(228 days)
The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.
Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, and pancreatic islet cell carcinoma.
The i-STAT 1 Analyzer is intended for use in the in vitro quantification of various analytes in whole blood or plasma in point of care or clinical laboratory settings. Analyzers and cartridges should be used by healthcare professionals trained and certified to use the system and should be used according to the facility's policies and procedures.
The i-STAT System is for in vitro diagnostics use. Caution: Federal law restricts this device to sale by or on the order of a licensed practitioner.
The i-STAT CG8+ cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System to measure glucose (Glu) in arterial, venous or capillary whole blood.
The i-STAT 1 System is an in vitro diagnostic (IVD) medical device intended for the quantitative determination of various clinical chemistry tests contained within i-STAT cartridges using whole blood. The i-STAT 1 System consists of a portable blood analyzer (i-STAT 1 analyzer), single-use disposable test cartridges (i-STAT cartridges), liquid quality control and calibration verification materials, and accessories (i-STAT 1 Downloader/Recharger, i-STAT Electronic Simulator and i-STAT 1 Printer). The i-STAT 1 System, including the i-STAT CG8+ cartridge, is designed for use by trained medical professionals in point of care or clinical laboratory settings and is for prescription use only.
The i-STAT CG8+ cartridge contains the required sensors, a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within cartridge. The i-STAT cartridge format allows all the tests in the cartridge to be performed simultaneously. All the test steps and fluid movement occur within the i-STAT CG8+ cartridge. Cartridges require two to three drops of whole blood, which are typically applied to the cartridge using a transfer device, by the trained user before the cartridge is placed within the analyzer.
The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The instrument interacts with the i-STAT CG8+ cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes).
The acceptance criteria and study proving the device meets those criteria are detailed below, based on the provided FDA 510(k) summary for the i-STAT CG8+ cartridge with the i-STAT 1 System.
Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly listed in a single table with corresponding performance values in the provided document. Instead, performance expectations are implied through the comparison to the predicate device and the presentation of various study results (precision, linearity, detection limits, interference, sensitivity, and method comparison) against established CLSI guidelines or internal thresholds.
However, based on the provided information, a summary of key performance characteristics and their reported values can be presented as follows:
1. Table of Acceptance Criteria and Reported Device Performance (Implied from Study Results)
| Performance Characteristic | Acceptance Criteria (Implied/Standard) | Reported Device Performance (i-STAT CG8+ Glucose Test) |
|---|---|---|
| Precision | Consistent and acceptable repeatability, between-run, between-day, between-operator, and within-laboratory/total precision across various glucose levels and sample types, demonstrated through statistical metrics like SD and %CV, in line with CLSI EP05-A3 and internal standards. | 20-Day Precision (Aqueous Materials - CV%): Ranges from 0.80% to 1.66% within-laboratory for various levels (28.4 to 576.8 mg/dL). Repeatability %CV is 0.53% to 1.25%.Multi-site/Operator Precision (Aqueous Materials - Overall %CV): Ranges from 0.64% to 1.59% for various levels (28.1 to 578.4 mg/dL).Whole Blood Precision (SD/ %CV): Ranges from 0.66% to 1.46% (%CV) for venous blood (73.9 to 544.3 mg/dL), 0.66% to 0.96% (%CV) for arterial blood (80.8 to 280.7 mg/dL), and 1.34% to 2.16% (%CV) for capillary blood (77.6 to 203.8 mg/dL). Values are well within typical acceptable ranges for point-of-care glucose testing. |
| Linearity/Reportable Range | Demonstration of linearity across the claimed reportable range (20 – 700 mg/dL), typically assessed by a high correlation coefficient (R²) and slope/intercept close to 1 and 0, respectively. | Reportable Range: 20 – 700 mg/dL.Range Tested: 17.1 – 795.4 mg/dL.Regression Summary: Slope = 0.994, Intercept = -1.385, R² = 0.9993. This indicates excellent linearity across and beyond the reportable range. |
| Detection Limits (LoQ, LoB, LoD) | Limit of Quantitation (LoQ) at or below the lower limit of the reportable range. Limit of Blank (LoB) and Limit of Detection (LoD) sufficiently low to ensure reliable detection of very low glucose levels. | Determined LoQ: 17 mg/dL (lower limit of reportable range is 20 mg/dL). Meets criteria.Determined LoB: 0.2 mg/dL.Determined LoD: 0.9 mg/dL. These values demonstrate the ability to accurately measure very low glucose concentrations. |
