(139 days)
The X-CUBE 50, X-CUBE 60 diagnostic ultrasound systems are intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Peripheral Vessel(PV); and Urology(including prostate).
And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF(Pulse Repetition Frequency) Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode.
The X-CUBE 50, X-CUBE 60 are intended to be used in a hospital or medical clinic.
The X-CUBE 50 and X-CUBE 60 products are general purpose ultrasound imaging system for medical diagnosis assistance. These products are used as an aid tool to diagnosis, such as a commonly used ultrasound diagnostic device.
Also X-CUBE 50 and X-CUBE 60 have the same operating principles, intended use, risk grade and design/manufacturing characteristics as reference/predicate devices.
This systems platform provide patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, image quality.
The provided document is a 510(k) Summary for the X-CUBE 50 and X-CUBE 60 diagnostic ultrasound systems. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a clinical study. Therefore, much of the requested information regarding acceptance criteria, sample sizes, ground truth establishment, expert involvement, and MRMC studies is not available in this document.
Here's an analysis of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not specify quantitative acceptance criteria for device performance. Instead, it states that the devices have been evaluated for biocompatibility, acoustic output, and thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. The performance is implicitly assessed against these standards and against the characteristics of the predicate devices.
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Biocompatibility | Conforms to AAMI/ANSI/ISO10993-1:2009(R)2013 |
| Acoustic Output | Conforms to NEMA UD 2-2004(R2009) and AIUM MUS, Third edition |
| Thermal Safety | Conforms to IEC 60601-1:2005+A1:2012 and IEC 60601-2-37:2007/AMD1:2015, NEMA UD 3-2004(R2009) |
| Electrical Safety | Conforms to IEC 60601-1:2005+A1:2012 and IEC 60601-2-37:2007/AMD1:2015 |
| Electromag. Safety | Conforms to IEC60601-1-2:2014 |
| Mechanical Safety | Conforms to IEC 60601-1:2005+A1:2012 and IEC 60601-2-37:2007/AMD1:2015 |
| Quality Management System | Conforms to 21 CFR 820 and ISO 13485 |
| Essential Performance | As safe and effective as the predicate device |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: “The subject of this premarket submission, X-CUBE 50 and X-CUBE 60, did not require clinical studies to support substantial equivalence.” Therefore, there is no specific test set or data provenance related to a clinical study. The evaluation appears to be based on adherence to standards and comparison of technical specifications with predicate devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable, as no clinical studies with a specific test set requiring expert-established ground truth were performed.
4. Adjudication Method for the Test Set
Not applicable, as no clinical studies with a test set requiring adjudication were performed.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The X-CUBE 50 and X-CUBE 60 are diagnostic ultrasound systems, not AI-powered interpretation software meant to assist human readers. The document does not describe any AI component or human reader effectiveness studies.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done
Not applicable, as the device is a diagnostic ultrasound system and not an AI algorithm.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
Not applicable, as no clinical studies with a specific test set were performed to assess diagnostic accuracy against a ground truth.
8. The Sample Size for the Training Set
Not applicable, as the device is a diagnostic ultrasound system and not an AI algorithm requiring a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable, as the device is a diagnostic ultrasound system and not an AI algorithm requiring a training set.
Summary of the Study that Proves the Device Meets Acceptance Criteria:
The "study" described in the 510(k) summary is not a traditional clinical trial or performance study against specific diagnostic metrics. Instead, it is an engineering and regulatory compliance assessment designed to demonstrate "substantial equivalence" to legally marketed predicate devices.
The proof that the device meets its implicit acceptance criteria (that it is as safe and effective as the predicate) is based on:
- Non-Clinical Tests: Evaluation for biocompatibility, acoustic output, and thermal, electrical, electromagnetic, and mechanical safety, confirming adherence to various international and national standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, AAMI/ANSI/ISO 10993-1, AAMI/ANSI/ISO 14971, AIUM MUS, NEMA UD 2, NEMA UD 3).
- Quality Management System Measures: Adherence to 21 CFR 820 and ISO 13485, including medical device risk management, requirements reviews, design reviews, component verification, integration review, performance testing, safety testing, and design validation.
- Comparison to Predicate Devices: A detailed comparison (table provided in the 510(k) summary) of the proposed devices (X-CUBE 50, X-CUBE 60) with predicate devices (K220857 X-CUBE 50, X-CUBE 60, and K221093 X-CUBE 70, X-CUBE 90). This comparison covers indications for use, physical dimensions, electrical power, imaging modes, imaging features, accessories, and compliance to safety standards. The document highlights that the new devices have minor changes (e.g., model name changes, addition of some transducers) and maintain the same operating principles, intended use, risk grade, and design/manufacturing characteristics as the predicate devices.
