(107 days)
Not Found
No
The device description and intended use focus on mechanical components and safety features for blood collection, with no mention of AI or ML.
No
The device is primarily intended for blood specimen collection and short-term, single infusions, not for treating a disease or condition.
No
Explanation: The device is a Safety Push Button Blood Collection Set, designed for collecting blood specimens and for short-term infusions. It is used to obtain blood, not to analyze or interpret the blood for diagnostic purposes.
No
The device description clearly lists multiple physical components (Needle, Spring, Butterfly wings, Hub, etc.) and describes physical characteristics and testing related to these components and their function. There is no mention of software as a component or function of the device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The primary intended use is for the collection of blood specimens from a patient's vascular system. While the collected blood may be used for in vitro diagnostic testing, the device itself is a tool for collecting the sample, not for performing the diagnostic test on the sample.
- Device Description: The components described are related to the physical act of accessing the bloodstream and collecting blood. There are no components or features mentioned that would indicate the device performs any analysis or testing of the blood itself.
- Performance Studies: The performance studies focus on the physical and functional aspects of the device (biocompatibility, strength, sterility, safety features) and its ability to collect blood safely and effectively. There are no studies related to the diagnostic performance of the device on blood samples.
- Lack of IVD Characteristics: IVD devices typically involve reagents, assays, or analytical components that interact with a biological sample to provide diagnostic information. This device lacks these characteristics.
In summary, the device is a blood collection tool, which is a necessary step for many IVD procedures, but it is not an IVD device itself.
N/A
Intended Use / Indications for Use
The Safety Push Button Blood Collection Set is intended to be used for insertion into a patient's vascular system for blood specimens. When used without the male adapter, the device allows the clinician to obtain blood sampling from the female hub with a syringe, if necessary, or can be used for short-term (up to 2 hours), single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and remain under the direct supervision of a clinician.
It aids in the prevention of accidental needle stick through the use of an active safety feature.
The device is a sterile, multiple sample, single-use.
Product codes
JKA, FPA
Device Description
The proposed device consists of these components: (1) Needle, (3) Spring, (4) Butterfly wings, (5) Hub, (6) Safety sheath, (7) Flexible tube, (8) Female luer, (9) Male luer hub, (10) Puncture needle, (11) Rubber sleeve, (12) Puncture needle protect cover or holder, (13) Safety push button protect cover.
The proposed device have three configurations: With holder, With holder, Without male adapter. For without male adapter, there is a a protective cap to cover female luer. The proposed device is available various needle size and tube length. The range of needle size is from 25G(0.5mm). The range of tube length is from 10cm(4") to 30cm(12").
The proposed device have sharps injury protection features. It aids in the prevention of accidental needle stick through the use of an active safety feature.
The proposed device is a sterile, single use device. It is sterilized by Ethylene Oxide Gas (EtO) to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.
No DEHP, BPA and Natural Rubber Latex are added in the proposed device.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
vascular system
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Healthcare facilities by trained medical professional
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies applicable parts of the following standards:
- ISO 10993-7:2008 AMD.1:2019 Biological evaluation of medical devices- Part 7: Ethylene Oxide Sterilization Residuals
- ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
- ISO 10993-11:2017 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity
- USP Pyrogen Test
- ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
- ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration.
- ASTM F88/F88M-21 standard method for seal strength of flexible barrier materials
- ASTM F1886/F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- USP Bacterial Endotoxins Test
- ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications Part 7:Connectors for intravascular or hypodermic applications
- ISO 8536-4:2019 Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed
- ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- ISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods
In addition, the following verification or testing is performed to meet the stated performance requirements:
- Extraction force testing in simulation using
- Simulated Clinical Study for safety feature
- Sharps injury protection testing in accordance with ISO 23908
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 862.1675 Blood specimen collection device.