| Analytical Specificity (Interference) | Minimal or no significant interference from common endogenous and exogenous substances at specified concentrations, as determined by the difference between control and test samples falling within allowable error (±Ea). For identified interferents, a dose-response study is required. | Most substances tested (e.g., Acetaminophen, Acetoacetate, Bilirubin, Cholesterol, Ethanol, Heparin) showed No Interference. Identified Interferents for Glu: • Bromide: Yes, "Use Another Method"• Hydroxyurea: Yes, "Increased results ≥ 0.08 mmol/L"• Isoniazid: Yes, "Increased results ≥ 0.29 mmol/L"(Note: Intralipid 20% showed "Increased results >", but without a specific concentration or 'Yes/No' for interference, it's less clear, though typically levels above ~300 mg/dL (triglycerides) can interfere with glucose assays). |
| Sensitivity (Oxygen, Hematocrit, Altitude) | Insensitivity to variations in oxygen levels, hematocrit levels, and altitude within specified clinical ranges, demonstrating comparable performance under these varied conditions. | Oxygen Sensitivity: Insensitive to oxygen levels between 20 and 503 mmHg. (95% CI of difference within ±Ea).Hematocrit Sensitivity: Insensitive to hematocrit levels between 15% to 75% PCV. (Difference vs. mid-hematocrit within ±Ea).Altitude: Equivalent performance at ~10,000 feet above sea level, with a correlation coefficient (r) of 1.00 and a slope of 0.96 (95% CI: 0.957 to 0.971), meeting acceptance criteria. |
| Method Comparison (vs. Predicate) | Substantial equivalence to the predicate device in arterial, venous, and capillary whole blood, demonstrated by Passing-Bablok regression with a high correlation coefficient (r) close to 1, and slope/intercept close to 1 and 0 respectively. | Venous/Arterial/Capillary Whole Blood (Pooled Data):N: 547Slope: 0.98Intercept: 1.62r: 1.00This demonstrates strong agreement with the comparative methods (i-STAT CHEM8+ and epoc Blood Analysis System), supporting substantial equivalence. |
| Matrix Equivalence | Demonstration of equivalence between non-anticoagulated and anticoagulated whole blood specimens for Glucose. | Non-anticoagulated vs. Anticoagulated Whole Blood:N: 297r: 1.00Slope: 1.00Intercept: 0.00This indicates excellent matrix equivalence. |
Note: The acceptance criteria are largely implied by the successful completion and positive results of studies designed according to CLSI (Clinical and Laboratory Standards Institute) guidelines, such as EP05-A3, EP06-Ed2, EP17-A2, EP07-ED3, EP37-ED1, EP09c-ED3, and EP35. Meeting the statistical benchmarks (e.g., specific SD, %CV, R², slope, intercept ranges, or being within "allowable error") for these guidelines indicates that the device's performance is acceptable for its intended use.
Study Details
Here's a breakdown of the study details based on the provided document:
2. Sample Sizes Used for the Test Set and Data Provenance
-
Precision Studies:
- 20-day Precision (Aqueous): N=80 samples per level.
- Multi-site/Operator Precision (Aqueous): N=90-96 samples per level (across 3 sites).
- Whole Blood Precision: Venous (N=29-102 per range), Arterial (N=5-105 per range), Capillary (N=15-107 per range).
-
Linearity: Whole blood samples of varying glucose levels. Specific N not provided for this section, but typically multiple points across the range are tested.
-
Detection Limits (LoQ, LoB, LoD): Whole blood samples (altered to low/blank glucose levels). Specific N not provided.
-
Analytical Specificity (Interference): Whole blood samples. Specific N for each substance not provided, but the study was extensive (Table 8 lists many substances).
-
Other Sensitivity Studies:
- Oxygen Sensitivity: Whole blood samples.
- Hematocrit Sensitivity: Whole blood samples (at low, mid, high hematocrit levels).
- Altitude: Whole blood samples.
-
Method Comparison (with predicate device):
- N = 547 (pooled data from arterial, venous, and capillary whole blood specimens).
-
Matrix Equivalence:
- N = 297 (venous and arterial whole blood specimens).
-
Data Provenance:
- The document states "collected across multiple point of care sites" for whole blood precision and method comparison studies. It also mentions "at one site" for the 20-day precision study and "at three (3) sites" for the multi-site precision study.
- The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. However, clinical studies for 510(k) submissions are typically prospective, especially those involving patient samples collected concurrently with the study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- For this type of in vitro diagnostic device (quantitative glucose measurement), the "ground truth" for the test set is established by comparative reference methods, not by expert consensus or interpretations of images by radiologists.
- The document mentions "a comparative method" (i-STAT CHEM8+ and epoc Blood Analysis System for glucose) for the method comparison study. These are legally marketed, validated laboratory or point-of-care devices that serve as the reference standard for measuring glucose.
- No human experts (like radiologists in an imaging study) are used to establish "ground truth" for quantitative lab tests in this context. Their role might be in collecting samples by healthcare professionals, but not in determining the true value of the analyte.
4. Adjudication Method for the Test Set
- Adjudication methods (e.g., 2+1, 3+1) are primarily relevant for imaging studies where human readers interpret data, and discrepancies need to be resolved.
