K Number
K222208
Date Cleared
2022-11-17

(115 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

These products are intended to be used in displaying radiological images for review, analysis and diagnosis by trained medical practitioners. They do not support the display of mammography images for diagnosis.

Device Description

The Hisense LCD monitors are intended for trained medical practitioners and provides the image viewing and medical diagnostic functions. HMD2C21A, HMD4C27S, HMD6C30D are developed with different resolution:1600 x 1200, 2560 x 1440, 3280 × 2080. So the LCD monitors can be used in different environment according to different resolution requirement. The three models also developed with same features such as energy saving, ambient light induction, front-facing sensor calibration, etc. In particular, HMD4C27S, HMD6C30D have body-inductive energy-saving and auto awake function.

AI/ML Overview

The provided text describes the acceptance criteria and the study conducted for the Hisense LCD monitors (HMD2C21A, HMD4C27S, HMD6C30D) to demonstrate substantial equivalence to predicate devices.

1. A table of acceptance criteria and the reported device performance

The acceptance criteria for the Hisense LCD monitors are based on demonstrating compliance with performance standards and being substantially equivalent to legally marketed predicate devices. The performance is reported through bench testing in accordance with AAPM Task Group 18 (TG18) guidelines.

Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Measured Guidance)Reported Device Performance (HMD2C21A, HMD4C27S, HMD6C30D Measurements)
a. Spatial resolutionMeasure Spatial resolution with TG18 Resolution
b. Pixel defects (maximum counts, allowed defect types, and locations)Measure Pixel defects with TG18
c. ArtifactsMeasure Artifacts with TG18
d. Temporal responseMeasure Temporal response with TG18
e. Luminance (maximum, minimum, achievable, and recommended)Measure Luminance with TG18
f. Conformance to a grayscale-to-luminance function (e.g., DICOM GSDF)Measure Conformance to a grayscale-to-luminance function with TG18
m. Color tracking (primary colors and color gamut)Measure Color tracking with TG18
Electrical SafetyCompliance with IEC 60601-1
Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2

Note: The document explicitly states "No" for several other TG18 measurements (g. Luminance at 30° and 45° in diagonal, horizontal, and vertical directions at center and four corners; h. Luminance uniformity or Mura test; i. Stability of luminance and chromaticity response with temperature and time of operation or on-time; j. Spatial noise; k. Reflection coefficient; l. Veiling glare or small-spot contrast; n. Gray tracking (gray shades and white point)), indicating these were not considered part of the acceptance criteria for this submission.

2. Sample size used for the test set and the data provenance

The document does not specify a separate "test set" in the context of clinical images or patient data. The performance evaluation was primarily through bench testing of the devices themselves. Therefore, the "sample size" refers to the number of device models tested, which are HMD2C21A, HMD4C27S, and HMD6C30D.

  • Sample size: 3 device models (HMD2C21A, HMD4C27S, HMD6C30D).
  • Data provenance: Not applicable in the context of patient data. The data originates from physical measurements and tests conducted on the manufactured medical display devices. The study is prospective in the sense that the tests were performed on the devices to support the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The study involved bench testing of display performance against established technical standards (AAPM TG18 guidelines, IEC 60601-1, IEC 60601-1-2), not the establishment of ground truth for medical images by human experts. The "ground truth" for the device's technical specifications is the physical characteristics and performance measurements of the monitors in comparison to the predefined standards and predicate devices.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. There was no test set involving human interpretation of medical images that would require an adjudication method. The evaluation was based on objective physical and electrical measurements of the display devices.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. The device is a medical LCD monitor, not an AI-powered diagnostic tool. The submission focuses on the display capabilities of the monitor itself, not on its assistance to human readers or AI performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm-only performance study was not done. The device is a monitor, which by its nature requires a human-in-the-loop (a medical practitioner) to interpret the displayed images. There is no algorithm for diagnostic purposes operating independently on the monitor itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is objective technical specifications and performance metrics defined by international standards and guidelines, such as:

  • AAPM Task Group 18 (TG18) guidelines: For display performance characteristics like spatial resolution, pixel defects, artifacts, temporal response, luminance, conformance to DICOM GSDF, and color tracking.
  • IEC 60601-1: For electrical safety.
  • IEC 60601-1-2: For electromagnetic compatibility (EMC).
  • Comparison to predicate devices: The "ground truth" for substantial equivalence is that the proposed device's characteristics and performance are comparable to (or better than, without raising new questions of safety or effectiveness) those of legally marketed predicate devices.

