K Number
K230954

Validate with FDA (Live)

Date Cleared
2023-11-17

(227 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Samfilcon B Contact Lens: Bausch + Lomb Samfilcon B Contact Lens for daily wear is indicated for the correction of refractive ametropia (myopia or hyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity.

Samfilcon B Contact Lens for Astigmatism: Bausch + Lomb Samfilcon B Contact Lens for Astigmatism for daily wear is indicated for the correction of refractive ametropia (myopia, hyperopia and/or astigmatism) in aphakic persons with non-diseased eyes with refractive astigmatism not exceeding 10.00 diopters.

Samfilcon B Contact Lens for Presbyopia: Bausch + Lomb Samfilcon B Contact Lens for Presbyopia for daily wear is indicated for the correction of refractive ametropia (myopia or hyperopia) and presbyopia in aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity.

Samfilcon B Contact Lens Multifocal for Astigmatism: Bausch + Lomb Samfilcon B Contact Lens Multifocal for Astigmatism for daily wear is indicated for the correction of refractive ametropia (myopia, and/or astigmatism) and presbyopia in aphakic and not aphakic persons with non-diseased eyes with refractive astigmatism not exceeding 10.00 diopters.

When prescribed for frequent planned replacement wear, the contact lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lens may be disinfected using a chemical disinfection system.

The Samfilcon B Contact Lens is available for a frequent/planned replacement modality.

Device Description

The Bausch + Lomb Samfilcon B Contact Lens is made from a silicone hydrogel material and is approximately 41% water by weight when immersed in a sterile phosphate buffered saline solution. This lens is tinted blue with Reactive Blue Dye 246. The color additive conforms with 21 CFR Part 73.3106.

The Bausch + Lomb Samfilcon B Contact Lens is to be prescribed for daily wear.

The custom-made lenses will be manufactured in spherical, toric, multifocal and toric multifocal designs with the following parameters:
Diameter: 10.0 mm to 16.0 mm
Base Curve: 6.5 mm to 9.7 mm
Power Range: +20.00D to -20.00D
Cylinder Power (Toric): 0D to -10.00D
Cylinder Axis: 1° to 180°
Add Power (Multifocal): +0.50D up to +4.00D

The lenses are packaged in borosilicate USP Type 1 glass vials containing phosphate buffered saline solution. A silicone vial stopper and aluminum crimp cap secure the lenses in the vial. Vials are labeled with lot number, expiration date and applicable lens parameters. Expiration dating is supported by product stability, package integrity, and validation of the sterilization process.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the Samfilcon B Contact Lens. It includes information about the device's technical specifications, comparison to predicate devices, and summaries of non-clinical and clinical testing. However, it does not explicitly detail a "study that proves the device meets the acceptance criteria" in the format of specific acceptance criteria.

Instead, the document states that "The primary safety endpoint was statistical non-inferiority with respect to the proportion of eyes with slit lamp findings Grade >2 at any follow-up visit between the test and control lenses. And the primary effectiveness endpoint was statistical non-inferiority with respect to mean distance high-contrast logMAR lens visual acuity for each eye between the test and control lenses." These are the endpoints used to demonstrate substantial equivalence, rather than a direct set of acceptance criteria with reported performance against them in a table format.

Therefore, I cannot generate the requested table of acceptance criteria and reported device performance from the provided text because these specific criteria and their direct outcomes are not explicitly listed in that manner.

Below is the information that can be extracted from the provided text regarding the clinical study:

1. Table of Acceptance Criteria and Reported Device Performance:

As noted above, a direct table of specific acceptance criteria and detailed reported performance is not explicitly provided in the text. However, the primary endpoints of the clinical study, which serve as the basis for proving effectiveness and safety, are stated.