- For quantitative in vitro diagnostic tests like glucose measurement, adjudication of results in the traditional sense is not applicable. The "ground truth" is determined by the output of a reference instrument or method, and performance is assessed by statistical agreement between the new device and the reference.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
- No, an MRMC comparative effectiveness study was not done.
- This device is an in vitro diagnostic for quantitative measurement of glucose, not an AI-based imaging or diagnostic device that assists human readers. Therefore, the concept of human readers improving with AI assistance is not relevant to this submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
- Yes, the performance characteristics described (precision, linearity, detection limits, interference, sensitivity) represent the standalone performance of the device (i-STAT CG8+ cartridge with i-STAT 1 System).
- The method comparison study also evaluates the device's performance against a reference method independently. Human involvement is primarily in operating the device and collecting samples, not in interpreting or enhancing the device's quantitative output.
7. The Type of Ground Truth Used
- The ground truth for the device's performance evaluation was established using comparative reference methods (e.g., i-STAT CHEM8+ glucose test, epoc Blood Analysis System) which are considered established and validated methods for glucose quantification.
- This is akin to using a gold standard laboratory test result for comparison.
8. The Sample Size for the Training Set
- The provided document is a 510(k) summary for a point-of-care in vitro diagnostic device, not an AI/Machine Learning device.
- Therefore, there is no "training set" in the context of machine learning model development. The device relies on electrochemical detection mechanisms and pre-calibrated algorithms, not on statistical models trained on large datasets in the way AI applications do.
- The studies described are for validation (test set), not training.
9. How the Ground Truth for the Training Set Was Established
- As explained in point 8, there is no "training set" for this device in the machine learning sense.
- The fundamental principles and calibration of the glucose sensor (glucose oxidase-based amperometric peroxide detection) are based on established chemical and electrochemical principles in analytical chemistry, refined and validated during product development.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of the HHS logo is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 28, 2023
Abbott Point of Care Inc. Jacquelyn Gesumaria Principal Specialist Regulatory Affairs 400 College Road East Princeton, New Jersey 08540
Re: K223710
Trade/Device Name: i-STAT CG8+ cartridge with the i-STAT 1 System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CGA, JJE Dated: June 28, 2023 Received: June 28, 2023
Dear Jacquelyn Gesumaria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Paula V. Caposino -S
Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K223710
Device Name
i-STAT CG8+ cartridge with the i-STAT 1 System
Indications for Use (Describe)
The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.
Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, and pancreatic islet cell carcinoma.
The i-STAT 1 Analyzer is intended for use in the in vitro quantification of various analytes in whole blood or plasma in point of care or clinical laboratory settings. Analyzers and cartridges should be used by healthcare professionals trained and certified to use the system and should be used according to the facility's policies and procedures.
The i-STAT System is for in vitro diagnostics use. Caution: Federal law restricts this device to sale by or on the order of a licensed practitioner.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Abbott logo. On the left is a blue stylized letter A. To the right of the letter is the word "Abbott" in black, bold font.
510(k) SUMMARY
The information in this 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
I. SUBMITTER INFORMATION
| Owner | Abbott Point of Care Inc.400 College Road EastPrinceton, NJ 08540 |
|---|---|
| Contact | Primary: Jacquelyn GesumariaPrincipal Specialist Regulatory AffairsPhone: 609-454-9384Secondary: Mojgan SoleimaniAssociate Director Regulatory AffairsPhone: 613-295-0932 |
| Date Prepared | June 28, 2023 |
II. DEVICE INFORMATION
| Proprietary Name | i-STAT CG8+ cartridge with the i-STAT 1 System |
|---|---|
| Common Name | Chemistry test, analyzer, handheld |
| 510(k) Number | K223710 |
| Product Code | Device ClassificationName | RegulationNumber | Class | Panel |
|---|---|---|---|---|
| CGA | Glucose Oxidase,Glucose | 862.1345 | II | Clinical Chemistry |
III. PREDICATE DEVICE
K210958 (K183678) 510(k) Number
| ProductCode | Device ClassificationName | RegulationNumber | Class | Panel |
|---|---|---|---|---|
| CGA | Glucose Oxidase, Glucose | 862.1345 | II | Clinical Chemistry |
{4}------------------------------------------------
IV. DEVICE DESCRIPTION
The i-STAT CG8+ cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System to measure glucose (Glu) in arterial, venous or capillary whole blood.
The i-STAT 1 System is an in vitro diagnostic (IVD) medical device intended for the quantitative determination of various clinical chemistry tests contained within i-STAT cartridges using whole blood. The i-STAT 1 System consists of a portable blood analyzer (i-STAT 1 analyzer), single-use disposable test cartridges (i-STAT cartridges), liquid quality control and calibration verification materials, and accessories (i-STAT 1 Downloader/Recharger, i-STAT Electronic Simulator and i-STAT 1 Printer). The i-STAT 1 System, including the i-STAT CG8+ cartridge, is designed for use by trained medical professionals in point of care or clinical laboratory settings and is for prescription use only.
