K Number
K230507

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-01-04

(314 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Zimed Distal Medial Tibial Plate and Screw System is indicated for fixation of complex extra-articular and intraarticular distal medial tibial fractures and osteotomies.

Device Description

Zimed Distal Medial Tibial Plate and Screw System is used for complex extra-articular and intraarticular distal tibial fractures, and osteotomies surgeries specifically designed for the anatomical alignment of the broken bones and to provide splinting with the ability to ensure the early mobility of the patient. Zimed plate and screw system is implantable device in long term contact with bones.

Zimed Distal Medial Tibial Plate and Screw System will be supplied non-sterile.

AI/ML Overview

This document is a 510(k) premarket notification for the Zimed Distal Medial Tibial Plate and Screw System. It describes the device, its intended use, and provides a comparative analysis to predicate devices to establish substantial equivalence.

Here's an analysis of the acceptance criteria and study information, based on the provided text:

This device is a metallic bone fixation appliance, not an AI/software device, therefore the typical acceptance criteria and study information for AI-driven solutions are not applicable. The information provided focuses on demonstrating substantial equivalence to previously cleared predicate devices through comparisons of technological characteristics and performance testing to established standards.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Standards)Reported Device Performance (Compliance/Results)
ASTM F382 (Standard Specification for Metallic Bone Plates)Performance tests "confirmed that the subject device... meet ASTM F382"
ASTM F136 (Standard Specification for Wrought Titanium-6Aluminum-4Vanadium Alloy for Surgical Implant Applications)Performance tests "confirmed that the subject device... meet ASTM F136"
ASTM F543 (Standard Specification and Test Methods for Metallic Medical Bone Screws)Performance tests "confirmed that the subject device... meet ASTM F543"
FDA guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway"Testing utilized this FDA guidance for screw testing.
FDA guidance "Orthopedic Fracture Fixation Plates - Performance Criteria for Safety and Performance Based Pathway"Testing utilized this FDA guidance for plate testing.

2. Sample Size for the Test Set and Data Provenance:

The document describes performance testing on the device itself (physical plates and screws), not a clinical study involving patients or data. Therefore, terms like "test set," "sample size," and "data provenance" (country of origin, retrospective/prospective) are not applicable in the context of this 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Since this is a physical device safety and performance evaluation rather than a diagnostic device, there are no "experts used to establish ground truth for the test set" in the way one would for an AI-driven image analysis tool. The "ground truth" is adherence to established engineering and material standards. This is determined through objective laboratory testing.

4. Adjudication Method for the Test Set:

Not applicable, as this is not a study involving human interpretation or subjective assessments that would require adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, a MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human reader performance (with and without aid) is evaluated. This 510(k) is for a physical orthopedic implant.

6. If a standalone (algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a physical implant, not an algorithm or software.

7. The Type of Ground Truth Used:

The ground truth used is adherence to established engineering and material standards (ASTM F382, ASTM F136, ASTM F543) and compliance with relevant FDA guidance documents for orthopedic implant performance.

8. The Sample Size for the Training Set:

Not applicable. This device is a physical implant, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As there is no training set for a physical device, no ground truth needed to be established for it.

{0}------------------------------------------------

January 4, 2024

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side by side.

Zimed Medikal % Barry Sands President and Founder RQMIS, Inc. 110 Haverhill Road Suite 524 Amesbury, Massachusetts 01913

Re: K230507

Trade/Device Name: Zimed Distal Medial Tibial Plate and Screw System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: December 5, 2023 Received: December 6, 2023

Dear Barry Sands:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

K230507 - Barry Sands

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230507

Device Name

Zimed Distal Medial Tibial Plate and Screw System

Indications for Use (Describe)

The Zimed Distal Medial Tibial Plate and Screw System is indicated for fixation of complex extra-articular and intraarticular distal medial tibial fractures and osteotomies.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K230507

