K Number
K231140
Manufacturer
Date Cleared
2023-05-19

(28 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The aprevo® transforaminal lumbar interbody fusion device is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® transforaminal lumbar interbody fusion device is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The device is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches.

The aprevo® transforaminal lumbar interbody fusion device is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® transforaminal lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implants may be implanted via a variety of open or minimally invasive approaches.

Device Description

The aprevo® transforaminal lumbar interbody fusion devices are designed to stabilize the lumbar spinal column and facilitate fusion. The personalized aprevo® devices incorporate patient specific features to allow the surgeon to tailor the deformity correction to the individual needs of the patient and include an aperture intended for the packing of bone graft. The individualized surgical correction plan and device configurations are developed using patient CT scans. The aprevo® devices are manufactured from titanium alloy (Ti-6A1-4V) per ASTM F3001 and are provided sterile.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (aprevo® transforaminal lumbar interbody fusion device). This type of document focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving the device meets specific acceptance criteria through a clinical or performance study with detailed metrics typically seen in AI/ML medical devices.

Therefore, the information required to answer your questions regarding acceptance criteria, performance studies, sample sizes, expert involvement, and ground truth establishment (common for AI/ML device submissions) is not present in this 510(k) summary.

Here's why and what can be inferred:

  • Acceptance Criteria & Device Performance: The document does not list specific numerical acceptance criteria for a device's "performance" in the way an AI/ML device would (e.g., sensitivity, specificity, AUC). Instead, it states that "Engineering analyses and confirmatory dynamic compression-shear testing per ASTM F2077 were conducted to support substantial equivalence to the predicate devices." This implies mechanical and material performance criteria, but the specific values or a direct comparison table are not provided.
  • Sample Size (Test Set/Training Set) & Data Provenance: This 510(k) pertains to a physical implantable device. There are no "test sets" or "training sets" in the context of data for an AI/ML algorithm. The "testing" mentioned refers to biomechanical testing of the device itself (e.g., strength, durability), not the analysis of medical images or patient data by an algorithm to generate a diagnosis or prediction.
  • Number of Experts, Ground Truth, Adjudication: These concepts are relevant for studies determining the accuracy of an AI/ML algorithm against human interpretation or confirmed diagnoses. They are not applicable to the clearance of a mechanical intervertebral fusion device. The "ground truth" for a device like this would be its physical properties meeting design specifications and regulatory standards.
  • MRMC Study / Standalone Performance: These are specific to AI/ML device clinical evaluations. This document describes a physical medical device, not an AI algorithm.
  • Ground Truth Type: Not applicable in the AI/ML sense. The "ground truth" for this device is its mechanical properties and biocompatibility.

In summary, based on the provided text, I cannot complete a table or answer the questions as they relate to the performance study of an AI/ML driven medical device.

The document details the device description, its indications for use, and states that "Engineering analyses and confirmatory dynamic compression-shear testing per ASTM F2077 were conducted to support substantial equivalence to the predicate devices." This means the device's mechanical properties were evaluated against an existing standard and shown to be comparable to another legally marketed device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 19, 2023

Carlsmed, Inc. Karen Liu VP Quality and Regulatory 1800 Aston Ave., Ste. 100 Carlsbad, California 92008

Re: K231140

Trade/Device Name: aprevo® transforaminal lumbar interbody fusion device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: April 21, 2023 Received: April 21, 2023

Dear Karen Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Katherine D. Kavlock -S

for

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K231140

Device Name

aprevo® transforaminal lumbar interbody fusion device

Indications for Use (Describe)

The aprevo® transforaminal lumbar interbody fusion device is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® transforaminal lumbar interbody fusion device is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The device is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches.

The aprevo® transforaminal lumbar interbody fusion device is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® transforaminal lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implants may be implanted via a variety of open or minimally invasive approaches.

Type of Use (Select one or both, as applicable)

( Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

4 510(k) Summary

Submitter:Carlsmed, Inc.1800 Aston Ave., Ste. 100Carlsbad, CA 92008
Official Correspondent:Karen LiuVP Quality and RegulatoryCarlsmed, Inc.1800 Aston Ave., Ste. 100Carlsbad, CA 92008regulatory@carlsmed.com
Date Prepared:April 20, 2023
Trade Name:aprevo® transforaminal lumbar interbody fusion device
Common Name:Intervertebral fusion device with bone graft, lumbar
Classification:Class II21 CFR §888.3080
Product Code:MAX
Primary Predicate Device:aprevo® transforaminal lumbar interbody fusion device(K222082)
Additional Predicate:aprevo® Anterior Lumbar Interbody Fusion Device withInterfixation (K222009)

Device Description:

The aprevo® transforaminal lumbar interbody fusion devices are designed to stabilize the lumbar spinal column and facilitate fusion. The personalized aprevo® devices incorporate patient specific features to allow the surgeon to tailor the deformity correction to the individual needs of the patient and include an aperture intended for the packing of bone graft. The individualized surgical correction plan and device configurations are developed using patient CT scans. The aprevo® devices are manufactured from titanium alloy (Ti-6A1-4V) per ASTM F3001 and are provided sterile.

Indications for Use:

aprevo® transforaminal lumbar interbody fusion devices (TLIF):

The aprevo® transforaminal lumbar interbody fusion device is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® transforaminal lumbar interbody fusion device is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The device

{4}------------------------------------------------

is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches.

The aprevo® transforaminal lumbar interbody fusion device is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® transforaminal lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implants may be implanted via a variety of open or minimally invasive approaches.

Performance Testing:

Engineering analyses and confirmatory dynamic compression-shear testing per ASTM F2077 were conducted to support substantial equivalence to the predicate devices.

Substantial Equivalence:

The subject device was demonstrated to be substantially equivalent to the predicate cited above with respect to indications, design, function, and performance.

Conclusion:

The aprevo® transforaminal lumbar interbody fusion device is substantially equivalent to the previously cleared devices with respect to its indications for use, design, function, and performance.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.