- Conclusion of Substantial Equivalence: The manufacturer concluded that the differences between the proposed and predicate devices would not affect safety, effectiveness, and essential performance, thus arguing for substantial equivalence.
In essence, the proof relies on a comprehensive engineering and regulatory checklist showing that the new devices meet established safety standards and are technically similar to previously cleared devices for their intended use. No new clinical performance data was deemed necessary by the FDA for this 510(k) clearance process.
{0}------------------------------------------------
April 17, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a stylized image of a human figure. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION".
Alpinion Medical Systems Co., Ltd. % Boyeon CHO Quality Management Representative 4F, 15, Magokjungang 14-ro, Gangseo-gu Seoul. Seoul 07789 KOREA
Re: K223564
Trade/Device Name: X-CUBE 50, X-CUBE 60 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: March 24, 2023 Received: March 27, 2023
Dear Boyeon CHO:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K223564
Device Name X-CUBE 50, X-CUBE 60
Indications for Use (Describe)
The X-CUBE 50, X-CUBE 60 diagnostic ultrasound systems are intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Peripheral Vessel(PV); and Urology(including prostate).
And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF(Pulse Repetition Frequency) Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode.
The X-CUBE 50, X-CUBE 60 are intended to be used in a hospital or medical clinic.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
510(k) Summary
In accordance with 21CFR807.92, the following summary of information is provided;
- Mar 24th, 2023 Date
- Submitter: ALPINION MEDICAL SYSTEMS Co., Ltd. Address: 4F, 15, Magokjungang 14-ro, Gangseo-gu, Seoul, 07789, Republic of Korea
- Primary Boyeon CHO Contact Person Quality Management Representative(QMR) Address: 4F, 15, Magokjungang 14-ro, Gangseo-qu, Seoul, 07789, Republic of Korea Phone: +82.2.3777.8600 Fax: +82.2.3777.8691 Email: ga ra@alpinion.com Secondary Kevin CHUN Contact Person Address: 21222 30th Dr SE Ste C-122, Bothell, WA 98021, United States Phone: 425 949 1059
Fax: 425 949 4910 Email: kevin.chun@alpinionusa.com
Device X-CUBE 50, X-CUBE 60
Trade Name:
- Common/ Ultrasonic Pulsed Doppler Imaging System Usual Name:
- Classification System, Imaging, Pulsed Doppler Ultrasonic Names
- Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550 90-IYN Product Code: Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO Diagnostic Ultrasound Transducer, 21CFR 892.1570, 90-ITX
Predicate Device K220857 X-CUBE 50, X-CUBE 60 Ultrasonic Pulsed Doppler Imaging System
- Reference K221093 X-CUBE 70, X-CUBE 90 Ultrasonic Pulsed Doppler Imaging System Devices
- Proposed New The new device are as follows compared to the cleared Predicate devices. Device - X-CUBE 50 Change model name: C1-7GT → C1-6C, Add model: C1-6CT - X-CUBE 60 Add models: L10-25H. SL3-19X. C1-6CT. C1-6C - Add accessories: EC2-11H/EV2-11H Disposable Biopsy needle guide
The X-CUBE 50 and X-CUBE 60 products are general purpose ultrasound Device imaging system for medical diagnosis assistance. Description: These products are used as an aid tool to diagnosis, such as a commonly used
{4}------------------------------------------------
ultrasound diagnostic device.
Also X-CUBE 50 and X-CUBE 60 have the same operating principles, intended use, risk grade and design/manufacturing characteristics as reference/predicate devices.
This systems platform provide patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, image quality.
-
Patient population Adult and Pediatric
-
Signal Mode:
2D(B) mode, Harmonic mode (HAR), M mode, Color M mode, Anatomical M mode, Color Flow Doppler(CF) Mode, Power Doppler(PD) Mode, Directional PD mode, Pulsed Wave Doppler(PWD) Mode, Continuous Wave Doppler(CWD) Mode, High PRF(Pulse Repetition Frequency) Doppler mode, Tissue Doppler Imaging(TDI) Mode, 3D/4D mode
-
Combination Mode: B/Color Doppler, B/PWD, B/Color Doppler/PWD
-
Acoustic output track: Track 3
Indications The X-CUBE 50. X-CUBE 60 diagnostic ultrasound systems are intended for For Use: use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Peripheral Vessel(PV); and Urology(including prostate).
And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Color M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode: Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF(Pulse Repetition Frequency) Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode.
The X-CUBE 50, X-CUBE 60 are intended to be used in a hospital or medical clinic.