The i-STAT CG8+ cartridge contains the required sensors, a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within cartridge. The i-STAT cartridge format allows all the tests in the cartridge to be performed simultaneously. All the test steps and fluid movement occur within the i-STAT CG8+ cartridge. Cartridges require two to three drops of whole blood, which are typically applied to the cartridge using a transfer device, by the trained user before the cartridge is placed within the analyzer.
The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The instrument interacts with the i-STAT CG8+ cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes).
V. INTENDED USE STATEMENT
The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.
Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
The i-STAT 1 Analyzer is intended for use in the in vitro quantification of various analytes in whole blood or plasma in point of care or clinical laboratory settings. Analyzers and cartridges should be used by healthcare professionals trained and certified to use the system and should be used according to the facility's policies and procedures.
The i-STAT System is for in vitro diagnostics use. Caution: Federal law restricts this device to sale by or on the order of a licensed practitioner.
{5}------------------------------------------------
VI. SUMMARY COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
| Table 1: Similarities and Differences (Test and Instrument) | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Feature orCharacteristic | Candidate Devices:i-STAT CG8+ cartridge with the i-STAT 1System, i-STAT Glucose test | Predicate Device:i-STAT CHEM8+ cartridge with the i-STAT 1System, i-STAT Glucose test (K210958) | ||||||||
| Intended Use | The i-STAT CG8+ cartridge with thei-STAT 1 System is intended for use in thein vitro quantification of glucose in arterial,venous or capillary whole blood in point ofcare or clinical laboratory settings.Glucose measurements are used in thediagnosis, monitoring, and treatment ofcarbohydrate metabolism disordersincluding, but not limited to, diabetesmellitus, neonatal hypoglycemia,idiopathic hypoglycemia, and pancreaticislet cell carcinoma. | The i-STAT CHEM8+ cartridge with thei-STAT 1 System is intended for use in thein vitro quantification of sodium,potassium, chloride, ionized calcium,glucose, blood urea nitrogen, creatinine,hematocrit, and total carbon dioxide inarterial or venous whole blood in point ofcare or clinical laboratory settings.Glucose measurements are used in thediagnosis, monitoring, and treatment ofcarbohydrate metabolism disordersincluding, but not limited to, diabetesmellitus, neonatal hypoglycemia, idiopathichypoglycemia, and pancreatic islet cellcarcinoma. | ||||||||
| DeviceClassification | Same | Class II | ||||||||
| Product Code | Same | CGA | ||||||||
| RegulationNo. | Same | 862.1345 | ||||||||
| ReportableRange | Same | 1.1 – 38.9 mmol/L20 – 700 mg/dL0.20 - 7.00 g/L | ||||||||
| Sample Type | Arterial, venous, or capillary whole blood | Arterial and venous whole blood | ||||||||
| SampleVolume | Same | 95 µL | ||||||||
| SamplePreparation | Same | Ready to Use | ||||||||
| Samplecollection | Same | • Without anticoagulant• With balanced heparin anticoagulant orlithium heparin anticoagulant | ||||||||
| Traceability | Same | NIST SRM965 | ||||||||
| Calibration | Same | 1-point on-board contained withincartridge |
{6}------------------------------------------------
| Table 1: Similarities and Differences (Test and Instrument) | ||
|---|---|---|
| Feature or Characteristic | Candidate Devices:i-STAT CG8+ cartridge with the i-STAT 1 System, i-STAT Glucose test | Predicate Device:i-STAT CHEM8+ cartridge with the i-STAT 1 System, i-STAT Glucose test (K210958) |
| Time to Test(Sample Stability) | Without anticoagulant:Arterial and venous within 3 minutesWith anticoagulant:Capillary within 3 minutesArterial and venous within 30 minutes | Without anticoagulant:Arterial and venous within 3 minutesWith anticoagulant:Arterial and venous within 30 minutes |
| Principle of Measurement | Same | Glu: Glucose oxidase based amperometric peroxide detection |
| Reagent Format | Same | Cartridge |
| Reagent Storage and Stability | Refrigerated at 2 to 8°C (35 to 46°F) until expiration dateRoom Temperature at 18-30°C (64-86°F) for 2 months | Refrigerated at 2 to 8°C (35 to 46°F) until expiration dateRoom Temperature at 18-30°C (64-86°F) for 14 days |
| Analyzer Type | Same | Handheld |
PERFORMANCE CHARACTERISTICS VII.
