(180 days)
AVVIGO™ Guidance System II K212490
Not Found
No
The document describes standard image processing and data acquisition techniques for IVUS and FFR/DFR. There is no mention of AI, ML, or related concepts like deep learning, neural networks, or automated interpretation beyond basic border tracing functionality. The focus is on hardware and software updates for existing modalities.
No
The text indicates that the device is used for diagnosis and obtaining anatomical information, but not for direct treatment or therapy.
Yes
Explanation: The "Intended Use / Indications for Use" section explicitly states that "FFR and DFR are indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters." This directly indicates a diagnostic purpose for the device's functions. Additionally, the IVUS modality is for "ultrasound examinations of intravascular pathology," which is also a diagnostic activity.
No
The device description explicitly mentions hardware components such as a V4 Acquisition PC, mobile pole, and Table Side Controller user interface, in addition to software.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to examine specimens taken from the human body (like blood, urine, or tissue) to provide information for diagnosis, monitoring, or screening.
- Device Function: The AVVIGO™+ Multi-Modality Guidance System uses intravascular ultrasound (IVUS) and pressure measurements (FFR/DFR) within the body to visualize and assess vascular and cardiac pathology and hemodynamics. It does not analyze specimens taken from the body.
- Intended Use: The intended use clearly describes procedures performed in vivo (within a living organism), such as angioplasty and atherectomy, and obtaining physiological parameters directly from vessels.
Therefore, the device's function and intended use fall outside the scope of In Vitro Diagnostics.
N/A
Intended Use / Indications for Use
The IVUS modality of the System is intended for ultrasound examinations of intravascular pathology. Intravascular ultrasound is indicated in patients who are candidates for transluminal procedures such as angioplasty and atherectomy.
FFR and DFR™ are intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. FFR and DFR are indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters.
Refer to the Catheter Instructions for Use provided with all Boston Scientific Ultrasound Imaging Catheters to determine compatibility with the System. All Ultrasound Imaging Catheters will be referred to as Imaging Catheters throughout the remainder of this User Guide.
The Imaging Catheters generate ultrasound images and are intended for ultrasound examination of vascular and cardiac pathology. Boston Scientific manufactures a wide variety of catheters for different applications. The recommended use of each of these catheters may vary depending on the size and type of the catheter. Please refer to the Imaging Catheter Instructions for Use, packaged with each catheter.
Indications for Auto Pullback Use (IVUS Only) Automatic Pullback is indicated when the following occurs:
- The physician/operator wants to standardize the method in which intravascular ultrasound images are obtained and documented: procedure-to-procedure, operator-to-operator.
- · The physician/operator wants to make linear determinations post-procedurally, which requires the imaging core of a catheter to be pulled back at a known uniform speed.
- Two-dimensional, longitudinal reconstruction of the anatomy is desired.
Product codes
DSK, IYO, DSK, ITX
Device Description
The AVVIGO™+ Multi-Modality Guidance System is used by physicians to obtain physiologic, intravascular, and/or intracardiac anatomical information. The system utilizes an interactive display that displays information and allows the physician to:
- Perform intravascular physiology assessment by making measurements with a pressure ● wire that invasively obtains pressure readings from the vessels, and an FFR Link that retrieves the information from the pressure wire and transfers the information wirelessly to the interactive display, and/or
- Perform intravascular or intracardiac ultrasound imaging using a specially designed . catheter with a miniaturized ultrasound probe attached to the distal end and a Motor Drive. A PC is used for data acquisition and processing to visualize ultrasound images of selected anatomical structures.
The AVVIGO™+ Multi-Modality Guidance System hardware has been modified to include a V4 Acquisition PC for IVUS image processing, updates to the mobile pole, and a Table Side Controller user interface for user control access near the patient bedside.
The AVVIGO™+ Multi-Modality Guidance System software includes updated vessel and lumen border trace functionality to improve ease of use. Additionally, the AVVIGO+ Multi-Modality Guidance System GUI includes visual display of frame markers as well as graphical display of DFR with pullback support.
The AVVIGO™+ Multi-Modality Guidance System underlying fundamental principles and technology supporting IVUS and Physiology modalities are unmodified from that of the predicate. IVUS and FFR/DFR remain identical to that of the predicate device, AVVIGO Guidance System II.
AVVIGO™ + Multi-Modality Guidance System modifications do not raise any new issues of safety and effectiveness.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Ultrasound
Anatomical Site
Intravascular, cardiac
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Physicians in catheterization and related cardiovascular specialty laboratories
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Substantial equivalence of the AVVIGO™ + Multi-Modality Guidance System is determined through non-clinical performance evaluation including hardware, software, electrical safety, packaqing verification and validation activities.
