K Number
K223718

Validate with FDA (Live)

Date Cleared
2023-06-05

(175 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HandX is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.

Device Description

The Self-Righting Needle Holder is a single-use, disposable, ETO sterilized instrument for use with the HandX™ device. The Self-Righting Needle Holder is connected to the HandX™ and transmits the motors' rotations to articulate the movement of the Self-Righting Needle Holder's end effector. Self-Righting Needle Holder orients the suturing needle vertically to 90° mounting. It is designed to address the surgeons' needs relating to suturing for various surgical purposes.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information provided, structured as requested:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (Implied from "predefined requirements")Reported Device Performance
Self-Righting Needle Holder Angle TestThe subject device's articulation angles (when connected to HandX device) should be comparable to the predicate device."The subject device tip articulation ability is comparable to the predicate device."
SRNH instrument Dimensional AttributeThe subject device's dimensions must be within the specified requirements."The subject device dimensions are within the defined requirements."
SRNH Suture Holding ForceThe subject device's suture holding force must withstand the defined requirements, comparable to the predicate device."The subject device suture holding force withstands the defined requirements as the predicate device."
SRNH Needle Pulling MomentThe subject device must maintain the needle in place under external moment, meeting the defined requirement."The subject device complies with the needle pulling moment requirement."
SRNH Needle Holding ForceThe subject device must maintain the needle in place under external force, meeting the defined requirement."The subject device complies with the needle holder holding force requirement."
SRNH Needle RightingThe subject device must demonstrate the ability to transition a surgical needle to a "right position" (implicitly defined as 90° mounting, as mentioned in the device description). This implies functionality analogous to the reference device's self-righting capability."The subject device demonstrated transition ability of surgical needles to right position, thus met the predefined requirements." (The device description specifies "Self-Righting Needle Holder orients the suturing needle vertically to 90° mounting," which would be the specific "right position.")

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for each specific bench test. It mentions "The Self-Righting Needle Holder was subjected to bench testing," but does not provide details on the number of units tested.

  • Data Provenance: The data is generated from bench testing conducted by the device manufacturer, Human Xtensions Ltd., to demonstrate that the design outputs meet the design input requirements. The country of origin of the testing is not explicitly stated, but the company is based in Israel. This is prospective testing for the purpose of the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

Not applicable. This evaluation involves bench testing against predetermined engineering and performance specifications, not a clinical study requiring expert assessment of ground truth.

4. Adjudication Method for the Test Set

Not applicable. As described above, the tests are objective bench tests, not assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (surgical instrument), not an AI/imaging diagnostic device. Therefore, an MRMC study is not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical surgical instrument, not an algorithm. The device itself (the needle holder) performs the self-righting function, but its operation is "Electromechanically operated, software controlled" by the HandX™ device. The bench tests evaluate the performance of this physical device.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the bench tests is the "predefined acceptance criteria" derived from engineering specifications and comparisons to the predicate and reference devices. These criteria relate to physical properties and functional performance (e.g., specific angles, force thresholds, dimensional tolerances, and the ability to achieve a "right position").

8. The Sample Size for the Training Set

Not applicable. This is a physical device, not an algorithm that requires a training set. The "software controlled" aspect refers to the HandX™ device controlling the mechanical movement, not an AI/ML algorithm that is "trained."

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 5, 2023

Human Xtensions Ltd. Orit Yaniv VP QA/RA Human Xtensions Ltd. Grand Netter Building 4 Meir Ariel Street Netanya, 4250574, Israel Telephone: + (972) 77 36 30 300 Email: orit@human-x.com

Re: K223718

Trade/Device Name: HandXTM Self-Righting Needle Holder Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: December 11, 2022 Received: December 12, 2022

Dear Orit Yaniv:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.06.05
11:23:01 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

Device Name HandX(TM) Self-Righting Needle Holder

Indications for Use (Describe)

The HandX is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

Premarket Notification 510(k) Summary

Submitter:Human Xtensions Ltd.Grand Netter Building4 Meir Ariel StreetNetanya, 4250574, IsraelTelephone: + (972) 77 36 30 300
Contact:Orit YanivHuman Xtensions Ltd.Grand Netter Building4 Meir Ariel StreetNetanya, 4250574, IsraelTelephone: + (972) 77 36 30 300Email: orit@human-x.com
Date Prepared:May 24, 2023
Subject Device:
Trade/Device Name:Common/Usual Name:Device Classification Name:Product Codes:Product Class:Regulation Numbers:HandX™ Self-Righting Needle HolderArticulated Laparoscopic InstrumentsEndoscope and AccessoriesGCJClass II21 CFR Parts 876.1500
Predicate Devices:HX Device, Needle Holder, Human Xtensions Ltd.(K173919)
Reference Device:ENDOPATH® Self-Righting Needle Holder (K972679)Ethicon. Used as a reference device for Self-Rightingfunction.

Device Description:

The Self-Righting Needle Holder is a single-use, disposable, ETO sterilized instrument for use with the HandX™ device. The Self-Righting Needle Holder is connected to the HandX™ and transmits the motors' rotations to articulate the movement of the Self-Righting Needle Holder's end effector. Self-Righting Needle Holder orients the suturing needle vertically to 90° mounting. It is designed to address the surgeons' needs relating to suturing for various surgical purposes.

