K Number
K222429

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-09-09

(29 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ellusa Reusable Bipolar Cable is intended to transfer power from an RF Generator to Bipolar Electrodes.

Device Description

The subject device, Ellusa Reusable Bipolar Cable is a bipolar accessory designed to transfer the electrosurgical power to the Bipolar electrode. It is connected to a High Frequency Generator via the male end and the female end termination connects to the standard Bipolar instrument. The cable is made of Silicone material and Polypropylene H1500 which is flexible and autoclavable. The length of the cable is 3 meters, and it is provided non-sterile.

AI/ML Overview

This document is a 510(k) summary for the Ellusa Reusable Bipolar Cable. It primarily focuses on demonstrating substantial equivalence to a predicate device through comparison of intended use, technological characteristics, and performance, rather than presenting a study demonstrating that the device meets specific performance acceptance criteria from a clinical or diagnostic standpoint.

Based on the provided text, here's what can be extracted:

1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria in terms of clinical or diagnostic performance (e.g., sensitivity, specificity, accuracy). Instead, the "acceptance criteria" are implied by compliance with established medical device standards and demonstrating performance equivalent to the predicate device.

Test / CharacteristicAcceptance Criteria (Implied)Reported Device Performance
Basic Safety & Essential Performance of High Frequency Surgical Equipment and AccessoriesCompliance with IEC 60601-2-2:2017"The following tests were successfully performed: IEC 60601-2-2: 2017 Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories"
Sterilization Validation (Moist Heat)Compliance with ISO 17665-1:2006"-ISO 17665-1:2006 - Sterilization of health care products Moist heat Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices" (Note: Document lists ISO 17665-5 in text, but ISO 17665-1 in the table on page 4, assuming ISO 17665-1:2006 is the correct reference as it is the more general standard for moist heat sterilization).
Substantial EquivalenceEquivalence in intended use, technological characteristics, operating principle, and performance characteristics compared to the predicate device (Ellusa Disposable Bipolar Cables, K202558)."The Ellusa Reusable Bipolar Cables are substantially equivalent in intended use, technological characteristics, operating principle, and performance characteristics to the predicate device by Ellusa (K202558)."The comparison table on page 4 details the similarities (e.g., Product Code, Device Classification, Functions, Operating principle, Energy Source, Cable Length, Activation Method).

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence." Therefore, there is no test set sample size, data provenance, or associated study as typically understood for AI/diagnostic devices. The testing performed was non-clinical (laboratory/engineering) to confirm compliance with safety and performance standards relevant to the device's function as an electrosurgical accessory.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical performance study involving ground truth established by experts was conducted or relied upon.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical performance study involving adjudication of ground truth was conducted or relied upon.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an electrosurgical cable, not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable, as no clinical performance study requiring ground truth was conducted or relied upon. The "ground truth" for this device's performance relies on engineering measurements and compliance with recognized standards (IEC, ISO) for electrical safety, functional performance, and sterilization effectiveness.

8. The sample size for the training set

Not applicable, as this device does not involve a training set (e.g., for machine learning).

9. How the ground truth for the training set was established

Not applicable, as this device does not involve a training set.

{0}------------------------------------------------

September 9, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Ellusa, LLC Ms. Suzanne Lucas Sr. Regulatory Affairs Specialist 2473 Grand Avenue Baldwin, New York 11510

Re: K222429

Trade/Device Name: Ellusa Reusable Bipolar Cable Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 25, 2022 Received: August 11, 2022

Dear Ms. Lucas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222429

Device Name Ellusa Reusable Bipolar Cables

Indications for Use (Describe)

The Ellusa Reusable Bipolar Cable is intended to transfer power from an RF Generator to Bipolar Electrodes.

Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "ellusa" in a stylized, bold, sans-serif font. A swooping graphic element is on the left side of the word, adding a dynamic touch to the overall design. The letters are evenly spaced and have a uniform thickness, contributing to a clean and modern aesthetic. The image is in black and white.

