(254 days)
The Evenflo Premium Double Electric Breast Pump (Model 4018) is an electrically powered suction device intended to express and collect milk from a lactating woman's breasts. It is intended for a single user.
The Evenflo Electric Breast Pump (Model 4018) is an electrically powered breast pump to be used in a home environment by a single user. The device is provided non-sterile and should be cleaned and disinfected prior to first use according to the instructions for use. The breast pump can be used on one breast (single pumping) or on both breasts at the same time (double pumping).
The device consists of a pump unit, AC power adapter, power cable, AdvancedFit flange kit and bottle kit. The AdvancedFit flange kit consists of a flange body, flange insert, check valve (valve and membrane), diaphragm, diaphragm cap, and tubing for each breast, a tubing adapter to connect the kit to the pump unit and a tubing adapter plug to convert for single pumping. It does not incorporate off-the-shelf (OTS) software or wireless technology/mobile app software functionality. The pump is powered by an internal, non-replaceable, rechargeable lithium-ion battery which is charged using the included AC power supply and cable. The subject device can be operated while plugged into AC power.
The breast pump uses cyclic negative pressure (suction) to mimic the suckling patterns of a feeding infant. A DC motor drives a membrane vacuum pump to generate the suction required to stimulate and express breast milk. The timing of this pattern is dependent upon the suction/speed settings selected by the user and is preprogrammed in the device. The device is capable of producing peak suction levels between -50 and -250 mmHg in double pumping mode and between -85 and -250 mmHg in single pumping modes at speeds between 30 and 80 cycles per minute. There are 9 distinct levels of suction and 6 speeds available for each suction level.
The breast pump expresses breast milk by creating a seal around the nipple using a flange and applying and releasing suction to the nipple. The milk is collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system.
All other components (i.e., motor unit) of the subject device are not in contact with the breast. All milk contacting components are compliant with 21 CFR 174-179.
The provided text is a 510(k) summary for the Evenflo Premium Double Electric Breast Pump (Model 4018). It focuses on demonstrating substantial equivalence to a predicate device based on similar intended use and technological characteristics, and non-clinical performance testing.
However, the document does NOT contain information about acceptance criteria for a device that uses AI/algorithm or a study involving human readers or expert consensus for ground truth establishment, which are typical for AI/ML-enabled medical devices.
Based on the provided text, I can answer some of your questions, but many will be marked as "Not applicable" or "Not provided" because the device is a breast pump, not an AI/ML diagnostic tool, and the summary focuses on equivalence to a predicate device rather than de novo performance studies as would be seen for an AI/ML device.
Here's the breakdown of the information from the document in response to your questions:
1. A table of acceptance criteria and the reported device performance
The document does not present acceptance criteria for AI model performance (e.g., sensitivity, specificity, AUC) or human reader performance. Instead, it describes performance testing for a breast pump, focusing on engineering specifications.
| Acceptance Criteria Category | Specific Test/Performance Metric | Reported Device Performance | Comments/Context |
|---|---|---|---|
| Biocompatibility | Cytotoxicity (ISO 10993-5:2009) | Non-cytotoxic | Tested on patient-contacting components |
| Sensitization (ISO 10993-10:2010) | Non-sensitizing | Tested on patient-contacting components | |
| Irritation (ISO 10993-10:2010) | Non-irritating | Tested on patient-contacting components | |
| Electrical Safety | ANSI/AAMI ES60601-1:2005/A2:2010 compliance | Compliant | General requirements for basic safety and essential performance |
| IEC 62133-2:2017 compliance (Lithium systems) | Compliant | Safety requirements for portable sealed secondary cells | |
| IEC 60601-1-11:2015 compliance (Home healthcare) | Compliant | Requirements for medical electrical equipment in home healthcare | |
| Electromagnetic Compatibility (EMC) | IEC 60601-1-2:2014 compliance | Compliant | Requirements and tests for EMC |
| Software | Software evaluation (Minor level of concern) | Evaluated per guidance | No specific performance metrics provided, but states "minor level of concern" implies it met thresholds for low risk. Note: This refers to the software controlling the pump, not AI/ML. |
| Mechanical/Functional Performance | Vacuum level verification | Meets mode/cycle specifications | Tested at each mode/cycle to confirm specified vacuum levels (-50 to -250 mmHg double; -85 to -250 mmHg single) |
| Backflow protection | Liquid does not backflow into tubing | Confirms physical separation to prevent milk from entering vacuum system | |
| Use life testing | Maintains specifications throughout proposed use life | Demonstrates durability over the expected lifespan | |
| Battery performance testing | Battery remains functional during stated use-life | Confirms battery operation duration | |
| Battery status indicator testing | Indicator remains functional during stated battery life | Confirms accuracy/functionality of low-battery warnings |
2. Sample sizes used for the test set and the data provenance
- Sample Size for Test Set: Not explicitly stated in terms of number of devices tested for each performance criterion. The document refers to "testing" conducted.
