K Number
K230884

Validate with FDA (Live)

Date Cleared
2023-09-26

(180 days)

Product Code
Regulation Number
870.1425
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IVUS modality of the System is intended for ultrasound examinations of intravascular pathology. Intravascular ultrasound is indicated in patients who are candidates for transluminal procedures such as angioplasty and atherectomy.

FFR and DFR™ are intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. FFR and DFR are indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters.

Refer to the Catheter Instructions for Use provided with all Boston Scientific Ultrasound Imaging Catheters to determine compatibility with the System. All Ultrasound Imaging Catheters will be referred to as Imaging Catheters throughout the remainder of this User Guide.

The Imaging Catheters generate ultrasound images and are intended for ultrasound examination of vascular and cardiac pathology. Boston Scientific manufactures a wide variety of catheters for different applications. The recommended use of each of these catheters may vary depending on the size and type of the catheter. Please refer to the Imaging Catheter Instructions for Use, packaged with each catheter.

Indications for Auto Pullback Use (IVUS Only) Automatic Pullback is indicated when the following occurs:

  • The physician/operator wants to standardize the method in which intravascular ultrasound images are obtained and documented: procedure-to-procedure, operator-to-operator.
  • · The physician/operator wants to make linear determinations post-procedurally, which requires the imaging core of a catheter to be pulled back at a known uniform speed.

· Two-dimensional, longitudinal reconstruction of the anatomy is desired.

Device Description

The AVVIGO™+ Multi-Modality Guidance System is used by physicians to obtain physiologic, intravascular, and/or intracardiac anatomical information. The system utilizes an interactive display that displays information and allows the physician to:

  • Perform intravascular physiology assessment by making measurements with a pressure ● wire that invasively obtains pressure readings from the vessels, and an FFR Link that retrieves the information from the pressure wire and transfers the information wirelessly to the interactive display, and/or
  • Perform intravascular or intracardiac ultrasound imaging using a specially designed . catheter with a miniaturized ultrasound probe attached to the distal end and a Motor Drive. A PC is used for data acquisition and processing to visualize ultrasound images of selected anatomical structures.

The AVVIGO™+ Multi-Modality Guidance System hardware has been modified to include a V4 Acquisition PC for IVUS image processing, updates to the mobile pole, and a Table Side Controller user interface for user control access near the patient bedside.

The AVVIGO™+ Multi-Modality Guidance System software includes updated vessel and lumen border trace functionality to improve ease of use. Additionally, the AVVIGO+ Multi-Modality Guidance System GUI includes visual display of frame markers as well as graphical display of DFR with pullback support.

The AVVIGO™+ Multi-Modality Guidance System underlying fundamental principles and technology supporting IVUS and Physiology modalities are unmodified from that of the predicate. IVUS and FFR/DFR remain identical to that of the predicate device, AVVIGO Guidance System II.

AVVIGO™ + Multi-Modality Guidance System modifications do not raise any new issues of safety and effectiveness.

AI/ML Overview

The user is asking for specific details about the acceptance criteria and the study proving the device meets these criteria. However, the provided document (FDA 510(k) clearance letter and summary for the AVVIGO+ Multi-Modality Guidance System) explicitly states:

"Clinical Performance Data: Not applicable; determination of substantial equivalence is not based on clinical performance data. Substantial equivalence is based on an assessment of non-clinical performance data."

This means that a clinical study with human patients, including elements like a test set, expert readers, ground truth establishment, MRMC studies, or standalone performance, was not performed or at least not required for the 510(k) clearance of this device. The modifications to the AVVIGO+ system were deemed substantially equivalent to its predicate (AVVIGO Guidance System II) based on non-clinical performance data (hardware, software, electrical safety, packaging verification and validation against recognized standards).

Therefore, I cannot provide the requested information for acceptance criteria related to a clinical study because the document states no clinical performance data was used for this clearance.

I can, however, extract information about the non-clinical performance data and the type of evaluation performed.

Here's a summary based on the provided document, addressing what information is available and indicating what is not:


Acceptance Criteria and Device Performance Study (as per provided document)

The provided documentation states that no clinical performance data was used for the determination of substantial equivalence for the AVVIGO+ Multi-Modality Guidance System. Therefore, the device performance was assessed through non-clinical performance data, focusing on compliance with recognized safety and performance standards.

