K Number
K231478
Device Name
Digital Blood Pressure Monitor (Model: YB-800)
Date Cleared
2023-12-15

(207 days)

Product Code
Regulation Number
870.1130
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adults via non-invasive oscillometric technique in which an inflatable CUFF (22-32cm) is wrapped around the upper arm at medical facilities or at home, it's supplied for OTC use.
Device Description
The arm-type electronic blood pressure monitor uses fuzzy logic intelligence to detect both upper and lower pressure value simultaneously. The personalized optimal inflation level determines the result of each measurement. This equipment is composed of a main body and a cuff. The main body is composed of a central processing unit, a pressure sensor, an air pump, a solenoid valve, a uniform speed vent valve, a PCB board, and an LCD liquid crystal display.
More Information

Not Found

Yes
The device description explicitly mentions "fuzzy logic intelligence," which is a form of artificial intelligence used for decision-making based on imprecise or uncertain data.

No
The device is intended to measure blood pressure and pulse rate, which are diagnostic/monitoring functions, not therapeutic.

Yes
The device is intended to measure systolic and diastolic blood pressure, which provides information about a patient's health status, aligning with the definition of a diagnostic device. While it doesn't diagnose a disease itself, it provides data essential for diagnosis.

No

The device description explicitly lists hardware components such as a main body, cuff, central processing unit, pressure sensor, air pump, solenoid valve, uniform speed vent valve, PCB board, and LCD liquid crystal display.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: The description clearly states that this device measures blood pressure and pulse rate non-invasively using an inflatable cuff wrapped around the upper arm. This is a physical measurement taken directly from the body, not a test performed on a sample outside the body.
  • Intended Use: The intended use is to measure blood pressure and pulse rate, which are physiological parameters, not to diagnose a condition based on analyzing a biological sample.

Therefore, this Digital Blood Pressure Monitor falls under the category of a non-invasive medical device, not an in vitro diagnostic device.

No
The input explicitly states "Control Plan Authorized (PCCP) and relevant text: Not Found," indicating no mention of PCCP authorization.

Intended Use / Indications for Use

The Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adults via non-invasive oscillometric technique in which an inflatable CUFF (22-32cm) is wrapped around the upper arm at medical facilities or at home, it's supplied for OTC use.

Product codes

DXN

Device Description

The arm-type electronic blood pressure monitor uses fuzzy logic intelligence to detect both upper and lower pressure value simultaneously. The personalized optimal inflation level determines the result of each measurement. This equipment is composed of a main body and a cuff. The main body is composed of a central processing unit, a pressure sensor, an air pump, a solenoid valve, a uniform speed vent valve, a PCB board, and an LCD liquid crystal display.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

The arm-type electronic blood pressure monitor uses fuzzy logic intelligence to detect both upper and lower pressure value simultaneously.

Input Imaging Modality

Not Found

Anatomical Site

Upper arm

Indicated Patient Age Range

Adults. The study included 85 adult subjects (41 females, 44 males) with an age range of 16 to 70 years.

Intended User / Care Setting

medical facilities or at home, it's supplied for OTC use.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-Clinical Data:

  • Biocompatibility testing: The biocompatibility evaluation for the Digital Blood Pressure Monitor and the NIBP CUFF were conducted in accordance with the International Standard ISO 10993-1. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. Tests included: Cytotoxicity, Skin Sensitization, Skin Irritation.
  • Electrical safety and electromagnetic compatibility (EMC): Testing conducted on the Arm-type Electronic Blood Pressure Monitor, complying with IEC 60601-1: 2012 and IEC 60601-1-2: 2014.
  • Bench Testing: Testing conducted on the Arm-type Electronic Blood Pressure Monitor, complying with IEC 60601-1-11: 2010 and ISO 80601-2-30: 2009.
  • Software Verification and Validation Testing: Conducted in accordance with FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Software considered a "major" level of concern.

Clinical data:

  • Study Type: Clinical validation of automated measurement type.
  • Sample Size: 85 adult subjects (41 females, 44 males).
  • Key Results: All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018. No adverse effect and/or complication is found in this study.