(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 5, 2023
Jiangsu Caina Medical Co.,Ltd Jianwei Pan Regulatory Affairs No.23, Huanxi Rd, Zhutang Town Jiangyin, Jiangsu 214415 China
Re: K222834
Trade/Device Name: Safety Push Button Blood Collection Set Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA, FPA Dated: November 28, 2022 Received: December 5, 2022
Dear Jianwei Pan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
1
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
David Walloschek
David Wolloscheck, Ph.D. For Joyce M. Whang, Ph.D. Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known)
Device Name Safety Push Button Blood Collection Set
The Safety Push Button Blood Collection Set is intended to be used for insertion into a patient's vascular system for blood specimens. When used without the male adapter, the device allows the clinician to obtain blood sampling from the female hub with a syringe, if necessary, or can be used for short-term (up to 2 hours), single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and remain under the direct supervision of a clinician.
It aids in the prevention of accidental needle stick through the use of an active safety feature.
The device is a sterile, multiple sample, single-use.
× Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/1 description: The image shows the logo for CAINA. The logo is in blue and green. The word "CAINA" is in large, bold, blue letters. Below the word "CAINA" is a tagline in smaller letters.
K222834 - 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K222834
-
- Date of Preparation: Jan. 03, 2023
-
- Sponsor Identification
Jiangsu Caina Medical Co., Ltd. No.23, Huanxi Road, Zhutang Town, Jiangyin, Jiangsu, 214415, China
Establishment Registration Number: 3005670221
Contact Person: Jun Lu Position: General Manager Tel: +86-510-86205183 Fax: +86-510-86215183 Email: jun.lu(@cainamed.com
-
- Designated Submission Correspondent
Mr. Jianwei pan (Primary Contact Person) Email: jianwei.pan@cainamed.com Ms. Tracy Gong (Alternative Contact Person) Email: tracy.gong@cainamed.com Tel: +86-510-86866666-8027 Fax: +86-510-86866666-8009
- Designated Submission Correspondent
-
Identification of Proposed Device 4.
Trade Name: Safety Push Button Blood Collection Set
Common name: Blood Collection Set
Regulatory Information Classification Name: Blood specimen collection device Class: II
Regulation Number: 21 CFR 862.1675 Product Code: JKA and FPA Review Panel: General Hospital
4
Image /page/4/Picture/1 description: The image shows the word "CAINA" in large, blue, sans-serif font. The "A" in the word is stylized with a green triangle in the middle. Below the word "CAINA", there is some text that is too small to read.
Indications for Use Statement:
The Safety Push Button Blood Collection Set is intended to be used for insertion into a patient's vascular system for blood specimens. When used without the male adapter, the device allows the clinician to obtain blood sampling from the female hub with a syringe, if necessary, or can be used for short-term (up to 2 hours), single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and remain under the direct supervision of a clinician. It aids in the prevention of accidental needle stick through the use of an active safety feature. The device is a sterile, multiple sample, single-use.
న్. Device Description
The proposed device consists of these components: (1) Needle, (3) Spring, (4) Butterfly wings, (5) Hub, (6) Safety sheath, (7) Flexible tube, (8) Female luer, (9) Male luer hub, (10) Puncture needle, (11) Rubber sleeve, (12) Puncture needle protect cover or holder, (13) Safety push button protect cover.
The proposed device have three configurations: With holder, With holder, Without male adapter. For without male adapter, there is a a protective cap to cover female luer. The proposed device is available various needle size and tube length. The range of needle size is from 25G(0.5mm). The range of tube length is from 10cm(4") to 30cm(12").
The proposed device have sharps injury protection features. It aids in the prevention of accidental needle stick through the use of an active safety feature.
The proposed device is a sterile, single use device. It is sterilized by Ethylene Oxide Gas (EtO) to achieve a SAL of 106 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.
No DEHP, BPA and Natural Rubber Latex are added in the proposed device.
