K Number
K223890
Manufacturer
Date Cleared
2023-06-16

(171 days)

Product Code
Regulation Number
876.5980
Panel
GU
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NAJA Gastrointestinal Balloon Catheter is intended to administer contrast medium to the small intestine.

Device Description

The NAJA Gastrointestinal Balloon Catheter ("NAJA device") is a single-use, single-operator, over the-wire, double balloon catheter that is compatible with 0.035" guidewires. The device is introduced orally and designed for temporary occlusion of the gastrointestinal (GI) lumen (small intestine), allowing infusion of fluids for enteroclysis studies (i.e., imaging tests of the small intestine). The device is designed to fit inside the working channel of a gastroscope and reaches its intended location by tracking over a 0.035" guidewire.

The NAJA device features a separate hub connector that is attachable and detachable from the catheter. This feature allows removal of the gastroscope from the catheter once the balloons are in the correct location and before the hub connector is attached for balloon inflation and infusion of fluids into the GI lumen. The hub connector is supplied with 1-way stopcocks on the inflation and infusion ports. The NAJA device is also provided with a 30cc syringe for balloon inflation and infusion of fluids. The balloons are inflated independently using air.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the NAJA Gastrointestinal Balloon Catheter, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The FDA 510(k) summary details various design verification tests. The "Specification" column represents the acceptance criteria, and the "Test Result" column indicates if the device met that criterion (mostly "Pass" or a specific measurement).

Design Verification TestAcceptance CriteriaReported Device Performance
Biocompatibility Testing
Cytotoxicity StudyNo evidence of causing cell lysis or toxicity; grade less than 2 (mild reactivity) per ISO 10993-5:2009.Pass
ISO Intracutaneous StudyDifference between test article extract overall mean score and corresponding control extract overall mean score was 0.0 per ISO 10993-23:2021.Pass
ISO Maximization Sensitization StudyNo evidence of causing delayed dermal contact sensitization in guinea pig per ISO 10993-10:2021.Pass
Bench and Performance Testing (Package)
Legibility of MarkingsIFU, pouch label, and carton label are legible.Pass
Seal WidthPouch seal width is ≥ 0.19".Pass
Pouch Seal Visual InspectionPouch seals are free of channels or other damages that could impair sterility.Pass
Bubble Emission TestNo evidence of leakage through any surface of pouch.Pass
Pouch Seal StrengthPouch seal can withstand a force ≥ 3.35N over a seal length of 25.4mm.Pass
Pouch Ease of OpeningPouch is easily opened with gloved hands, while keeping internal contents stable. Components inside pouch are easily removable from backboard.Pass
Bench and Performance Testing (Component & Device)
Component Visual InspectionDevice components are free of damages that could affect performance of product (kinks, cracks, separated components, etc.).Pass
Surface DefectsAll surfaces and open channels of device are free of loose foreign matter, pores, cracks, and remainders of tooling agents.Pass
Atraumatic TipDistal tip is smooth, rounded, and free from sharp edges.Pass
Atraumatic Infusion HolesInfusion holes are smooth and free from sharp edges.Pass
Radiopaque MarkersFive (5) markers are present on device.Pass
0.035" GW Loading - Front/BackDevice accommodates front- and back-loading of a 0.035" GW.Pass
Proximal Extrusion DimensionsDimensions suitable for insertion into hub connector.Pass
10.5Fr Channel CompatibilityDistal 30cm of device fits through a 10.5Fr channel.Pass
Working LengthDevice working length is 2360 ± 20mm.Pass
Tip LengthTip length is 20 ± 2mm.Pass
Scaffold LengthScaffold length is 120 ± 2mm.Pass
Catheter Outer DiameterOuter diameter is ≤ 3.5mm for the catheter shaft, balloon welds, distal tip.Pass
Simulated Use Testing
Number of Balloons and SizeDevice has two balloons that can be inflated to a diameter of 50mm.Pass
Hub Connector DetachabilityHub connector is detachable from device and can be attached for fluid delivery.Pass
Shaft Insertion FeatureShaft marker is no longer visible once inserted into hub connector.Pass
Gastroscope CompatibilityDevice fits through working channel of gastroscope.Pass (implied, as other aspects of simulated use passed)
Simulated Use TestingThe device remains functional throughout the lifecycle of one clinical procedure.Pass (implied, as other aspects of simulated use passed)
Unique Hub Connector IdentificationProximal terminations of hub assembly are appropriately identified.Pass (implied, as other aspects of simulated use passed)
Legibility of Device MarkingsWording on hub assembly labels is legible.Pass (implied, as other aspects of simulated use passed)
Functional Testing
Hub Connector and Hub Assembly PressurePressure decay specification is met.Pass
Balloon Pressure and CrosstalkPressure decay specification is met.Pass
Balloon OD/L Ratio – unconfinedUnconfined balloon shape specification is met.Pass
Balloon Diameter and Inflation Volume RelationshipRelationship between balloon diameter and inflation volume shall be determined.Balloon Compliance: 9mL @ 20mm OD, 20mL @ 30mm OD, 44mL @ 40mm OD, 76mL @ 50mm (MAX)
Balloon OD/L Ratio – confinedConfined balloon shape specification is met.Pass
Inflation TimeInjection of 30mL of air in ≤ 4s.Pass
Deflation TimeRemoval of 30mL of air in ≤ 17s.Pass
Infusion TimeInfusion of 30mL of water in ≤ 19s.Pass
Channel OcclusionDevice is able to occlude channels 20-50mm.Pass
Destructive Testing
Rated Burst VolumeBalloons have a rated burst volume (RBV) ≥ 81.5mL.Pass
Bend RadiusNo shaft kink and/or fracture when wrapped 180° around a 20mm bend radius.Pass
Shaft Buckling ForceShaft buckle force specification is met.Pass
Hub to Shaft Slip ForceHub connector slip force specification is met.Pass
Tensile – Proximal Shaft to ExtrusionTensile force of proximal shaft to extrusion is ≥ 15N.Pass
Tensile - Distal Shaft to ExtrusionTensile force of distal shaft to extrusion is ≥ 15N.Pass
Tensile - Extrusion at ScaffoldTensile force of scaffold shall be ≥ 15N.Pass
Shear – Balloon Distal NeckShear force of distal balloon neck to extrusion bond is ≥ 15N.Pass
Peel - Balloon Proximal NeckPeel force of proximal balloon neck to extrusion bond is ≥ 15N.Pass
Tensile - Extension Tube to Hub ConnectorTensile force of extension tube to hub connector bond is ≥ 15N.Pass
Tensile - Luer to Extension TubeTensile force of luer to extension tube bond is ≥ 15N.Pass

