K Number
K230627
Manufacturer
Date Cleared
2023-11-20

(259 days)

Product Code
Regulation Number
886.1120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Zilia Ocular is a non-contact imaging device for capturing, displaying and storing images of the retina under mydriatic conditions.

Zilia Ocular is intended for use as an aid to clinicians in the evaluation and documentation of ocular health.

Device Description

The Zilia Ocular Fundus Camera (Zilia Ocular FC) is a non-contact, stand-alone device designed to image the eye fundus.

The device includes a base, a head support, an onboard computer with a 15.6'' touch-screen display and an optical head. A joystick, situated at the base of the device, controls the movements of the optical head to enable adequate positioning. The touch-screen display also includes a user interface allowing the user to control the device and acquire and display patient data and images.

The Zilia Ocular FC uses near-infrared (NIR) light-emitting diodes (LEDs) as illumination during alignment to the patient's pupil, and a white LED during focusing and imaging of the eye fundus. The region of the eye fundus being imaged can be modified by changing or moving the fixation target. The images are saved and stored on the onboard computer.

The dimensions of the Zilia Ocular FC are 620 mm (L) by 360 mm (W) by 500 mm (H) and it weighs about 32 kg.

AI/ML Overview

The provided document is a 510(k) Summary for the Zilia Ocular Fundus Camera (Zilia Ocular FC). It describes the device, its intended use, and performance testing conducted to demonstrate substantial equivalence to predicate and reference devices.

However, the document does not contain specific acceptance criteria, reported device performance data against those criteria, or the details of a clinical study (such as sample size, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance).

The document explicitly states: "Clinical Performance Data: This section is not applicable because clinical data was not provided for this 510(k) submission."

Therefore, I cannot provide the information requested in points 1, 2, 3, 4, 5, 6, 7, 8, and 9 based on the provided text. The submission relied on non-clinical performance testing (electrical safety, EMC, recognized standards, risk management, biocompatibility, processing of healthcare products, software evaluation, and usability) to demonstrate substantial equivalence, rather than clinical performance data.

{0}------------------------------------------------

November 20, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Zilia Inc. % Christina Henza Regulatory Consultant Ultra Life Science Solutions Inc. 2811 Milton Ave #409 Janesville, Wisconsin 53545

Re: K230627

Trade/Device Name: Zilia Ocular FC (ZIL-10002) Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic camera Regulatory Class: Class II Product Code: HKI Dated: October 27, 2023 Received: October 27, 2023

Dear Christina Henza:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Elvin Y. Na -S

Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230627

Device Name Zilia Ocular

Indications for Use (Describe)

Zilia Ocular is a non-contact imaging device for capturing, displaying and storing images of the retina under mydriatic conditions.

Zilia Ocular is intended for use as an aid to clinicians in the evaluation and documentation of ocular health.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "zilia" in a sans-serif font. The color of the text is a light green, almost pastel. The background is a solid, light gray color, providing a subtle contrast to the text.

125 Charest Blvd. East, Suite 400 Quebec City, Quebec G1K 3G5

□ 1 833 501-6996

510(k) SUMMARY Zilia Inc. Zilia Ocular Fundus Camera (Zilia Ocular FC)

This 510(k) summary is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.

SUBMITTER
Company NameZilia Inc.
Company Address125 Charest Boulevard East, Suite 400Quebec City, Quebec, CanadaG1K 3G5
Contact PersonPatrick Sauvageau, O.D., M.Sc., CEO
Phone1 833 501-6996
E-mailpatrick.sauvageau@ziliahealth.com
Websiteziliahealth.com
Date prepared2023-10-10
DEVICE INFORMATION
Trade NameZilia Ocular Fundus Camera
Common NameZilia Ocular FC
Classification NameOphthalmic Camera, AC powered
510(k) NumberK230627
Regulatory ClassClass II

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "zilia" in a light green color. The font is sans-serif and appears to be a custom design. The letters are evenly spaced and the overall impression is clean and modern.

Product CodeHKI
Regulation Number21 CFR 886.1120
PREDICATE DEVICE
Trade NameNFC-700 Non-Mydriatic Auto Fundus Camera
Classification NameOphthalmic Camera, AC powered
510(k) NumberK182199
REFERENCE DEVICE
Trade NameMydriatic Hyperspectral Retinal Camera(MHRC-C1)
Classification NameOphthalmic Camera, AC powered
510(k) NumberK200254

DEVICE DESCRIPTION

The Zilia Ocular Fundus Camera (Zilia Ocular FC) is a non-contact, stand-alone device designed to image the eye fundus.

