K Number
K222453
Date Cleared
2022-11-04

(81 days)

Product Code
Regulation Number
882.1320
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Self-Adhesive Electrode is intended for as a reusable, conductive interface between the patient's skin and the marketed electrical stimulators (i.e.TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF or IFC(Interferential current),or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current, for OTC (Over-The-Counter) or Prescription use.

Device Description

Self-Adhesive Electrodes are composed of conductive carbon film and adhesive Electrodes composed of Aluminum foil film,Adhesive Electrodes manufactured by Comytens are multi-layer reusable, flexible structures composed of laminated materials commonly used in this application:

  1. Top layer: Fabric/Foam/ Tan fabric
  2. Middle layer: Conductive film (Silver coated Carbon Film/Carbon film/Aluminum foil film)
  3. Bottom layer: Biocompatible self-adhesive conductive hydrogel
  4. Connection: Leadwire/snap button/magnetic button
  5. double sides adhesive tape: which is used for attaching the non-woven fabrics and conducting film
  6. Release liner: plastic film is a protective layer for the hydrogel.
    The electrodes are designed for single-patient/multiple application use. Because of the adhesive nature of the biocompatible conductive hydrogel, no securing materials are required to secure the device to the patient's skin.
    For the electrical connection, Comytens provides lead wire type and snap button/Magnetic button type.
AI/ML Overview

The medical device in question is a Self-Adhesive Electrode manufactured by Guangdong Comytens Medical Technology Co.,Ltd. This device is intended for use as a reusable, conductive interface between the patient's skin and marketed electrical stimulators (e.g., TENS, EMS, IF/IFC, PGF) for transmitting electrical current, for both OTC and Prescription use.

Here's a breakdown of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document provides a comparison table between the subject device (Self-Adhesive Electrode) and a predicate device (ZMI Self-adhesive electrodes, K180865). This table implicitly defines the acceptance criteria by stating that the subject device's performance is "Same" or "similar" to the predicate, and explicitly lists some technical specifications.

Acceptance Criteria (Measured against Predicate Device)Reported Device Performance (Subject Device)
Intended Use (reusable, conductive interface for electrical stimulators: TENS, EMS, IF/IFC, PGF, transmitting electrical current, OTC/Prescription use)Same
OTC or Prescription UseOTC and Prescription
Target Population (General [Adult])General [Adult]
Shape (compared to various shapes of predicate)Square (40x40mm)
ColorWhite, Red, Black, and Tan
Regulation Number (21 CFR 882.1320)882.1320
Product Code (GXY)GXY
Classification Name (Cutaneous electrode)Cutaneous electrode
Design Feature: Construction Materials (Insulation backing material, Conductive film, Conductive hydrogel)Similar (Subject device has additional layers beyond the three basic ones mentioned for the predicate: conductive carbon fiber lead wire (or snap for Magnetic Style Electrode), double sides adhesive tape, and plastic film)
Electrical Connection (Leadwire, Snap button, Magnetic button)Same
Lead Wire Connector (.080"/0.1" (2mm/2.5mm) female socket or Pin connector)Similar (.080" (2mm) female socket connector for predicate, subject device lists .080"/0.1" (2mm/2.5mm) female socket or Pin connector)
SterilityNon-sterile
ReusabilityReusable
PackagingRe-sealable bag packed
Adhesive TypeSelf-adhesive
Biocompatibility (Complies with ISO10993)Complies with ISO10993
A.C. Impedance (<200 ohms)<200 ohms
Force required to remove wire from electrode (>6 pounds of force)More than 6 pounds of force
Single Patient UseYes

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily describes bench testing and biocompatibility testing.

