K Number
K230499
Device Name
ArtiSential Laparoscopic Instruments-Electrodes
Manufacturer
Date Cleared
2023-04-26

(61 days)

Product Code
Regulation Number
878.4400
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Indications for use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures.
Device Description
The ArtiSential Laparoscopic Instruments – Electrodes, Bipolar series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) are sterile, single-use, invasive instruments that used in laparoscopic surgery. There are four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series. Three versions are same except for jaw shape. This product is a specific component, but not the entire electrosurgical device. The device is not intended to be marketed with multiple components, accessories, and as part of a system.
More Information

No
The summary describes a standard electrosurgical instrument and does not mention any AI or ML capabilities.

No.
The device is used for surgical procedures (coagulation, dissection, grasping) and does not inherently treat diseases or conditions.

No

The device is an electrosurgical instrument for coagulation, dissection, and grasping tissue during surgical procedures. Its function is therapeutic and interventional, not diagnostic.

No

The device description explicitly states it is a sterile, single-use, invasive instrument used in laparoscopic surgery, and mentions performance studies related to biocompatibility, electrical safety, sterilization, and thermal effects on tissue, all of which are characteristic of hardware medical devices.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is for "electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures." This describes a surgical tool used directly on the patient's body for treatment and manipulation of tissue.
  • Device Description: The description confirms it's a "sterile, single-use, invasive instrument that used in laparoscopic surgery." This further reinforces its role as a surgical instrument.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue samples) to provide information for diagnosis, monitoring, or compatibility testing.

IVD devices are used in vitro (outside the body) to analyze samples for diagnostic purposes. This device is used in vivo (inside the body) for surgical procedures.

N/A

Intended Use / Indications for Use

Indications for use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures.

Product codes

GEI

Device Description

The ArtiSential Laparoscopic Instruments – Electrodes, Bipolar series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) are sterile, single-use, invasive instruments that used in laparoscopic surgery. There are four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series. Three versions are same except for jaw shape. This product is a specific component, but not the entire electrosurgical device. The device is not intended to be marketed with multiple components, accessories, and as part of a system.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

The device has been evaluated for its biological safety according to ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1 : Evaluation and Testing Within a Risk Management Process". Following endpoints have been assessed during the evaluation:

  • Cytotoxicity -
  • -Intracutaneous reactivity
  • -Skin Sensitization
  • Acute systemic toxicity -
  • -Pyrogenicity

The ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) have been tested according to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18 and IEC 60601-2-2. The test setup included:
The device had passed all performed tests.

ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series. ABD01 series, ABD02 series and ABD04 series) are provided sterile, intended to be single-use. This product is EO-Sterilization in accordance with ISO-11135.

The proposed expiration date is 3 years from the manufacturing date. The real-time testing will be performed to confirm the shelf-life for 3 years.

The device had passed all performed tests.

  1. Appearance: Implement the visual inspection for surface of the device
  • Test requirement: There should be no defects in the appearance of the product and there should be no problem in use.
  • Results-Remarks: No crack, stain or no substances on the surface of the product
  1. Dimension: Measure by ruler and vernier calipers
  • Test requirement: It shall be within +/- 5% of the indicated value of the dimensional term.
  • Results-Remarks: Pass. Refer to [Test result] on 9-50 page at attachment 12
  1. Operational test: Manipulating the grip and control ring, and measure the angle at bending and rotation by goniometer.
  • Test requirement: The jaw must be smoothly opened and closed and free from jamming, the jaw and hub can be bent up, down, left, and right a range of above +/- 80° and are capable of 360° rotation.
  • Results-Remarks: The jaw and hub are bent up, down, left and right within above 80° and can rotate 360°.
  1. Tensile strength: Hold the jaw and shaft connections respectively and apply a force of 20 N using Push pull gauge.
  • Test requirement: The jaw and shaft connections shall not be damaged from pulling of 20 N.
  • Results-Remarks: No damage to the connection when applying a force of 20N
  1. Feedthrough test: Electrical conduction between the electrode tip and the connector is tested using a DMM (digital multi meter).
  • Test requirement: Electricity should be transmitted between the electrode tip and the connector.
  • Results-Remarks: The resistance value between the electrode tip and the connector is less than 1Ω

Thermal effects on tissue were also tested. A histological analysis was performed on thermal effect to porcine tissues(liver, kidney and abdominal muscle) through an electrosurgical device.