The difference between the X-CUBE 50 and X-CUBE 60 is the number of Differences Tx/Rx channels. X-CUBE 50 operates with 64 channels, but X-CUBE 60 between Proposed operates with 128 channels. The component constituting the Tx/Rx channel is Devices. the FE(Front End) board. Both devices are in the same except for the FE(Front End) board.
{5}------------------------------------------------
|--|
| Model | ProposedX-CUBE 50,X-CUBE 60ALPINION MedicalSystems Co., Ltd. | PredicateX-CUBE 50,X-CUBE 60ALPINION MedicalSystems Co., Ltd. | ReferenceX-CUBE 70,X-CUBE 90ALPINION MedicalSystems Co., Ltd. |
|---|---|---|---|
| Feature | K223564 | K220857 | K221093 |
| Indications for Use | |||
| - Fetal | √ | √ | √ |
| - Abdominal(Renal&GYN/Pelvic) | √ | √ | √ |
| - Intra-operative(Specify, Neuro) | |||
| - Pediatric | √ | √ | √ |
| - Small Organ(breast, testes, thyroid) | √ | √ | √ |
| - Neonatal Cephalic | √ | √ | √ |
| - Adult Cephalic | √ | √ | √ |
| - Trans-rectal | √ | √ | √ |
| - Trans-vaginal | √ | √ | √ |
| - Musculo-skeletal(Conventional) | √ | √ | √ |
| - Musculto skeletal(Superficial) | √ | √ | √ |
| - Cardiac (Adult) | √ | √ | √ |
| - Cardiac (Pediatric) | √ | √ | √ |
| - Peripheral Vessel | √ | √ | √ |
| - Urology(including prostate) | √ | √ | √ |
| Dimensions and Weight | |||
| Weight(Excluding options) | 70kg | 70kg | 85kg |
| Height | 1310/1670mm | 1310/1670mm | 1,440/1,605 mm |
| Width | 560mm | 560mm | 580 mm |
| Depth | 780mm | 780mm | 835 mm |
| Electrical Power | |||
| Voltage | 100-120V~,200-240V~ | 100-120V~,200-240V~ | 100-120V~,200-240V~ |
| Frequency | 50-60 Hz | 50-60 Hz | 50-60 Hz |
| Power | Max. 600VA | Max. 600VA | Max. 700VA |
| Imaging Modes | |||
| - 2D(B) mode | √ | √ | √ |
| - Harmonic mode | √ | √ | √ |
| - M mode | √ | √ | √ |
| - Color M mode | √ | √ | √ |
| - Anatomical M mode | √ | √ | √ |
| - Color Flow Doppler(CF) mode | √ | √ | √ |
| - Power Doppler (PD)mode | √ | √ | √ |
| - Microvascular Imaging(MVI) | √ | √ | √ |
| - Directional PD mode | √ | √ | √ |
| - Pulsed wave Doppler(PWD) mode | √ | √ | √ |
| - Continuous waveDoppler (CWD) mode | √ | √ | √ |
| - High PRF(PulseRepetition Frequency)Doppler mode | √ | √ | √ |
| - Tissue Dopplerimaging (TDI) mode | √ | √ | √ |
| - 3D/4D mode | √ | √ | √ |
| Imaging Features | |||
| - Xpeed™ | √ | √ | √ |
| - Full SRI™ | √ | √ | √ |
| -Spatial CompoundingImage (SCI) | √ | √ | √ |
| - Panoramic | √ | √ | √ |
| - Stress Echo | √ | √ | √ |
| - Cube Strain™ | √ | √ | √ |
| - Live HQ™ | √ | √ | √ |
| - Needle Vision™/Needle Vision™ Plus | √ | √ | √ |
| - Elastography | √ | √ | √ |
| - Cube view™ | √ | √ | √ |
| - Contrast EnhancedUltrasound (CEUS) | √ | √ | √ |
| - Cube Note | √ | √ | √ |
| - B-STIC(STIC) | √ | √ | √ |
| - Auto EF | √ | √ | √ |
| - AnySlice™ | √ | √ | √ |
| - X + Compare | √ | √ | √ |
| - X + Assistant | √ | √ | √ |
| - Time Intensity Curve(TIC) Analysis | √ | √ | √ |
| - Auto NTMeasurements | √ | √ | √ |
| - X + Auto Biometry | √ | √ | √ |
| - Auto IMTMeasurements | √ | √ | √ |
| Accessories or Kits | |||
| Color printer | √ | √ | √ |
| B/W printer | √ | √ | √ |
| DVD-RW | √ | √ | √ |
| Foot switch | √ | √ | √ |
| Wireless LAN | √ | √ | √ |
| SC1-6 Biopsy guide kit | √ | √ | √ |
| L3-12 Biopsy guide kit | √ | √ | √ |
| L3-12X Biopsy guide kit | √ | √ | √ |
| EV2-11H ReusableBiopsy needle guide | √ | √ | √ |
| EC2-11H/EV2-11HDisposable Biopsyneedle guide | √ | ||
| EN3-10 ReusableBiopsy needle guide | √ | √ | |
| EN3-10 DisposableBiopsy needle guide | √ | √ | |
| VE3-10H ReusableBiopsy needle guide | √ | √ | √ |
| VE3-10H DisposableBiopsy needle guide | √ | √ | √ |
| ECG module / cable | √ | √ | √ |
| Disinfectant & Ultrasound Gel | |||
| Ultrasonic gel | √ | √ | √ |