8. The sample size for the training set

Not applicable. The device is a medical display monitor, not a machine learning or AI algorithm. Therefore, there is no "training set" in the context of machine learning.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

Qingdao Hisense Medical Equipment Co., Ltd. % Lu Zhonghao Quality Engineer No. 399 Songling Road, Laoshan District Oingdao, Shandong, 266100 CHINA

November 17, 2022

Re: K222208

Trade/Device Name: Hisense LCD monitor HMD2C21A, HMD4C27S, HMD6C30D Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: October 7, 2022 Received: October 18, 2022

Dear Lu Zhonghao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb. Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K222208

Device Name Hisense

LCD monitor( HMD2C21A,HMD4C27S,HMD6C30D)

Indications for Use (Describe)

These products are intended to be used in displaying radiological images for review, analysis and diagnosis by trained medical practitioners. They do not support the display of mammography images for diagnosis.

Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section5: 510(k) Summary

(K22208)

1. In accordance with 21 CFR 807.92 the following summary of information is provided:

DateOct. 7, 2022
SubmitterQingdao Hisense Medical Equipment Co.,Ltd.
Address: No. 399 Songling Road, Laoshan District
266100, Qingdao, Shandong, P. R. China
Contact Person:Primary Contact Person
Lu zhonghao
Quality Engineer
Qingdao Hisense Medical Equipment Co. Ltd.
Mail:luzhonghao@hisense.com
Tel: +86-532-83091111
Fax:+86-532-83091111
Secondary Contact Person
Wu yalan
RA Manager
Qingdao Hisense Medical Equipment Co. Ltd.
Mail: wuyalan@hisense.com
Tel: +86-532-83091111
Fax:+86-532-83091111
Device Trade Name:Hisense LCD monitor HMD2C21A, HMD4C27S,HMD6C30D
Common/Usual Name:2M/4M/6M Color LCD monitor
Classification Name:Display, Diagnostic Radiology 21CFR 892.2050
Product Code:PGY
No.Trade Name510(k) NumberProduct CodeManufacturer
Predicate Device(s):1JUSHA-C 270GK183498PGYNanjing Jusha Display Technology Co., Ltd
2C44W+K202374PGYshenzhen Beacon Display Technology Co., Ltd.
3RadiForce RX660K163335PGYEIZO Corporation

{4}------------------------------------------------

Device Description:The Hisense LCD monitors are intended fortrained medical practitioners and provides theimage viewing and medical diagnostic functions.HMD2C21A, HMD4C27S, HMD6C30D aredeveloped with different resolution:1600 x 1200,2560 x 1440, 3280 × 2080. So the LCD monitorscan be used in different environment according todifferent resolution requirement.The three models also developed with samefeatures such as energy saving, ambient lightinduction, front-facing sensor calibration, etc. Inparticular, HMD4C27S, HMD6C30D havebody-inductive energy-saving and auto awakefunction.The Hisense LCD monitor complies withstandards as following:1. EN 60601-1:2006+A1:2013+A12:2014 &IEC60601-1:2005+CORR.1:2006+CORR.2:2007+AM1:2012 Medical electrical equipmentPart 1: General requirements for basic safety andessential performance2. IEC 60601-1-2:2014 & EN 60601-1-2:2015Medical electrical equipment - Part 1-2: Generalrequirements for basic safety and essentialperformance - Collateral standard:Electromagnetic disturbances - Requirements andtests.
Intended Use:These products are intended to be used indisplaying radiological images for review, analysisand diagnosis by trained medical practitioners.They do not support the display of mammographyimages for diagnosis.
Technology:1)HMD2C21A is a 2M color LCD diagnosticmonitor. It strictly complies with the DICOMstandard and has the features of high-resolution,high-brightness display. It uses OSD menusoftware, monitor control software, and qualitycontrol software to achieve functional stability.2) HMD4S27S is a 4M color, one-screen anddual-display diagnostic monitor. It strictly complies
Determination of SubstantialEquivalence:with the DICOM standard and integrates thefunctions of image browsing, reading and writingclinical reports, etc. It uses OSD menu software,monitor control software, and quality controlsoftware to achieve functional stability.3)HMD6C30D is a 6M color, one-screen anddual-display diagnostic monitor. The built-inDICOM standard LUT ensures the long-termstability and consistency of the display, and trulyrestores the original information of the image. Ituses OSD menu software, monitor controlsoftware, and quality control software to achievefunctional stability.
Summary of Non-Clinical Tests:The Hisense LCD monitor (HMD2C21A,HMD4C27S, HMD6C30D) complies withstandards as following:1. EN 60601-1:2006+A1:2013+A12:2014 &IEC60601-1:2005+CORR.1:2006+CORR.2:2007+AM1:2012 Medical electrical equipment Part 1:General requirements for basic safety andessential performance2. IEC 60601-1-2:2014 & EN 60601-1-2:2015Medical electrical equipment - Part 1-2: Generalrequirements for basic safety and essentialperformance - Collateral standard:Electromagnetic disturbances - Requirements andtests.The Hisense LCD monitor HMD2C21A,HMD4C27S, HMD6C30D are substantiallyequivalent to JUSHA-C270G, C44W+, RadiForceRX660. They have equivalent characteristics andfunctions according to comparison table, pleaserefer to 2. Product ComparisonThe following quality assurance measures wereapplied to the development of the system:·Electrical Safety & EMC test·Performance test
Risk Management & Analysis Design Reviews Raw materials verification Final acceptance test etc. Summary of Clinical Tests:The subject of this premarket submission,Hisense LCD monitor, don't require clinicalstudies to support substantial equivalence.The proposed device is Substantially Equivalent(SE) to the predicate device which is US legallymarket device. Therefore, the subject device isdetermined as safe and effectiveness.
ConclusionQingdao Hisense Medical Equipment Co., Ltd.Considers the Hisense HMD2C21A, HMD4C27Sand HMD6C30D LCD monitor to be as safe, aseffective, and performance is substantiallyequivalent to the predicate devices.