A. Analytical Performance
a. Precision/Reproducibility:
- i. Precision 20 days (Aqueous materials)
The precision of the i-STAT Glucose (Glu) test in the i-STAT CG8+ cartridge on the i-STAT 1 System was evaluated using five (5) levels of aqueous material. This 20day multi-day precision testing was based on CLSI document EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition. The study was conducted using multiple analyzers and two (2) test cartridge lots over 20 days at one site. Repeatability, between-run, between-day, and within-laboratory precision were estimated for each level. The results of the 20-day precision study for the i-STAT CG8+ cartridge on the i-STAT 1 System are shown in Table 2.
| Table 2: Results of 20-Day Precision of the i-STAT CG8+ Cartridge on the i-STAT 1 Analyzer | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Test(units) | FluidLevel | N | Mean | Repeatability | Between-run | Between-day | Within-Laboratory | ||||
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | ||||
| Glu(mg/dL) | CV L1 | 80 | 28.4 | 0.36 | 1.25 | 0.20 | 0.69 | 0.24 | 0.85 | 0.47 | 1.66 |
| CV L2 | 80 | 42.0 | 0.36 | 0.87 | 0.29 | 0.70 | 0.23 | 0.56 | 0.52 | 1.25 | |
| CV L3 | 80 | 124.7 | 0.66 | 0.53 | 0.55 | 0.44 | 0.52 | 0.42 | 1.00 | 0.80 | |
| CV L4 | 80 | 279.4 | 1.47 | 0.53 | 1.79 | 0.64 | 1.26 | 0.45 | 2.63 | 0.94 | |
| CV L5 | 80 | 576.8 | 4.70 | 0.81 | 2.35 | 0.41 | 2.91 | 0.50 | 6.01 | 1.04 |
{7}------------------------------------------------
ii. Multi-site and operator-to-operator precision (Aqueous materials)
Multi-day precision testing was performed at three (3) sites using a panel of aqueous solutions containing five (5) levels of glucose. At each site, each level was tested once a day by two (2) operators for five (5) days on six (6) i-STAT 1 Analyzers using i-STAT CG8+ cartridges. Within-run, between-day, betweenoperator and within-site (total) variance components were calculated by site. These components were also calculated for all sites combined and provided in the Table 3 below.
{8}------------------------------------------------
| Table 3: Multi-Day Precision of the i-STAT CG8+ Cartridge on the i-STAT 1 Analyzer | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Test(units) | FluidLevel | N | Mean | Within-Run | Between-Day | Between-Operator | Within-Site (Total) | Between-Site | Overall | ||||||
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | ||||
| Glu(mg/dL) | CV L1 | 91 | 28.1 | 0.39 | 1.40 | 0.16 | 0.55 | 0.08 | 0.29 | 0.43 | 1.54 | 0.11 | 0.41 | 0.45 | 1.59 |
| CV L2 | 90 | 41.4 | 0.50 | 1.21 | 0.12 | 0.30 | 0.17 | 0.41 | 0.54 | 1.32 | 0.00 | 0.00 | 0.54 | 1.32 | |
| CV L3 | 96 | 124.5 | 0.64 | 0.52 | 0.47 | 0.38 | 0.08 | 0.06 | 0.8 | 0.64 | 0.00 | 0.00 | 0.80 | 0.64 | |
| CV L4 | 90 | 280.3 | 1.42 | 0.51 | 0.33 | 0.12 | 0.22 | 0.08 | 1.47 | 0.53 | 0.44 | 0.16 | 1.54 | 0.55 | |
| CV L5 | 90 | 578.4 | 4.25 | 0.73 | 1.78 | 0.31 | 0.00 | 0.00 | 4.61 | 0.80 | 0.00 | 0.00 | 4.61 | 0.80 |
{9}------------------------------------------------
iii. Precision (Whole Blood)
Whole blood precision of the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 System was evaluated using arterial, venous, and capillary whole blood specimens collected with lithium heparin. The whole blood precision was assessed using the duplicate test results collected across multiple point of care sites. The mean values for each sample were divided into subintervals for each sample type across the reportable range for the i-STAT Glucose test. The results are summarized in Table 4.