Indications for Use:

{4}------------------------------------------------

The HandX™ is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.

Comparison of Technological Characteristics with the Predicate Devices:

The Self-Righting Needle Holder is substantially equivalent to the predicate device, Human Xtensions' HX Device (K173919), in its intended use, contraindications, fundamental technology, packaging, sterilization method and general characteristics. Its Self-Righting function is the same as Ethicon's ENDOPATH® Self-Righting Needle Holder (K972679).

A summary of the technological characteristics of the Self-Righting Needle Holder device in comparison to those of the predicate device and reference device is presented in the table below.

DescriptionHandX™ Self-Righting NeedleHolder(Subject device)HX Device, Needle Holder(Predicate device)ENDOPATH® Self-Righting NeedleHolder(Reference device)
510(k) NumberK223718K173919K972679
Product CodeGCJGCJGAT
CFR21 CFR 876.150021 CFR 876.150021 CFR 876.4730
Indications for UseThe HandX™ is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.The HX Device is intended to assist in the accurate control of HX laparoscopic instruments including needle holder and grasper, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, during laparoscopic surgical procedures. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.The ENDOPATH® Endoscopic Tissue Fastening System (ETFS) is intended for use in minimally invasive surgical applications where soft tissue is being approximated with the interrupted stitches.
Contra- indicationsThis device is intended for use only as indicated.This device is intended for use only as indicated.NA
SpecialtyLaparoscopicLaparoscopicMinimally invasive surgical application
Mode of operationElectromechanically operated, software controlledElectromechanically operated, software controlledManually
Insertion MethodVia 5mm TrocarVia 5mm TrocarVia 5mm Trocar
ArticulationYesYesNo

{5}------------------------------------------------

DescriptionHandX™ Self-Righting NeedleHolder(Subject device)HX Device, Needle Holder(Predicate device)ENDOPATH® Self-Righting NeedleHolder(Reference device)
Articulationrange0 to 80°±10°0 to 85°±5°NA
Articulationlocking featureYesYesNA
Tip axialrotation feature360°360°NA
Shaft materialStainless steelStainless steelStainless steel
Instrument typeSelf-Righting Needle HolderNeedle HolderSelf-Righting NeedleHolder
Material forelectrode tipStainless steelStainless steelStainless steel
Shaft length330 mm ±0.1 mm330 mm ±0.1 mmNA
Shaft diameter5.5 mm ±0.2 mm5.5 mm ±0.1 mmNA
Single Use /ReusableSingle use instrumentSingle use instrumentRe-usable
PackagingBlisterBlisterNA
ProvidedsterilizedYesYesNA
SterilizationmethodEtOEtONA
Sterilityassurance level10-610-6NA

lt can be concluded that the Self-Righting Needle Holder is substantially equivalent to its predicate. No new questions of safety and effectiveness were raised.

Performance Data:

The following performance data were provided in support of the substantial equivalence determination:

Biocompatibility Testing:

The materials of the Self-Righting Needle Holder are identical to those of the predicate device. The predicate device has been tested and is considered biocompatible for its intended use. Therefore, from a biocompatibility and toxicological perspective, the device remains unchanged, and no additional

{6}------------------------------------------------

testing is required for the HandX™ Self-righting Needle Holder device since only the jaws geometry of the needle was changed.

Bench Testing:

The Self-Righting Needle Holder was subjected to bench testing to evaluate its performance and demonstrate that the design outputs meet the design input requirements, and that the device is safe and effective for its intended use.

The verification bench testing is described in the table below, all tests met the predefined acceptance criteria. The testing identified no new questions of safety and effectiveness.

TestTest SummaryConclusions
Self-Righting Needle HolderAngle TestThe purpose of the test wasto measure the subjectdevice articulation angleswhile connected to theHandX device.The subject device tiparticulation ability iscomparable to the predicatedevice.
SRNH instrumentDimensional AttributeThe purpose of the test wasto confirm the subjectdevice dimensions to thespecificationsThe subject devicedimensions are within thedefined requirements.
SRNH Suture Holding ForceThe purpose of the test wasto measure the subjectdevice suture holding force.The subject device sutureholding force withstands thedefined requirements as thepredicate device.
SRNH Needle PullingMomentThe purpose of this test wasto determine whether thesubject device maintainsthe needle in place underexternal moment.The subject device complieswith the needle pullingmoment requirement.
SRNH Needle Holding ForceThe purpose of this test wasto determine whether thesubject device maintainsthe needle in place underexternal force.The subject device complieswith the needle holderholding force requirement
SRNH Needle RightingThe purpose of the test wasto verify the subject devicetransition ability of asurgical needle to a rightposition.The subject devicedemonstrated transitionability of surgical needles toright position, thus met thepredefined requirements.

Conclusion:

Based upon the intended use, indications, technological characteristics and performance testing, as well as comparison to the predicate device, it can be concluded that the Self-Righting Needle Holder is substantially equivalent to its predicate device. The minor differences between the subject and predicate devices do not raise any questions of safety and effectiveness.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.