Non-clinical performance verification was performed on the complete AVVIGO™+ Multi-Modality Guidance System, including where applicable compatible equipment, according to recognized performance and safety standards.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
AVVIGO™ Guidance System II K212490
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1425 Programmable diagnostic computer.
(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
September 26, 2023
Boston Scientific Corporation Kevin Catalano Principal Regulatory Affairs Specialists Two Scimed Place Maple Grove, Minnesota 55311
Re: K230884
Trade/Device Name: AVVIGO+ Multi-Modality Guidance System Regulation Number: 21 CFR 870.1110 Regulation Name: Blood Pressure Computer Regulatory Class: Class II Product Code: DSK, IYO, DSK, ITX Dated: August 23, 2023 Received: August 24, 2023
Dear Kevin Catalano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
1
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K230884
Device Name AVVIGO+ Multi-Modality Guidance System
Indications for Use (Describe)
The IVUS modality of the System is intended for ultrasound examinations of intravascular pathology. Intravascular ultrasound is indicated in patients who are candidates for transluminal procedures such as angioplasty and atherectomy.
FFR and DFR™ are intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. FFR and DFR are indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters.
Refer to the Catheter Instructions for Use provided with all Boston Scientific Ultrasound Imaging Catheters to determine compatibility with the System. All Ultrasound Imaging Catheters will be referred to as Imaging Catheters throughout the remainder of this User Guide.
The Imaging Catheters generate ultrasound images and are intended for ultrasound examination of vascular and cardiac pathology. Boston Scientific manufactures a wide variety of catheters for different applications. The recommended use of each of these catheters may vary depending on the size and type of the catheter. Please refer to the Imaging Catheter Instructions for Use, packaged with each catheter.
Indications for Auto Pullback Use (IVUS Only) Automatic Pullback is indicated when the following occurs:
- The physician/operator wants to standardize the method in which intravascular ultrasound images are obtained and documented: procedure-to-procedure, operator-to-operator.
- · The physician/operator wants to make linear determinations post-procedurally, which requires the imaging core of a catheter to be pulled back at a known uniform speed.
· Two-dimensional, longitudinal reconstruction of the anatomy is desired.
Type of Use (Select one or both, as applicable)
X | Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
3
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
Image /page/4/Picture/1 description: The image shows the logo for Boston Scientific. The text is in a dark blue color. The word "Boston" is on the top line, and the word "Scientific" is on the bottom line.
510(k) Summary
per 21 CFR 807.92
| Sponsor | Boston Scientific Corporation
300 Boston Scientific Way
Marlborough, Massachusetts 01752
USA |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name and
Information | Kevin Catalano
Prin. Regulatory Affairs Specialist
Phone: 612.817.2987
E-mail: Kevin.Catalano@bsci.com |
| Date Prepared | March 31, 2023 |
| Proprietary Name | AVVIGO™+ Multi-Modality Guidance System |
| Common Name | Computer Diagnostic Programmable
Blood Pressure Computer
Ultrasonic Pulsed Echo Imaging System
Diagnostic Ultrasonic Transducer |
| Classification Name | Primary:
21 CFR 870.1425 (Programmable Diagnostic Computer)
Subsequent:
21 CFR 870.1110 (Blood Pressure Computer)
21 CFR 892.1560 (Ultrasonic Pulsed Echo Imaging System)
21 CFR 892.1570 (Diagnostic Ultrasonic Transducer) |
| Product Code | Primary:
DQK (Computer, Diagnostic, Programmable)
Subsequent:
DSK (Computer, Blood-Pressure)
IYO (System, Imaging, Pulsed Echo, Ultrasonic)
ITX (Transducer, Ultrasonic, Diagnostic) |
| Classification | Class II, 21 CFR 870.1425 |
| Predicate Device | AVVIGO™ Guidance System II
K212490 cleared 03 OCT 2021 |
Device Description
The AVVIGO™+ Multi-Modality Guidance System is used by physicians to obtain physiologic, intravascular, and/or intracardiac anatomical information. The system utilizes an interactive display that displays information and allows the physician to:
- Perform intravascular physiology assessment by making measurements with a pressure ● wire that invasively obtains pressure readings from the vessels, and an FFR Link that retrieves the information from the pressure wire and transfers the information wirelessly to the interactive display, and/or
- Perform intravascular or intracardiac ultrasound imaging using a specially designed . catheter with a miniaturized ultrasound probe attached to the distal end and a Motor Drive. A PC is used for data acquisition and processing to visualize ultrasound images of selected anatomical structures.
5
Image /page/5/Picture/1 description: The image shows the logo for Boston Scientific. The text is in a blue serif font. The word "Boston" is on the top line, and the word "Scientific" is on the bottom line.
The AVVIGO™+ Multi-Modality Guidance System hardware has been modified to include a V4 Acquisition PC for IVUS image processing, updates to the mobile pole, and a Table Side Controller user interface for user control access near the patient bedside.