  • Sample Size for Bench Testing: Not explicitly stated. The document mentions "The Self-Adhesive Electrode was tested..." in several instances (e.g., for electrical & impedance, adhesion, connector & mechanical, packaging, accelerated aging), implying a sample of the manufactured device.
  • Sample Size for Biocompatibility Testing: Not explicitly stated. Biocompatibility tests usually involve a defined number of samples for each test (e.g., in vitro cytotoxicity, irritation, sensitization).
  • Data Provenance: The document does not explicitly state the country of origin of the data. However, the submitter is Guangdong Comytens Medical Technology Co., Ltd. from China. The testing itself is described as "Non-Clinical Data" indicating it was conducted in a controlled environment (labs) rather than on human subjects. This is prospective testing, as it's performed specifically to demonstrate compliance for regulatory submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the device is a cutaneous electrode and the acceptance criteria are based on physical, electrical, and biological performance characteristics measured through bench testing and biocompatibility testing, not on human-interpreted data that requires expert ground truth establishment.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable for the reasons stated above (bench and biocompatibility testing). Adjudication methods are typically used for clinical studies involving interpretation of medical images or outcomes where expert consensus is needed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices that assist human readers (e.g., radiologists interpreting images). The Self-Adhesive Electrode is a physical medical device, not an AI diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone (algorithm only) performance study was not done. The device is not an algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is established by objective measurements against recognized international and national standards.

  • Bench Testing: The "ground truth" for electrical, mechanical, and adhesive properties is derived from the specifications outlined in standards (e.g., IEC60601-1, ASTM F1980) and the predicate device's established performance. The device must meet or exceed these values.
  • Biocompatibility Testing: The "ground truth" for biocompatibility is whether the device material elicits a harmful biological response, as assessed by the criteria defined in ISO 10993-1 and subsequent specific tests (Cytotoxicity, Skin Sensitization, Skin Irritation). Failure to pass these tests would indicate a deviation from acceptable biological safety.

8. The sample size for the training set

This information is not applicable. The device is not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 4, 2022

Guangdong Comytens Medical Technology Co., Ltd Mr. Amos Zou RA Manager Building A-102, No. 24, Jiangjunmao Industrial Zone Wulian, Longgang District Shenzhen, Guangdong 518116 China

Re: K222453

Trade/Device Name: Self-adhesive Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: August 12, 2022 Received: August 15, 2022

Dear Mr. Amos Zou:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tushar Bansal -S

Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222453

Device Name Self-adhesive Electrode

Indications for Use (Describe)

Self-Adhesive Electrode is intended for as a reusable, conductive interface between the patient's skin and the marketed electrical stimulators (i.e.TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF or IFC(Interferential current),or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current, for OTC (Over-The-Counter) or Prescription use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K)SUMMARY

1. Submitter of 510(K):

Sponsor
Company Name:Guangdong Comytens Medical Technology Co.,Ltd
Address:Building A-102, No. 24, Jiangjunmao Industrial Zone, Wulian, Longgang District ,Shenzhen, Guangdong, China 518116
Contact person:Deng Jiuzhen
TEL:+86-13418878095
FAX:/
E-mail:info@comytens.com 276148005@qq.com
Date of Prepared:Nov, 3 2022

Application Correspondent:

Company Name:Guangdong Comytens Medical Technology Co.,Ltd
Address:Building A-102, No. 24, Jiangjunmao Industrial Zone, Wulian, Longgang District ,Shenzhen, Guangdong, China 518116
Contact person:Amos Zou
TEL:+86-15015249549
E-mail:Amos.zou@139.copm
Date of Prepared:Nov, 3 2022

2. Proposed Device and code:

Device Trade Name:Self-Adhesive Electrode
Model SeriesSelf-Adhesive Electrodes with lead wire;Self-Adhesive Electrodes with Snap/Magnetic connector;Self-Adhesive Electrode Pads
Regulation Medical SpecialtyNeurology

{4}------------------------------------------------

Product Code:GXY
Regulation number21 CFR 882.1320
Device ClassII
510(K) NumberK222453

Predicate Device: 3.

510(K)Trade NameSubmitterProduct Code
K180865ZMI Self-adhesiveZMI Electronics, Ltd.GXY

Description of Proposed Device: 4.

Self-Adhesive Electrodes are composed of conductive carbon film and adhesive Electrodes composed of Aluminum foil film,Adhesive Electrodes manufactured by Comytens are multi-layer reusable, flexible structures composed of laminated materials commonly used in this application:

No.DescriptionMaterial Composition
1Top layerFabric/Foam/ Tan fabric
2Middle layerConductive film (Silver coated Carbon Film/Carbon film/Aluminum foil film)
3Bottom layerBiocompatible self-adhesive conductive hydrogel
4ConnectionLeadwire/snap button/magnetic button
5double sidesadhesive tapewhich is used for attaching the non-woven fabrics and conducting film
6Release linerplastic film is a protective layer for the hydrogel.

