K Number
K221873
Device Name
NeoGen PSR System
Manufacturer
Date Cleared
2023-01-12

(198 days)

Product Code
Regulation Number
878.4400
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
When used with the 5mm nozzle the NeoGen PSR System is intended for treatment of the following dermatological conditions : - · Treatment of wrinkles and rhytids - · Superficial skin lesions - · Actinic Keratosis - Viral Papillomata - · Seborrheic Keratosis - · Acne Scars When used with the 25mm nozzle, the NeoGen PSR System is intended for coagulation of tissue in dermatological procedures.
Device Description
The system comprises the following major elements: - . The Energist NeoGen PSR generator. This incorporates a source of Ultra High Frequency (UHF) energy and the gas control system required to support creation of nitrogen plasma; - . A Footswitch – single pedal type for activation of the plasma output. - . A handpiece that connects to the generator via an umbilical, which incorporates electrical and gas connections. The handpiece incorporates a finger-operated button switch that can be used to activate the plasma as an alternative to the footswitch. - . A Procedure Pack, comprising one or more limited life disposable tips (nozzles) through which the plasma is delivered. These are a snap fit to the distal end of the handpiece; - A gas supply system - principally a nitrogen gas cylinder and a regulator assembly; - . A bespoke trolley which provides a secure, mobile platform for the generator, handpiece, gas cylinder (plus a spare cylinder) and gas regulator assembly.
More Information

No
The device description and performance studies focus on the energy source (UHF), gas control, and mechanical components. There is no mention of AI, ML, image processing, or data analysis that would suggest the use of such technologies.

Yes

The device is intended for the treatment of various dermatological conditions such as wrinkles, rhytids, superficial skin lesions, actinic keratosis, viral papillomata, seborrheic keratosis, and acne scars, and for the coagulation of tissue in dermatological procedures, all of which are therapeutic applications.

No

The NeoGen PSR System is intended for treatment of various dermatological conditions and for coagulation of tissue. It is a therapeutic device, not a diagnostic one.

No

The device description clearly lists multiple hardware components including a generator, footswitch, handpiece, procedure pack with disposable tips, gas supply system, and a trolley. While there is mention of software validation, the device is fundamentally a hardware system that utilizes software for functionality.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly describes treatment of dermatological conditions and coagulation of tissue. These are therapeutic procedures performed directly on the patient's body.
  • Device Description: The device components (generator, handpiece, nozzles, gas supply) are designed for delivering energy to the skin for treatment, not for analyzing samples taken from the body.
  • Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.), reagents, or any process related to diagnosing a condition based on in vitro analysis.

IVD devices are specifically designed to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. This device's function is entirely focused on therapeutic intervention on the skin.

N/A

Intended Use / Indications for Use

When used with the 5mm nozzle the NeoGen PSR System is intended for treatment of the following dermatological conditions:

  • Treatment of wrinkles and rhytids
  • Superficial skin lesions
  • Actinic Keratosis
  • Viral Papillomata
  • Seborrheic Keratosis
  • Acne Scars

When used with the 25mm nozzle, the NeoGen PSR System is intended for coagulation of tissue in dermatological procedures.

Product codes

GEI

Device Description

The system comprises the following major elements:

  • . The Energist NeoGen PSR generator. This incorporates a source of Ultra High Frequency (UHF) energy and the gas control system required to support creation of nitrogen plasma;
  • . A Footswitch – single pedal type for activation of the plasma output.
  • . A handpiece that connects to the generator via an umbilical, which incorporates electrical and gas connections. The handpiece incorporates a finger-operated button switch that can be used to activate the plasma as an alternative to the footswitch.
  • . A Procedure Pack, comprising one or more limited life disposable tips (nozzles) through which the plasma is delivered. These are a snap fit to the distal end of the handpiece;
  • A gas supply system - principally a nitrogen gas cylinder and a regulator assembly;
  • . A bespoke trolley which provides a secure, mobile platform for the generator, handpiece, gas cylinder (plus a spare cylinder) and gas regulator assembly.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Skin

Indicated Patient Age Range

Not Found

Intended User / Care Setting

The system is intended for professional use by trained and duly qualified medical practitioners in hospitals and clinics. It must not be used in domestic, industrial or outdoor locations, or in critical care environments.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance Testing
Representative samples of the NeoGen PSR system have been tested by Energist Limited UK to demonstrate compliance with the user and functional performance requirements. Numerous documents are available, the "headline" ones being:

DD0133R NeoGen System Validation Report ;Overall System Requirements SD0618R NeoGen System V&V Report :Software determined functionality

In respect of the 25mm nozzle specifically, please refer to the following document:

DD0246 NeoGen 25mm Nozzle - Safety & Efficacy Review.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K132754, K072394

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.

