Intended Use
Bonelogic software is to be used by orthopaedic healthcare professionals for diagnosis and surgical planning in a hospital or clinic environment. Bonelogic software provides: - Semi-automatic segmentation with manual or assisted input of bony structure identification from CT imaging input, - Three-dimensional mathematical models of the anatomical structures of foot and ankle, - Measurement templates containing radiographic measures of foot and ankle, and tools for manually obtaining linear and angular measurements, - Surgical planning application for foot and ankle using three-dimensional models of the anatomical structures and radiographic measures. The three-dimensional models of the anatomical structures combined with the measurements can be used for the diagnosis of orthopaedic healthcare conditions. The surgical planning application containing the three-dimensional structural models combined with the measurements can be used for the planning of treatments and operations to correct orthopaedic healthcare conditions of foot and ankle.
Device Description
The Bonelogic is a software tool that segments bone anatomy using dedicated semiautomatic tools and fully automatic algorithms. More specifically, Bonelogic is intended to segment foot and ankle bones from computed tomography (CT) images. The segmented structures may then be used to create 3D models of their respective bones and replicate the anatomy of a patient. The semi-automatic tools of the software require a healthcare professional to mark the different bones in an initial 3D rendered model prior to when the segmentation process is initialized. This method is called the semi-automatic workflow with manual input. The software also comprises an optional semiautomatic workflow with assisted input that replaces the required user input with an estimate based on a locked artificial neural network (ANN) model. The fully automatic algorithm processes the final result in the same way based on input generated by the semi-automatic workflow with user input and with ANN model. The user still needs to mark the laterality and as a new step acknowledge the bones discovered by the ANN model.
More Information

Not Found

Yes
The device description explicitly states that the software comprises an optional semiautomatic workflow with assisted input that replaces required user input with an estimate based on a locked artificial neural network (ANN) model. ANN is a type of machine learning algorithm.

No.
The Bonelogic software is intended for diagnosis and surgical planning, not for providing direct therapy or treatment.

Yes

The "Intended Use / Indications for Use" section explicitly states that the "Bonelogic software is to be used by orthopaedic healthcare professionals for diagnosis and surgical planning" and that "The three-dimensional models of the anatomical structures combined with the measurements can be used for the diagnosis of orthopaedic healthcare conditions."

Yes

The device is explicitly described as a "software tool" and its function is entirely based on processing and analyzing medical images and providing outputs for diagnosis and surgical planning. There is no mention of any associated hardware components that are part of the device itself.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening.
  • Bonelogic's Function: Bonelogic software processes medical images (CT scans) of the human body. It does not analyze biological specimens.
  • Intended Use: The intended use clearly states it's for "diagnosis and surgical planning" using imaging data, not laboratory analysis of samples.

Therefore, while Bonelogic is a medical device used for diagnosis, it falls under the category of medical imaging software rather than an In Vitro Diagnostic device.

No
The letter does not explicitly state that the FDA has reviewed and approved or cleared a PCCP for this specific device.

Intended Use / Indications for Use

Bonelogic software is to be used by orthopaedic healthcare professionals for diagnosis and surgical planning in a hospital or clinic environment.
Bonelogic software provides:

  • Semi-automatic segmentation with manual or assisted input of bony structure identification from CT imaging input,
  • Three-dimensional mathematical models of the anatomical structures of foot and ankle,
  • Measurement templates containing radiographic measures of foot and ankle, and tools for manually obtaining linear and angular measurements,
  • Surgical planning application for foot and ankle using three-dimensional models of the anatomical structures and radiographic measures.
    The three-dimensional models of the anatomical structures combined with the measurements can be used for the diagnosis of orthopaedic healthcare conditions. The surgical planning application containing the three-dimensional structural models combined with the measurements can be used for the planning of treatments and operations to correct orthopaedic healthcare conditions of foot and ankle.

Product codes

QIH

Device Description

The Bonelogic is a software tool that segments bone anatomy using dedicated semiautomatic tools and fully automatic algorithms. More specifically, Bonelogic is intended to segment foot and ankle bones from computed tomography (CT) images. The segmented structures may then be used to create 3D models of their respective bones and replicate the anatomy of a patient. The semi-automatic tools of the software require a healthcare professional to mark the different bones in an initial 3D rendered model prior to when the segmentation process is initialized. This method is called the semi-automatic workflow with manual input. The software also comprises an optional semiautomatic workflow with assisted input that replaces the required user input with an estimate based on a locked artificial neural network (ANN) model. The fully automatic algorithm processes the final result in the same way based on input generated by the semi-automatic workflow with user input and with ANN model. The user still needs to mark the laterality and as a new step acknowledge the bones discovered by the ANN model.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Yes

Input Imaging Modality

CT imaging

Anatomical Site

foot and ankle

Indicated Patient Age Range

Not Found

Intended User / Care Setting

orthopaedic healthcare professionals for diagnosis and surgical planning in a hospital or clinic environment.