The device includes a base, a head support, an onboard computer with a 15.6'' touch-screen display and an optical head. A joystick, situated at the base of the device, controls the movements of the optical head to enable adequate positioning. The touch-screen display also includes a user interface allowing the user to control the device and acquire and display patient data and images.

The Zilia Ocular FC uses near-infrared (NIR) light-emitting diodes (LEDs) as illumination during alignment to the patient's pupil, and a white LED during focusing and imaging of the eye fundus. The region of the eye fundus being imaged can be modified by changing or moving the fixation target. The images are saved and stored on the onboard computer.

The dimensions of the Zilia Ocular FC are 620 mm (L) by 360 mm (W) by 500 mm (H) and it weighs about 32 kg.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "zilia" in a light green color. The font is sans-serif and the letters are rounded. The word is written in lowercase.

The detailed technological characteristics and specifications of the Zilia Ocular FC are listed in the section Comparison Of Technological Characteristics With The Predicate Device located on page 6 of this summary.

INDICATIONS FOR USE

The Zilia Ocular FC is a non-contact imaging device for capturing, displaying, and storing images of the eye fundus under mydriatic conditions.

The Zilia Ocular FC is intended for use as an aid to clinicians in the evaluation and documentation of ocular health.

PERFORMANCE TESTING

All necessary performance testing was conducted on the Zilia Ocular FC and demonstrated that the subject device is substantially equivalent to the predicate device, the NFC-700, in terms of safety and effectiveness.

Electrical Safety and Electromagnetic Compatibility

It was verified that the Zilia Ocular FC complies with the standard ANSI/AAMI ES60601-1:2005/(R)2012+A1:2012 for electrical safety and with the standard IEC 60601-1-2:2014 for electromagnetic compatibility under the ASCA pilot program.

Recognized Consensus Standards For Ophthalmic Cameras

The Zilia Ocular FC was found to comply with the following recognized consensus standards:

  • ISO 15004-1:2020 Ophthalmic instruments Fundamental requirements and test methods . - Part 1: General requirements applicable to all ophthalmic instruments;
  • ISO 15004-2:2007 Ophthalmic instruments Fundamental requirements and test methods . - Part 2: Light hazard protection:
  • ANSI Z80.36-2021 American National Standard: Ophthalmics Light Hazard Protection . for Ophthalmic Instruments;
  • ISO 10940:2009 Ophthalmic instruments Fundus cameras; and .
  • IEC 60601-1-6:2010 Medical electrical equipment Part 1-6: General requirements for . basic safety and essential performance - Collateral standard: Usability.

Risk-Based Approach

The risk management report was prepared to document the evaluation and the decisions made and all necessary safety measures in the design and manufacturing of the Zilia Ocular FC. All documentation in accordance with ISO 14971:2019 Medical devices - Application of risk management to medical devices was created including a Risk Management Plan, a Hazard and Risk Analysis, a Production Failure Mode and Effect Analysis (PFMEA), and a Risk Management Report. The risks related to all applicable hazards which were identified for the Zilia Ocular FC have been reduced to the acceptable level by mitigation.

{6}------------------------------------------------

The Zilia Ocular FC is shown to be at least as safe and effective as the predicate device and the inherent risks are believed to be overcome by the benefits of the device use as indicated. Therefore, all residual risks post-mitigation have been deemed acceptable for this design.

Biocompatibility

The intact skin from the chin and forehead of the patient is intended to contact the Zilia Ocular FC for a short period of time. The biocompatibility of the Zilia Ocular FC was assessed in accordance with ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.

Processing of Health Care Products

The Zilia Ocular FC Instructions for Use document was assessed to comply with ISO 17664:2017 Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices.

Software Evaluation

The Zilia Ocular FC software was evaluated in accordance with IEC 62304:2006/AMD 1:2015 Medical device software - Software life cycle processes. The software for this device is considered to be a "Moderate Level of Concern" since a failure or latent flaw in the software could directly result in minor injury to the patient or user and due to the use of Off-the-shelf Software.

Labeling

The Zilia Ocular labeling was assessed to verify compliance to ISO 15223-1:2021 and ISO 20417:2021 and all applicable local requirements.

Usabilitv

Zilia applied the recommendations from the FDA's guidance Applying human factors and usability engineering to medical device:2016 and IEC 60601-1-6:2010 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability:2010+AMD1:2013+AMD2:2020 to maximize the likelihood that the Zilia Ocular FC is safe and effective for the intended users, uses and use environments.

Clinical Performance Data

This section is not applicable because clinical data was not provided for this 510(k) submission.