Key Metrics

Not Found

Predicate Device(s)

K200875, K220384

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 26, 2023

LivsMed Inc. Dong Wook Lee QMR (Quality Management Representative) #304, D-dong, 700, Pangyo-ro, Bundang-gu Seongnam-si. Gyeonggi-do 13516 Korea, South

Re: K230499

Trade/Device Name: ArtiSential Laparoscopic Instruments-Electrodes, Bipolar series (four versions, ABF01series, ABD01 series, ABD02 series and ABD04 series)

Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: February 24, 2023 Received: February 24, 2023

Dear Dong Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Image /page/2/Picture/0 description: The image shows the logo for LIV'SMED. The logo is in a dark blue color, with the apostrophe in LIV'SMED being a lighter blue color. The font is sans-serif and the letters are all capitalized.

3. Indications for Use Statement

3

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K230499

Device Name

ArtiSential Laparoscopic Instruments-Electrodes, Bipolar series (four versions, ABF01 series, ABD02 series and ABD04 series)

Indications for Use (Describe)

Indications for use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

4

510(k) Summary

1. General Information

Applicant/Submitter: LivsMed Inc.
Address:#304, D-dong, 700, Pangyo-ro, Bundang-gu, 13516
Seongnam-si, Gyeonggi-do, Republic of Korea
Tel) +82-70-4282-7652
Fax) +82-31-706-3211
Contact Person:Dong Wook Lee / QMR
(Quality Management Representative)
Address:#304, D-dong, 700, Pangyo-ro, Bundang-gu, 13516
Seongnam-si, Gyeonggi-do, Republic of Korea
Tel) +82-70-7709-4993
Fax) +82-31-706-3211
Email) dongwook.livsmed@gmail.com
Preparation Date:02-24-2023

2. Device Name and Code

| Device Trade Name | ArtiSential Laparoscopic Instruments-Electrodes,
Bipolar Series (four versions, ABF01 series, ABD01
series, ABD02 series and ABD04 series) |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Electrosurgical Instruments |
| Classification Name | Electrosurgical, cutting & coagulation & accessories |
| Product Code | GEI |
| Regulation Number | 21 CFR 878.4400 |
| Classification | Class II |
| Review Panel | General & Plastic Surgery |

3. Predicate Devices

ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) are substantially equivalent to the following devices

5

Table 3.1 Predicate device 1

ApplicantDevice Name510(k) Number
LivsMed Inc.ArtiSential Bipolar
Fenestrated ForcepsK200875

Table 3.2 Predicate device 2

ApplicantDevice Name510(k) Number
LivsMed Inc.ArtiSential Bipolar
Fenestrated ForcepsK220384

4. Device Description

The ArtiSential Laparoscopic Instruments – Electrodes, Bipolar series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) are sterile, single-use, invasive instruments that used in laparoscopic surgery. There are four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series. Three versions are same except for jaw shape. This product is a specific component, but not the entire electrosurgical device. The device is not intended to be marketed with multiple components, accessories, and as part of a system.

5. Indications for use

  • 5.1 Indications for use
    Indications for use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures.