0

January 12, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two main elements: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Energist Limited Graham Booth Chief Technical Officer 2 Park Pavilions, Clos Llyn Cwm Enterprise Park Swansea, SA6 8QY United Kingdom

Re: K221873

Trade/Device Name: NeoGen PSR System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: December 08, 2022 Received: December 13, 2022

Dear Graham Booth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm = identifies = combination = product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical

1

device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and -CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Colin K. Chen -S

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K221873

Device Name

NeoGen PSR Plasma Skin Resurfacing Device

Indications for Use (Describe)

When used with the 5mm nozzle the NeoGen PSR System is intended for treatment of the following dermatological conditions :

  • · Treatment of wrinkles and rhytids
  • · Superficial skin lesions
  • · Actinic Keratosis
  • Viral Papillomata
  • · Seborrheic Keratosis
  • · Acne Scars

When used with the 25mm nozzle, the NeoGen PSR System is intended for coagulation of tissue in dermatological procedures.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

K221873: Energist Limited NeoGen PSR 510(k) Application - 510(k) Summary

Revision:BDate:12 January 2023
Document Title:510(k) Summary (Per 21 CFR 807.92)
Subject:NeoGen PSR System 510(k) Application
510(k) NumberK221873
Submitter Name:Energist Limited
Establishment Registration No:3007353760
Submitter Address:2 Park Pavilions, Clos Llyn Cwm, Swansea
Enterprise Park, Swansea, SA6 8QY, United Kingdom
Contact Person:Graham Booth – Chief Technical Officer
Email: gbooth@energist.com
Telephone Number:+44 1792 798768
Revision Date:11 January 2023
Device Trade Name:NeoGen PSR System
Common Name:Electrosurgical Device
Device Classification:Electrosurgical cutting and coagulation device and
accessories Class II
Product Code:GEI (21 CFR 878.4400)

DEVICE DESCRIPTION

System Elements

The system comprises the following major elements:

  • . The Energist NeoGen PSR generator. This incorporates a source of Ultra High Frequency (UHF) energy and the gas control system required to support creation of nitrogen plasma;
  • . A Footswitch – single pedal type for activation of the plasma output.
  • . A handpiece that connects to the generator via an umbilical, which incorporates electrical and gas connections. The handpiece incorporates a finger-operated button switch that can be used to activate the plasma as an alternative to the footswitch.
  • . A Procedure Pack, comprising one or more limited life disposable tips (nozzles) through which the plasma is delivered. These are a snap fit to the distal end of the handpiece;
  • A gas supply system - principally a nitrogen gas cylinder and a regulator assembly;
  • . A bespoke trolley which provides a secure, mobile platform for the generator, handpiece, gas cylinder (plus a spare cylinder) and gas regulator assembly.

4

Principle of Operation

The NeoGen PSR delivers controlled thermal energy to the outer layers of the skin so that part or all of the epidermis becomes non-viable and there is controlled thermal modification to the underlying dermis. The body's natural healing process causes the damaged skin to be replaced with new, fresh, growth – thus resulting in improvement of the appearance and effects of the indicated conditions.

