K Number
K222096
Date Cleared
2023-03-23

(248 days)

Product Code
Regulation Number
872.4850
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For root canal cleaning. The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners.

Device Description

The Endo Ultrasonic Activator is mainly used for root canal cleaning.It mainly contain handpiece, sleeve, wrench, adapter and tip. The ultrasonic frequency generated by the CPU is driven by the drive circuit to drive the piezoelectric ceramic transducer, using the reverse piezoelectric effect to produce the ultrasonic vibration, motivating the working tip to produce the resonance, and using the cavitation and sound flow effect generated by the ultrasonic vibration to wash the root canal swing.

AI/ML Overview

The provided text describes the regulatory clearance of a medical device, the "Endo Ultrasonic Activator," and its substantial equivalence to a predicate device. It contains a comparison table of characteristics and analysis of differences, but it does not describe specific acceptance criteria and a study proving the device meets those criteria in the typical format one would expect for an AI algorithm's performance.

This document is a 510(k) summary for a physical medical device (an ultrasonic activator for root canal cleaning), not an AI/ML powered device. Therefore, the requested information components related to AI/ML device performance (like sample sizes for test/training sets, data provenance, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable and are not present in the provided text.

The "Non-Clinical Test Conclusion" and "Substantially Equivalent (SE) Comparison" sections describe how the device meets various safety and performance standards relevant to non-AI medical devices.

Here's an attempt to extract relevant information while acknowledging the non-AI nature of the document:

1. Table of Acceptance Criteria and Reported Device Performance

For a physical medical device, "acceptance criteria" are typically compliance with recognized standards. The "reported device performance" is demonstrated through testing against these standards and comparison to a predicate device.

Acceptance Criteria (e.g., Standard Compliance)Reported Device Performance (Compliance Status)
ANSI/AAMI ES60601-1:2005 + A1 (Medical Electrical Equipment Safety)Complies (Verified through non-clinical tests)
IEC 60601-1-2:2014 (EMC)Complies (Verified through non-clinical tests)
IEC 80601-2-60:2019 (Dental Equipment Safety)Complies (Verified through non-clinical tests)
ISO 18397:2016 (Dentistry-Powered scaler)Complies (Functions verified, comparable performance demonstrated, e.g., operating frequency)
IEC 62133-2 Edition1.0 2017-02 (Li-ion battery safety)Complies (For the Li-ion battery)
ISO 10993-5: 2009 (In Vitro Cytotoxicity)Complies (Biocompatibility testing performed for direct contact materials)
ISO 10993-10: 2010 (Irritation & Skin Sensitization)Complies (Biocompatibility testing performed for direct contact materials)
ISO 17665-1:2006 (Moist Heat Sterilization)Complies (Sterilization validation testing performed)
Substantial Equivalence to Predicate DeviceAchieved (Differences analyzed and deemed not to affect safety/effectiveness; device operates on the same principle of ultrasonic vibration for root canal cleaning, and shares similar or acceptable ranges for key technical characteristics like primary tip vibration excursion, operating temperature, etc.)

2. Sample size used for the test set and the data provenance
Not Applicable (N/A). This document refers to the testing of a physical medical device, not an AI/ML algorithm that would typically have a "test set" of data. The testing involved compliance with various engineering and safety standards, likely through bench testing and material evaluations.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
N/A. No ground truth establishment by experts for a test set is described as this is not an AI/ML device.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
N/A. No test set or corresponding adjudication method is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
N/A. This is not an AI/ML device, so no MRMC study or AI assistance effect size is discussed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
N/A. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
N/A. No ground truth is described in the context of an AI/ML algorithm. The "truth" for this device's performance is its compliance with engineering, safety, and biocompatibility standards, and demonstrating the same fundamental operating principle as a legally marketed predicate device.