The electrodes are designed for single-patient/multiple application use. Because of the adhesive nature of the biocompatible conductive hydrogel, no securing materials are required to secure the device to the patient's skin.

For the electrical connection, Comytens provides lead wire type and snap button/Magnetic button type:

  • Leadwire assembly:
    The electrode is connected to the electrical stimulator by lead wire, at least 40 mm long wire with a standard connector: 0.08"/0.1"(2mm/2.5mm) diameter female socket connector with insulating outer. By design, the insulated outer jacket prevents the connection to earth or hazardous voltages. The lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR part 898 by testing under ANSI/AAMI ES60601-1, subclause 8.5.2.3.

  • Snap button/Magnetic button assembly:
    With 2.5~5mm diameter male socket. The snap button and the magnetic button provide different types of physical contact with the source of stimulation current via spring tension or magnetic force among male and femaleconnectors. Either method is similar to the lead wire connection providing the friction to create the physicalcontact.

5. Intended for Use

Self-Adhesive Electrode is intended for as a reusable, conductive interface between the patient's skin and the marketed electrical stimulators (i.e.TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF or IFC(Interferential current), or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current, for OTC (Over-The-Counter) or Prescription use.

{5}------------------------------------------------

6. Technical and Performance

The following table compares the device to the predicate device with basic technological characteristics.

Comparison to predicate device:

Element ofComparisonSubject DevicePredicate DeviceComment
510(k)K222453K180865N/A
CompanyGuangdong Comytens MedicalTechnology Co.,LtdZMI Electronics, Ltd.N/A
Device NameSelf-Adhesive ElectrodesZMI Self-adhesive electrodesN/A
Model name● Self-Adhesive Electrodes with leadwire;● Self-Adhesive Electrodes with Snap/Magnetic connector;● Self-Adhesive Electrode PadsSAE TypeN/A
ShapeSquare(40*40mm)Round, Rectangular, Square, Oval, andButterflysimilar1#
ColorWhite, Red, Black, and TanWhite, Red, Black, and TanSame
RegulationNumber882.1320882.1320Same
Product CodeGXYGXYSame
ClassificationNameCutaneous electrodeCutaneous electrodeSame
Intended UseSelf-Adhesive Electrode is intended for asa reusable, conductive adhesive interfacebetween the patient's skin and themarketed electrical stimulators (i.e.TENS(Transcutaneous Electrical NerveStimulation), EMS (Electrical MuscularStimulation), IF or IFC(Interferentialcurrent), RUSS(Russian stimulation),MIC(Microcurrent) or PGF (PulsedGalvanic Stimulation) for transmittingelectrical current, for OTC (Over-The-Counter) or Prescription use.ZMI Self-adhesive electrodes areintended for as a reusable, conductiveadhesive interface between the patient'sskin and the marketed electricalstimulators (i.e. TENS (TranscutaneousElectrical Nerve Stimulation), EMS(Electrical Muscular Stimulation), IF(Interferential) or PGF (Pulsed GalvanicStimulation) for transmitting electricalcurrent, for OTC (Over-The-Counter) orPrescription use.Same
OTC orPrescriptionOTC and PrescriptionOTC and PrescriptionSame
TargetPopulationGeneral [Adult]General [Adult]Same
DesignFeatureSix layers:1. Insulation backing material:Fabric/Foam/Tan fabric2. Conductive film: Aluminum foil filmThree layers:1. Insulation backing material:Fabric/Foam/Tan fabric2.Conductive film: Aluminum foilsimilar2#
/Carbon film/Carbon film coated withsilver/3.Conductive hydrogel4.Connection5.double sides adhesive tape6.Release linerfilm/Carbon film/Carbon film coated withsilver/3.Conductive hydrogel
ElectricalConnectionLeadwireSnap buttonMagnetic buttonLeadwireSnap buttonMagnetic buttonSame
Lead Wireconnector.080"/0.1"(2mm/2.5mm) female socketor Pin connector.080" (2mm) female socket connectorsimilar3#
Non-sterileNon-sterileNon-sterileSame
ReusableReusableReusableSame
PackagingRe-sealable bag packedRe-sealable bag packedSame
AdhesiveTypeSelf-adhesiveSelf-adhesiveSame
BiocompatibilityComplies with ISO10993Complies with ISO10993Same
A.C.Impedance<200 ohms<200 ohmsSame
Forcerequired toremovewire fromelectrodeMore than 6 pounds of forceMore than 6 pounds of forceSame
SinglePatient UseYesYesSame

{6}------------------------------------------------

Comparison in Detail(s):

1#: The shape and size are related to the treatment area may cause higher current density, which leads to discomfort and burns. The current densities for any shaped electrodes exceeding 2mA r.m.s/cm2 may require the special attention of the user. The color is mainly for identification and marketing purpose, and does not raise any new concerns of safety or effectiveness.