1. Table of Acceptance Criteria and Reported Device Performance

As no clinical study was performed for this clearance, there isn't a table of clinical performance metrics like sensitivity, specificity, or accuracy derived from a test set. Instead, the acceptance criteria are based on compliance with various engineering, safety, and quality standards for medical devices.

Acceptance Criteria CategorySpecific Standard/Requirement MetReported Device Performance/Verification
Electrical SafetyANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)Non-clinical performance verification was performed on the complete system, demonstrating compliance.
Electromagnetic Compatibility (EMC)ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)]Non-clinical performance verification was performed, demonstrating compliance.
Blood Pressure MonitoringIEC 60601-2-34 Edition 3.0 2011-05 Medical electrical equipment Part 2-34: Particular requirements for the basic safety, including essential performance, of invasive blood pressure monitoring equipmentNon-clinical performance verification was performed, demonstrating compliance.
Ultrasonic Medical EquipmentIEC 60601-2-37 Edition 2.1 2015 Medical electrical equipment Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipmentNon-clinical performance verification was performed, demonstrating compliance.
Software Lifecycle ProcessesIEC 62304 Medical Device Software Software Lifecycle Processes, (edition 1.1 2015-06)Non-clinical performance verification was performed, demonstrating compliance.
Wireless CoexistenceAAMI TIR69:2017/(R2020) Technical Information Report Risk management of radiofrequency wireless coexistence for medical devices and systems.Non-clinical performance verification was performed, demonstrating compliance.
General ComplianceFDA Guidance: Radio Frequency Wireless Technology in Medical Devices – Guidance for Industry and FDA Staff published August 14, 2013Verified during non-clinical performance evaluation.
CybersecurityFDA Guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff published October 2, 2014Verified during non-clinical performance evaluation.
EMC (Guidance)FDA Guidance: Electromagnetic Compatibility (EMC) of Medical Devices Guidance for Industry and Food and Drug Administration Staff June 6, 2022Verified during non-clinical performance evaluation.
Hardware, Software, PackagingNot specifically enumerated standards, but stated as part of evaluation."hardware, software, electrical safety, packaging verification and validation activities" were performed.

2. Sample size used for the test set and the data provenance:

  • Not Applicable. No clinical test set data was used for this 510(k) clearance determination. The evaluation was based on non-clinical performance testing and compliance with recognized standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. No clinical test set and thus no ground truth established by experts was used for this 510(k) clearance.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. No clinical test set was used for this 510(k) clearance.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. An MRMC study was not stated as being performed for this 510(k) clearance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable / No. The document focuses on the system's compliance with safety and performance standards for its intended functions (IVUS imaging and FFR/DFR physiological parameters), not on an "algorithm only" performance study in a clinical context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not Applicable. No clinical ground truth was used for this 510(k) clearance. The "ground truth" for the non-clinical evaluation was adherence to the specifications defined by recognized medical device standards and the device's own design documentation.

8. The sample size for the training set:

  • Not Applicable. This 510(k) clearance is for a medical device that includes imaging and physiological measurement capabilities, but the document does not describe it as an AI/ML device requiring a distinct "training set" in the context of learned models. The software updates mentioned are for "vessel and lumen border trace functionality" and GUI improvements, but no details of an AI/ML model's training process are provided or implied to be part of the substantial equivalence determination.

9. How the ground truth for the training set was established:

  • Not Applicable. (See point 8).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

September 26, 2023

Boston Scientific Corporation Kevin Catalano Principal Regulatory Affairs Specialists Two Scimed Place Maple Grove, Minnesota 55311

Re: K230884

Trade/Device Name: AVVIGO+ Multi-Modality Guidance System Regulation Number: 21 CFR 870.1110 Regulation Name: Blood Pressure Computer Regulatory Class: Class II Product Code: DSK, IYO, DSK, ITX Dated: August 23, 2023 Received: August 24, 2023

Dear Kevin Catalano:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230884

Device Name AVVIGO+ Multi-Modality Guidance System

Indications for Use (Describe)

The IVUS modality of the System is intended for ultrasound examinations of intravascular pathology. Intravascular ultrasound is indicated in patients who are candidates for transluminal procedures such as angioplasty and atherectomy.