The thermal energy emitted by the system is in the form of a nitrogen gas plasma. This is created by subjecting pressurised nitrogen gas to a high level of Ultra High Frequency (UHF) energy. This process takes place in the handpiece of the device, more specifically in a resonator tube, or "nozzle", the distal end of which is directed at the skin surface. The handpiece is connected to the main generator unit via an "umbilical", which incorporates a gas supply tube and a cable capable of conducting the UHF energy. It also incorporates a cable for control signals between the handpiece and the generator unit to support the action of a control button and gas solenoid valve, plus secure identification and "counter" elements in the nozzle and the handpiece itself. The plasma is created in short pulses, approximately 5 to 15 milli-seconds in duration. Automatic adjustment of the UHF power level and pulse duration to the handpiece enables control of tissue effects by altering the amount of energy delivered by the plasma to the skin surface for each pulse. In practice, the energy per pulse can be varied from 1.0 to 4.0 Joules when the system is used with the 5mm nozzle, and from 0.5 to 0.8 Joules when used with the 25mm nozzle. The system can be configured to automatically generate a series of pulses at set intervals to allow the operator to move the handpiece over the surface of the skin at a steady speed to achieve a uniform effect. Alternatively, the pulses can be manually triggered - either by pressing the control button on the handpiece, or by depressing a footswitch, according to operator preference.

The nozzles are limited-life components which are a snap-fit onto the distal end of the handpiece and are easily replaced. A stand-off incorporated in the nozzle allows the distance from the skin surface to be accurately determined. The separation distance influences the amount of energy impacting the skin surface. NeoGen PSR systems are capable of supporting the use of nozzles with 5mm standoff, as well as 25mm. This is as per the Rhytec Portrait PSR3, which was the legally marketed predicate device cited when the NeoGen PSR was originally cleared under K132754.

Handpiece usage is also monitored and the system limits use to guard against unexpected failure due to degradation with use over time. The handpiece is user replaceable.

The generator incorporates a large colour touchscreen display supporting a Graphical User Interface that clearly conveys information to the operator and allows settings to be made readliy and conveniently.

5

Intended Use/ Indications For Use

When used with the 5mm nozzle the NeoGen PSR System is intended for treatment of the following dermatological conditions:

  • . Treatment of wrinkles and rhytids
  • Superficial skin lesions
  • Actinic Keratosis .
  • Viral Papillomata
  • Seborrheic Keratosis .
  • · Acne Scars

When used with the 25mm nozzle, the NeoGen PSR System is intended for coagulation of tissue in dermatological procedures.

Intended Use Environment

The system is intended for professional use by trained and duly qualified medical practitioners in hospitals and clinics. It must not be used in domestic, industrial or outdoor locations, or in critical care environments.

6

TECHNOLOGICAL CHARACTERISTICS

The technological characteristics of the current Energist NeoGen PSR system with 25mm nozzle option that is the subject of this K221873 submission and the predicate PSR system originally cleared under K132754 are summarised and compared in the table below:

ParameterK132754 Predicate NeoGen PSR systemK222873 Subject NeoGen PSR system with 25mm option
Energy setting Displayed Values Precision0.1 Joule0.1 Joule
Max Output Energy4 Joules4 Joules
Treatment Pulse Repetition RateRepeat Pulse mode: 1.0 to 2.5 Hz
Single Pulse mode: User determined.Repeat Pulse mode: 1.0 to 2.5 Hz
Single Pulse mode: User determined.
Output Pulse Width5.2 to 15.4 ms5.2 to 15.4 ms
Safety Classification
(per EN/ISO 60601-1)Class 1 (the Generator requires connection to a Protective Earth).Class 1 (the Generator requires connection to a Protective Earth).
FDA Regulatory ClassificationClass IIClass II
FDA Regulation Number21 CFR 878.440021 CFR 878.4400
FDA Regulation NameElectrosurgical cutting and coagulation device and accessories.Electrosurgical cutting and coagulation device and accessories.
FDA Product CodeGEIGEI
EU MDR ClassificationRisk Class IlbRisk Class Ilb
Applied PartsType BF(not defibrillator proof)Type BF(not defibrillator proof)
Duty Cycle20 seconds on, 10 seconds off, for a maximum of 50 plasma pulses at the 2.5Hz maximum rate20 seconds on, 10 seconds off, for a maximum of 50 plasma pulses at the 2.5Hz maximum rate
Operational Ambient Temperature Range10 to 30 °C.10 to 30 °C.
Operational Humidity30 to 75% RH30 to 75 RH
Atmospheric Pressure700 to 1060 hPa.700 to 1060 hPa.
Power
RequirementsVoltage: 100-230 Vrms, 50-60Hz, 1 ph.
Current: 6.5A max.
Power Consumption: ≤650 VAVoltage: 100-230 Vrms, 50-60Hz, 1 ph.
Current: 6.5A max.
Power Consumption: ≤650 VA
Inlet Safety
FusesQuantity, Size: 2x, 20mm x 5mm
Type: Time Lag (T) type
Rating: 6.3A, 250 VAC
Approvals: UL Listed
Breaking Capacity: 1500A @250VACQuantity, Size: 2x, 20mm x 5mm
Type: Time Lag (T) type
Rating: 6.3A, 250 VAC
Approvals: UL Listed
Breaking Capacity: 1500A @250VAC
ParameterK132754 Predicate NeoGen PSR systemK222873 Subject NeoGen PSR system with 25mm option
Mains Connection
Inlet type:
Power Cord:
Mains Switch:IEC 60320 type C14
IEC 60320 type C13 (local plug a/r)
Double Pole, integral with inletIEC 60320 type C14
IEC 60320 type C13 (local plug a/r)
Double Pole, integral with inlet
RF Output Frequency2450 to 2480 MHz (2470 MHz typ).
No sub-harmonics2450 to 2480 MHz (2470 MHz typ).
No sub-harmonics
ModulationPulsed CW (Carrier Wave)Pulsed CW (Carrier Wave)
Gas SystemCommercially available medical
grade nitrogen is supplied via a
Regulator assembly included with
the system to provide a feed to the
Generator at 1.7Bar (nom.)Commercially available medical
grade nitrogen is supplied via a
Regulator assembly included with
the system to provide a feed to the
Generator at 1.7Bar (nom.)
FootswitchMedical type, rated 60V, 5A DCMedical type, rated 60V, 5A DC
TrolleyCustom made 5-wheel trolley with
locking wheels accommodates the
Generator and 2 gas cylinders
(including a spare). Meets stability
requirements of IEC 60601-1.Custom made 5-wheel trolley with
locking wheels accommodates the
Generator and 2 gas cylinders
(including a spare). Meets stability
requirements of IEC 60601-1.
HandpieceConnecting to the Generator via a
flexible umbilical, the Handpiece
accommodates the nozzle, which is
where the plasma is generated.
Total number of shots is limited
(electronically) to 250,000 after
which the handpiece must be
replaced.Connecting to the Generator via a
flexible umbilical, the Handpiece
accommodates the nozzle, which is
where the plasma is generated.
Total number of shots is limited
(electronically) to 250,000 after
which the handpiece must be
replaced.
Nozzles5mm standoff version allows for
delivery of plasma energy in the
range 1.0 to 4.0 Joules.5mm standoff version allows for
delivery of plasma energy in the
range 1.0 to 4.0 Joules.
25mm standoff option allows for
delivery of plasma energy in the
range 0.5 to 0.8 Joules.

7

With the exception of the 25mm nozzle option (highlighted in red) – clearance of which is the subject of this 510(k) submission – the technological characteristics of the predicate and subject devices are identical.

8

TESTS PERFORMED

A full description of tests carried out to confirm conformity to applicable standards is given in Energist document no. DD0254 NeoGen PSR Compliance Summary Report [USA], included with this submission.

Note that the 25mm nozzle has been included in the scope of the assessments conducted (it has been made available in other markets since 2014). However, the choice of 5mm or 25mm nozzle does not materially impact safety or EMC.

A summary is given below.

Basic Safety

Representative samples of the NeoGen PSR system have been tested by an accredited thirdparty test organisation to demonstrate compliance with the following medical device safety standards: These tests have included assessment

60601-1 Basic Safety

| Standard | ANSI/AAMI ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 &
A2:2010/(R)2012 – Medical electrical equipment Part 1: General
requirements for basic safety and essential performance. |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Consensus
Standard Recognition
Number | 19-4 |
| Report Ref. | 071-75945748-000TRF_CE_E (inc. attachment 1) |
| Issued By | TÜV SÜD, United Kingdom |

60601-2-2 HF Surgical Equipment

| Standard | IEC 60601-2-2:2017 - Medical electrical equipment – Part 2-2:
Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical
accessories |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Consensus
Standard Recognition
Number | 6-389 |
| Report Ref. | 071-75945748-000TRF_CE_E (attachment 2) |
| Issued By | TÜV SÜD, United Kingdom |

