K Number
K222864

Validate with FDA (Live)

Date Cleared
2022-12-21

(90 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CL-R813 is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel.

It is not meant to be used in digital mammography.

Device Description

32-inch 8MP Color LCD Monitor CL-R813, 3840 x 2160 (landscape)

  • Color LCD panel, which has wide view angle, is used for this product. It is designed for medical image display.
  • Luminance stabilization function composed with luminance sensor and luminance control circuit always observes the luminance and makes it stable.
  • Images are faithfully displayed along grayscale characteristics (DICOM GSDF) based on the calibrated data stored to the lookup table of the monitor.
  • Luminance mura correction functions will help achieve uniformity on the whole screen.
AI/ML Overview

The provided text describes a 510(k) summary for the JVCKENWOOD 8MP Color LCD Monitor CL-R813. It is a submission for a display device, not an AI/ML medical device. Therefore, many of the requested categories related to AI/ML device studies (such as sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and specific ground truth types like pathology or outcomes data) are not applicable to this document.

However, I can extract information regarding the device's technical performance and the tests conducted to demonstrate its characteristics.

Here's a summary of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The document lists performance test items from the "Guidance for Industry and FDA Staff: Display Devices for Diagnostic Radiology" and the methods used. It states that the CL-R813 met "predefined criteria" but does not explicitly provide a table of acceptance criteria with numerical thresholds nor does it report specific numerical performance results for the CL-R813 against those criteria. It only states a qualitative finding of equivalence.

Performance Test ItemTest Method(s)Reported Device Performance
a. Spatial resolutionThe bar pattern is displayed and captured by a digital camera equipped with a macro lens. The MTF is calculated with the captured data.Confirmed equivalent display characteristics to predicate device. Meets predefined criteria.
b. Pixel defects (maximum counts, allowed defect types, and locations)ISO 13406-2; IDMS 1.03, 7.6 DEFECTIVE PIXELS. Pixel defects are counted based on the ISO13406-2, 3.4.13 table 3.Confirmed equivalent display characteristics to predicate device. Meets predefined criteria.
c. ArtifactsAAPM-TG18, 4.9 Miscellaneous Tests; IDMS 1.03, 4.6 ARTIFACTS & IRREGULARITIESConfirmed equivalent display characteristics to predicate device. Meets predefined criteria.
d. Temporal responseIDMS 1.03, 10.2.3 GRAY-TO-GRAY RESPONSE TIME. Rise and fall time constants at four grayscale intervals (0-100%, 5-95%, 10-90%, 40-60%) are provided by the panel manufacturer.Confirmed equivalent display characteristics to predicate device. Meets predefined criteria.
e. Luminance (maximum, minimum, achievable, and recommended)Lmin and Lmax on the calibrated luminance are confirmed.Confirmed equivalent display characteristics to predicate device. Meets predefined criteria.
f. Conformance to a grayscale to luminance function (e.g., DICOM GSDF)AAPM-TG18, 4.3.5 Advanced Luminance Response. Luminance response for 256 levels are measured.Confirmed equivalent display characteristics to predicate device. Meets predefined criteria.
(For color displays) m. Color tracking (primary colors and color gamut)Color scale: IDMS 1.03, 6. Gray- and Color-Scale Measurement & IDMS 1.03, 5.4 Color-Signal White. Color gamut volume: IDMS 1.03, 5.31 Volume-Color-Reproduction Capability.Confirmed equivalent display characteristics to predicate device. Meets predefined criteria.
(For color displays) n. Gray tracking (gray shades and white point)AAPM-TG196 Gray Tracking; IEC 62563-1: 2009+AMD1:2016+AMD2:2021, 7.4.9 Greyscale chromaticity evaluation.Confirmed equivalent display characteristics to predicate device. Meets predefined criteria.

2. Sample size used for the test set and the data provenance

Not applicable. This is a display device, not an AI/ML algorithm that processes a "test set" of medical images from patients. The testing involved evaluating the physical characteristics of the monitor hardware.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. Ground truth for an AI/ML algorithm is not relevant for a display hardware submission.

4. Adjudication method for the test set

Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a display device, not an AI assistance tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable.