6. Technical Characteristics in Comparison to Predicate Devices

Proposed devicePredicate Device 1Predicate Device 2
510(K)
NumberIn processK200875K220384
ManufactureLivsMed, Inc.LivsMed, Inc.LivsMed, Inc.
Device NameArtiSential Laparoscopic
Instruments-ElectrodesArtiSential Laparoscopic
Instruments-ElectrodesArtiSential Laparoscopic
Instruments-Electrodes
Clearance DateN/A05-21-202002-24-2022
Classification /
RegulationClass 2 / 878.4400Class 2 / 878.4400Class 2 / 878.4400
Product CodeGEIGEIGEI
Intended forPrescription UsePrescription UsePrescription Use
Indications for
UseElectrosurgical
coagulation, dissection,
and grasping of tissue
during the performance
of laparoscopic and
general surgical
procedures.Electrosurgical
coagulation, dissection,
and grasping of tissue
during the performance
of laparoscopic and
general surgical
procedures.Electrosurgical
coagulation, dissection,
and grasping of tissue
during the performance
of laparoscopic and
general surgical
procedures.
Principles ofThis product is a single-This product is a single-This product is a single-
operationuse instrument used inuse instrument used inuse instrument used in
electrosurgical units toelectrosurgical units toelectrosurgical units to
hold soft tissues orhold soft tissues orhold soft tissues or
coagulate and make ancoagulate and make ancoagulate and make an
incision (tissueincision (tissueincision (tissue
dissection) duringdissection) duringdissection) during
general laparoscopicgeneral laparoscopicgeneral laparoscopic
surgery, which uses thesurgery, which uses thesurgery, which uses the
principle of applyingprinciple of applyingprinciple of applying
high-frequency currentshigh-frequency currentshigh-frequency currents
from the electrode to thefrom the electrode to thefrom the electrode to the
human body to generatehuman body to generatehuman body to generate
heat by bioimpedanceheat by bioimpedanceheat by bioimpedance
when radio frequencywhen radio frequencywhen radio frequency
(RF) energy from the(RF) energy from the(RF) energy from the
electrosurgical unitelectrosurgical unitelectrosurgical unit
applies an electricapplies an electricapplies an electric
current to the electrodecurrent to the electrodecurrent to the electrode
part, and using thepart, and using thepart, and using the
generated heat to incisegenerated heat to incisegenerated heat to incise
cellular tissues and causecellular tissues and causecellular tissues and cause
coagulation.coagulation.coagulation.
It is composed of a jaw,It is composed of a jaw,It is composed of a jaw,
Φ8 diameter shaft, gripΦ8 diameter shaft, gripΦ8 diameter shaft, grip
(including a control(including a control(including a control
ring), and electrosurgicalring), and electrosurgicalring), and electrosurgical
unit connectionunit connectionunit connection
electrode connector.electrode connector.electrode connector.
During a procedure withDuring a procedure withDuring a procedure with
this product, the jawthis product, the jawthis product, the jaw
opens if the control ringopens if the control ringopens if the control ring
opens, and jaw closes ifopens, and jaw closes ifopens, and jaw closes if
the control ring closes.the control ring closes.the control ring closes.
In addition, the jaw isIn addition, the jaw isIn addition, the jaw is
also bent up, down, leftalso bent up, down, leftalso bent up, down, left
and right within a rangeand right within a rangeand right within a range
of ±80° or more byof ±80° or more byof ±80° or more by
moving the grip up,moving the grip up,moving the grip up,
down, left and right, and
the jaw can also turndown, left and right, and
the jaw can also turndown, left and right, and
the jaw can also turn
360° when rotating the360° when rotating the360° when rotating the
grip.grip.grip.
Energy TypeRadiofrequencyRadiofrequencyRadiofrequency
Electrode typeBipolarBipolarBipolar
(monopolar or
bipolar)
Physical- Shaft diameter: 8mm- Shaft diameter: 8mm- Shaft diameter: 8mm
dimensions- Shaft Length: 250mm,- Shaft Length: 250mm,- Shaft Length: 250mm,
and design380mm, 450mm380mm, 450mm380mm, 450mm
(size, length)
Rated voltage200Vp200Vp200Vp
MaterialsStainless steelStainless steelStainless steel
(electrode)
Materials
(insulation)PolyetherimidePolyetherimidePolyetherimide
Materials
(Shaft)Glass fiberGlass fiberGlass fiber
Articulating
featurePitch:±80° or more,
Yaw:±80° or more and
Open-ClosePitch:±80° or more,
Yaw:±80° or more and
Open-ClosePitch:±80° or more,
Yaw:±80° or more and
Open-Close
Tip rotation360°360°360°
SterilizationEOEOEO