- Data Provenance: Not specified directly for the testing. It's implied that the testing was conducted by Evenflo Feeding, Inc. or their designated testing facilities.
- Retrospective or Prospective: Not applicable or specified for this type of device testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Ground truth as understood for AI/ML clinical studies (e.g., radiologist reads) is not relevant for the testing of a breast pump's mechanical and electrical performance. The "ground truth" here is the design specification or established engineering standard.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are typically used in clinical studies with human readers or image analysis, not for mechanical/electrical testing of a breast pump.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a breast pump, not an AI system intended to assist human readers in, for example, diagnosing medical conditions from images.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device does not contain an AI/ML algorithm for diagnostic purposes. Its software is for operational control (e.g., suction/speed settings).
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. The "ground truth" for this device's performance testing is based on engineering specifications, adherence to international standards (e.g., ISO, IEC), and functional requirements for a breast pump (e.g., vacuum levels, backflow prevention).
8. The sample size for the training set
Not applicable. This document is for a physical medical device (breast pump), not an AI/ML model that requires training data.
9. How the ground truth for the training set was established
Not applicable. (See #8)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
November 3, 2023
Evenflo Feeding, Inc. Michael Sewak Director of Quality 9277 Centre Pointe Drive. Suite 160 West Chester, OH 45069
Re: K230481
Trade/Device Name: Evenflo Premium Double Electric Breast Pump (Model 4018) Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: October 1, 2023 Received: October 3, 2023
Dear Michael Sewak:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K230481
Device Name
Evenflo Premium Double Electric Breast Pump (Model 4018)
Indications for Use (Describe)
The Eventlo Premium Double Electric Breast Pump (Model 4018) is an electrically powered suction device intended to express and collect milk from a lactating woman's breasts. It is intended for a single user.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K230481 Page 1 of 4
510(k) Summary - K230481
1. Submitter Information
| Applicant: | Evenflo Feeding, Inc. |
|---|---|
| Phone: | (513) 870-1620 |
| Address: | 9277 Centre Pointe Drive, Suite 160 |
| West Chester, OH 45069. |
2. Correspondent Information
| Contact: | Michael Sewak |
|---|---|
| Director of Quality | |
| Email: | Michael.sewak@kcc.com |
| Phone: | Office Phone: (513) 745-4625 |
| Mobile Phone: (513) 773-8188 |
3. Date prepared: November 1, 2023
4. Device Information
| Device Name: | Evenflo Premium Double Electric Breast Pump (Model 4018) |
|---|---|
| Common Name: | Powered breast pump |
| Regulation Number: | 21 CFR 884.5160 |
| Regulation Name: | Powered Breast Pump |
| Product Code: | HGX (Pump, Breast, Powered) |
| Regulatory Class: | Class II |
5. Predicate Device Information
| Device Name: | Evenflo Model 2951 Advanced Double Electric Breast Pump |
|---|---|
| 510(k) Number: | K131153 |
| Manufacturer: | Evenflo Feeding Inc. |
The predicate device has not been subject to a design-related recall.
6. Device Description
The Evenflo Electric Breast Pump (Model 4018) is an electrically powered breast pump to be used in a home environment by a single user. The device is provided non-sterile and should be cleaned and disinfected prior to first use according to the instructions for use. The breast pump can be used on one breast (single pumping) or on both breasts at the same time (double pumping).
The device consists of a pump unit, AC power adapter, power cable, AdvancedFit flange kit and bottle kit. The AdvancedFit flange kit consists of a flange body, flange insert, check valve (valve and membrane), diaphragm, diaphragm cap, and tubing for each breast, a tubing adapter to connect the kit to the pump unit and a tubing adapter plug to convert for single pumping. It does not incorporate off-the-shelf (OTS) software or wireless technology/mobile app software functionality. The pump is powered by an internal, non-replaceable, rechargeable lithium-ion battery which is charged using the included AC power supply and cable. The subject device can be operated while plugged into AC power.
The breast pump uses cyclic negative pressure (suction) to mimic the suckling patterns of a feeding infant. A DC motor drives a membrane vacuum pump to generate the suction required to stimulate and express breast
{4}------------------------------------------------
milk. The timing of this pattern is dependent upon the suction/speed settings selected by the user and is preprogrammed in the device. The device is capable of producing peak suction levels between -50 and -250 mmHg in double pumping mode and between -85 and -250 mmHg in single pumping modes at speeds between 30 and 80 cycles per minute. There are 9 distinct levels of suction and 6 speeds available for each suction level.
The breast pump expresses breast milk by creating a seal around the nipple using a flange and applying and releasing suction to the nipple. The milk is collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system.
All other components (i.e., motor unit) of the subject device are not in contact with the breast. All milk contacting components are compliant with 21 CFR 174-179.
7. Indications for Use
The Evenflo Premium Double Electric Breast Pump (Model 4018) is an electrically powered suction device intended to express and collect milk from a lactating woman's breasts. It is intended for a single user.
8. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
| Evenflo PremiumDouble Electric BreastPump, Model 4018K230481Subject Device | Evenflo Model 2951Advanced DoubleElectric Breast PumpK131153Predicate Device | Comparison | |
|---|---|---|---|
| Product Code | HGX | HGX | Same |
| Regulation Number | 21 CFR 884.5160 | 21 CFR 884.5160 | Same |
| Regulatory Class | Class II | Class II | Same |
| Patient Population | Lactating Women | Lactating Women | Same |
| Indications for Use | The Evenflo Premium DoubleElectric Breast Pump (Model4018) is an electrically poweredsuction device intended toexpress and collect milk from alactating woman's breasts. It isintended for a single user. | The Evenflo Advanced DoubleElectric Breast Pump is anelectrically powered suctiondevice intended to express andcollect milk from lactatingwoman's breasts. It is intendedfor a single user. | Same Intended Use |
| Pump Options | Single or Double | Single or Double | Same |
| Cycling controlmechanism | Microcontroller | Microcontroller | Same |
| Backflow Protection | Yes | Yes | Same |
| Mobile Application | No | No | Same |
| Indicators | Yes, (2) 7-segment Digits(Suction & Speed) | (12) Discrete LEDs(8 Suction, 4 Speed) | Different. Differences in userinterface/indicators do not raisedifferent questions of safety and |
Table 1: Comparator Table for Subject and Predicate Devices
{5}------------------------------------------------
| effectiveness. | |||
|---|---|---|---|
| Single User | Yes | Yes | Same |
| Media separation(backflowprotection) | Yes | Yes | Same |
| Expression pattern | 2-Phase | 2-Phase | Same |
| Suction levels(Single pumping) | 85-250 mmHg | 25-270 mmHg | Different. Differences in massagesuction levels do not raisedifferent questions of safety andeffectiveness |
| Suction levels(Double pumping) | 50-250 mmHg | 25-254 mmHg | Different. Differences inpumping suction levels do notraise different questions of safetyand effectiveness |
| Cycles per Minute | 30-80 cpm | 30-80 cpm | Same |
| Suction levels | 9 vacuum levels | 8 vacuum levels | Different. Differences inavailable suction levels do notraise different questions of safetyand effectiveness. |
| User Interface | 8 pushbuttons | 4 pushbuttons | Different. Differences in userinterface do not raise differentquestions of safety andeffectiveness. |
| Adjustable SuctionLevels | Yes | Yes | Same |
| Design | Tabletop Milk Collector andFlange | Tabletop Milk Collector andFlange | Same |
| IncludedAccessories | Flanges, Soft Flange Inserts,Bottles, Tubing Check valves,Diaphragms, Bottle Stands | Flanges, Soft Flange Inserts,Bottles, Tubing Check valves,Diaphragms | Similar |
| Power Supply | 7.4 VDC 2200 mAh Battery | 6 X AA (LR6) Primary Battery | Different. Differences inavailable power supply do notraise different questions of safetyand effectiveness as performanceis independently assessed foreach available powerconfiguration. |
The indications for use of the subject and predicate device are similar, and they have the same intended use (i.e., for collection of breast milk from the breasts of lactating women).
The subject and predicate devices have similar technological features, including available accessories, cycle speed levels, and overall design. However, as shown in the table above, there are technological differences between the subject and predicate device, including different overall vacuum specifications, power supply, user interface, and vacuum levels. The different technological characteristics of the subject device, as compared to the predicate device, do not raise different questions of safety and effectiveness.
9. Summary of Non-Clinical Performance Testing
Biocompatibility
Biocompatibility studies on the patient-contacting components of the subject device, including irritation, cytotoxicity, and sensitization testing were performed in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical
{6}------------------------------------------------
Devices - Part 1: Evaluation and testing within a risk management process", as follows:
- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010) ●
- Irritation (ISO 10993-10:2010) .
The user-contacting materials were shown to be non-cytotoxic, non-irritating, and non-sensitizing.
Electrical Safety
Testing was conducted in accordance with ANSI/AAMI ES60601- 1:2005/A2:2010 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance), IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems, and IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
Electromagnetic Compatibility
Testing was conducted in accordance with IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
Software
Software was evaluated for a minor level of concern as recommended in the 2005 FDA guidance document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Performance Testing
Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:
- Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications.
- . Backflow protection testing was conducted to verify liquid does not backflow into the tubing.
- . Use life testing was conducted to demonstrate that the device maintains its specifications throughout its proposed use life.
- . Battery performance testing was conducted to demonstrate that the battery remains functional during its stated battery use-life.
- . Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.
10. Conclusion
The results of the performance testing described above demonstrate that the Evenflo Premium Double Electric Breast Pump (Model) 4018 is as safe and effective as the predicate device and supports a determination of substantial equivalence.
§ 884.5160 Powered breast pump.
(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).