-
- Identification of Predicate Device
Predicate Device
- Identification of Predicate Device
510(k) Number: K153309
Product Name: BD Vacutainer UltraTouch Push Button Blood Collection Set
7. Substantially Equivalent (SE) Comparison
Table 1 Comparison of Technology Characteristics with K153309
ITEM | Proposed Device | Predicate Device | Comment |
---|---|---|---|
Product code | JKA and FPA | JKA and FPA | Same |
5
Image /page/5/Picture/1 description: The image shows the word "CAINA" in blue letters. The "A" in the word has a green triangle in the middle of it. The font is sans-serif and the letters are bold.
| Regulation
No. | 21 CFR 862.1675 | 21 CFR 862.1675 | Same |
---|---|---|---|
Regulation | |||
Name | Blood specimen collection device | Blood specimen collection device | Same |
Class | II | II | Same |
Indications | |||
for use | The Safety Push Button Blood | ||
Collection Set is intended to be used | |||
for insertion into a patient's vascular | |||
system for blood specimens. When | |||
used without the male adapter, the | |||
device allows the clinician to obtain | |||
blood sampling from the female hub | |||
with a syringe, if necessary, or can | |||
be used for short-term (up to 2 | |||
hours), single infusions with | |||
consideration given to patient size | |||
and appropriateness for the solution | |||
being infused. The device is not to | |||
be left in place and remain under the | |||
direct supervision of a clinician. It | |||
aids in the prevention of accidental | |||
needle stick through the use of an | |||
active safety feature. The device is a | |||
sterile, multiple sample, single-use. | The BD Vacutainer® UltraTouch™ | ||
Push Button Blood Collection Set is | |||
a sterile, multiple sample, single-use | |||
fixed winged blood collection set | |||
intended for venipuncture to obtain | |||
blood specimens from patients. | |||
When used without the male | |||
adapter, the device allows the | |||
clinician to obtain blood sampling to | |||
the female hub with a syringe, if | |||
necessary, or can be used for | |||
short-term, single infusions with | |||
consideration given to patient size | |||
and appropriateness for the solution | |||
being infused. The device is not to | |||
be left in place and remain under the | |||
direct supervision of a clinician. | |||
The recommended use of the device | |||
is to activate the needle safety | |||
feature prior to removal from the | |||
venipuncture site. The retraction of | |||
the intravenous (IV) end of the | |||
needle aids in the prevention of | |||
accidental needlestick injury. | Different | ||
See | |||
Comment 1 | |||
Configuration | Three configuration: | ||
Without holder, | |||
With holder, | |||
Without male adapter | Two configuration: | ||
with an integrated male luer adapter | |||
which connects to a holder, | |||
without the male luer adapter | Different | ||
See | |||
Comment 2 | |||
Consisted of | (1) Needle cap, (2) Needle, (3) | ||
Spring, (4) Butterfly wings, (5) | |||
Hub, (6) Safety sheath, (7) Flexible | |||
tube, (8) Female luer, (9) Male luer | |||
hub, (10) Puncture needle, (11) | |||
Rubber sleeve, (12) Puncture needle | |||
protect cover or holder, (13) Safety | Needle protector, | ||
Stainless steel cannula (Intravenous | |||
end and Non-patient end of | |||
cannula), | |||
Stainless steel spring, | |||
Hub, front and rear barrel, | |||
Wings, | |||
Tubing, | |||
Female luer connector and an | Same | ||
push button protect cover. | optional male luer adapter, | ||
Luer Cap | |||
patient | |||
contact | |||
material | Stainless steel SUS304, | ||
Methyl methacrylate acrylonitrile | |||
butadiene styrene plastics(MABS), | |||
Acrylonitrile Butadiene Styrene | |||
Copolymer(ABS), | |||
Polyvinyl chloride(PVC), | |||
Isoprene Rubber, | |||
Polydimethysiloxane, | |||
Color additive | Polypropylene, | ||
Stainless Steel 304, | |||
PVC, | |||
Silicone, | |||
ABS, | |||
Isoprene Rubber | See | ||
Comment 3 | |||
Expiration | |||
Date | 3 years | 2 years | See |
Comment 4 | |||
Sterile | Sterile | Sterile | Same |
Sterile | |||
method | EtO Sterilized | Gamma | See |
Comment 5 | |||
SAL | 10-6 | 10-6 | Same |
Single use | Yes | Yes | Same |
Environment | |||
of use | Healthcare facilities by trained | ||
medical professional | Healthcare facilities by trained | ||
medical professional | Same | ||
Needle size | Needle gauge 25G,23G,22G,21G, | ||
20G, | |||
Needle length 1/2", 5/8", 3/4" | Needle gauge 25G,23G,21G | ||
Needle length 3/4" | See | ||
Comment 6 | |||
Tube length | 10cm(4"), 180cm(7"), 30cm(12") | 7" and 12" | See |
Comment 6 | |||
principles of | |||
operation | When vacuum blood collection tube | ||
is use, blood flow into blood | |||
collection tube via the device under | |||
the action of differential pressure. | When vacuum blood collection tube | ||
is use, blood flow into blood | |||
collection tube via the device under | |||
the action of differential pressure. | Same | ||
Activation | |||
mode | Manual, active safety feature, | ||
retraction by spring | Manual, active safety feature, | ||
retraction by spring | Same |
6
Image /page/6/Picture/1 description: The image shows the logo for CAINA. The logo is in blue and green. The text "CAINA" is in large, bold, blue letters. Below the text is the website address "www.cainamed.com".