Animal Studies

This section is not applicable as there is no data related to animal studies for this 510(k) submission.

{7}------------------------------------------------

ZI IC

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE AND REFERENCE DEVICE

The proposed Zilia Ocular FC is substantially equivalent to the NFC-700 Non-Mydriatic Auto Fundus Camera (Crystalvue Medical Corporation) cleared in K182199 on 2019-01-02, certain technical differences are supported by the reference device Mydriatic Hyperspectral Retinal Camera (MHRC-C1) cleared in K200254 on 04/27/2020. The proposed indications for use are equivalent to the predicate and reference device, as the three devices are intended to capture, display, and store images of the eye fundus. The technological characteristics of the proposed Zilia Ocular FC encompass the functionalities of the NFC-700 and MHRC-C1.

FeatureZilia OcularFundus Camera(Zilia Ocular FC)(K230627)NFC-700(K182199)MHRC-C1(K200254)
Regulation number21 CFR 886.112021 CFR 886.112021 CFR 886.1120
Product codeHKIHKIHKI, NFJ
Regulation nameOphthalmicCameraOphthalmicCameraOphthalmicCamera
Regulatory classClass IIClass IIClass II
IntendedUse /Indicationfor useZilia Ocular is anon-contactimaging devicefor capturing,displaying, andstoring images ofthe eye fundusunder mydriaticconditions.Zilia Ocular isintended for useas an aid toclinicians in theevaluation anddocumentation ofocular health.NFC-700 is an AC-powered ophthalmiccamera to takephotographs of theeye and thesurrounding area.NFC-700 is a non-contact, highresolution digitalimaging devicewhich is suitablefor photographing,displaying andstoring images ofthe retina andexternal areas ofthe eye to beevaluated undernon-mydriaticThe MHRC-C1 isintended to captureimages of the retinaat multiplewavelengths(colors) undermydriaticconditions.
FeatureZilia OcularFundus Camera(Zilia Ocular FC)(K230627)NFC-700(K182199)MHRC-C1(K200254)
conditions.NFC-700 isindicated for in-vivoviewing of theposterior andexternal area of theeye and the imagesare intended for useas an aid toclinicians in theevaluation,diagnosis anddocumentation ofocular health.
PrescriptionuseRx OnlyRx OnlyRx Only
FundusimageMydriatic, colorimageNon-mydriatic,color imageMydriatic,monochrome
RecordmediaDigitalDigitalDigital
Contactareas withthe patientsChin andforeheadChin and foreheadChin and forehead
Mean foralignmentManualpositioningFully automatic 3DtrackingManualpositioning
MaintechnologyAC-poweredophthalmic cameraon a support framewith:AC-poweredophthalmic cameraon a support framewith:AC-poweredophthalmic cameraon a support framewith:
• A patient chinand forehead restfor patient head• A patient chinand forehead restfor patient head• A patient chinand forehead restfor patient head
FeatureZilia OcularFundus Camera(Zilia Ocular FC)(K230627)NFC-700(K182199)MHRC-C1(K200254)
placement.placement.placement.
• Visual fixationtarget for patientto maintainpatient's gazeand the pupil'salignment.• Visual fixationtarget for patientto maintainpatient's gazeand the pupil'salignment.• Visual fixationtarget for patientto maintainpatient's gazeand the pupil'salignment.
• Camera opticalhead alignmentfeatures to allowthe user toobserve, align,and target thedesired image ofthe patient's eyearea.• Camera opticalhead alignmentfeatures to allowthe user toobserve, align,and target thedesired image ofthe patient's eyearea.• Camera opticalhead alignmentfeatures to allowthe user toobserve, align,and target thedesired image ofthe patient's eyearea.
• NIR lightingduring alignmentand white LEDlighting duringfocusing andimaging.• NIR lightingduring alignmentand initialobservation tonot cause pupilconstriction.• Monochromaticlight (700 nm)during alignment.
• Camera to recordthe given imagesprovides lightingwith mask toilluminate theselected area forthe imaging.• Camera to recordthe given imagesprovides lightingwith mask toilluminate theselected area forthe imaging.• Camera to recordthe given imagesprovides lightingwith mask toilluminate theselected area forthe imaging.
FeatureZilia OcularFundus Camera(Zilia Ocular FC)(K230627)NFC-700(K182199)MHRC-C1(K200254)
· Software andcontrol userinterfaces tooperate thedevice.· Software andcontrol userinterfaces tooperate thedevice.· Software andcontrol userinterfaces tooperate device.
OperationPrincipleThe optical designof the funduscamera is based onthe principle ofmonocular indirectophthalmoscopy.1. Fundusobservation:Alignment to thepupil is madeusing two NIRLEDs. Indicatorsshown on thescreen help theuser center thedevice with thepupil. A whiteLED is used tofocus and imagethe eye fundus.The optical designof the funduscamera is based onthe principle ofmonocular indirectophthalmoscopy.1. Fundusobservation:A built-in lightray from theinfrared lightLED source toilluminate thefundus.Alignment ofthe device isperformed bybuilt-in eyetrackingindicator andworkingdistanceindicator toautomaticallyadjust thesystem to thebest XYZpositionautomatically.The optical designof fundus camera isbased on theprinciple ofmonocular indirectophthalmoscopy.1. Fundusobservation:A positioningsystem is used toalign the camerarelative to thepatient's eye.Monochromaticlight (700 nm) isused duringalignment. Thefundus isobserved througha video feed onan LCD monitor.
FeatureZilia OcularFundus Camera(Zilia Ocular FC)(K230627)NFC-700(K182199)MHRC-C1(K200254)
2. Image capture:System uses real-time imageanalysis toperform imagefocus adjustmentautomatically tocapture the bestquality of image.White lightemitted fromLEDs irradiatesthe eye fundus.The lightreflected from theeye fundus formsan image, and theimage is capturedby a built-in colorCMOS cameramodule forfundus imagecapture.2. Image capture:System usessplit-imagetechnique to doimage focusadjustmentautomatically tocapture the bestquality of image.White light fromLEDs Flashmodule irradiatesthe fundus. Thelight reflectedfrom eye portionsforms an image,and the image iscaptured by abuilt-in colorCMOS cameramodule forfundus imagecapture.2. Image capture:Focus is achievedmanually througha focus wheel.System uses atunable lightsource thatsequentiallypresents theretina withmonochromaticlight in thespectral range905 nm to 450nm in steps of 5nm.
Alignment/visualizationlight sourceNear-infrared LEDduring pupilalignmentWhite LED duringfocusing andimagingNear-infrared LEDduring alignmentWhite LED duringimagingMonochromaticlight (700 nm)during alignmentMonochromaticlight (sequentialfrom 905 nm to 450nm in steps of 5 nm)during imaging
Field ofview25°45°31.5°
FeatureZilia OcularFundus Camera(Zilia Ocular FC)(K230627)NFC-700(K182199)MHRC-C1(K200254)
Workingdistance55 mmfrom lens tocornea25 mmfrom lens tocornea46.7 mmfrom lens tocornea
Minimumpupildiameter4 mm4 mm6 mm
FocusManual orautomaticManual orautomaticManual
ExposureManualAutomaticAutomatic
Compensation forametropia-15D to +15D-15D to + 10D(withoutcompensation lens)-30D to -10D or+5D to +30D (withcompensation lens)-15D to +15D
Internalfixationtargets5 points10 pointsNone
Externalfixationtarget1 point,mobileNone1 point,mobile
Cameraresolution -pixelsize/pitchCMOS 5 MP -3.45 $μ$ m (pixelpitch)CMOS 12 MP -5.12 $μ$ m (pixelpitch)Not available