8. The sample size for the training set
N/A. This is not an AI/ML device.

9. How the ground truth for the training set was established
N/A. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 23, 2023

Foshan COXO Medical Instrument Co.,Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui. No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 CHINA

Re: K222096

Trade/Device Name: Endo Ultrasonic Activator Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: Class II Product Code: ELC Dated: February 24, 2023 Received: February 24, 2023

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael E. Adjodha -2

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222096

Device Name Endo Ultrasonic Activator

Indications for Use (Describe) For root canal cleaning. The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

|_ | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

    1. Date of Preparation:2023/02/24
    1. Sponsor Identification

Foshan COXO Medical Instrument Co., Ltd.

BLDG 4, District A, Guangdong New Light Source Industrial Base, South of Luocun Avenue, Nanhai District, Foshan, 528226 Guangdong, China

Contact Person: Yongjian Zheng Position: Legal Representative Tel: +86-13702544788 Fax: +86-0757-81800058 Email: 13702544788@163.com

  • Designated Submission Correspondent 3.
    Mr. Ray Wang

Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China

Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name:Endo Ultrasonic Activator Common Name: Scaler,Ultrasonic

Regulatory Information Classification Name: Scaler,Ultrasonic Classification: 2 Product Code: ELC Regulation Number: 872.4850 Review Panel: Dental

Indication For Use Statement:

For root canal cleaning.

The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners.

Device description:

The Endo Ultrasonic Activator is mainly used for root canal cleaning.It mainly contain handpiece, sleeve, wrench, adapter and tip. The ultrasonic frequency generated by the CPU is driven by the drive circuit to drive the piezoelectric ceramic transducer, using the reverse piezoelectric effect to produce the ultrasonic vibration, motivating the working tip to produce the resonance, and using the cavitation and sound flow effect generated by the ultrasonic vibration to wash the root canal swing.

న. Identification of Predicate Device(s)

Predicate Device 510(k) Number: K053555 Product Name: Ultrasonic Scaler(Model: UDS-L) Manufacturer: Guilin Woodpecker Medical Instrument Company, Limited

  • Non-Clinical Test Conclusion 6.
    Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ANSI/AAMI ES60601-1:2005 + A1, Medical Electrical Equipment Part 1: General requirements > for basic safety and essential performance.

  • EC 60601-1-2:2014, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Tests

  • IEC 80601-2-60:2019 Medical electrical equipment Part 2-60: Particular requirements for the >

{5}------------------------------------------------

basic safety and essential performance of dental equipment

  • ISO 18397:2016 Dentistry-Powered scaler

  • IEC 62133-2 Edition1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems

  • ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity

  • ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

  • ISO 17665-1:2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices

    1. Clinical Test Conclusion

No clinical study is included in this submission.