Comment 1
The proposed indications for use, are similar to the predicate indications for use. The difference is commented as follows:
-the short term is clearly defined for up to 2 hours in indications for use of proposed device. It does not raise new questions of safety/effectiveness, however users are more aware of the safe use time.
-in predicate indications for use, the recommended use is to activate the needle safety feature prior to removal from the venipuncture site. The needle safety feature can be activated before or after the needle leaves the body, which will not cause new safety/effectiveness. It aids in the prevention of accidental needle stick through the use of an active safety feature.
So the differences are not a significant change.
7
Comment 2
The predicate device is two configuration: with an integrated male luer adapter which connects to a holder(without holder) and without the male luer adapter is intended to installed the holder and held during operation. The proposed device is expanded into two configurations, pre installed holder and without holder. This gives users more choices. The pre installed holder do not change any function of proposed device. So the differences are not a significant change.
Comment 3
The patient contact materials for the proposed device are different from predicate device. According to guidance, Use of International Standard ISO 10993-1 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", the proposed device is external communicating device of Blood path, indirect and limited contact. The Cytotoxicity test, Sensitization test, Irritation test, Acute systemic toxicity test, Pyrogen test, Hemolysis test have been performed for proposed device. Therefore, this material difference does not affect substantially equivalence on safety and effectiveness.
Comment 4
The expiration date for the proposed device is different from predicate device. The proposed devices have been performed 3 years accelerated aging and demonstrated that the aged samples also complied with the requirements of relevant performance standards. The ability of immediate package of the proposed device to maintain the device in a sterile state for a period of 3 years has been validated in accordance with ISO 11607 and ISTA 3A. Therefore, this expiration date difference does not affect substantially equivalence on safety and effectiveness.
Comment 5
The sterile method of proposed device is Eto, the sterile method of predicate device is Gamma. Both of them achieve a Sterility Assurance Level (SAL) of 10-5. Examination of the Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) residuals have met with ISO 10993-7,AMD1:2019, this sterilization difference does not affect substantially equivalence on safety and effectiveness.
Comment 6
The needle size and tube length is different between proposed device and predicate device. The proposed device have expanded of needle and tube range. These extended devices have been verified according to ISO 9626, ISO 7864, ISO 8536-4. The difference does not affect substantially equivalence on safety and effectiveness.
8. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies applicable parts of the following standards:
8
-
ISO 10993-7:2008 AMD.1:2019 Biological evaluation of medical devices- Part 7: Ethylene Oxide Sterilization Residuals
-
ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity
-
ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
- ISO 10993-11:2017 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity >
-
USP Pyrogen Test
-
ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
-
ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration.
-
ASTM F88/F88M-21 standard method for seal strength of flexible barrier materials
-
ASTM F1886/F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
-
USP Bacterial Endotoxins Test
-
ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications Part 7:Connectors for intravascular or hypodermic applications
-
ISO 8536-4:2019 Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed
-
ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
-
ISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods
In addition, the following verification or testing is performed to meet the stated performance requirements:
-
Extraction force testing in simulation using
-
Simulated Clinical Study for safety feature
-
Sharps injury protection testing in accordance with ISO 23908
-
- Clinical Test Conclusion
No clinical study was included in this submission.
-
- Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device in intended use, principles of operation, technology, design, materials and performance.
- Substantially Equivalent (SE) Conclusion