Summary:
Based on the non-clinical and clinical performance, the subject devices were found to have a safety and effectiveness profile that is same to the predicate device.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K220886

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

December 15, 2023

Wenzhou Yosun Medical Technology Co.,Ltd Xiao Yuan Lan, Regulatory Affair Manager No.17 Shahong Road, Lingmen Beibaixiang Town Yueaing Wenzhou, Zhejiang 325603 China

Re: K231478

Trade/Device Name: Digital Blood Pressure Monitor (Model: YB-800) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 21, 2023 Received: November 21, 2023

Dear Xiao Yuan Lan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

1

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

2

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

3

Indications for Use

510(k) Number (if known) K231478

Device Name

Digital Blood Pressure Monitor (Model: YB-800)

Indications for Use (Describe)

The Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adults via non-invasive oscillometric technique in which an inflatable CUFF (22-32cm) is wrapped around the upper arm at medical facilities or at home, it's supplied for OTC use.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

4

510(k) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

| 1. Applicant | Wenzhou Yosun Medical Technology Co.,Ltd
No.17, Shahong Road, Lingmen Beibaixiang Town, Yueqing Wenzhou 325603
Zhejiang P.R. China
086-0577-61896987 |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | XiaoYuan Lan, 610826399@qq.com |
| Prepare date | 2023-04-15 |
| 2. Device name
and
classification | Device Name: Digital Blood Pressure Monitor
Models: YB-800
Classification Name:
21 CFR 870.1130
Noninvasive Blood Pressure Measurement System
Product code: DXN
Regulatory Class: Class II |
| 3.Predicate
Device(s) | Shenzhen Jamr Technology Co., Ltd. Upper Arm Type Blood Pressure
Monitor, Model F1701T, K220886 |
| 4. Device
Description | The arm-type electronic blood pressure monitor uses fuzzy logic intelligence to
detect both upper and lower pressure value simultaneously. The personalized
optimal inflation level determines the result of each measurement.
This equipment is composed of a main body and a cuff. The main body is
composed of a central processing unit, a pressure sensor, an air pump, a
solenoid valve, a uniform speed vent valve, a PCB board, and an LCD liquid
crystal display. |
| 5. Indications for
Use | The Digital Blood Pressure Monitor is intended to measure the systolic and
diastolic blood pressure, as well as the pulse rate of adults via non-invasive
oscillometric technique in which an inflatable CUFF (22-32cm) is wrapped
around the upper arm at medical facilities or at home, it's supplied for OTC
use. |

5

6. Predicate Device Comparison

Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device.

Please refer to following table to find differences between the subject device and predicate device.

Table 1 Comparison between the predicate PG-800B36 and the subject device

| ITEM | Proposed Device
YB-800 | Predicate Device
K22086/ F1701T | Comparison
Result | |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|------------------------|
| Device Name | Digital Blood Pressure Monitor | Upper Arm Type Blood Pressure
Monitor | -- | |
| Appearance | Image: Digital Blood Pressure Monitor | Image: Upper Arm Type Blood Pressure Monitor | -- | |
| Manufacturer | Wenzhou Yosun Medical
Technology Co.,Ltd | Shenzhen Jamr Technology Co.,Ltd | -- | |
| Indications for Use | The Digital Blood Pressure Monitor
is intended to measure the systolic
and diastolic blood pressure, as
well as the pulse rate of adults via
non-invasive oscillometric
technique in which an inflatable
CUFF (22-32cm) is wrapped
around the upper arm at medical
facilities or at home, it's supplied
for OTC use.

The Upper Arm Type Blood Pressure
Monitor is intended to measure the
systolic and diastolic blood pressure
as well as the pulse rate of adult by
using the arm cuff (22-42cm), it can be
used in medical facilities or at home. It
is supplied for OTC use. | | Same | |
| Prescription & OTCOTC | OTC | OTC | Same | |
| Contraindication | Not Known | Use of this instrument on patients under
dialysis therapy or on anticoagulant,
antiplatelets, or steroids could cause
internal bleeding. | Different¹
(Note 1) | |
| Clinical Use | Medical Facilities and Home Use | Medical Facilities and Home Use | Same | |
| Patient Population | Adults | Adults | Same | |
| Measurement site | Upper arm | Upper arm | Same | |
| Measurement
Principle | Oscillometric | Oscillometric | Same | |
| Components | LCD / Key / Cuff / MCU / Pump / Batteries | LCD / Key / Cuff / MCU / Pump / Batteries | Same | |
| Memory | 120 memories for 2 users (SYS, DIA, Pulse) | Automatically stores the last 120 measurements for measure | Same | |
| Mode of operation | Continuous operation | Continuous operation | Same | |
| Power Source | 4x1.5V AA Alkaline Batteries | 3x1.5V AA Alkaline Batteries | Different1
(Note 1) | |
| Mode of operation | Continuous operation | Continuous operation | Same | |
| Safety classifications | Type BF applied part | Type BF applied part | Same | |
| Measurement Range | Blood Pressure | 0 ~ 300 mmHg | Blood Pressure 0 ~ 295 mmHg | Different2
(Note 2) |
| | Pulse Rate | 40 -199bpm | Pulse Rate 40-180 bpm | |
| Accuracy | Static Pressure | $\pm$ 3 mmHg | Static Pressure $\pm$ 3 mmHg | Same |
| | Pulse Rate | $\pm$ 5% | Pulse Rate $\pm$ 5% | |
| Arm Circumference | 22 cm32cm | 22 cm42 cm | Different2
(Note 2) | |
| Patient Contact Material | Cuff -Flannelette/Sticky fabric | Cuff -Nylon | Different3
(Note 3) | |
| Physical Dimensions (LengthWidthHeight) | Approx:
131mm111mm79mm | 109*121.5 * 52.5mm | Different4
(Note 4) | |
| Weight | About 265g (including battery and cuff) | 290g (batteries and AC adapter is not included) | | |
| Safety and essential performance | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
ISO 80601-2-30
ISO 81060-2 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
ISO 80601-2-30
ISO 81060-2 | Same | |
| Biocompatibility | ISO 10993-1;
ISO 10993-5;
ISO 10993-23; | ISO 10993-1;
ISO 10993-5;
ISO 10993-10 | Same | |
| Operation Environments | + 5℃~ + 40℃,
15%RH80%RH,80 kPa105 kPa | + 5℃~ + 40℃, 15%RH93%RH,70 kPa106 kPa | Different5
(Note 5) | |
| Storage Environments | - 20℃~ + 55℃,
10%RH90%RH,80 kPa106 kPa | - 25℃~ + 70℃, ≤93%RH, 70 kPa~106 kPa | | |