{5}------------------------------------------------

{6}------------------------------------------------

2 Product Comparison

This comparison identifies the similarities and differences of the proposed HMD2 C21A, HMD4C27S and HMD6C30D LCD monitor device to the legally marketed predicate JUSHA-C270G, C44W+ and RadiForce RX660 LCD Monitor.

2.1 HMD2C21A and its predicate

Tabel1 HMD2C21A Detailed Comparison with Predicate

Elements ofComparisonProposed devicePredicate Device
ID510(k) NumberK222208K183498
ManufacturerQingdao Hisense MedicalEquipment Co., LtdNanjing Jusha DisplayTechnology Co., Ltd
Devicetype/modelHMD2C21AJUSHA-C270G
1.Display Performance/Specifications
1.1Screen size21,3 inches (540 mm)21.3"
1.2ScreenTechnologyColor LCD panel (IPS)Color TFT LCD Panel
1.3Viewing AngleHorizontal: 178°; vertical: 178°(CR ≥ 10)Horizontal 178°, Vertical 178°
1.4Pixel array1600 pixels (H) x 1200 pixels(V)1600 x 1200/1200 x 1600

{7}------------------------------------------------

IDElements ofComparisonProposed devicePredicate Device
510(k) NumberK222208K183498
ManufacturerQingdao Hisense MedicalEquipment Co., LtdNanjing Jusha DisplayTechnology Co., Ltd
Devicetype/modelHMD2C21AJUSHA-C270G
1.5Display Area432 (H) x 324 (V) (mm)432.0 (H) x 324.0(V) mm
1.6Pixel Pitch0.270 (H) x 0.270 (V) (mm)0.270x0.270 mm
1.7Subpixel drivingNot ApplicableNot Applicable
1.8Contrast Ratio1800:1 (typ.)1400:1
1.9Frame rate60Hz37.9~75kHz;60Hz
1.10MaximumBrightness(typ)1000 cd/m²1000cd/m²
1.11Recommendedbrightness400cd/m²350cd/m²
1.12Backlight typeLEDLED
1.13AmbientlightsensingBuilt-in ambient light sensorAmbient luminanceself-adaptation(ABA)
1.14Response Time16 ms (typ.)16 ms(8ms+8ms)
1.15Aspect Ratio4:34:3
1.16LuminancecalibrationFront-facing sensorUpper Computer SoftwareBuilt in calibration sensorprovided
1.17Touch-screenNot ApplicableNot Applicable
2.Video Signals
2.1InputVideoSignalDVI-D x 1DisplayPort x 1DVI-D x 1,DisplayPort x 1
2.2Output SignalDisplayPort x 1DisplayPort x1
2.3Video bandwidthDVI: 215MHzDisplayPort : 215MHzDVI: 215MHzDisplayPort : 215MHz
3.Power Related Specification
3.1PowerRequirements24 V DC, 1.9ADC 24V
3.2PowerConsumption/save mode45.6W/Below 1W50W/less than 0.5W
3.3PowerManagementDVI DMPMDisplayPort 1.2aDVI DMPMDisplayPort 1.1a
4.Miscellaneous Features/Specifications
IDElements ofComparisonProposed devicePredicate Device
510(k) NumberK222208K183498
ManufacturerQingdao Hisense MedicalEquipment Co., LtdNanjing Jusha DisplayTechnology Co., Ltd
Devicetype/modelHMD2C21AJUSHA-C270G
4.1USB ports1 uplink port and2 downlink ports/ Rev. 2.01 upstream (endpoint),2 downstream/ Rev. 2.0
4.2Display color10bit16-bit
4.3User controlsOff the shelfOff the shelf
4.4Software/Firmware:Built-in embedded softwareBuilt-in embedded software
4.5Dimensions w/ostand(W x H x D)Without base:384 mm x 492mm x 70mmWith base:384 mm x517.5-637.5 mm x 273.5 mmWithout stand: 382mmx490mm x77mmWith stand: 382mm x635mmx238mm
4.6Net weight4.3 kg (excluding the base)7.5 kg(without stand)
4.7VESA standard100 x 100 (mm)100 x 100 (mm)
5. Intended use
5.1Indication for useThis product is intended to beused in displaying radiologicalimages for review, analysis anddiagnosis by trainedmedical practitioners.The product does not supportthe display of mammographyimages for diagnosis.JUSHA-C270G LCD Monitoris intended to be used indisplaying and viewing digitalimages for diagnosis of X-rayor MRI, etc. by trainedmedical practitioners.The device does not supportthe display of mammographyimages for diagnosis.
