K Number
K230597
Date Cleared
2023-04-28

(56 days)

Product Code
Regulation Number
890.5500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aduro Comb (Model: SZ-22A) is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of Ila - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.

Device Description

The Aduro Comb is (Model: SZ-22A) a comb-shaped device that emits led light with the intention to promote hair growth. The device provides distributed led to the scalp at 650nm (±10nm) while the comb teeth simultaneously part the user's hair to ensure the user's scalp. The device is designed as a handheld product, and it consists of the main unit, the charging dock and a power cable, as well as it is powered by the built-in rechargeable lithium battery. The device has only one key for switching on and off the device and it will automatically shut down after a 10-minute treatment is completed.

AI/ML Overview

The Aduro Comb (Model: SZ-22A) is indicated to treat Androgenetic Alopecia and promote hair growth. The following information outlines its acceptance criteria and the studies performed.

1. Table of Acceptance Criteria and Reported Device Performance

CriteriaAcceptanceReported Performance
Electrical SafetyCompliance with AAMI/ANSI ES60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-1-2Complies with all listed standards
Photobiological SafetyCompliance with IEC 62471Complies with IEC 62471
Battery SafetyCompliance with IEC 62133-2Complies with IEC 62133-2
Biocompatibility (for patient-contacting materials)Compliance with ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), ISO 10993-10 (Irritation)Complies with all listed ISO 10993 standards, based on equivalence to previously cleared devices.
Software Verification and ValidationCompliance with IEC 62304 and FDA Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"Documentation provided and considered sufficient for a "moderate" level concern software.

2. Sample size used for the test set and the data provenance

The document explicitly states, "Clinical testing was not needed for this 510(k)." Therefore, there was no clinical test set in the traditional sense, and thus no sample size or data provenance from a clinical study for the device's efficacy. The performance evaluation relied solely on non-clinical testing and comparison to a predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as no clinical test set was used to assess the device's efficacy or establish ground truth from experts. The evaluation focused on non-clinical safety and performance.

4. Adjudication method for the test set

Not applicable, as no clinical test set was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an infrared lamp for hair growth, not an AI-assisted diagnostic or imaging device that would typically involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is a physical infrared light therapy device, not an algorithm.

7. The type of ground truth used

For the non-clinical tests, the "ground truth" was compliance with established international and national standards for electrical safety, photobiological safety, battery safety, and biocompatibility.

8. The sample size for the training set

Not applicable, as this device does not involve a "training set" in the context of machine learning or AI.

9. How the ground truth for the training set was established

Not applicable, as there was no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 28, 2023

Light Tree Ventures Europe B.V. Alain Dijkstra CEO Laan van Ypenburg 108, 2497 GC Hague, Netherlands

Re: K230597

Trade/Device Name: Aduro Comb (Model SZ-22A) Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: March 2, 2023 Received: March 3, 2023

Dear Alain Dijkstra:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230597

Device Name Aduro Comb (Model: SZ-22A)

Indications for Use (Describe)

The Aduro Comb (Model: SZ-22A) is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of Ila - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale 1-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)X

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of K230579

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

Sponsor Name: Light Tree Ventures Europe B.V. Establishment Registration Number: 3017422691 Address: Laan van Ypenburg 108, 2497 GC, The Hague, The Netherlands Contact Person (including title): Alain Dijkstra (Manager) Tel: +86-135-10378748 Fax: +86-755-25024651 E-mail: regulation@kaiyanmedical.com

Application Correspondent

Contact Person: Alain Dijkstra Company: Light Tree Ventures Europe B.V. Address: Laan van Ypenburg 108, 2497 GC, The Hague, The Netherlands Tel: +86 755 82129361 Fax: +86 755 25024651 Email: requlation@kaiyanmedical.com

Manufacture

Company Name: Shenzhen Kaiyan Medical Equipment Co., Ltd Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China

Summary Prepared Date: 2023-04-03

2. Subject Device Information

Trade Name: Aduro Comb, model: SZ-22A Classification Name: Laser, Comb, Hair (OAP) Review Panel: Plastic Surgery Product Code: OAP Regulation Number: 21 CFR 890.5500 Regulation Class: II

3. Predicate Device Information

Predicate Device (K211038) Sponsor: Biophotas Inc Trade Name: Biophotas Celluma RESTORE Classification Name: Laser, Comb, Hair 510(k) Number: K211038 Review Panel: General & Plastic Surgery Product Code: OAP

{4}------------------------------------------------

Regulation Number: 21 CFR 890.5500 Regulation Class: II

4. Device Description

The Aduro Comb is (Model: SZ-22A) a comb-shaped device that emits led light with the intention to promote hair growth. The device provides distributed led to the scalp at 650nm (±10nm) while the comb teeth simultaneously part the user's hair to ensure the user's scalp. The device is designed as a handheld product, and it consists of the main unit, the charging dock and a power cable, as well as it is powered by the built-in rechargeable lithium battery. The device has only one key for switching on and off the device and it will automatically shut down after a 10-minute treatment is completed.

