries though the voltage of proposed device is bigger than the predicate. The device is verified through IEC 60601-1, IEC 60601-1-2 and IEC 80601-2-30 and no additional safety and effectiveness issues was introduced. (Note 1)

The range of pressure and the pulse rate is wider than the predicate device while the arm circumference is narrower than predicate device. They have the same accuracy. The differences are very slightly and they both contain the blood pressure and pulse range of most people, and the measurement range of proposed device is fully verified according to IEC 80601-2-30. Therefore, the differences do not bring additional risks. (Note 2)

Though the cuff material is different from the predicate, the cuff of proposed device was biocompatible according to ISO 10993 tests. No new safety and effectiveness issues raised. (Note 3)

Minor differences in the dimensions and weight do not impact the safety or performance of blood pressure or pulse rate measurements (Note 4).

Although the operating humidity and atmospheric pressure of the proposed device is slightly narrower than the predicate device, and the storage temperature, humidity and atmospheric pressure of the proposed device are also little narrower, they all comply with IEC 60601-1, IEC 60601-1-11, and IEC 80601-2-30, so this difference will not raise any safety or effectiveness issue. (Note 5)

Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the proposed device is substantially equivalent to the predicate device.

{8}------------------------------------------------

Performance Testing:

Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.

Non-Clinical Data:

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the Digital Blood Pressure Monitor and the NIBP CUFF were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests:

  • 그 Cytotoxicity
  • Skin Sensitization
  • Skin Irritation

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Arm-tvpe Electronic Blood Pressure Monitor, consisting of all the modules and accessories in the system complies with the IEC 60601-1: 2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2014 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances – Requirements and tests standard for EMC.

Bench Testing

Bench testing was conducted on the Arm-type Electronic Blood Pressure Monitor, consisting of all the accessories in the system. The system complies with the IEC 60601-1-11: 2010 Medical ELECTRICAL EQUIPMENT –Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, ISO 80601-2-30: 2009 Medical electrical equipment -- Part 2-30: Requirements for basic safety and essential performance of automated non-invasive sphygmomanometers.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

Clinical data

YB-800 was tested to ISO 81060-2: 2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. The study included 85 adult subjects (41 females, 44 males) with an age range of 16 to 70 years. All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018. No adverse effect and/or complication is found in this study.

Summary

Based on the non-clinical and clinical performance as documented in the device development, the subject devices were found to have a safety and effectiveness profile that is same to the predicate device.

Conclusion

Verification and validation testing was conducted on the subject device and all testing passed pre-specified criteria. This premarket notification submission demonstrates that the Digital Blood Pressure Monitor (YB-800) is substantially equivalent to the predicate devices(K220

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).