Description of the training set, sample size, data source, and annotation protocol

The AI algorithm for bone identification was developed using 145 CT image studies and metal identification was developed using 130 CT image studies. The 357 CT image studies were from individual patients with few studies from same patient with different foot alignments. The CT image series' were collected from various sites across USA and Europe with a minimum of 50% of the images originating from the USA.
The ground truths for bone and metal identification were independently established by three (3) U.S. Orthopedic surgeons with a 3rd party software. Each clinicians reviewed each of the DICOM series through axial/sagittal/coronal views and/or 3D reconstruction and marked on a spreadsheet the presence of a bone and metal in the image series. Based on the majority vote of three, two same responses were required to establish a ground truth on each of the DICOM series.
The Bonelogic software machine learning algorithm training and tuning data used during the algorithm development were independent data sets. Each CT image study was allowed to be allocated to only data set.

Description of the test set, sample size, data source, and annotation protocol

Testing was carried out using 82 CT image studies. The CT image series' were collected from various sites across USA and Europe with a minimum of 50% of the images originating from the USA.
The CT image studies were from patients with different ages and racial groups, with minimum of 35% male/female within each dataset, with mean age approximately 47 years (SD 15 years), and representatives from White, (Non-)Hispanic. African American, and Native racial groups. Each dataset was balanced in terms of subjects with different foot alignment, demographics, imaging devices and with subjects from clinical subgroups ranging from control/normal feet (44% with test data) to pre-/post-operative clinical conditions such as Hallux Valgus, Progressive Collapsing Foot Deformity, fractures, or with metal implants (40% of the test data).
The 82 images in the test data were acquired using Curvebeam PedCAT (61%), Planmed Verify (11%), and Carestream OnSight 3D Extremity (26%), and Toshiba Somatom. Typical imaging protocol is disclosed within the IFU, however, the test data contains wider range of parameters for generalization (tube voltages between 90-120 kV, tube currents 5-8 mA, and slice thickness/pixel spacing 0.37-1.5mm).
The ground truths for bone and metal identification were independently established by three (3) U.S. Orthopedic surgeons with a 3rd party software. Each clinicians reviewed each of the DICOM series through axial/sagittal/coronal views and/or 3D reconstruction and marked on a spreadsheet the presence of a bone and metal in the image series. Based on the majority vote of three, two same responses were required to establish a ground truth on each of the DICOM series.
The test data used in the standalone software performance assessment study was an independent data set. Each CT image study was allowed to be allocated to only data set.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Software verification and validation were carried out based on the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" at the unit, integration, and system levels to determine substantial equivalence to the predicate device. The predicate device meets the subject device's established acceptance criteria of 95% model conformance within 1.0mm distance to reference model and 2.0 degrees standard deviation for angular measurements.
Bench testing Software verification and validation for Bonelogic software demonstrated its substantial equivalence to the predicate device. An additional clinical data based software performance assessment study was carried out to validate the standalone performance of AI algorithms from a clinical perspective. The testing for 82 CT image series presented 100% correctly identified bones of foot and ankle.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

specificity 98%, sensitivity 100% (for metal identification)

Predicate Device(s)

K203290

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Disior Ltd % Alex Cadotte Associate Director, Software and Digital Health Regulatory Affairs Mcra. LLC 1050 K St NW Suite 1000 Washington, DC 20001

Re: K223757

December 8, 2023

Trade/Device Name: Bonelogic Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: November 7, 2023 Received: November 13, 2023

Dear Alex Cadotte:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

1

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

2

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb Assistant Director Iamging Sofware Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

3

Indications for Use

510(k) Number (if known) K223757

Device Name Bonelogic 2.2

Indications for Use (Describe)

Bonelogic software is to be used by orthopaedic healthcare professionals for diagnosis and surgical planning in a hospital or clinic environment.

Bonelogic software provides:

  • · Semi-automatic segmentation with manual or assisted input of bony structure identification from CT imaging input,
  • · Three-dimensional mathematical models of the anatomical structures of foot and ankle,

• Measurement templates containing radiographic measures of foot and tools for manually obtaining linear and angular measurements,

• Surgical planning application for foot and ankle using three-dimensional models of the anatomical structures and radiographic measures.