  • Substantially Equivalent (SE) Comparison 8.
ItemProposed DevicePredicate DeviceK053555(Model:UDS-L)Reference DeviceK213947Remark
Device NameEndo UltrasonicActivatorUltrasonic ScalerUltrasonicEndoActivation Device/
ClassificationRegulation872.4850872.4850872.4850SAME
ClassificationPanelDentalDentalDentalSAME
Product CodeELCELCELCSAME
Common NameScaler,UltrasonicScaler,UltrasonicScaler,UltrasonicSAME
IndicationforuseFor root canalcleaning.The instrumentmust only beused in hospitalenvironments,clinics or dentaloffices, byqualifiedpractitioners.1. Removing supraand sub gingivalcalculus deposits andstains from the teeth2. Periodontal pocketlavage withsimultaneousultrasonic tipmovement3. Preparing,cleaningand irrigating rootThe Ultrasonic EndoActivation Device(Model:Actor I pro ) is anultrasonic-handpiecewhich is intended use forroot canal cleaning andpreparation.The Ultrasonic EndoActivation Device(Model:Actor I pro ) isintended for use bySAME
canals4. Retrogradepreparation of rootcanalstrained dentalprofessionals inprofessional health carefacilities on patients thatneedroot-canal-treatment.
InputAdapterAC100-240V50/60HzMain Unit 24V 50Hz1.3AAC100-240V, 50/60HzAnalysis(1)
Input Power10VA38VANot AvailableAnalysis(2)
Power supplyBattery 3.7V1200mAh(Li-ionBattery)Power source 220V50HZRechargeable Li-ionBattery Capacity1600mAh,3.7VAnalysis(3)
Operatingfrequency30kHz $\pm$ 10%28kHz $\pm$ 3kHz30 $\pm$ 3kHz(27-33 kHz)Analysis(4)
Primary tipvibrationexcursion<100 μ m1 μ m-100 μ mNot AvailableSAME
Operation modeNon-continuousContinuousContinuousAnalysis(5)
Classified ofprotectionagainst ElectricShockClass IIClass IINot AvailableSAME
Protectionagainst ElectricShockType B appliedpartType BF applied partNot AvailableAnalysis(6)
Degree ofProtectionIPX0General equipmentIPX0Not AvailableSAME
Operatingtemperature+5°C - +40°C+5°C - +40°CNot AvailableSAME
Operatinghumidity20% - 80%30%-75%Not AvailableSimilar
Atmosphericpressure86kPa -106kPa70kPa -106kPaNot AvailableSimilar
Storagetemperature-10°C - +55°C-20°C - +55°CNot AvailableSimilar
Storagehumidity≤93%10%-93%Not AvailableSimilar
Atmosphericpressure50kPa -106kPa70kPa -106kPaNot AvailableSimilar
CordlessYesNoYesAnalysis(7)
Device typeHandpieceUnitHandpieceAnalysis(7)
BiocompatibilityISO10993-5ISO10993-10Not AvailableISO 10993-5:2009ISO 10993-10:2010Analysis(8)
ISO 10993-11:2017
SterilizationISO 17665-1Not AvailableNot AvailableAnalysis(8)
Electrical SafetyIEC 60601-1IEC 80601-2-60Not AvailableIEC 60601-1:2012IEC 80601-2-60:2019Analysis(8)
EMCIEC60601-1-2Not AvailableIEC 60601-1-2:2014Analysis(8)
DentalISO 18397Not AvailableNot AvailableAnalysis(8)

Table 1 Comparison of Technology Characteristics

{6}------------------------------------------------

{7}------------------------------------------------

Analysis :

Analysis(1):

There is difference on the input of proposed device is adapter, the input of predicate device is main unit, however, both of them could meet the electricity safety standard IEC 60601-1, the difference does not affect the substantial equivalence in effectiveness and safety.

Analysis(2):

There is difference on the input power, however, both of them could meet the electricity safety standard IEC 60601-1, the difference does not affect the substantial equivalence in effectiveness and safety.

Analysis(3):

The proposed device is powered by rechargeable 3.7V Li-ion battery, and the predicate device is powered by AC power source, there is difference on the power supply mode. In principle, no matter what kind of power supply mode, piezoelectric ceramic transducer is driven electrically to drive the working tip to vibrate, and the intended use of root canal cleaning is achieved by using acoustic flow effect and thermal effect generated by ultrasonic. In terms of type, the subject device the transducer through battery power, while the transucer of the predicate device is powered by AC power source, and it may need to drive the transducer through a transformer. But the vibration generated by the current driven transducer and the ultrasonic are essentially the same. So the difference would not affect its effectiveness.The proposed device comply with electrical safety standard IEC 60601-1, and its Li-ion battery comply with the battery safety standard IEC 62133-2, so the power supply mode difference would not affect its safety.

In addition, powered by battery is a very common power supply mode. For example, the Reference Device K213947 we determined uses the battery to supply power.

This reference device K213947 is consistent on the indication for use, technological characteristics and other relevant information with the subject device.

By comparing with the reference device, we can find that the intended use of the subject device is consistent with the reference device. Both the subject device and the reference device are based on the principle of ultrasonic vibration for root canal cleaning. And the reference device, like the subject device, is also a cordless, battery-powered device.