| Table 4: Whole Blood Precision of arterial, venous, and capillary whole blood for the i-STATGlucose test in the i-STAT CG8+ cartridge on the i-STAT 1 Analyzer | ||||||
|---|---|---|---|---|---|---|
| Test(units) | Sample Type | Sample Range | N | Mean | SD | %CV |
| Glu(mg/dL) | Venous Whole Blood | 20-90 | 29 | 73.9 | 0.86 | 1.17 |
| >90-150 | 102 | 111.7 | 1.08 | 0.96 | ||
| >150-250 | 27 | 173.4 | 1.82 | 1.05 | ||
| >250-400 | 13 | 308.5 | 2.08 | 0.67 | ||
| >400-700 | 9 | 544.3 | 7.92 | 1.46 | ||
| Glu(mg/dL) | Arterial Whole Blood | 20-90 | 5 | 80.8 | 0.77 | 0.96 |
| >90-150 | 105 | 113.6 | 0.75 | 0.66 | ||
| >150-250 | 35 | 178.0 | 1.55 | 0.87 | ||
| >250-400 | 8 | 280.7 | 2.19 | 0.78 | ||
| Glu(mg/dL) | Capillary Whole Blood | 20-90 | 32 | 77.6 | 1.08 | 1.39 |
| >90-150 | 107 | 108.3 | 2.34 | 2.16 | ||
| >150-700 | 15 | 203.8 | 2.74 | 1.34 |
b. Linearity/assay reportable range:
i. Linearity
The study was designed based on CLSI EP06-Ed2: Evaluation of the Linearity of Quantitative Measurement Procedures - Second Edition.
The linearity of the i-STAT Glucose test in the i-STAT CG8+ cartridge with the i-STAT 1 System was evaluated by preparing whole blood samples of varying glucose levels. The i-STAT Glucose test in the i-STAT CG8+ cartridge demonstrated linearity over the reportable range for each i-STAT test. Regression summary of the response for the i-STAT Glucose test versus the concentration of the whole blood samples of varying glucose levels is provided in Table 5.
| Table 5: Regression Summary for the i-STAT Glucose test in the i-STAT CG8+ Cartridge on thei-STAT 1 Analyzer | ||||||
|---|---|---|---|---|---|---|
| Test | Units | Reportable Range | Range Tested | Slope | Intercept | R2 |
| Glu | mg/dL | 20 – 700 | 17.1 – 795.4 | 0.994 | -1.385 | 0.9993 |
c. Detection Limit
i. Limit of Quantitation (LoQ)
The study was based on the CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition.
The LoQ of the i-STAT Glucose test in the i-STAT CG8+ cartridge was evaluated on the i-STAT 1 analyzer using two (2) i-STAT CG8+ cartridge lots and whole blood
{10}------------------------------------------------
that was altered to a low glucose level. The LoQ for the i-STAT Glucose test in the i-STAT CG8+ cartridge was determined to be at or below the lower limit of the reportable range for the i-STAT Glucose test as shown in Table 6.
| Table 6: Summary of LoQ Results for i-STAT Glucose Test in the i-STAT CG8+Cartridge | ||
|---|---|---|
| Test (units) | Lower limit of thereportable range | Determined LoQ |
| Glu (mg/dL) | 20 | 17 |
ii. Limit of Blank and Detection (LoB/LoD)
The study was based on CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition.
The LoB and LoD of the i-STAT Glucose (Glu) test in the i-STAT CG8+ cartridge was evaluated on the i-STAT 1 analyzer using two (2) i-STAT CG8+ cartridge lots. Whole blood was altered to a blank glucose level for LoB testing. Whole blood was altered to two (2) low levels of glucose for LoD testing.
The LoB and LoD were determined based on the maximal LoB or LoD value obtained for each lot tested.
The determined LoB and LoD for the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 analyzer is shown in the Table 7.
| Table 7: Summary of LoB and LoD Results | ||
|---|---|---|
| Test (units) | LoB | LoD |
| Glu (mg/dL) | 0.2 | 0.9 |
d. Analytical Specificity
i. Interference
The study was based on CLSI EP07-ED3: Interference Testing in Clinical Chemistry, Third Edition.
The interference performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 analyzer with the i-STAT 1 System was evaluated using whole blood samples based on CLSI EP07-ED3: Interference Testing in Clinical Chemistry, Third Edition. The effect of each substance was evaluated by comparing the performance of a control sample, spiked with blank solvent solution, with the test results from a test sample spiked with the potentially interfering substance at the toxic/pathological concentration based on CLSI EP37-ED1: Supplemental Tables for Interference Testing in Clinical Chemistry, First Edition, as applicable. A substance was identified as an interferent if the difference between the control and test samples was outside of the allowable error (+Ea) for the i-STAT Glucose test. For an identified interferent, a dose-response was performed to determine the degree of interference as a function of the substance concentration.
Table 8 contains the lists of potentially interfering substances tested and the interference results for the i-STAT CG8+ cartridge.