6 Applicable standard
6.1ApplicablestandardCompliance with IEC 60601-1Compliance with IEC 60601-1-2Compliance with IEC 60601-1Compliance with IEC 60601-1-2
IDElements ofComparisonProposed devicePredicate Device
510(k) NumberK222208K202374
ManufacturerQingdao Hisense MedicalEquipment Co., Ltdshenzhen Beacon DisplayTechnology Co., Ltd.
Devicetype/modelHMD4C27SC44W+
1.Display Performance/Specifications
1.1Screen size27-inch772 mm (30.4")
1.2ScreenTechnologyColor LCD panel (IPS)TFT LCD panel
1.3Viewing AngleHorizontal: 178°; vertical: 178°(CR ≥ 10:1) (typ.)H:178°, V:178°,(CR>10)
1.4Pixel array2560 x 14404MP(2560 x 1600)
1.5Display Area596.74 (H) x 335.66 (V) (mm)641.28 (H) x 400.8 (V) mm
1.6Pixel Pitch0.2331 x0. 2331 (mm)0.2505 x 0.2505 mm
1.7Subpixel drivingNot ApplicableNot Applicable
1.8Contrast Ratio1000:1 (typ.)1000:1
1.9Frame rate /refresh rate49-76Hz,71.3~112.6kH31~140kHz, 29-60Hz
1.10MaximumBrightness(typ)550cd/m²700cd/m²
1.11Recommendedbrightness400 cd/m²500cd/m²
1.12Backlight typeLEDLED
1.13AmbientlightsensingBuilt-in ambient light sensorYes
1.14Response Time16 ms (typ.)6ms
1.15Aspect Ratio16:1016:10
1.16Luminancenon-uniformitycompensationNot ApplicableNot Applicable
1.17Touch-screenNot ApplicableNot Applicable
2.Video Signals
2.1Input VideoSignalDVI-D x 2DisplayPort x 2HDMIDVI-D x 2DisplayPort x 2VGA×1
2.2Output SignalDisplayPortNA
IDElements ofComparisonProposed devicePredicate Device
510(k) NumberK222208K202374
ManufacturerQingdao Hisense MedicalEquipment Co., Ltdshenzhen Beacon DisplayTechnology Co., Ltd.
Devicetype/modelHMD4C27SC44W+
2.3Video bandwidthDVI: 215MHzDisplayPort : 215MHzDVI: 215MHzDisplayPort : 215MHz
3.Power Related Specification
3.1PowerRequirements24 V DC, 6.25ADC 24 V/9A
3.2PowerConsumption/save mode70W/Below 0.5W80W/ less than 5W
3.3PowerManagementDVI DMPMDisplayPort 1.2aDVI DMPM,DisplayPort 1.2a
4.Miscellaneous Features/Specifications
4.1USB ports1 uplink port and2 downlink ports/ Rev. 2.01 upstream2 downstream/ Rev. 2.0
4.2Display color8bit10-bit,1,073,741,824color
4.3Software/Firmware:Built-in embedded softwareBuilt-in embedded software
4.4Dimensions w/ostand(W x H x D)Without base:653.8 mm x520mm x 60mmWith base: 653.8 mm x400mm x 60 mm699.3 x 513-633 x251.4mmWithout packing864 x600 x334mm(withpacking)
4.5Net weight8.5 kg (excluding the base)13±0.5kg(wet)
4.6VESA standard100 x 100 (mm)VESA (100 mm)
5. Intended use
5.1Indication for useThis product is intended to beused in displaying radiologicalimages for review, analysisand diagnosis by trainedmedical practitioners. Theproduct does not support thedisplayofmammographyThese products are intendedto be used in displaying andviewing digital images forreview and analysis bytrained medical practitioners.They do not support thedisplaying of mammography
Elements ofComparisonProposed devicePredicate Device
510(k) NumberK222208K202374
IDManufacturerQingdao Hisense MedicalEquipment Co., Ltdshenzhen Beacon DisplayTechnology Co., Ltd.
Devicetype/modelHMD4C27SC44W+
images for diagnosis.images for diagnosis.
6 Applicable standard
6.1ApplicablestandardCompliance with IEC 60601-1Compliance with IEC 60601-1-2Compliance with IEC 60601-1Compliance with IEC 60601-1-2

