(25 days)
The Salvo® Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
The Salvo® Spine System is a thoracolumbosacral pedicle screw system designed to be implanted through a posterior surgical approach. The implants are manufactured from titanium alloy (Ti-6Al-4V) per ASTM F136, commercially pure titanium per ASTM F67, cobalt-chromium alloy per ASTM F1537 and PEEK-OPTIMA. The Salvo® Spine System consists of a variety of screws, rods, and connectors that create a rigid construct as an adjunct to fusion for internal fixation and stabilization of the thoracic, lumbar, and sacral spine.
The provided text is a 510(k) summary for the Salvo® Spine System, which is a medical device. This type of document is a notification to the FDA of the intent to market a medical device, claiming substantial equivalence to an already legally marketed device.
Crucially, this document focuses on mechanical performance and substantial equivalence to a predicate device, not on AI-driven diagnostic or treatment algorithms that would typically involve acceptance criteria and study designs in the way your request outlines.
Therefore, most of the information you're asking for (sample sizes for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth types) is not applicable or present in this document.
The "Performance Data" section explicitly states: "The Salvo® Spine System demonstrated substantially equivalent mechanical performance to the predicate device through dynamic compression bending mechanical testing with reference to ASTM F1717 and a risk analysis compared to the predicate device."
Here's what can be extracted based on the document:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical performance, specifically dynamic compression bending based on ASTM F1717 | Achieved "substantially equivalent mechanical performance" to the predicate device. |
| Risk analysis | Demonstrated substantial equivalence compared to the predicate device. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable/Provided. The performance data relates to mechanical testing of the physical device, not a test set of data like in an AI study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable/Provided. Ground truth for mechanical testing would involve engineering standards and measurements, not expert clinician consensus on medical images or patient outcomes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable/Provided.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is a mechanical device, not an AI diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. There is no algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For mechanical testing: Reference to ASTM F1717 standards would serve as the "ground truth" for evaluating the mechanical properties like dynamic compression bending. The "risk analysis" would also have its own internal criteria for evaluating safety.
8. The sample size for the training set
- Not applicable/Provided. The device is a physical pedicle screw system, not an AI model requiring a training set.
9. How the ground truth for the training set was established
- Not applicable/Provided.
{0}------------------------------------------------
August 29, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
Spine Wave, Inc. Ronald Smith Executive Vice President - Quality, Regulatory & Clinical Affairs Three Enterprise Drive, Suite 210 Shelton, Connecticut 06484
Re: K222362
Trade/Device Name: Salvo® Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: August 3, 2022 Received: August 4, 2022
Dear Ronald Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K222362
Device Name Salvo® Spine System
Indications for Use (Describe)
The Salvo® Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ✓ Prescription Use (Part 21 CFR 801 Subpart D) | ❍ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the words "Spine Wave" in a bold, sans-serif font. The words are arranged horizontally, with "Spine" on the left and "Wave" on the right. A blue wave-like design is located below the words, adding a visual element to the logo. The text is black, providing a strong contrast against the white background.
510(k) Summary Salvo® Spine System
1. Submitter Information
| Submitter: | Spine Wave, Inc. |
|---|---|
| Address: | Three Enterprise DriveSuite 210Shelton, CT 06484 |
| Telephone: | 203-712-1846 |
| Telefax: | 203-944-9493 |
| Contact: | Ronald K. Smith |
| Date Prepared: | August 3, 2022 |
2. Device Information
| Trade Name: | Salvo® Spine System |
|---|---|
| Common Name: | Thoracolumbosacral Pedicle Screw System |
| Classification: | Class II per 21 CFR 888.3070 |
| Classification Name: | Thoracolumbosacral Pedicle Screw System |
| Classification Code: | NKB |
3. Purpose of Submission
The purpose of this submission is to gain clearance for additional connectors to the Salvo® Spine System.
4. Predicate Device Information
The Salvo® Spine System described in this submission is substantially equivalent to the following predicate:
| Primary Predicate Device | Manufacturer | 510(k) No. |
|---|---|---|
| Salvo® Spine System | Spine Wave, Inc. | K202476 |
5. Device Description
The Salvo® Spine System is a thoracolumbosacral pedicle screw system designed to be implanted through a posterior surgical approach. The implants are manufactured from titanium alloy (Ti-6Al-4V) per ASTM F136, commercially pure titanium per ASTM F67, cobalt-chromium alloy per ASTM F1537 and PEEK-OPTIMA. The Salvo® Spine System consists of a variety of screws, rods, and connectors that create a rigid construct as an adjunct to fusion for internal fixation and stabilization of the thoracic, lumbar, and sacral spine.
{4}------------------------------------------------
6. Indications for Use
The Salvo® Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
7. Comparison of Technological Characteristics
The subject Salvo® Spine System has technological characteristics similar to the predicate device, including intended use and indications for use, performance, design, and material composition.
8. Performance Data
The Salvo® Spine System demonstrated substantially equivalent mechanical performance to the predicate device through dynamic compression bending mechanical testing with reference to ASTM F1717 and a risk analysis compared to the predicate device.
9. Conclusion
The indications for use, technological characteristics, and performance data show that the subject Salvo® Spine System is substantially equivalent to the predicate device identified in this submission and does not present any new issues of safety or effectiveness.
N/A