{11}------------------------------------------------
| Table 8: Potentially Interfering Substances and Test Concentrations for the i-STAT Glucose test inthe i-STAT CG8+ Cartridge | ||||||
|---|---|---|---|---|---|---|
| Test Concentration | ||||||
| Substance 1 | mmol/L(unlessspecified) | mg/dL(unlessspecified) | i-STATTest | Interference(Yes/No) | Comments | |
| Acetaldehyde 2 | 0.045 | 0.2 | Glu | No | ||
| Acetaminophen | 1.03 | 15.6 | Glu | No | ||
| Acetoacetate(LithiumAcetoacetate) | 2.0 | 20 | Glu | No | ||
| Acetyl Cysteine (N-Acetyl-L-Cysteine) | 0.92 | 15 | Glu | No | ||
| Ammonium(AmmoniumChloride) 2 | 2.0 | 10.7 | Glu | No | ||
| Ascorbic Acid (L-Ascorbic Acid) | 0.298 | 5.25 | Glu | No | ||
| β-HydroxybutyricAcid 2 | 6.0 | 62.46 | Glu | No | ||
| Bilirubin | 0.684 | 40 | Glu | No | ||
| 2.5 | 21.7 | Glu | No | |||
| Bromide 2(Lithium Bromide) | 37.5 | 325.7 | Glu | Yes | Use AnotherMethod | |
| Cholesterol | 10.3 | 400 | Glu | No | ||
| Creatinine | 1.326 | 15 | Glu | No | ||
| Dopamine(DopamineHydrochloride) | 4.06μmol/L | 0.0621 | Glu | No | ||
| Ethanol | 130 | 600 | Glu | No | ||
| Fluoride (LithiumFluoride) | 0.0632 | 0.12 | Glu | No | ||
| Formaldehyde 2 | 0.133 | 0.399 | Glu | No | ||
| Fructose | 1 | 18 | Glu | No | ||
| Galactose | 3.33 | 60 | Glu | No | ||
| Gentamicin(Gentamicin Sulfate) | 0.0628 | 3 | Glu | No | ||
| Gentisic Acid | 0.0973 | 1.5 | Glu | No | ||
| Glucosamine(GlucosamineHydrochloride) 2 | 0.030 | 0.647 | Glu | No | ||
| Glutathione, reduced 3 | 3 | 3 mEq/L | Glu | No | ||
| Glycolic Acid 2 | 10.0 | 76.05 | Glu | No | ||
| Guaifenesin | 0.0227 | 0.45 | Glu | No | ||
| Hemoglobin | 10 g/L | 1000 | Glu | No | ||
| Heparin (SodiumHeparin) | 3.30 U/mL | 330 U/dL | Glu | No | ||
| Hydroxyurea | 0.405 | 3.08 | Glu | Yes | Increased results ≥0.08 mmol/L | |
| Ibuprofen | 1.06 | 21.9 | Glu | No | ||
| Table 8: Potentially Interfering Substances and Test Concentrations for the i-STAT Glucose test in | ||||||
| the i-STAT CG8+ Cartridge | ||||||
| Substance 1 | Test Concentrationmmol/L(unlessspecified) | mg/dL(unlessspecified) | i-STATTest | Interference(Yes/No) | Comments | |
| Intralipid 20% | N/A | 2891 | Glu | No | Increased results > | |
| Isoniazid | 0.438 | 6 | Glu | Yes | Increased results ≥0.29 mmol/L | |
| Lactate (LithiumLactate) | 10 | 90 | Glu | No | ||
| Maltose | 10.5 | 360 | Glu | No | ||
| Mannose 2 | 1 | 18.02 | Glu | No | ||
| Nithiodote (SodiumThiosulfate) 2 | 16.7 | 264.04 | Glu | No | ||
| pH | 8.0 pHunits | N/A | Glu | No | ||
| Pyruvate (LithiumPyruvate) | 0.570 | 5 | Glu | No | ||
| Salicylate (LithiumSalicylate) | 0.207 | 2.86 | Glu | No | ||
| Thiocyanate (LithiumThiocyanate) | 0.898 | 5.22 | Glu | No | ||
| Triglyceride | 16.94 | 1500 | Glu | No | ||
| Uric Acid | 1.4 | 23.5 | Glu | No | ||
| Xylose 2 | 3 | 45.04 | Glu | No |
1 The compound tested to evaluate the interfering substance is presented in parenthesis.
² The test concentration for this substance is not included in CLSI guideline EP37 1st edition.
{12}------------------------------------------------
ii. Other sensitivity studies
1) Oxygen Sensitivity Study
The effect of oxygen on the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 System was evaluated with low and high oxygen levels using whole blood samples altered to four (4) glucose levels the reportable range of the i-STAT Glucose test. The equivalency between the high and low oxygen conditions was determined if the 95% confidence interval (CI) of the difference in means (or medians) was within the allowable error (±Ea). The study demonstrated that the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 System perform is insensitive to oxygen levels between 20 and 503 mmHg.