Criteria Type (Primary Endpoint)Acceptance Goal (Implicit)Reported Device Performance (Summary)
Primary Safety Endpoint: Proportion of eyes with slit lamp findings Grade >2 at any follow-up visitStatistical non-inferiority between test (Samfilcon B) and control lenses.Non-inferior to the control lens with respect to the incidence of moderate/severe slit lamp findings.
Primary Effectiveness Endpoint: Mean distance high-contrast logMAR lens visual acuityStatistical non-inferiority for each eye between test (Samfilcon B) and control lenses.Non-inferior to the control lens with respect to distance lens visual acuity.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set (Clinical Study):
    • Enrollment: 56 for Investigational Device (Samfilcon B Contact Lens), 28 for Control Device, Total = 84 subjects.
    • Primary Safety Endpoint Analysis: 56 for Investigational Device, 28 for Control Device, Total = 84 subjects.
    • Primary Effectiveness Endpoint Analysis: 46 for Investigational Device, 21 for Control Device, Total = 67 subjects.
  • Data Provenance: The clinical study was a "multicenter, clinical study in the US." It was active-controlled, randomized, and double-masked.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the text.

4. Adjudication method for the test set:

This information is not provided in the text. The study was double-masked, which suggests a controlled environment, but a specific adjudication method beyond that is not mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This was not an MRMC study, and it does not involve AI assistance. The study compared the Samfilcon B contact lens to a control lens.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This was not a standalone (algorithm only) study, as it involved human subjects wearing contact lenses.

7. The type of ground truth used:

For the clinical study, the "ground truth" was established through clinical observations and measurements by eye care professionals, specifically regarding:

  • Slit lamp findings (for safety)
  • Distance high-contrast logMAR lens visual acuity (for effectiveness)

8. The sample size for the training set:

There is no mention of a "training set" in the context of this contact lens clinical study. This term is typically used in machine learning or AI development, which is not applicable here. The study involved a test set of human subjects.

9. How the ground truth for the training set was established:

As there was no "training set" in the machine learning sense, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 17, 2023

Bausch + Lomb, Incorporated Barbara Klube-Falso Director, Regulatory Affairs 1400 North Goodman Street Rochester, NY 14609

Re: K230954

Trade/Device Name: Samfilcon B Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: October 12, 2023 Received: October 13, 2023

Dear Barbara Klube-Falso:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketingsafety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

J Angelo Green -S

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230954

Device Name samfilcon B

Indications for Use (Describe)

Samfilcon B Contact Lens: Bausch + Lomb Samfilcon B Contact Lens for daily wear is indicated for the correction of refractive ametropia (myopia or hyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity.

Samfilcon B Contact Lens for Astigmatism:

Bausch + Lomb Samfilcon B Contact Lens for Astigmatism for daily wear is indicated for the correction of refractive ametropia (myopia, hyperopia and/or astigmatism) in aphakic persons with non-diseased eyes with refractive astigmatism not exceeding 10.00 diopters.

Samfilcon B Contact Lens for Presbyopia:

Bausch + Lomb Samfilcon B Contact Lens for Presbyopia for daily wear is indicated for the correction of refractive ametropia (myopia or hyperopia) and presbyopia in aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity.

Samfilcon B Contact Lens Multifocal for Astigmatism:

Bausch + Lomb Samfilcon B Contact Lens Multifocal for Astigmatism for daily wear is indicated for the correction of refractive ametropia (myopia, and/or astigmatism) and presbyopia in aphakic and not aphakic persons with non-diseased eyes with refractive astigmatism not exceeding 10.00 diopters.

When prescribed for frequent planned replacement wear, the contact lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lens may be disinfected using a chemical disinfection system.

The Samfilcon B Contact Lens is available for a frequent/planned replacement modality.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY K230954

Submitter Information:

Date Prepared:November 17, 2023
Name:Bausch & Lomb Incorporated
Address:1400 North Goodman StreetRochester, NY 14609
Contact Person:Barbara Klube-FalsoDirector, Regulatory Affairs
Phone Number:(585) 338-8503
Email:Barbara.Klube-Falso@bausch.com

Device Information:

Common Name:Soft (hydrophilic) Contact Lens
-----------------------------------------------
  • To Be Determined Trade Name:
    Classification: Ophthalmic Medical Device

Device Classification: Class II (21 CFR 886.5925)

  • LPL Product Code:

Predicate Device:

Bausch + Lomb Ultra (samfilcon A) Contact Lens, K131208 for the material, and

Alden HP 54 Sphere, Toric, Multifocal and Toric Multifocal Soft Contact Lenses, K091327 for the Indications for Use.