{8}------------------------------------------------

{9}------------------------------------------------

2.2 HMD4C27S and its predicate

Tabel2 HMD4C27S Detailed Comparison with Predicate

{10}------------------------------------------------

{11}------------------------------------------------

510(k) Premarket Notification Submission

3.2 HMD6C30D and its predicate

Tabel3 HMD6C30D Detailed Comparison with Predicate

IDElements ofComparisonProposed devicePredicate Device
510(k) NumberK222208K163335
ManufacturerQingdao Hisense MedicalEquipment Co., LtdEIZO Corporation
Devicetype/modelHMD6C30DRadiForce RX660
1.Display Performance/Specifications
1.1Screen size30 inches30"
1.2ScreenTechnologyColor LCD panel (IPS)IPS
1.3Viewing AngleHorizontal: 178°; vertical: 178°(CR ≥ 10)H: 176°, V: 176°
1.4Pixel array3280 × 20806MP (3,280 x 2,048)
1.5Display Area645.504 (H) × 409.344 (V)(mm)645.5 mm x 403.0 mm
1.6Pixel Pitch0.197 (H) x 0.197 (V) (mm)0.1968 mm x 0.1968 mm
1.7Subpixel drivingNot ApplicableNot Applicable
1.8Contrast Ratio2000:1 (typ.)1500:1
1.9Frame rate48-62Hz31 - 127 kHz / 22 - 61 Hz(VGA Text: 69 - 71 Hz)Frame synchronous mode:29.5 - 30.5 Hz, 59 - 61 Hz
1.10MaximumBrightness(typ)1300cd/m²1000 cd/m²
1.11Recommended400 cd/m²500 cd/m²
Elements ofComparisonProposed devicePredicate Device
510(k) NumberK222208K163335
IDManufacturerQingdao Hisense MedicalEquipment Co., LtdEIZO Corporation
Devicetype/modelHMD6C30DRadiForce RX660
brightness
1.12Backlight typeLEDLED
1.13AmbientlightsensingBuilt-in ambient light sensorYes
1.14Response Time30 ms (typ.)25ms
1.15Aspect Ratio16:1016:10
1.16Luminancenon-uniformitycompensationNADigital Uniformity Equalizer
1.17Touch-screenNot ApplicableNot Applicable
2.Video Signals
2.1VideoInputSignalDVI-D x 2,DisplayPort x 2,HDMI x 1DVI-D (dual link) x 1,DisplayPort x 2
2.2Output SignalDisplayPort connectorDisplayPort x 1 (daisy chain
2.3Video bandwidthDVI: 215MHzDisplayPort : 215MHzDVI: 215MHzDisplayPort : 215MHz
3.Power Related Specification
3.1PowerRequirements24 V DC, 6.25 AAC 100 - 240 V: 50 / 60 Hz
3.2PowerConsumption/save mode150W/Below 0.5W190 W / Less than 1.6 W
3.3PowerManagementDVI DMPMDisplayPort 1.2aDVI DMPM, DisplayPort 1.2a
4.Miscellaneous Features/Specifications
4.1USB ports1 uplink port and2 downlink ports/ Rev. 2.02 upstream,3 downstream / Rev. 2.0
4.2Display color10bit10-bit
4.3Software/FirmwareBuilt-in embedded softwareBuilt-in embedded software
IDElements ofComparisonProposed devicePredicate Device
510(k) NumberK222208K163335
ManufacturerQingdao Hisense MedicalEquipment Co., LtdEIZO Corporation
Devicetype/modelHMD6C30DRadiForce RX660
4.4Dimensions w/ostand(W x H x D)Without base:700 mm x478mm x 68mmWith base:700 mm x 516-626mm x 274 mm
4.5Net weight17.1 kg (excluding the base)10.1kg(without stand)
4.6VESA standard100 x 100 (mm)100 x 100 (mm)
5. Intended use
5.1Indication for useThis product is intended to beused in displaying radiologicalimages for review, analysisand diagnosis by trainedmedical practitioners.Theproduct does not support thedisplay of mammographyimages for diagnosis.The display is intended to beused for displayingandviewing digital images(excluding digitalmammography) for reviewand analysis by trainedmedical practitioners.
6 Applicable standard
6.1ApplicablestandardCompliance with IEC 60601-1Compliance with IEC 60601-1-2Compliance with IEC 60601-1Compliance with IEC 60601-1-2