2) Hematocrit Sensitivity
A hematocrit sensitivity study was conducted to evaluate the performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge with the i-STAT 1 Sustem to assess the effect of hematocrit. Three (3) hematocrit levels (low, mid and high) were evaluated at four (4) glucose levels across the reportable range of the i-STAT Glucose test in the i-STAT CG8+ cartridge. The hematocrit sensitivity at each glucose level was assessed by comparing the results at the low and high hematocrit level to the mid hematocrit level. Equivalency was assessed by determining whether the difference between the low and high hematocrit level and the mid hematocrit level was within the allowable error (±Ea). The study demonstrated that i-STAT Glucose test in the i-STAT CG8+ cartridge with the i-STAT 1 System is insensitive to hematocrit levels between 15% to 75% packed cell volume (PCV).
{13}------------------------------------------------
3) Altitude
The performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 analyzer at an altitude of approximately 10,000 feet above sea level was evaluated using whole blood samples at relevant glucose levels across the reportable range for the i-STAT Glucose test. The glucose test results obtained from the i-STAT CG8+ cartridges (candidate device) were compared to the glucose test results obtained from the i-STAT CHEM8+ (blue) cartridges on the i-STAT 1 analyzer (comparator device). Passing-Bablok regression analyses between the first replicate of the candidate device (y-axis) and mean of the comparator device (x-axis) were performed based on the CLSI EP09c: Measurement Procedure Comparison and Bias Estimation using Patient Samples, 3rd ed. The results of the correlation coefficient and slope met acceptance criteria and demonstrated equivalent performance between the candidate and comparator conditions at approximately 10,000 feet above sea level. The results are summarized in Table 9.
| Table 9: Summary of Altitude Study Results for the i-STAT CG8+ Cartridge | ||||
|---|---|---|---|---|
| Test | Correlation Coefficient (r) | Slope | ||
| r | 95% CI | Slope | 95% CI | |
| Glu | 1.00 | 1.000 to 1.000 | 0.96 | 0.957 to 0.971 |
B. Comparison Studies
a. Method Comparison with predicate device
Method comparison for arterial, venous, and capillary whole blood specimens on the i-STAT CG8+ cartridge with the i-STAT 1 System was demonstrated in studies based on CLSI EP09c-ED3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples - Third Edition.
Lithium heparin venous and arterial whole blood specimens collected across multiple point of care sites were evaluated using i-STAT CG8+ cartridges on the i-STAT 1 analyzer against whole blood specimens tested on a comparative method. The first replicate glucose result from the i-STAT 1 analyzer was compared to the mean glucose result from the comparative method.
Two (2) capillary whole blood specimens collected from skin puncture with balanced heparin capillary tubes from each study subject across multiple point of care sites were evaluated and analyzed in singlicate on the i-STAT 1 analyzer against the comparative method. A Passing-Bablok linear regression analysis for glucose was performed using the singlicate result from the i-STAT 1 analyzer versus the singlicate result of the comparative method.
The venous, arterial, and capillary whole blood data were pooled, and a Passing-Bablok linear regression analysis was performed using the results from the i-STAT CG8+ cartridges on the i-STAT 1 analyzer versus the comparative method results.
Method comparison results for arterial, venous, and capillary whole blood specimens are shown in Table 10. In the table, N is the number of specimens in the data set, and r is the correlation coefficient.
| Table 10: Method Comparison Results for i-STAT CG8+ Cartridge with i-STAT 1 System | ||||||
|---|---|---|---|---|---|---|
| Test | Comparative Method | N | Slope | Intercept | r | |
| Arterial/Venous | Capillary | |||||
| Glu | i-STAT CHEM8+ | epoc Blood AnalysisSystem | 547 | 0.98 | 1.62 | 1.00 |
{14}------------------------------------------------
b. Matrix Equivalence
A matrix equivalence study was conducted to evaluate the performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 System using nonanticoagulated venous and arterial whole blood specimens. The study design and analysis method were based on recommendations from the Clinical and Laboratory Standards Institute (CLSI) guideline EP35: Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures, 1st ed. The matrix equivalence of the i-STAT Glucose test in the i-STAT CG8+ cartridge was assessed by comparing arterial or venous whole blood specimens collected without anticoagulant (candidate specimen type) to samples collected with balanced heparin or lithium heparin anticoagulant (primary specimen type). Each specimen was tested in duplicate using two (2) i-STAT CG8+ cartridges with two (2) i-STAT 1 analyzers. A Passing-Bablok linear regression analysis was performed using the first replicate result from the candidate (y-axis) versus the mean result from the primary specimen (x-axis). The regression analysis results are summarized in Table 11. In the table, N is the number of specimens in the data set, and r is the correlation coefficient.
| Table 11: Matrix Equivalence Results | ||||||
|---|---|---|---|---|---|---|
| Test (units) | N | Candidate SpecimenRange | Primary SpecimenRange | r | Slope | Intercept |
| Glu (mg/dL) | 297 | 39-688 | 39-671 | 1.00 | 1.00 | 0.00 |
CONCLUSION VIII.
The results of these studies demonstrate that performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge with the i-STAT 1 System are substantially equivalent to the predicate device.
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.