Device Description:

The Bausch + Lomb Samfilcon B Contact Lens is made from a silicone hydrogel material and is approximately 41% water by weight when immersed in a sterile phosphate buffered saline solution. This lens is tinted blue with Reactive Blue Dye 246. The color additive conforms with 21 CFR Part 73.3106.

The Bausch + Lomb Samfilcon B Contact Lens is to be prescribed for daily wear.

The physical properties of the lens are:

Refractive index1.4191
Light transmission97%
Water Content41%
Specific Gravity1.052
Oxygen Permeability126 x 10-11[cm³O2(STP) x cm]/(sec x cm² x mmHg)@35°C(polarographic method)

{5}------------------------------------------------

The custom-made lenses will be manufactured in spherical, toric, multifocal and toric multifocal designs with the following parameters:

Diameter10.0 mm to 16.0 mm
Base Curve6.5 mm to 9.7 mm
Power Range+20.00D to -20.00D
Cylinder Power (Toric)0D to -10.00D
Cylinder Axis1° to 180°
Add Power (Multifocal)+0.50D up to +4.00D

The lenses are packaged in borosilicate USP Type 1 glass vials containing phosphate buffered saline solution. A silicone vial stopper and aluminum crimp cap secure the lenses in the vial. Vials are labeled with lot number, expiration date and applicable lens parameters. Expiration dating is supported by product stability, package integrity, and validation of the sterilization process.

Indications for Use:

Samfilcon B Contact Lens

Bausch + Lomb Samfilcon B Contact Lens for daily wear is indicated for the correction of refractive ametropia (myopia or hyperopia) in aphakic persons with nondiseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity.

Samfilcon B Contact Lens for Presbyopia

Bausch + Lomb Samfilcon B Contact Lens for Presbyopia for daily wear is indicated for the correction of refractive ametropia (myopia or hyperopia) and presbyopia in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 1,50 diopters or less where the astigmatism does not interfere with visual acuity.

Samfilcon B Contact Lens for Astigmatism

Bausch + Lomb Samfilcon B Contact Lens for Astigmatism for daily wear is indicated for the correction of refractive ametropia (myopia, hyperopia, and/or astigmatism) in aphakic and not aphakic persons with non-diseased eves with refractive astigmatism not exceeding 10.00 diopters.

Samfilcon B Contact Lens Multifocal for Astigmatism

Bausch + Lomb Samfilcon B Contact Lens Multifocal for Astigmatism for daily wear is indicated for the correction of refractive ametropia (myopia, hyperopia, and/or astigmatism) and presbyopia in aphakic and not aphakic persons with non-diseased eyes with refractive astigmatism not exceeding 10.00 diopters.

When prescribed for frequent planned replacement wear, the contact lens is to be cleaned. rinsed, and disinfected each time it is removed from the patient's eve and discarded after the recommended wearing period prescribed by the eye care practitioner. The lens may be disinfected using a chemical disinfection system.

Samfilcon B Contact Lenses for daily wear are available for a frequent/planned replacement modality.

{6}------------------------------------------------

Technological comparison to predicate devices

IndicationStatementsPredicate DeviceAlden HP Contact Lenses, K091327Subject DeviceBausch + LombSamfilcon BContact Lens
SphericalIndicated for the correction of refractive ametropia(myopia or hyperopia) in aphakic and not aphakicpersons with non-diseased eyes. The lens may be wornby persons who exhibit refractive astigmatism of 1.50diopters or less where the astigmatism does notinterfere with visual acuity.Same as predicate
MultifocalIndicated for the correction of refractive ametropia(myopia or hyperopia) and presbyopia in aphakic andnot aphakic persons with non-diseased eyes. The lensmay be worn by persons who exhibit refractiveastigmatism of 1.50 diopters or less where theastigmatism does not interfere with visual acuity.Same as predicate
ToricIndicated for the correction of refractive ametropia(myopia, hyperopia, and/or astigmatism) in aphakic andnot aphakic persons with non-diseased eyes withrefractive astigmatism not exceeding 10.00 diopters.Same as predicate
Multifocal ToricIndicated for the correction of refractive ametropia(myopia, hyperopia, and/or astigmatism) and presbyopiain aphakic and not aphakic persons with non-diseasedeyes.Same as predicate
PropertyPredicate DeviceSubject Device
FunctionalityBausch + LombUltra (samfilcon A) Contact Lens, K131208Bausch + LombSamfilcon BContact Lens
FunctionalityThe contact lenses act as a refractive medium that focuslight rays from near and distant objects on the retina.Same as predicate
ModalityDaily wear contact lensSame as predicate
ManufacturingMethodCast MoldedCustom made,Lathed
Material GroupSilicone hydrogelSame as predicate
USAN Namesamfilcon Asamfilcon B
Water Content46%41%
OxygenPermeability1(edge corrected)114126
Specific Gravity1.0481.052