5. Intended Use / Indications for Use

The Aduro Comb (Model: SZ-22A) is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.

6. Comparison to Predicate Device

Compared with the predicate device, the subject device has similar in the design principle, the intended use, the indications for use, functions and applicable standards. The differences between the subject device and the predicate device do not raise any new questions of safety or effectiveness.

Elements of ComparisonSubject DevicePredicate DeviceRemark
CompanyLight Tree Ventures EuropeB.V.Biophotas Inc--
Trade NameAduro Comb (Model: SZ-22A)Biophotas CellumaRESTORE--
Classification NameLaser, Comb, Hair (OAP)Laser, Comb, Hair (OAP)--
510(k) NumberK230579K211038--
Product CodeOAPOAPSame
Intended Use / Indicationsfor UseThe Aduro Comb (Model: SZ-22A) is indicated to treatAndrogenetic Alopecia andpromote hair growth in maleswho have Norwood- HamiltonClassifications of Ila - Vpatterns of hair loss and totreat Androgenetic Alopeciaand promote hair growth infemales who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2or frontal patterns of hair loss;both with Fitzpatrick SkinThe BioPhotasCelluma RESTORE isindicated to treatAndrogenetic Alopeciaand promote hairgrowth in males whohave Norwood-HamiltonClassifications of Ila -V patterns of hair lossand to treatAndrogenetic Alopeciaand promote hairSame
Elements of ComparisonSubject DevicePredicate DeviceRemark
Types I - IV.growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.
Wavelengths650±10nm640nm (±25nm)SimilarNote 1
Treatment timeEach Treatment: 30 minTotal Treatment: every other day, for 16 weeksEach Treatment:30 min Total Treatment: every other day, for 16 weeksSame
Treatment area31.5 cm²475 cm²DifferentNote 2
Energy density (mw/cm²)2.77 mw/cm²2.77 mw/ cm²Same
Treatment Dose(J/cm²)4.98 J/ cm²4.98 J/ cm²Same
Energy SourceLight emitting diodesLight emitting diodesSame
Number of LEDs22 LEDsNo publicly availableDifferentNote 2
Power supplyInput: 100-240V~, 50/60Hz, 0.35AmaxOutput:5V, 2ABattery: DC 3.7 V, 600 mAh, 2.22 Wh110-120VSimilarNote 3
Location for UseOTCOTCSame
Safety and EMCIEC 60601-1IEC 60601-1-11IEC 62471IEC 60601-2-57IEC 60601-1-2IEC 62133-2IEC 60601-1IEC 60601-1-2SimilarNote 4
BiocompatibilityAll patient contacting materials comply with ISO 10993-5, ISO 10993-10All patient contacting materials comply with ISO 10993-5, ISO 10993-10Same

{5}------------------------------------------------

Note 1: Although the "Wavelength" of the subject device has a slight difference from the predicate devices, the wavelength of the subject device is within the error range of the predicate devices, so the wavelength of the subject device can be fully covered by the predicate devices. Therefore, this difference between the subject device and predicate devices will not raise any safety or effectiveness issues.

{6}------------------------------------------------

Note 2: Although the "Treatment area" and the "Number of LEDs" are different from the predicate devices because of the different design shapes, both the subject device and predicate devices have the same treatment parameters. Therefore, these slight differences between the subject device and predicate devices will not raise any safety or effectiveness issues.

Note 3: Although the "Power supply" of the subject device is not exactly the same as the predicate devices, both the subject device and the predicate devices conduct the safety test according to the IEC 60601 series standards, and the test results are in compliance with safety standards' requirements. So, this difference between the subject device and the predicate devices will not raise any safety or effectiveness issues.

Note 4: Although the description in "Safety and EMC" of the subject device is slightly different from the predicate devices, both the subject device and the predicate devices conducted the electrical safety and electromagnetic compatibility tests according to the international series standards, and the test results are in compliance with the standards' requirements. So, this difference between the subject device and the predicate devices will not raise any safety or effectiveness issues.

7. Test Summary

7.1 Non-Clinical Tests Performed

1) Electrical safety, and electromagnetic compatibility

Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:

  • AAMI/ANSI ES60601-1 Medical electrical equipment Part 1: General requirements for basic safety � and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)
    • IEC 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • � IEC 60601-2-57 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
  • ◆ IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • � IEC 62471 Photobiological safety of lamps and lamp systems
  • � IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid electrolytes -Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems

2) Biocompatibility

The component materials of the subject device are identical to the corresponding component materials of the e previously cleared devices (K203271 and K202390) in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents).

{7}------------------------------------------------

There is no change in biocompatibility since the previously cleared devices. Therefore, based on this information, the subject device can comply with the biocompatibility requirements of ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), and ISO 10993-10 (Irritation).

3) Software verification and validation testing

Software verification and validation testing were conducted and documentation was provided as recommended by the IEC 62304 and FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.

7.2 Summary of Clinical Performance

Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.

Final Conclusion:

The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated devices K211038.

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.