The AVVIGO™+ Multi-Modality Guidance System software includes updated vessel and lumen border trace functionality to improve ease of use. Additionally, the AVVIGO+ Multi-Modality Guidance System GUI includes visual display of frame markers as well as graphical display of DFR with pullback support.
The AVVIGO™+ Multi-Modality Guidance System underlying fundamental principles and technology supporting IVUS and Physiology modalities are unmodified from that of the predicate. IVUS and FFR/DFR remain identical to that of the predicate device, AVVIGO Guidance System II.
AVVIGO™ + Multi-Modality Guidance System modifications do not raise any new issues of safety and effectiveness.
Intended Use/indications for Use
The IVUS modality of the System is intended for ultrasound examinations of intravascular pathology. Intravascular ultrasound is indicated in patients who are candidates for transluminal interventional procedures such as angioplasty and atherectomy.
FFR and DFR™ are intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices.
FFR and DFR are indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters.
Refer to the Catheter Instructions for Use provided with all Boston Scientific Ultrasound Imaging Catheters to determine compatibility with the System. All Ultrasound Imaging Catheters will be referred to as Imaging Catheters throughout the remainder of this User Guide.
The Imaging Catheters generate ultrasound images and are intended for ultrasound examination of vascular and cardiac pathology. Boston Scientific manufactures a wide variety of catheters for different applications. The recommended use of each of these catheters may vary depending on the size and type of the catheter. Please refer to the Imaging Catheter Instructions for Use, packaged with each catheter.
Indications for Auto Pullback Use (IVUS Only)
Automatic Pullback is indicated when the following occurs:
- The physician/operator wants to standardize the method in which intravascular . ultrasound images are obtained and documented: procedure-to-procedure, operatorto-operator.
- . The physician/operator wants to make linear distance determinations postprocedurally, which requires the imaging core of a catheter to be pulled back at a known uniform speed.
6
Two-dimensional, longitudinal reconstruction of the anatomy is desired.
Comparison of Technological Characteristics
The AVVIGO™ + Multi-Modality Guidance System has the same intended use, operational principles and technological characteristics supporting its intended use as the predicate device, AVVIGO™ Guidance System II. The AVVIGO™+ Multi-Modality Guidance System is substantially equivalent to the predicate device in intended use, IVUS and Physiology modalities, fundamental scientific technology and performance features supporting IVUS and Physiology modalities. Modifications within the AVVIGO™+ Multi-Modality Guidance System do not raise any new issues of safety and effectiveness as demonstrated via non-clinical performance data.
Non-Clinical Performance Data
Substantial equivalence of the AVVIGO™ + Multi-Modality Guidance System is determined through non-clinical performance evaluation including hardware, software, electrical safety, packaqing verification and validation activities.
Non-clinical performance verification was performed on the complete AVVIGO™+ Multi-Modality Guidance System, including where applicable compatible equipment, according to recognized performance and safety standards including:
- ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and ● A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) FDA Recognition Number 19-4
- ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] Medical electrical equipment -. Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)] FDA Recognition Number 19-36
- IEC 60601-2-34 Edition 3.0 2011-05 Medical electrical equipment Part 2-34: Particular . requirements for the basic safety. including essential performance. of invasive blood pressure monitoring equipment FDA Recognition Number 3-115
- IEC 60601-2-37 Edition 2.1 2015 Medical electrical equipment Part 2-37: Particular ● requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment FDA Recognition Number 12-293
- IEC 62304 Medical Device Software Software Lifecycle Processes, (edition 1.1 2015-. 06) FDA recognition: 13-79
- . AAMI TIR69:2017/(R2020) Technical Information Report Risk management of radiofrequency wireless coexistence for medical devices and systems. FDA Recognition Number 19-22
To support a determination of substantial equivalence, non-clinical performance verification and validation was conducted in accordance with the following FDA guidance documents:
- Radio Frequency Wireless Technology in Medical Devices – Guidance for Industry and FDA Staff published August 14, 2013
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff published October 2, 2014
- Electromagnetic Compatibility (EMC) of Medical Devices Guidance for Industry and Food . and Drug Administration Staff June 6, 2022
7
Clinical Performance Data
Not applicable; determination of substantial equivalence is not based on clinical performance data. Substantial equivalence is based on an assessment of non-clinical performance data.
Conclusion
Based on a comparison of intended use, operating principles, and underlying fundamental design technology, including non-clinical performance verification and validation, the AVVIGO™+ Multi-Modality Guidance System is substantially equivalent to the predicate device, AVVIGO™ Guidance System II Multi-Modality Guidance System. A comparison of the modified device and predicate, including non-clinical performance verification tests performed, support a determination of substantial equivalence and raise no new issues of safety and effectiveness.