{8}------------------------------------------------

Zilia Ocular FC Premarket Notification

Image /page/8/Picture/1 description: The image shows the word "zilia" in a sans-serif font. The color of the text is a light green. There are two dots above the "i" letters, which are also light green.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the word "zilia" in a sans-serif font. The color of the text is a light green. The letters are lowercase and evenly spaced.

{10}------------------------------------------------

Zilia Ocular FC Premarket Notification

zilia

{11}------------------------------------------------

zilia

{12}------------------------------------------------

Zilia Ocular FC Premarket Notification

zilia

Considering the above, the Zilia Ocular FC proposes an equivalent technology and methods as the predicate and reference devices to address the same indications. The noted differences in their detailed specifications do not present any new questions of safety or effectiveness, whereby the devices are substantially equivalent.

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image contains the word "zilia" in a sans-serif font. The color of the text is a light green. The letters are all lowercase.

CONCLUSION

As described in this 510(k) Summary, comprehensive testing and analysis was conducted on the Zilia Ocular FC to ensure that the device is equivalent to the predicate and reference devices cleared under K182199 and K200254 when used in accordance with its Instructions for Use.

Based on the information in this submission, Zilia Ocular FC has the same intended use, technological characteristics, and operation principles as its predicate and reference devices. Therefore, Zilia Ocular FC is substantially equivalent to the predicate device.

§ 886.1120 Ophthalmic camera.

(a)
Identification. An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.(b)
Classification. Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.