510(K) SUMMARY
Submitter information
Date PreparedJanuary 2, 2024
Applicant/ManufacturerZimed Medikal Sanayi ve Ticaret Limited SirketiAydinlar Mah.03070 Cad.No:04 Sehitkamil GaziantepTurkiyeGaziantep, Sehitkamil, 27580, Turkiye
Primary ContactBarry SandsPresident and Founder, RQMISPhone: (978) 358-7307Email: regulatorysubmissions@rqmis.com

Device information

Trade Name of Subject DeviceZimed Distal Medial Tibial Plate and Screw System
Common NamePlate, Fixation, Bone; Screw, Fixation, Bone
Device classClass II
Product CodeHWC, HRS (primary)
Regulation Number888.3030 Single/multiple component metallic boneFixation appliances and accessories (Primary)888.3040 Smooth or threaded metallic bone fixationfastener
Predicate Devices:Primary predicate: Synthes (USA) LCP Distal TibiaPlate (K013248)Reference predicate: HEMC BRAND Locking Bone Platesand Screws Osteosynthesis Plating System, HEMC BRANDof DHS Plating system (K222282)

{4}------------------------------------------------

Device Description

Zimed Distal Medial Tibial Plate and Screw System is used for complex extra-articular and intraarticular distal tibial fractures, and osteotomies surgeries specifically designed for the anatomical alignment of the broken bones and to provide splinting with the ability to ensure the early mobility of the patient. Zimed plate and screw system is implantable device in long term contact with bones.

Zimed Distal Medial Tibial Plate and Screw System will be supplied non-sterile.

Indications for Use

The Zimed Distal Medial Tibial Plate and Screw System is indicated for fixation of complex extraarticular and intra-articular distal medial tibial fractures and osteotomies.

Intended Use – Comparison of the subject device with Primary & Reference Predicate Devices

The subject device, Zimed Distal Medial Tibial Plate and Screw System, has similar intended use and indication for use to the primary predicate Synthes (USA) LCP Distal Tibia Plate (K013248), and reference predicate, HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, HEMC BRAND of DHS Plating system (K222282)

Technological Characteristics - Comparison with the primary predicate, Synthes (USA) LCP Distal Tibia Plate (K013248) and reference predicate, HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, HEMC BRAND of DHS Plating system (K222282)

The technological characteristics comparison demonstrated that the subject device, the Plate and Screw System are substantially equivalent in intended uses, designs, materials, and operational principles to the previously cleared predicate devices, respectively.

Basis of Substantial Equivalence

The substantial equivalence of the subject device was determined per the FDA guidance document, "The 510(k) program: evaluating substantial equivalence in premarket Notification [510(k)]." The analysis of the technological characteristics, which include design, material composition and other device features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2) (ii)(A), demonstrated that the Plate and Screw System are substantially equivalent to the identified predicate devices, respectively.

Zimed Distal Medial Tibial Plate and Screw System are substantially equivalent to the respective predicates. Comparison to the predicates does not raise different questions of safety and effectiveness. The performance testing, device comparisons, and feature analysis

{5}------------------------------------------------

demonstrate that the subject device is substantially equivalent to the identified predicate devices, respectively.

Performance Testing

The performance tests on the subject device were conducted in accordance with ASTM F382, ASTM F136 and ASTM F543 standards. Performance test reports confirmed that the subject device, Zimed Distal Medial Tibial Plate and Screw System, and the reference predicate, HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, HEMC BRAND of DHS Plating system (K222282) meet ASTM F136 and ASTM F543 and used the same test methodology as the required specifications. We utilized the FDA guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway" for the screw testing, and the FDA guidance, "Orthopedic Fracture Fixation Plates - Performance Criteria for Safety and Performance Based Pathway" for the plate testing. Therefore, the subject device is substantially equivalent to the respective predicates' devices identified throughout this submission.

Substantial Equivalence Conclusion

Zimed Distal Medial Tibial Plate and Screw System has similar intended use, material comparison, design, operational principle, and equivalent technological characteristics to the respective predicate devices. The non-clinical performance test reports support the safety, effectiveness and performance of the subject device and demonstrate that any difference in technological characteristics and material composition do not raise any new questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the respective predicate devices.

N/A