FFR and DFR™ are intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. FFR and DFR are indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters.

Refer to the Catheter Instructions for Use provided with all Boston Scientific Ultrasound Imaging Catheters to determine compatibility with the System. All Ultrasound Imaging Catheters will be referred to as Imaging Catheters throughout the remainder of this User Guide.

The Imaging Catheters generate ultrasound images and are intended for ultrasound examination of vascular and cardiac pathology. Boston Scientific manufactures a wide variety of catheters for different applications. The recommended use of each of these catheters may vary depending on the size and type of the catheter. Please refer to the Imaging Catheter Instructions for Use, packaged with each catheter.

Indications for Auto Pullback Use (IVUS Only) Automatic Pullback is indicated when the following occurs:

  • The physician/operator wants to standardize the method in which intravascular ultrasound images are obtained and documented: procedure-to-procedure, operator-to-operator.
  • · The physician/operator wants to make linear determinations post-procedurally, which requires the imaging core of a catheter to be pulled back at a known uniform speed.

· Two-dimensional, longitudinal reconstruction of the anatomy is desired.

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Boston Scientific. The text is in a dark blue color. The word "Boston" is on the top line, and the word "Scientific" is on the bottom line.

510(k) Summary

per 21 CFR 807.92

SponsorBoston Scientific Corporation300 Boston Scientific WayMarlborough, Massachusetts 01752USA
Contact Name andInformationKevin CatalanoPrin. Regulatory Affairs SpecialistPhone: 612.817.2987E-mail: Kevin.Catalano@bsci.com
Date PreparedMarch 31, 2023
Proprietary NameAVVIGO™+ Multi-Modality Guidance System
Common NameComputer Diagnostic ProgrammableBlood Pressure ComputerUltrasonic Pulsed Echo Imaging SystemDiagnostic Ultrasonic Transducer
Classification NamePrimary:21 CFR 870.1425 (Programmable Diagnostic Computer)Subsequent:21 CFR 870.1110 (Blood Pressure Computer)21 CFR 892.1560 (Ultrasonic Pulsed Echo Imaging System)21 CFR 892.1570 (Diagnostic Ultrasonic Transducer)
Product CodePrimary:DQK (Computer, Diagnostic, Programmable)Subsequent:DSK (Computer, Blood-Pressure)IYO (System, Imaging, Pulsed Echo, Ultrasonic)ITX (Transducer, Ultrasonic, Diagnostic)
ClassificationClass II, 21 CFR 870.1425
Predicate DeviceAVVIGO™ Guidance System IIK212490 cleared 03 OCT 2021

Device Description

The AVVIGO™+ Multi-Modality Guidance System is used by physicians to obtain physiologic, intravascular, and/or intracardiac anatomical information. The system utilizes an interactive display that displays information and allows the physician to:

  • Perform intravascular physiology assessment by making measurements with a pressure ● wire that invasively obtains pressure readings from the vessels, and an FFR Link that retrieves the information from the pressure wire and transfers the information wirelessly to the interactive display, and/or
  • Perform intravascular or intracardiac ultrasound imaging using a specially designed . catheter with a miniaturized ultrasound probe attached to the distal end and a Motor Drive. A PC is used for data acquisition and processing to visualize ultrasound images of selected anatomical structures.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Boston Scientific. The text is in a blue serif font. The word "Boston" is on the top line, and the word "Scientific" is on the bottom line.

The AVVIGO™+ Multi-Modality Guidance System hardware has been modified to include a V4 Acquisition PC for IVUS image processing, updates to the mobile pole, and a Table Side Controller user interface for user control access near the patient bedside.

The AVVIGO™+ Multi-Modality Guidance System software includes updated vessel and lumen border trace functionality to improve ease of use. Additionally, the AVVIGO+ Multi-Modality Guidance System GUI includes visual display of frame markers as well as graphical display of DFR with pullback support.

The AVVIGO™+ Multi-Modality Guidance System underlying fundamental principles and technology supporting IVUS and Physiology modalities are unmodified from that of the predicate. IVUS and FFR/DFR remain identical to that of the predicate device, AVVIGO Guidance System II.

AVVIGO™ + Multi-Modality Guidance System modifications do not raise any new issues of safety and effectiveness.