The three-dimensional models of the anatomical structures combined with the measurements can be used for the diagnosis of orthopaedic healthcare conditions. The surgication containing the three-dimensional structural models combined with the measurements can be used for the planning of treatments and operations to correct orthopaedic healthcare conditions of foot and ankle.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

4

510(k) Summary

Device Trade Name:Bonelogic 2.2
Manufacturer:Disior Ltd
HTC Helsinki, Building PINTA, 4th floor
Tammasaarenkatu 3
Helsinki, Finland 00180
Contact:Aarno Jussila
Director of Enabling Technologies
Phone: +358 40 6734939
Email: aarno@disior.com
Prepared by:Alex Cadotte
MCRA, LLC
803 7th St NW
Washington, DC 20001
Office: 202.552.5800
Date Prepared:December 7, 2023
Classifications:21 CFR 892.2050 Medical image management and processing
system.
Class:II
Product Code:QIH
Primary Predicate:K203290

Indications For Use:

Bonelogic software is to be used by orthopaedic healthcare professionals for diagnosis and surgical planning in a hospital or clinic environment.

Bonelogic software provides:

  • Semi-automatic segmentation with manual or assisted input of bony structure ● identification from CT imaging input,
  • Three-dimensional mathematical models of the anatomical structures of foot and ankle, ●
  • Measurement templates containing radiographic measures of foot and ankle, and tools ● for manually obtaining linear and angular measurements,
  • Surgical planning application for foot and ankle using three-dimensional models of the . anatomical structures and radiographic measures.

The three-dimensional models of the anatomical structures combined with the measurements can be used for the diagnosis of orthopaedic healthcare conditions. The surgical planning application containing the three-dimensional structural models combined with the

5

measurements can be used for the planning of treatments and operations to correct orthopaedic healthcare conditions of foot and ankle.

Device Description:

The Bonelogic is a software tool that segments bone anatomy using dedicated semiautomatic tools and fully automatic algorithms. More specifically, Bonelogic is intended to segment foot and ankle bones from computed tomography (CT) images. The segmented structures may then be used to create 3D models of their respective bones and replicate the anatomy of a patient. The semi-automatic tools of the software require a healthcare professional to mark the different bones in an initial 3D rendered model prior to when the segmentation process is initialized. This method is called the semi-automatic workflow with manual input. The software also comprises an optional semiautomatic workflow with assisted input that replaces the required user input with an estimate based on a locked artificial neural network (ANN) model. The fully automatic algorithm processes the final result in the same way based on input generated by the semi-automatic workflow with user input and with ANN model. The user still needs to mark the laterality and as a new step acknowledge the bones discovered by the ANN model.

Predicate Device:

Disior submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices. K223757 is substantially equivalent in indications, design principles, and performance to the following subject device:

Predicate Device K203290Subject Device K223757
Information(Bonelogic)(Bonelogic 2.2)Comparison
Classification NameMedical image management
and processing systemMedical image management
and processing systemIdentical
Service TypeSoftwareSoftwareIdentical
Classification21 CFR 892.205021 CFR 892.2050Identical
ClassIIIIIdentical
Product CodeLLZQIHIdentical
Indications for UseBonelogic software is intended
to be used by specialized
medical practitioners to assist
in the characterization of
human anatomy with 3D
visualization and specific
measurements. The medical
image modalities intended to
be used in the software are
computed tomography (CT)
images, cone beam computed
tomography (CBCT) images
and weight-bearing cone beam
CT (WBCT) images. The
intended patient population is
adults over 16 years of age.
Bonelogic software contains
the measurement templateBonelogic software is to be
used by orthopaedic healthcare
professionals for diagnosis and
surgical planning in a hospital
or clinic environment.
Bonelogic software provides:
Semi-automatic segmentation
with manual or assisted input
of bony structure identification
from CT imaging input,
Three-dimensional
mathematical models of the
anatomical structures of foot
and ankle,
Measurement templates
containing radiographic
measures of foot and ankle,
and tools for manuallyRemoval of wrist
and hand related
functionality and
addition of an
optional
semiautomatic
segmentation
workflow with
assisted bone
identification
process in the
subject device

Substantial Equivalence:

6

| Information | Predicate Device K203290
(Bonelogic) | Subject Device K223757
(Bonelogic 2.2) | Comparison |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| | with a set of distance and
angular measures. The
measurements can be used for
diagnostic purposes. The
three-dimensional (3D)
models are displayed and can
be manipulated in the
software. Together, the
information from the
measurements and the 3D
visualization can be used for
treatment planning in the field
of orthopedics (foot and ankle,
and hand and wrist). The 3D
models can be outputted from
the software for traditional or
additive manufacturing. The
physical models generated
based on the 3D digital models
are not intended for diagnostic
use. | obtaining linear and angular
measurements,
Surgical planning application
for foot and ankle using three-
dimensional models of the
anatomical structures and
radiographic measures.
The three-dimensional models
of the anatomical structures
combined with the
measurements can be used for
the diagnosis of orthopaedic
healthcare conditions. The
surgical planning application
containing the three-
dimensional structural models
combined with the
measurements can be used for
the planning of treatments and
operations to correct
orthopaedic healthcare
conditions of foot and ankle. | |
| Input | Computed tomography
DICOM images | Computed tomography
DICOM Computed
tomography | Identical |
| Image processing | Segmentation of bone
structures | Segmentation of bone
structures | Identical |
| Output | 3D model of patient anatomy | 3D model of patient anatomy | Identical |
| Measuring and
planning | Perform measurements for
presurgical planning | Perform measurements for
presurgical planning | Identical |
| Bone identification | Manual process | Manual and an optional
semiautomatic workflow with
assisted input process
performed by artificial neural
network | Optional
semiautomatic
bone
identification
process in |

Performance Testing Summary:

  • Software verification and validation were carried out based on the "Guidance for . the Content of Premarket Submissions for Software Contained in Medical Devices" at the unit, integration, and system levels to determine substantial equivalence to the predicate device. The predicate device meets the subject device's established acceptance criteria of 95% model conformance within 1.0mm distance to reference model and 2.0 degrees standard deviation for angular measurements.
  • Bench testing Software verification and validation for Bonelogic software ● demonstrated its substantial equivalence to the predicate device. An additional clinical data based software performance assessment study was carried out to validate the standalone performance of AI algorithms from a clinical perspective. The testing for 82 CT image series presented 100% correctly identified bones of foot and ankle. The existence of metal was identified correctly for 98.8% of the images (specificity 98%, sensitivity 100%).

7

Study subjects:

.

The AI algorithm for bone identification was developed using 145 CT image studies and metal identification was developed using 130 CT image studies. Testing was carried out using 82 CT image studies. Out of 357 CT image studies, 340 were from individual patients with few studies from same patient with different foot alignments. The CT image series' were collected from various sites across USA and Europe with a minimum of 50% of the images originating from the USA. The CT image studies were from patients with different ages and racial groups, with minimum of 35% male/female within each dataset, with mean age approximately 47 years (SD 15 years), and representatives from White, (Non-)Hispanic. African American, and Native racial groups. Each dataset was balanced in terms of subjects with different foot alignment, demographics, imaging devices and with subjects from clinical subgroups ranging from control/normal feet (44% with test data) to pre-/post-operative clinical conditions such as Hallux Valgus, Progressive Collapsing Foot Deformity, fractures, or with metal implants (40% of the test data).

  • Imaging Systems:
    The 357 image studies were collected using CT imaging system made by five (5) manufacturers (7 different models in total). From the test data of 82 images, 61% of the images were acquired using Curvebeam PedCAT, 11% with Planmed Verify, and 26% with Carestream OnSight 3D Extremity. In addition, system test data contains images acquired with Toshiba Somatom. Typical imaging protocol is disclosed within the IFU, however, the test data contains wider range of parameters for generalization (tube voltages between 90-120 kV, tube currents 5-8 mA, and slice thickness/pixel spacing 0.37-1.5mm).

  • Ground Truth: ●
    The ground truths for bone and metal identification were independently established by three (3) U.S. Orthopedic surgeons with a 3rd party software. Each clinicians reviewed each of the DICOM series through axial/sagittal/coronal views and/or 3D reconstruction and marked on a spreadsheet the presence of a bone and metal in the image series. Based on the majority vote of three, two same responses were required to establish a ground truth on each of the DICOM series.

  • Training, Tuning, and Validation Data Independence:
    The Bonelogic software machine learning algorithm training and tuning data used during the algorithm development, as well as test data used in the standalone software performance assessment study, were all independent data sets. Each CT image study was allowed to be allocated to only data set.

8

Conclusion:

The subject device and the predicate devices have intended use and have similar technological characteristics. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. Bonelogic 2.2 is as safe, as effective, and performs as well as, or better, than the predicate devices.