{12}------------------------------------------------

{13}------------------------------------------------

510(k) Premarket Notification Submission

3.Performance data

The following performance data were provided in support of the substantial equivalence determination.

3.1Bench testing:

The bench tests were performed on the proposed devices HMD2C21A, HMD4C27S, HMD6C30D in accordance with Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline). Bench testing was conducted to demonstrate the HMD2C21A, HMD4C27S, HMD6C30D meets all performance standards as follows:

{14}------------------------------------------------

510(k) Premarket Notification Submission

Table4 Bench test
Measurements GuidanceRecommended forNon-mammographyDisplay SubmissionsHMD2C21A,HMD4C27S,HMD6C30DMeasurements
a. Spatial resolutionYesMeasure Spatial resolutionwith TG18 Resolution
b. Pixel defects (maximumcounts, allowed defect types,and locations)YesMeasure Pixel defects withTG18
c. ArtifactsYes (Limited)Measure Artifacts withTG18
d. Temporal responseYesMeasure Temporalresponse with TG18
e. Luminance (maximum,minimum, achievable, andrecommended)YesMeasure Luminance withTG18
f. Conformance to agrayscale-to-luminancefunction (e.g., DICOM GSDF)YesMeasure Conformance to agrayscale-to-luminancefunction with TG18
g. Luminance at 30° and 45° indiagonal, horizontal, andvertical directions at center andfour cornersNo-
h. Luminance uniformity orMura testNo-
i. Stability of luminance andchromaticity response withtemperature and time ofoperation or on-timeNo-
j. Spatial noiseNo-
k. Reflection coefficientNo-
I. Veiling glare or small-spotcontrastNo-
m. Color tracking (primarycolors and color gamut)YesMeasure Color trackingwith TG18
n. Gray tracking (gray shadesand white point)No-

3.2 Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the HMD2C21A, HMD4C27S, HMD6C30D. The devices comply with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.

3.3Animal and clinical study

The subject of this premarket submission, Hisense LCD monitor HMD2C21A, HMD4C27S, HMD6C30D do not require animal or clinical studies to support substantial equivalence.

{15}------------------------------------------------

4. Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Qingdao Hisense Medical Equipment Co., Ltd. concludes that:

  • O The intended uses of proposed devices (HMD2C21A, HMD4C27S, HMD6 C30D) are equivalent to the predicate devices.
  • The technological characteristics differences between proposed devices ● (HMD2C21A, HMD4C27S, HMD6C30D) and the predicate devices do not affect the safety and effectiveness, so no new risk is raised.
  • . Demonstrated by the bench tests, the display characteristics of proposed devices (HMD2C21A, HMD4C27S, HMD6C30D) are equivalent to those of the predicate devices.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).