Table 6.1 Predicate Device

6

7

7. Performance Data

7.1 Biocompatibility

The device has been evaluated for its biological safety according to ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1 : Evaluation and Testing Within a Risk Management Process". Following endpoints have been assessed during the evaluation:

  • Cytotoxicity -
  • -Intracutaneous reactivity
  • -Skin Sensitization
  • Acute systemic toxicity -
  • -Pyrogenicity

7.2 Electrical Safety

The ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) have been tested according to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18 and IEC 60601-2-2. The test setup included:

The device had passed all performed tests.

7.3 Sterilization

ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series. ABD01 series, ABD02 series and ABD04 series) are provided sterile, intended to be single-use. This product is EO-Sterilization in accordance with ISO-11135.

  • 7.4 Shelf life
    The proposed expiration date is 3 years from the manufacturing date. The real-time testing will be performed to confirm the shelf-life for 3 years

7.5 Performance test

The device had passed all performed tests.

8

Test clause and specificationTest requirementResults- Remarks
1. Appearance
: Implement the visual
inspection for surface of the
deviceThere should be no defects in
the appearance of the product
and there should be no
problem in use.No crack, stain or no
substances on the surface
of the product
2. Dimension
: Measure by ruler and vernier
calipersIt shall be within $\pm$ 5% of the
indicated value of the
dimensional term.Pass
Refer to [Test result] on
9-50 page at attachment
12
3. Operational test
: Manipulating the grip and
control ring, and measure the
angle at bending and rotation
by goniometer.The jaw must be smoothly
opened and closed and free
from jamming, the jaw and
hub can be bent up, down,
left, and right a range of
above $\pm$ 80° and are capable
of 360° rotation.The jaw and hub are bent
up, down, left and right
within above 80° and can
rotate 360°.
4. Tensile strength
: Hold the jaw and shaft
connections respectively and
apply a force of 20 N using
Push pull gauge.The jaw and shaft
connections shall not be
damaged from pulling of 20
N.No damage to the
connection when applying
a force of 20N
5. Feedthrough test
: Electrical conduction
between the electrode tip and
the connector is tested using a
DMM (digital multi meter).Electricity should be
transmitted between the
electrode tip and the
connector.The resistance value
between the electrode tip
and the connector is less
than 1Ω

7.6 Thermal effect

Thermal effects on tissue were also tested. A histological analysis was performed on thermal effect to porcine tissues(liver, kidney and abdominal muscle) through an electrosurgical device.

Based on these performance characteristics, the results demonstrate that the performance requirements were met, the device performs as intended and that the subject device has substantially equivalent performance characteristics to the predicate devices.

8. Substantial Equivalence

ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) indication for use is same to the predicate device 1 (K200875) and predicate device 2 (K220384). The energy type, electrode type, sterilization as well as physical characteristics are the same. Although there are some minor differences with each product, these differences between the ArtiSential Bipolar Series and the predicate device do not raise new or different questions of safety and efficacy. There is no new technology and no difference that would raise new or different questions of safety or efficacy.

9

9. Conclusions

In conclusion, the comparison carried out covers all products, models, sizes, and the entire intended purpose of the device under evaluation. The subject device which is the ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) are same to the predicate device in principles of operation, technological characteristics, as well as performance characteristics. The testing was conducted to evaluate the performance of subject device in comparison to the predicate device. Results of validation and verification activities in design control that include testing/certification to designated standards and performance testing of the devices has demonstrated substantial equivalence of the subject device to the predicate in terms of safety and effectiveness for requested intended use.