K Number
K222703
Manufacturer
Date Cleared
2023-05-19

(254 days)

Product Code
Regulation Number
890.3800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.

Device Description

ATTO Mobility Scooter is an electrical battery powered, three-wheeled scooter intended to provide mobility for elderly or disabled individuals in a variety of indoor and outdoor settings. The ATTO is a collapsible electric scooter which has a front-wheel brushless DC hub motor. The scooter can be collapsed in 12 seconds and can also be disassembled with ease. The scooter controlled through a thumb throttle and protected by an electronic-release brake system, has a driving range up to 20 km between charges. It is capable of carrying a driver weighing up to 136 (for ATTO MAX) kg. It moves both directions and handles a 6° slope. It can come to a full stop within 1.1m when on a horizontal plane and within 2.2 m when on a 6° slope.

AI/ML Overview

This document is a 510(k) summary for the ATTO Mobility Scooter and its variations (ATTO SPORT, ATTO SPORT MAX). It describes the device, its intended use, a comparison to a legally marketed predicate device, and the testing performed to demonstrate substantial equivalence.

Here's the breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document states that "ATTO performance and safety testing is conducted according to EN 12184:2014. All Performance testing met the predetermined acceptance values." However, specific acceptance criteria values and the device's reported performance for each criterion are not explicitly listed in a table format within this submission. Instead, there's a comparison table (Section G) that shows some specifications for the proposed device and its predicate.

To extract acceptance criteria and reported performance, we can infer some from the predicate comparison and the standards mentioned. The document doesn't explicitly state "acceptance criteria" for each performance parameter but rather implies that the device met the requirements of the standards and performed as expected.

Feature / Performance ParameterPredicate ATTO Mobility Scooter (K160909)Proposed ATTO Mobility Scooter (K222703)Implied Acceptance Criteria (Based on comparison/standards compliance)Reported Device Performance (K222703)
Max Load100 kg120 kg (ATTO, ATTO Sport), 136 kg (ATTO SPORT MAX)Must meet or exceed predicate, and comply with safety standards for load capacity.120 kg (ATTO, ATTO Sport), 136 kg (ATTO SPORT MAX)
Maximum speed (forward)6.4 km/h6.4 km/h (ATTO, ATTO Max), 9.6 km/h (Max SPORT)Must function as expected and be safe. Consistency or improvement over predicate.6.4 km/h (ATTO, ATTO Max), 9.6 km/h (Max SPORT)
Maximum speed (reverse)4 km/h4 km/hMust function as expected and be safe. Consistent with predicate.4 km/h
Travel distance15 km20/17 kmMust meet or exceed predicate.20/17 km
Braking TimeNot publicly available0.5sMust be safe and within acceptable limits for mobility scooters.0.5s
Braking DistanceNot publicly available1.1m (horizontal plane), 2.2m (6° slope)Must be safe and within acceptable limits for mobility scooters.1.1m (horizontal plane), 2.2m (6° slope)
Slope handling(Implied from general use)Handles a 6° slopeMust enable safe operation on slopes as specified.Handles a 6° slope
Full stop withinN/A1.1m (horizontal), 2.2m (6° slope)Must comply with safety standards for braking effectiveness.1.1m (horizontal), 2.2m (6° slope)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a "sample size used for the test set" or the "data provenance" (country of origin, retrospective/prospective). The testing described is non-clinical performance and safety testing of the device itself, rather than a study involving human subjects or a dataset derived from patient information.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable. The submission describes non-clinical performance and safety testing of a mobility scooter against engineering standards (ISO 7176 series and EN 12184:2014). It does not involve "experts" in the sense of medical professionals establishing a "ground truth" for a diagnostic or AI-driven decision. The "ground truth" here is adherence to technical specifications and safety standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable. As explained above, this is non-clinical device testing, not a diagnostic study requiring expert adjudication of ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The submission is for a mobility scooter, which is not an AI-assisted diagnostic device or one that involves human "readers" or image interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is adherence to national and international safety and performance standards (ISO 7176 series and EN 12184:2014) and the manufacturer's own design specifications, which are benchmarked against the predicate device. It's based on objective measurements and engineering principles, not medical expert consensus, pathology, or outcomes data.

8. The sample size for the training set

This section is not applicable. There is no "training set" as this is not an AI/machine learning device. The testing pertains to the physical device.

9. How the ground truth for the training set was established

This section is not applicable. As there is no training set, there is no ground truth for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.

May 19, 2023

Moving Life Ltd Igor Naroditsky RA Adviser Havat Alenby Netzer Sereni, 7039500 Israel

Re: K222703

Trade/Device Name: ATTO Mobility Scooter (ATTO):ATTO Mobility Scooter (ATTO SPORT);ATTO Mobility Scooter (ATTO SPORT MAX) Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: April 17, 2023 Received: April 17, 2023

Dear Igor Naroditsky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tushar Bansal -S

Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222703

Device Name

ATTO Mobility Scooter (ATTO) ATTO Mobility Scooter (ATTO SPORT) ATTO Mobility Scooter (ATTO SPORT MAX)

Indications for Use (Describe)

ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY (K222703)

ATTO Mobility Scooter

A. Contact Details

Applicant Name: Moving Life Ltd Havat Allenby, Netzer Sereni, 7039500 Israel Contact Name: Igor Naroditsky, Regulatory Adviser Phone: + 972 54 4384386 Date Prepared: Aug 15, 2022

B. Device Name:

Trade Name: ATTO Mobility Scooter

Models: ATTO, ATTO SPORT, ATTO SPORT MAX

Common Name: Vehicle, Motorized 3-Wheeled

Classification Name: Motorized three-wheeled vehicle; INI; 890.3800

C. Legally Marketed Predicate Device(s)

510(k) NumberProduct CodeTrade NameManufacturer
K160909INIATTO Mobility ScooterMoving Life Ltd

D. Device Description

ATTO Mobility Scooter is an electrical battery powered, three-wheeled scooter intended to provide mobility for elderly or disabled individuals in a variety of indoor and outdoor settings. The ATTO is a collapsible electric scooter which has a front-wheel brushless DC hub motor. The scooter can be collapsed in 12 seconds and can also be disassembled with ease.