7. The type of ground truth used

Not applicable. For a display device, the "ground truth" is typically defined by industry standards and objective physical measurements (e.g., spectral characteristics of light, spatial resolution patterns). The document references standards like ISO, IDMS, AAPM-TG18, and IEC.

8. The sample size for the training set

Not applicable. This device does not involve a training set for an AI/ML algorithm.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

JVCKENWOOD Corporation Hideki Tengeiji Senior Manager 3-12, Moriya-cho, Kanagawa-ku Yokohama, Kanagawa 221-0022 Japan

December 21, 2022

Re: K222864

Trade/Device Name: 8MP Color LCD Monitor CL-R813 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: PGY Dated: December 14, 2022 Received: December 15, 2022

Dear Hideki Tengeiji:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,
Jessica Low

Jessica Lamb. Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222864

Device Name 8MP Color LCD Monitor CL-R813

Indications for Use (Describe)

CL-R813 is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel.

It is not meant to be used in digital mammography.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K222864

510(k) SUMMARY

I. Submitter

Submitter:JVCKENWOOD Corporation
3-12, Moriya-cho, Kanagawa-ku,
Yokohama-shi, Kanagawa, 221-0022 Japan
Contact Person:Hideki Tengeiji, Senior Manager
Email: tengeiji.hideki@jvckenwood.com
Tel: +81-45-450-2715
Fax: +81-45-450-1926
Date Prepared:September 22, 2022

II. Device

Name of Device:8MP Color LCD Monitor CL-R813
Common or Usual Name:Display, diagnostic radiology
Classification Name:Medical Image Management and Processing System(Part 892 Radiology Devices Sec. 892.2050)
Regulatory ClassII
Product CodePGY

III. Predicate Device

Name of Device:21.3 inch (54.0cm) Color LCD Monitor CL-R211 (K182539)
Common or Usual Name:Display, diagnostic radiology
Classification Name:Medical Image Management and Processing System(Part 892 Radiology Devices Sec. 892.2050)
Regulatory ClassII
Product CodePGY

JVCKENWOOD

{4}------------------------------------------------

IV. Device Description

32-inch 8MP Color LCD Monitor CL-R813, 3840 x 2160 (landscape)

  • Color LCD panel, which has wide view angle, is used for this product. It is designed for ■ medical image display.
  • Luminance stabilization function composed with luminance sensor and luminance control circuit always observes the luminance and makes it stable.
  • Images are faithfully displayed along grayscale characteristics (DICOM GSDF) based on the calibrated data stored to the lookup table of the monitor.
  • Luminance mura correction functions will help achieve uniformity on the whole screen.

V. Indications for use

CL-R813 is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel. It is not meant to be used in digital mammography.

VI. Cybersecurity

FDA quidance located at https://www.fda.gov/media/86174/download, is followed for cybersecurity concerns.

{5}------------------------------------------------

VII. Comparison of Technological Characteristics with the predicate device

The comparison table below list information obtained from the product catalogs and measurements and different technological characteristics are discussed in it.

