K Number
K223839
Date Cleared
2023-05-22

(151 days)

Product Code
Regulation Number
892.5050
Panel
RA
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TrueBeam-TrueBeam STx-Edge: The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere radiation therapy is indicated for adults and pediatric patients. The TrueBeam, TrueBeam STx, and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, nasopharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carvinoma), gentourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, avernous malformation, and skull base tumors), and medically refractory essential tremor (indicated for adults only). VitalBeam: VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pitutary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal turnors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin) and pediatric tumors (such as glioma, pituitary tumors, hemangioblastoma, craniopharyngioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors).

Device Description

The TrueBeam and VitalBeam Radiotherapy System is a medical linear accelerator that delivered therapeutic radiation to patient in accordance with the physician's prescription. The system consists of two major components – a photon, electron and diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located outside the treatment room.

AI/ML Overview

The provided text is a 510(k) Premarket Notification summary for the Varian Medical Systems TrueBeam™, TrueBeam STx™, Edge™, and VitalBeam radiotherapy delivery systems. This submission primarily focuses on expanding the indications for use to include "medically refractory essential tremor" for the TrueBeam™, TrueBeam STx™, and Edge™ Systems.

The document states that the subject device has the same technological characteristics as its previously cleared predicate device (K213977). Therefore, the provided text does not describe a new study to prove the device meets acceptance criteria for a newly designed or significantly modified product. Instead, it relies on the previously established performance, safety, and effectiveness of the predicate device and a literature review to support the expanded indication.

Here's an analysis based on the information provided, recognizing that it's a 510(k) for an expanded indication rather than a de novo submission for a new device:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not provide a specific table of acceptance criteria with numerical performance targets (e.g., accuracy, precision) for the device itself and its reported performance as these were established during the clearance of the predicate device (K213977).

Instead, the "Summary of Performance Testing" for the current submission K223839 states:

CriterionReported Performance
Performance SpecificationsEquivalent to predicate device (K213977)
Principle of OperationSame as predicate device (K213977)
Safety ProfileNo changes demonstrated by verification, validation, and safety standards testing (for predicate device K213977)
Expanded Indication (medically refractory essential tremor) for TrueBeam™, TrueBeam STx™, Edge™Literature review shows similar outcomes and complication rates for SRS treatment of tremor conditions, comparable to traditional surgical options and Gamma Knife. Linear accelerator-based SRS is reported to be safe and efficacious for this indication when other options are unavailable/contraindicated.

The "acceptance criteria" here implicitly refer to the continued safety and effectiveness of the device for its core functions, as established by the predicate device, and the literature-supported safety and efficacy of Stereotactic Radiosurgery (SRS) for medically refractory essential tremor when delivered by a linear accelerator.

2. Sample Size Used for the Test Set and Data Provenance:

  • For the device's core performance: Not applicable to this specific submission, as it relies on previous testing of the predicate device (K213977).
  • For the expanded indication (medically refractory essential tremor): The "Summary of Clinical Data" mentions a literature review of over 1300 patients treated with SRS for tremor conditions.
    • Data Provenance: The data comes from published literature, implying a retrospective compilation of studies that likely originated from various countries and institutions. The document does not specify particular countries or whether the studies were prospective or retrospective, but typical clinical literature reviews will encompass both.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • For the device's core performance: Not applicable to this specific submission as it relies on the predicate.
  • For the expanded indication: The "ground truth" for the efficacy and safety of SRS for tremor conditions is established through published clinical outcomes. The decision to include "medically refractory essential tremor" as an indication is based on a review of these published studies. The document does not explicitly state how many experts conducted this literature review for Varian, nor their specific qualifications. However, such reviews are typically conducted by regulatory affairs professionals, sometimes with input from clinical experts (e.g., radiation oncologists, neurosurgeons) who would understand the clinical context and outcomes.

4. Adjudication Method for the Test Set:

Not applicable. This submission relies on a literature review of outcomes for an existing therapy, not on a new test set requiring multi-reader adjudication of images or data.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done as described in this document. The submission is not about human readers improving with AI assistance; it's about the safety and efficacy of a medical linear accelerator system for an expanded treatment indication.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

Not applicable. This device is a radiotherapy delivery system, a hardware and software system that always involves human operators (radiation oncologists, medical physicists, radiation therapists) in its use for treatment planning and delivery. It is not an "algorithm-only" diagnostic AI device.

7. Type of Ground Truth Used:

  • For the device's core performance: Not explicitly stated in this document but would typically involve physical measurements, phantom studies, and possibly clinical outcome data from previous device versions/studies.
  • For the expanded indication (medically refractory essential tremor): Clinical outcomes data from a literature review. This includes reported tremor improvement and complication rates from published studies using both linear accelerators and Gamma Knife for SRS treatment of tremor.

8. Sample Size for the Training Set:

Not applicable. This is not an AI/ML device that requires a training set in the typical sense. The "training" for the device's functionality would have been part of its original development and testing (for the predicate device), but this submission does not describe an AI training process.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no mention of a training set for an AI/ML model in this submission.

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.