K222429

510(k) SUMMARY

(As required by 21 CFR 807.92(a))

Date Prepared September 7, 2022

Submitter's Information (807.92(a)(1))

Company Name and Address:

Ellusa, LLC 2473 Grand Avenue Baldwin, NY 11510 Phone: (516) 634-1370 Establishment Registration #: 3016087306

Contact Information:

Ms. Suzanne Lucas, BS Sr. Regulatory Affairs Specialist Ellusa, LLC 2473 Grand Avenue Baldwin, NY 11510 Phone: (516) 634-1370 Email: slucas@ellusa.com

Device Information (807.92(a)(2))

Trade Name: Ellusa Reusable Bipolar Cables

Common/Usual Name: Electrosurgical, Cutting & Coagulation Device & Accessories

Classification Name and Requlation: Electrosurgical Cutting and Coaqulation Device and Accessories, 21 CFR 878.4400

Classification Panel: General and Plastic Surgery

Device Class/Product Code FDA Classification: Class 2 FDA Product Code: GEI

Predicate Devices (807.92(a)(3))

  • Ellusa Bipolar Electrodes (K202558) ●
    ellusa LLC. 2473 Grand Avenue, Baldwin, NY 11510 Main: (516) 654-4000 • Fax: (516) 654-8000 www.ellusa.com

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "ellusa" in a stylized, sans-serif font. A curved, swooping line extends from the left, visually connecting to the "e" and adding a dynamic element to the logo. The letters are bold and evenly spaced, creating a clean and modern appearance. The overall design is simple yet distinctive, making it easily recognizable.

K222429

Device Description (807.92(a)(4))

The subject device, Ellusa Reusable Bipolar Cable is a bipolar accessory designed to transfer the electrosurgical power to the Bipolar electrode. It is connected to a High Frequency Generator via the male end and the female end termination connects to the standard Bipolar instrument. The cable is made of Silicone material and Polypropylene H1500 which is flexible and autoclavable. The length of the cable is 3 meters, and it is provided non-sterile.

Intended Use:

The Ellusa Reusable Bipolar Cable is intended to transfer power from an RF Generator to Bipolar Electrodes.

Substantial Equivalence Comparison (807.92(a)(6))

The Ellusa Reusable Bipolar Cables are substantially equivalent in intended use, technological characteristics, operating principle, and perf ormance characteristics to the predicate device by Ellusa (K202558).

Predicate InformationEllusa Reusable Bipolar Cable(SUBJECT DEVICE)Ellusa Disposable Bipolar Cables(K202558)PREDICATE
Intended UseThe Ellusa Reusable Bipolar Cable isintended to transfer power from an RFGenerator.to Bipolar Electrodes.The Ellusa Bipolar Cables are intended totransfer power from an RF Generator toBipolar Electrodes.
Product CodeGEIGEI
Device ClassificationClass IIClass II
FunctionsConnect to High Frequency Generator toprovide power to Bipolar Electrodes.Connect to Electrosurgical Generator toprovide power to Bipolar Electrodes.
Operating principleBipolarBipolar
Energy SourceRF EnergyRF Energy
Cable Length (m)3m3m
Activation MethodFootswtichFootswtich
MaterialsSilicon and Polypropylene H1500PVC
Testing standards usedIEC 60601-2-2:2017IEC 60601-2-2:2017
Steam SterilizationValidationISO 17665-1:2006Not applicable. Device is disposable

ellusa LLC. 2473 Grand Avenue, Baldwin, NY 11510 Main: (516) 654-4000 • Fax: (516) 654-8000 www.ellusa.com

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "ellusa" in a stylized, sans-serif font. To the left of the word, there is a curved, swooping design element that adds a dynamic and modern touch to the overall logo. The text and design are both in black, contrasting with the white background.

K222429

Non-Clinical Testing (807.92(b)(1))

The Ellusa Reusable Bipolar Cables will be manufactured in accordance with the design control requirements of 21 CFR 820.30. Appropriate non-clinical verification and validation activities were planned and conducted to address identified risks and ensure the safety and effectiveness of the device. The following tests were successfully performed:

  • IEC 60601-2-2: 2017 Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surqical accessories
  • -ISO 17665-5:2006 - Sterilization of health care products Moist heat Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices

Technological Characteristics

The devices are substantially equivalent to the predicate devices based on a comparison of physical and performance characteristics.

Clinical Testing (807.92(b)(2))

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

Conclusion (807.92(b)(3))

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the similarity to the predicate devices in terms of technology, performance, and indications for use, Ellusa, LLC concludes that the subject device, Ellusa Reusable Bipolar Cables does not raise any issues of safety or effectiveness and are substantially equivalent to the predicate device as described above.

ellusa LLC. 2473 Grand Avenue, Baldwin, NY 11510 Main: (516) 654-4000 • Fax: (516) 654-8000 www.ellusa.com

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.