| Cidex OPA(Disinfectant agaents) | √ | √ | √ |
| Cidex Plus(Disinfectant agaents) | √ | √ | √ |
| Gigasept FF(Disinfectant agaents) | √ | √ | √ |
| Virkon(Disinfectant agaents) | √ | √ | √ |
| Wavicide-01(Disinfectant agaents) | √ | √ | √ |
| AIDAL PLUS(Disinfectant agaents) | √ | √ | √ |
| Cetylcide-G(Disinfectant agaents) | √ | √ | √ |
| Sporicidin(Disinfectant agaents) | √ | √ | √ |
| Thermal, mechanical and electrical safety | |||
| - NEMA UD2, UD3 | √ | √ | √ |
| - AIUM MedicalUltrasound Safety | √ | √ | √ |
| - IEC 60601-1 | √ | √ | √ |
| - IEC 60601-1-2 | √ | √ | √ |
| - IEC 60601-2-37 | √ | √ | √ |
| Transducer Comparison (related to K223564) | |||
| C1-6C | √ | √ | √ |
| C1-6CT | √ | ||
| L10-25H | √ | √ | |
| SL3-19X | √ | √ | |
| - For C1-6C, only the model name has been changed from C1-7GT(K220857).- C1-6CT is the new model. However, except for the slight difference inApplicable frequency, C1-6CT and C1-6C are the same Convex typeprobe and provide the same applicable mode, indications for use, elementsize, element spacing.- L10-25H and SL3-19X are the same transducers as the predicatedevice(K221093). |
{6}------------------------------------------------
510(k) X-CUBE 50 & 60
{7}------------------------------------------------
510(k) X-CUBE 50 & 60
{8}------------------------------------------------
Summary of Non-Clinical Tests:
X-CUBE 50 and X-CUBE 60 have been evaluated for biocompatibility, acoustic output as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. X-CUBE 50, X-CUBE 60 and its application comply with voluntary standards as detailed in this premarket submission.
- IEC60601-1:2005(Third Edition)+CORR.1:2006+CORR.2:2007+A1:2012, � Medical Electrical Equipment - Part 1: General Requirements for Safety
- � IEC60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromaqnetic compatibility - Requirements and tests
- � IEC60601-2-37:2007/AMD1:2015, Medical Electrical Equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- � AAMI/ANSI/ISO10993-1:2009(R)2013, Biological Evaluation of Medical Devices - Part 1:Evaluation and Testing within a risk management process
- � AAMI/ANSI/ISO14971:2007/(R)2010, Medical devices-Application of risk management to medical devices
- � AIUM MUS, Third edition, Medical Ultrasound Safety
- � NEMA UD 2-2004(R2009), Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- � NEMA UD 3-2004(R2009), Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic ultrasound Equipment
The following quality management system measures were applied to the development of X-CUBE 50 and X-CUBE 60:
- Medical Device Risk Management �
- � Requirements Reviews
{9}------------------------------------------------
- � Design Reviews
- � Component Verification
- � Integration Review (System Verification)
- � Performance Testing (System Verification)
- � Safety Testing (Compliance Test)
- � Design Validation
Transducer materials and other patient contact materials are biocompatible.
Summary of Clinical Tests:
The subject of this premarket submission, X-CUBE 50 and X-CUBE 60, did not require clinical studies to support substantial equivalence.
Discussion:
X-CUBE 50 and X-CUBE 60 were compared with the predicate device. The subject devices are in conformance with applicable safety standards.
Therefore, the differences between X-CUBE 50 and X-CUBE 60, and the predicate device would not affect the safety, effectiveness and essential performance.
- Conclusion: The design, development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The devices are designed to conform to applicable medical device safety standards and compliance. Therefore, ALPINION MEDICAL SYSTEMS Co., Ltd. considers X-CUBE 50 and X-CUBE 60 to be as safe, and effective as the predicate device.
N/A