6

7

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]

The proposed device and the predicate device have similar technological characteristics. Both devices are battery-operated arm blood pressure monitors intended for adult use and employ the oscillometric method for measuring blood pressure and pulse. The devices share the same accuracy range of pressure reading of ±3 mmHg and the same accuracy range for pulse rate reading of ± 5%.

There are minor differences in technical specifications and features between the YB-800 proposed device and predicate device would be discussed below:

SUBSTANTIAL EQUIVALENCE

The contraindication of the proposed device is different form the predicate device. And they both work with alkaline batteries though the voltage of proposed device is bigger than the predicate. The device is verified through IEC 60601-1, IEC 60601-1-2 and IEC 80601-2-30 and no additional safety and effectiveness issues was introduced. (Note 1)

The range of pressure and the pulse rate is wider than the predicate device while the arm circumference is narrower than predicate device. They have the same accuracy. The differences are very slightly and they both contain the blood pressure and pulse range of most people, and the measurement range of proposed device is fully verified according to IEC 80601-2-30. Therefore, the differences do not bring additional risks. (Note 2)

Though the cuff material is different from the predicate, the cuff of proposed device was biocompatible according to ISO 10993 tests. No new safety and effectiveness issues raised. (Note 3)

Minor differences in the dimensions and weight do not impact the safety or performance of blood pressure or pulse rate measurements (Note 4).

Although the operating humidity and atmospheric pressure of the proposed device is slightly narrower than the predicate device, and the storage temperature, humidity and atmospheric pressure of the proposed device are also little narrower, they all comply with IEC 60601-1, IEC 60601-1-11, and IEC 80601-2-30, so this difference will not raise any safety or effectiveness issue. (Note 5)

Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the proposed device is substantially equivalent to the predicate device.

8

Performance Testing:

Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.

Non-Clinical Data:

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the Digital Blood Pressure Monitor and the NIBP CUFF were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests:

  • 그 Cytotoxicity
  • Skin Sensitization
  • Skin Irritation

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Arm-tvpe Electronic Blood Pressure Monitor, consisting of all the modules and accessories in the system complies with the IEC 60601-1: 2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2014 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances – Requirements and tests standard for EMC.

Bench Testing

Bench testing was conducted on the Arm-type Electronic Blood Pressure Monitor, consisting of all the accessories in the system. The system complies with the IEC 60601-1-11: 2010 Medical ELECTRICAL EQUIPMENT –Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, ISO 80601-2-30: 2009 Medical electrical equipment -- Part 2-30: Requirements for basic safety and essential performance of automated non-invasive sphygmomanometers.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

Clinical data

YB-800 was tested to ISO 81060-2: 2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. The study included 85 adult subjects (41 females, 44 males) with an age range of 16 to 70 years. All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018. No adverse effect and/or complication is found in this study.

Summary

Based on the non-clinical and clinical performance as documented in the device development, the subject devices were found to have a safety and effectiveness profile that is same to the predicate device.

Conclusion

Verification and validation testing was conducted on the subject device and all testing passed pre-specified criteria. This premarket notification submission demonstrates that the Digital Blood Pressure Monitor (YB-800) is substantially equivalent to the predicate devices(K220