And the reference device is also use the 3.7V Li-ion battery which is same with the subject device. Thus, powered by a battery is a very mature technology.

Analysis(4):

The proposed devices and the predicate device are different operation frequency. Operation frequency of

{8}------------------------------------------------

the subject device is only slightly different with the predicate device.

And we have also identified the reference device K213947 which is consistent with the operation frequency of the subject device.

We believe that the difference of the operating frequency of the subject device and predicate will not affect the substantial equivalence of safety and effectiveness.

Analysis(5):

The proposed devices and the predicate device are difference operation mode. The operating mode of the proposed devices are comply with the 60601-1, so this could not effect the effectiveness and safety.

Analysis(6):

The proposed devices and the predicate device are difference Protection against Electric Shock.The Protection against Electric Shock of the proposed devices are comply with the 60601-1, so this could not effect the effectiveness and safety.

Analysis(7):

There are some design differences between the subject device and the predicate device, for example, cordless and hanpiece. However, we believe that the design differences will not affect the safety and efficacy of the subject device.

First, in principle, the unit of the predicate device is connected to a power supply to power the CPU, and the software is built into the unit.

The unit and the handpiece through the cable connection. This cable is used to power the handpiece and transmit signals to the handpiece, and thus realize the functions of the product.

The subject device, the software is built into the handpiece. After battery power, it drives the transducer to control the CPU and set the function.

No matter how the power supply and signal transmission, the power supply ultimately drives the CPU to work and realize various functional settings of the product. The electrical structure is consistent in essence, only the appearance changes and the connection mode is not consistent, leading to the different selection of CPU, which will not affect the product to achieve the final intended use. So from a principle point of view we are the same as the predicate device, just the form of power supply and the signal transmission have changed.

Because of the connection mode changes, we introduced software validation to prove that functionality of the subject device can be fully implemented. And we introduced ISO 18397 testing to prove that our performance implementation meets the requirements.

In addition, the reference device K213947 we determined is also a cordless handpiece device, and both the subject device and the reference device are based on the principle of ultrasonic vibration for root canal cleaning, which shows that this design form is a very mature technology and will not affect the safety and effectiveness of the subject device.

Analysis(8):

We did not obtain relevant test information for the predicate device, We discuss here to make it clear that these minor differences are not concerns for the equivalency of the proposed device and predicate device.

{9}------------------------------------------------

And will not affect the performance of safety and effectiveness of the proposed device.

a. We performed applicable tests to support biocompatibility, including Cytotoxicity, Sensitization and Irritation.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. Biocompatibility testing of in vitro cytotoxicity was performed to verify the equivalent safety of the materials that are in direct contact with mucosal membrane and teeth in the mouth, and skin including tip and sleeve.

ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin

Sensitization Biocompatibility testing for Irritation and Skin Sensitization was performed to verify the equivalent safety of the materials that are in direct contact with mucosal membrane and teeth in the mouth, and skin including tip and sleeve.

b.Sterilization validation testing is performed in accordance with ISO 17665-1:2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices. But Reprocessing methods also meet the FDA Guidance Document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."

c.Electrical safety testing per IEC 60601-1 and IEC 80601-2-60 required testing.

d.Electromagnetic Compatibility testing per IEC 60601-1-2 required testing.

e. The functions of Endo Ultrasonic Activator were verified according to ISO 18397: 2016 Dentistry -Powered scaler. Performance test performed on proposed device aims to support the substantial equivalence of the proposed device to the predicate device. Test's setup and execution was in accordance with applicable standards. Results of the testing demonstrate the comparable performance (eg: Operating frequency) between proposed device and predicate device.

9. Substantially Equivalent (SE) Conclusion

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Ultrasonic Scaler cleared under K053555.

§ 872.4850 Ultrasonic scaler.

(a)
Identification. An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.(b)
Classification. Class II.