Predicate deviceLCD Monitor CL-R211Proposed deviceLCD Monitor CL-R813Explanation ofDifferences
510(k) NumberK182539--
Indication for useCL-R211 is intended to beused in displaying and viewingmedical images for diagnosisby trained medicalpractitioners or certifiedpersonnel. It is not meant tobe used in digitalmammography.CL-R813 is intended to beused in displaying andviewing medical images fordiagnosis by trained medicalpractitioners or certifiedpersonnel. It is not meant tobe used in digitalmammography.-
Display TechnologyIPS LCD panel with TFTactive-matrix array with LEDbacklightIPS LCD panel with TFTactive-matrix array with LEDbacklight-
Screen sizeDiagonal: 21.3"Aspect ratio: 4:3Diagonal: 32"Aspect ratio: 16:9The display area of theproposed device is largerthan that of the predicatedevice.
Backlight typeLEDLED-
Frame rate and refresh rate60Hz60Hz-
Resolution / Pixel array2MP (1200 x 1600)8MP (3840 × 2160)The resolution of theproposed device is higherthan that of the predicatedevice.
Pixel PitchHorizontal: 0.270mmVertical: 0.270mmHorizontal: 0.1845mmVertical: 0.1845mmPixel pitch of theproposed device aresmaller than those of thepredicate device.
Subpixel patternStripe RGBStripe RGB-
Pixel aperture ratio57 %57 %-
Subpixel driving (spatial andtemporal dithering)N/AN/A-
Display InterfaceInput: DVI-D x1DisplayPort x1Output: DisplayPort x1Input: DVI-D x1DisplayPort x2Output: DisplayPort x1-
Video bandwidthDot clock: 162 MHzDot clock: 533.25 MHzSo the resolution of theproposed device is higherthan that of the predicatedevice, the dot clock ofthe proposed device ishigher than that of thepredicate device
User controlDICOM CONFORMANCE TESTConfiguration switchInput signal switchDynamic gammaAUTO TEXTHuman presence sensorEDID switchTest patternUSB PowerConfig SelectBacklightContrastSource SelectHuman SensorAlthough the usercontrols are different, theperformance of theproposed device is thesame as the predicatedevice.
Ambient light sensingBuilt-in Sensor (For correctionduring calibration)Built-in Sensor (Forcorrection during calibration)-
Touch-screen technologyN/AN/A-
Luminance calibration tools /Quality-control proceduresHardware:Integrated sensorExternal sensorSoftware:QA Medivisor/Medivisor NXFCALHardware:Integrated sensorExternal sensorSoftware:QA Medivisor AgentFCAL-
Additional Software/FirmwareN/AN/A-

The above differences in technical characteristics do not affect the safety and the effectiveness of the CL-R813.

JVCKE OD

{6}------------------------------------------------

VIII. Performance Test

The following performance tests were performed on the CL-R813 in accordance with the instructions in "Guidance for Industry and FDA Staff: Display Devices for Diagnostic Radiology" issued on October 2, 2017.

Performance test itemsin the guidanceTest method(s)
a. Spatial resolutionThe bar pattern is displayed and captured by a digitalcamera equipped with a macro lens. The MTF is calculatedwith the captured data.
b. Pixel defects (maximumcounts, allowed defecttypes, and locations)ISO 13406-2IDMS 1.03, 7.6 DEFECTIVE PIXELSPixel defects are counted based on the ISO13406-2, 3.4.13table 3.
c. ArtifactsAAPM-TG18, 4.9 Miscellaneous TestsIDMS 1.03, 4.6 ARTIFACTS & IRREGULARITIES
d. Temporal responseIDMS 1.03, 10.2.3 GRAY-TO-GRAY RESPONSE TIMERise and fall time constants at four grayscale intervals (0-100%, 5-95%, 10-90%, 40-60%) are provided by thepanel manufacturer.
e. Luminance (maximum,minimum, achievable, andrecommendedLmin and Lmax on the calibrated luminance are confirmed.
f. Conformance to a grayscaleto luminance function (e.g.,DICOM GSDF)AAPM-TG18, 4.3.5 Advanced Luminance ResponseLuminance response for 256 levels are measured.
(For color displays)m. Color tracking (primarycolors and color gamut)Color scale:IDMS 1.03, 6. Gray- and Color-Scale MeasurementIDMS 1.03, 5.4 Color-Signal WhiteColor gamut volume:IDMS 1.03, 5.31 Volume-Color-Reproduction Capability
(For color displays)n. Gray tracking (gray shadesand white point)AAPM-TG196 Gray TrackingIEC 62563-1: 2009+AMD1:2016+AMD2:2021, 7.4.9Greyscale chromaticity evaluation

As a result of the performance test, it was confirmed that the CL-R813 has the equivalent display characteristics as the predicate device, CL-R211. The display characteristics of the CL-R813 meet the predefined criteria.

The CL-R813 has not been tested on animal or clinical testing.

IX. Conclusions

As shown above, the intended use of the subject device and the predicate device is identical, the technical characteristics are similar, and any differences in the characteristics do not affect the safety or effectiveness of the device. The results of the performance testing and the verification and validation demonstrate that the subject device is substantially equivalent to the predicate device currently on the market.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).