Intended Use/indications for Use

The IVUS modality of the System is intended for ultrasound examinations of intravascular pathology. Intravascular ultrasound is indicated in patients who are candidates for transluminal interventional procedures such as angioplasty and atherectomy.

FFR and DFR™ are intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices.

FFR and DFR are indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters.

Refer to the Catheter Instructions for Use provided with all Boston Scientific Ultrasound Imaging Catheters to determine compatibility with the System. All Ultrasound Imaging Catheters will be referred to as Imaging Catheters throughout the remainder of this User Guide.

The Imaging Catheters generate ultrasound images and are intended for ultrasound examination of vascular and cardiac pathology. Boston Scientific manufactures a wide variety of catheters for different applications. The recommended use of each of these catheters may vary depending on the size and type of the catheter. Please refer to the Imaging Catheter Instructions for Use, packaged with each catheter.

Indications for Auto Pullback Use (IVUS Only)

Automatic Pullback is indicated when the following occurs:

  • The physician/operator wants to standardize the method in which intravascular . ultrasound images are obtained and documented: procedure-to-procedure, operatorto-operator.
  • . The physician/operator wants to make linear distance determinations postprocedurally, which requires the imaging core of a catheter to be pulled back at a known uniform speed.

{6}------------------------------------------------

Two-dimensional, longitudinal reconstruction of the anatomy is desired.

Comparison of Technological Characteristics

The AVVIGO™ + Multi-Modality Guidance System has the same intended use, operational principles and technological characteristics supporting its intended use as the predicate device, AVVIGO™ Guidance System II. The AVVIGO™+ Multi-Modality Guidance System is substantially equivalent to the predicate device in intended use, IVUS and Physiology modalities, fundamental scientific technology and performance features supporting IVUS and Physiology modalities. Modifications within the AVVIGO™+ Multi-Modality Guidance System do not raise any new issues of safety and effectiveness as demonstrated via non-clinical performance data.

Non-Clinical Performance Data

Substantial equivalence of the AVVIGO™ + Multi-Modality Guidance System is determined through non-clinical performance evaluation including hardware, software, electrical safety, packaqing verification and validation activities.

Non-clinical performance verification was performed on the complete AVVIGO™+ Multi-Modality Guidance System, including where applicable compatible equipment, according to recognized performance and safety standards including:

  • ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and ● A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) FDA Recognition Number 19-4
  • ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] Medical electrical equipment -. Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)] FDA Recognition Number 19-36
  • IEC 60601-2-34 Edition 3.0 2011-05 Medical electrical equipment Part 2-34: Particular . requirements for the basic safety. including essential performance. of invasive blood pressure monitoring equipment FDA Recognition Number 3-115
  • IEC 60601-2-37 Edition 2.1 2015 Medical electrical equipment Part 2-37: Particular ● requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment FDA Recognition Number 12-293
  • IEC 62304 Medical Device Software Software Lifecycle Processes, (edition 1.1 2015-. 06) FDA recognition: 13-79
  • . AAMI TIR69:2017/(R2020) Technical Information Report Risk management of radiofrequency wireless coexistence for medical devices and systems. FDA Recognition Number 19-22

To support a determination of substantial equivalence, non-clinical performance verification and validation was conducted in accordance with the following FDA guidance documents:

  • Radio Frequency Wireless Technology in Medical Devices – Guidance for Industry and FDA Staff published August 14, 2013
  • Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff published October 2, 2014
  • Electromagnetic Compatibility (EMC) of Medical Devices Guidance for Industry and Food . and Drug Administration Staff June 6, 2022

{7}------------------------------------------------

Clinical Performance Data

Not applicable; determination of substantial equivalence is not based on clinical performance data. Substantial equivalence is based on an assessment of non-clinical performance data.

Conclusion

Based on a comparison of intended use, operating principles, and underlying fundamental design technology, including non-clinical performance verification and validation, the AVVIGO™+ Multi-Modality Guidance System is substantially equivalent to the predicate device, AVVIGO™ Guidance System II Multi-Modality Guidance System. A comparison of the modified device and predicate, including non-clinical performance verification tests performed, support a determination of substantial equivalence and raise no new issues of safety and effectiveness.

§ 870.1425 Programmable diagnostic computer.

(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).