The scooter controlled through a thumb throttle and protected by an electronic-release brake system, has a driving range up to 20 km between charges. It is capable of carrying a driver weighing up to 136 (for ATTO MAX) kg. It moves both directions and handles a 6° slope. It can come to a full stop within 1.1m when on a horizontal plane and within 2.2 m when on a 6° slope.

The differences between the ATTO Mobility Scooter and its predicates are:

  • । Introduction of the two additional models, ATTO SPORT, and ATTO SPORT MAX with
  • । Additional Manual Front Wheal brake for extra safety

{4}------------------------------------------------

  • Software Speed control feature (in ATTO SPORT, and ATTO SPORT MAX only) ।
  • -Enhanced LCD display
  • -Extra wide rear wheel
  • -Rear light handle
  • Powerful LED Headlights -
  • -Shock-absorbing airless NPT tires
  • -Regenerative braking control system
  • Modified controller -
  • -Modified charger and battery

ய் Intended Use/indications for use

The Moving Life ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.

F. Substantial Equivalence Comparison

The proposed ATTO Mobility Scooter is as safe and effective as the ATTO Mobility Scooter (K160909) by Moving Life. The Scooter has the same intended uses and same indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Moving Life ATTO Mobility Scooter and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Moving Life ATTO Mobility Scooter is as safe and effective as the ATTO Mobility Scooter (K160909) by Moving Life. Thus, the proposed ATTO Mobility Scooter is substantially equivalent.

ModelPredicate ATTO MobilityScooter (K160909)Proposed ATTO Mobility Scooter(K222703)
Indication for UseThe Moving Life ATTO MobilityScooter is an indoor/outdoorscooter that providestransportation for a disabled orelderly person.The Moving Life ATTO MobilityScooter is an indoor/outdoorscooter that providestransportation for a disabled orelderly person.
Front wheel size203 X 27 mm8 in
Rear wheel size228 X 57 mm9 in
Ground clearance100 mm100 mm
Number of wheels33
Length1200 mm1200 mm
Width560 mm560 mm
Max load100 kg120 kg for ATTO and ATTO Sport.136 kg for ATTO SPORT MAX
Required width ofangled corridor1000 mm1000 mm

G. Substantial Equivalence Chart

{5}------------------------------------------------

ModelPredicate ATTO MobilityScooter (K160909)Proposed ATTO Mobility Scooter(K222703)
Turn radius1350 mm1350 mm
BatteryLi-Ion battery 48V 5.2 ahLi-ion Battery 5.2 Ah; DC 48 V;249.6 Wh
ChargerAC input.100-240 volt,50/60Hz, DC output 54.6 volt.2 Amp for lithium-ion batteryAC input. 100-240 volt,50/60Hz, DC output 54.6 volt. 2Amp for lithium-ion battery
Type of controllerMicroprocessor based,proportional and integralcontroller which continuallymonitors the ATTO systems.Microprocessor based,proportional and integral controllerwhich continually monitors theATTO systems.
Maximum speed(forward)6.4 km/h6.4 km/h for ATTO and ATTOMax. 9.6 km/h for Max SPORT
Maximum speed(reverse)4 km/h4 km/h
Travel distance15 km20/17 km
Electrical System:48-volt DC48-volt DC
Drive systemFront wheel, direct drive.Front wheel, direct drive.
Motor48-volt DC, Brushless electricmotor48-volt DC, Brushless electric motor
BrakeSpring Applied ElectricalReleased Brakes withfreewheel modeSpring Applied ElectricalReleased Brakes with freewheelmode with regenerative brakingcontrol system. Additional frontdisk brake for extra safety
TillerAdjustable locking for drivingcomfortAdjustable locking for drivingcomfort
Braking TimeNot publicly available.5s
Braking DistanceNot publicly available1.1m
Speed ControlVelocity and current controlVelocity and current control
SeatFolding seatFolding seat

H. Non-clinical Testing

Moving Life ATTO Mobility Scooter is designed in accordance with the ISO7176 series including all relevant sections i.e., Section 2, Section 3, Section 4, Section 5, Section 6, Section 9, Section 10, Section 11, Section 14, Section 14, Section 16, Section 21 and Section 25, etc. ATTO performance and safety testing is conducted according to EN 12184:2014. All Performance testing met the predetermined acceptance values. The Moving Life ATTO Mobility Scooter functioned as intended and its functionality observed was as expected.

l. Clinical Testing

No clinical testing is included in this submission.

J. Technological Characteristics

ATTO Mobility Scooter is a same as the commercially available ATTO Mobility Scooter (K160909) by Moving Life and based on the same fundamental technology as the predicate devices. Both are

{6}------------------------------------------------

electric scooters that are battery operated and have automatic braking systems. Batteries and battery chargers are provided with each scooter. Performance parameters of the devices are very similar. There are some minor technology differences between the proposed ATTO Mobility Scooter and the commercially available ATTO Mobility Scooter (K160909) by Moving Life. None of these differences raises new issues of safety and effectiveness.

K. Conclusion

The safety and effectiveness of the ATTO Mobility Scooter were demonstrated by the testing in compliance with national and international standards. The intended use, basic technology, and the features of the ATTO Mobility Scooter are similar to the predicate device. No new issues of safety and effectiveness is raised by the differences between the proposed ATTO Mobility Scooter and its predicate devices.

N/A