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes bench and performance testing (also referred to as design verification testing), and biocompatibility testing. These are typical for medical device submissions and do not involve human subjects or clinical data in the form of a "test set" from patients.

  • Sample Size: Not explicitly stated for each individual bench or biocompatibility test, but device verification testing typically uses a statistically relevant number of samples to ensure robust performance across manufacturing batches.
  • Data Provenance: The tests are conducted in a laboratory setting using manufactured devices. There is no information regarding country of origin of data in terms of patient demographics, nor is it retrospective or prospective in a clinical sense.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This question is not applicable to the type of studies presented here.

  • The studies are engineering and laboratory tests (bench, performance, and biocompatibility). They do not involve interpretation of medical images or patient data that would require "experts" to establish "ground truth" in a clinical context. The "ground truth" for these tests is defined by the technical specifications and standards (e.g., ISO, internal design specifications).

4. Adjudication Method for the Test Set

This question is not applicable. As explained above, there is no "test set" in a clinical or diagnostic sense requiring human adjudication. The results are based on objective measurements against engineering specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The submission is a 510(k) for a physical medical device (catheter), not a diagnostic algorithm or AI-powered device. The studies focus on the physical and functional aspects of the catheter itself, not its impact on human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm performance study was not done. This device is a physical medical instrument, not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for the tests performed is defined by:

  • Engineering Specifications: Numerical ranges, physical properties, and functional requirements established during device design and development (e.g., "Pouch seal width is ≥ 0.19"").
  • International Standards: Specifically, ISO standards for biocompatibility (e.g., ISO 10993-5, ISO 10993-23, ISO 10993-10).
  • Observed Performance: Direct observation of device functionality and integrity (e.g., "Device components are free of damages").

These are objective, measurable criteria, not subjective interpretations requiring clinical ground truth like pathology, expert consensus, or outcomes data.

8. The Sample Size for the Training Set

This question is not applicable. The NAJA Gastrointestinal Balloon Catheter is a physical medical device, not an AI or machine learning model that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

This question is not applicable. As stated above, there is no "training set" for this type of device.

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.