If the subject device is manufactured from the identical raw materials manufacturing processes as a predicate device with the same type and duration of tissue contact, and any changes in geometry are not expected to impact the biological response, this is typically sufficient to establish substantially equivalent biocompatibility.

2#: The subject device and predicate device are similar in construction. They both contain three basic components: a nonconductive top layer, a patient contacting layer, a carbondispersion pad middle layer, and; while the Self-Adhesive Electrode also contains three additional layer:conductive carbon fiber lead wire (or snap for Magnetic Style Electrode),double sides adhesive tape which is used for attaching the non-woven fabrics and conducting film,and plastic film is a protective layer for the hydrogel. Which the differences will not affectiveness of the Self-Adhesive Electrode.

3#: The snap button and the magnetic button provide different types of physical contact with the source of stimulation current via spring tension or magnetic force among male and female connectors. Either method is similar to the lead wire connection providing the friction to create the physical contact.

{7}------------------------------------------------

7. Performance Testing:

Test Summary:

To establish substantial equivalence to the identified predicate devices, we performed the following tests on the subject device, Adhesive Electrodes TS serial,and the testing results provide evidence that the device complies with the applicable standards requirement and it is substantially equivalent to the predicate devices. Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.

7.1 Non-Clinical Data:

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrate that the proposed device complies with the following standards:

7.2 Biocompatibility testing

The biocompatibility evaluation for the Self-Adhesive Electrode conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is contacting for duration of less than 24 hours. And the testing included the following tests:

  • Cytotoxicity
  • Skin Sensitization
  • . Skin Irritation

7.3 Safety and EMC

To verify the basic safety and essential performance of the subject device, weperformed the test noted below: - IEC60601-1:2005+AMD1:2012+AMD2:2020; Medical electrical equipment - Part 1: General requirementsfor basic safety and essential performance

The submitted sample(s) complied with the requirements of IEC 60601-1:2005+AMD1:2012+AMD2:2020, Cl.8.5.2.3

7.4 Bench Testing

Bench tests were conducted on Self-Adhesive Electrode to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  1. Electrical & Impedance Tests

The Self-Adhesive Electrode was tested to assure the current is equally distributed. The Self-Adhesive Electrode was also tested for Impedance to assure proper electrical connection to the skin.

  1. Adhesion tests

The Self-Adhesive Electrode was tested for initial adhesion force after using multiple times. 3. Connector & Mechanical tests

The Self-Adhesive Electrode has a Snap or magnetic surface with the purpose to clamp the Self-Adhesive Electrode. This connection was tested to assure multiple use and also to assure the TENS stay fixed to the Self-Adhesive Electrode during use.

  1. Packaging testsReal-Time Shelf Life Test Report

The Self-Adhesive Electrode was tested for storage conditions:

· The Self-Adhesive Electrode should not be stored at temperatures lower than 41°E(+5°C) or higher than 80.6°F (+27°C). 5. Accelerated Aging

The Self-Adhesive Electrode was tested for accelerate aging in worst storage condition to assure within the expiration date the electrode function as expected.

  • ASTM F1980 (Reapproved 2016), Standard guide for accelerated aging of sterile barrier systems for medical devices. (Sterility)

7.5 Clinical data:

No clinical testing was performed

7.6 Summarv

Based on the non-clinical and clinical performance as documented in the device devices were found to have a safety and effectiveness profile that is similar to the predicate device.

8. Conclusions:

The proposed device has the same intended use and similar characteristics as the predicate device, Meanwhile, performance testing, bench testing, and safety report documentation supplied in this submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Guangdong Comytens Medical Technology Co.,Ltd. maintains that the Self-Adhesive Electrode is substantially equivalent to the

{8}------------------------------------------------

predicate device in indications for use, patient population, and environment for use, technology characteristics,
specifications / performance and compliance with internation

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).