1 Oxygen Permeability shown was determined by the polarographic method: x 10-11[cm³O₂(STP) x cm]/(sec x cm² x mmHg)@35°C

Summary of Non-Clinical Testing:

The testing performed on the Bausch + Lomb samfilcon B Contact Lens demonstrated that the device functions in a safe and effective manner. Performance testing included conformance to predetermined specifications, functional test results verify that the device performs as expected and is equivalent to the predicate without creating additional risk to the user.

{7}------------------------------------------------

In addition. Bausch + Lomb followed the Premarket Notification (510(k)) Guidance Document for Daily Wear Contact Lenses, May 1994, and ISO 10993-1 and completed the following tests of the Samfilcon B Contact Lens:

  • . In-Vitro Cytotoxicity
    • Direct Contact per ISO 10993-5 and ISO 18189 о
    • Agarose Overlay per ISO 10993-5 O
    • Elution Method per ISO 10993-5 O
  • Guinea pig maximization skin sensitization per ISO 10993-10 .
  • . Ocular Irritation Study
    • Acute Ocular Irritation per ISO 10993-10 o
    • 5-Day study in rabbits per ISO 9394 (modified) O
    • 22-Day study in rabbits per ISO 9394 o
  • . Systemic Toxicity
    • o Acute systemic toxicity per ISO 10993-11

The following tests were performed according to GLP on the silicone vial stopper:

  • . Cytotoxicity
    • o Elution Method per ISO 10993-5
  • . Ocular Irritation Study per ISO 10993-10
  • . Systemic Toxicity
    • o Acute systemic toxicity per ISO 10993-11

Performance Testing was conducted on the Bausch + Lomb Samfilcon B Contact Lens which included:

  • . Physical, Chemical and Spectral Properties
  • . Leachable Monomer and Additives

Summary of Clinical Performance Data

Bausch + Lomb conducted an active-controlled, randomized, double-masked, multicenter, clinical study in the US, evaluating the safety and effectiveness of the Bausch + Lomb Samfilcon B Contact Lens. Subjects were adapted soft contact lens wearers between the ages of 18 and 40 years old.

Both the primary safety and primary effectiveness endpoints of the study were met. The primary safety endpoint was statistical non-inferiority with respect to the proportion of eyes with slit lamp findings Grade >2 at any follow-up visit between the test and control lenses. And the primary effectiveness endpoint was statistical non-inferiority with respect to mean distance highcontrast logMAR lens visual acuity for each eye between the test and control lenses. These

{8}------------------------------------------------

results demonstrate that the Samfilcon B Contact Lens is noninferior to the control lens with respect to the incidence of moderate/severe slit lamp findings and distance lens visual acuity.

There were no reports of serious adverse events related to the test or control lens.

StageInvestigational Device TotalControl Device TotalTotal
Enrolment562884
Treatment462167
Primary SafetyEndpoint Analysis562884
PrimaryEffectivenessEndpoint Analysis462167

Patient Accountability (number of patients reported at each stage)

Substantial Equivalence Conclusion:

The cumulative results of laboratory, in vivo testing as well as the clinical study sponsored by Bausch + Lomb demonstrate that the safety, effectiveness, and performance of the Bausch + Lomb Samfilcon B Contact Lens are substantially equivalent to the Bausch + Lomb (Samfilcon A) Contact Lens.

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.