9

Electro-Magnetic Compatibility

Representative samples of the NeoGen PSR system have been tested by an accredited thirdparty test organisation to demonstrate compliance with the following medical device EMC standards:

| Standard | IEC 60601-1-2 - Medical electrical equipment -- Part 1-2: General
requirements for basic safety and essential performance -- Collateral
Standard: Electromagnetic disturbances -- Requirements and tests |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Consensus
Standard Recognition
Number | 19-8 |
| Report Ref. | R18-4514A Is. 2 |
| Issued By | Kiwa, United Kingdom |

Performance Testing

Representative samples of the NeoGen PSR system have been tested by Energist Limited UK to demonstrate compliance with the user and functional performance requirements. Numerous documents are available, the "headline" ones being:

DD0133R NeoGen System Validation Report ;Overall System Requirements SD0618R NeoGen System V&V Report :Software determined functionality

In respect of the 25mm nozzle specifically, please refer to the following document:

DD0246 NeoGen 25mm Nozzle - Safety & Efficacy Review.

10

TECHNOLOGY COMPARISON

This is a review of the differences between the 25mm standoff arrangements as cleared in K072394 for use with the Rhytec PSR3. (the original predicate as cited in the K132754 submission for the NeoGen PSR), and those of the current 25mm nozzle that is the subject of this K221873 submission.

Rhytec 25mm Standoff

This took the form of an adapter that could be fitted over the standard nozzle. The below excerpt from Rhytec's instruction sheet for the standoff shows the fitting instructions, and also gives an idea of the physical appearance and characteristics.

Image /page/10/Picture/5 description: In the image, a person wearing purple gloves is holding a white nozzle. The top surface of the nozzle is visible. Above the nozzle, there is a handle with a purple design on it. The text in the image labels the top surface of the handle and nozzle.

Top surface of Stand-off

Orientate Stand-off to the Nozzle noting alignment of top surfaces of each part

Image /page/10/Picture/8 description: In this image, a hand wearing a purple glove is holding a white and purple device with two prongs extending from the bottom. The device appears to be some sort of medical or scientific instrument. The prongs are translucent and have a yellowish tint. The hand is gripping the device firmly, and the background is a plain, light-colored surface.

Engage the Stand-off onto the Nozzle

Image /page/10/Picture/10 description: In the image, a gloved hand is holding a white plastic device with yellow tubes extending from the bottom. The device appears to be some sort of medical or laboratory instrument. The hand is holding the device near the top, and the tubes are pointing downwards. A black arrow is pointing to the top of the white plastic device.

Stand-off correctly fitted onto the Nozzle

Align protrusion on the Stand-off in relation to the joint between the Handpiece and Nozzle

Essential characteristics of the Rhytec 25mm standoff are as follows:

Standoff Distance:25mm
Service Life:To be replaced after use with 10 nozzles
Energy Reduction Factor:0.65
Patient Contacting Material:Ultem
Spot Separation:12mm
Energy Range:0.65 to 1.0 Joule*
  • The lower limit is determined by the minimum output level of the Portrait PSR3 (1 Joule) and the Energy Reduction Factor (0.65). The maximum is that recommended in the IFU to prevent overheating of the tips, which could result in patient discomfort and damage to the standoff.

The Rhytec Portrait PSR was a relatively basic device and was not able to detect the presence or absence of the adapter. It was therefore necessary for the operator to perform their own calculation of the effective energy reaching the skin surface, based on the reading on the numeric display.

11

The range of energy outputs available were:

PSR3 Energy Setting (J):1.01.11.21.31.41.5
Energy at Skin Surface (J):0.650.70.80.850.91

NeoGen 25mm Nozzle

On the NeoGen device, the standoff is an integral part of the nozzle. It attaches to the distal end of the handpiece in the same manner as the 5mm nozzle. A picture of the device is shown below.

Image /page/11/Picture/5 description: The image shows a black electrical plug with three prongs. The plug has a clear glass tube extending from one side, with a visible filament inside. The prongs are a light tan color and appear to be made of a flexible material. The plug is resting on a light-colored surface.

Essential characteristics of the NeoGen 25mm nozzle are as follows:

Standoff Distance:25mm
Service Life:9000 shots
Energy Reduction Factor:N/A
Patient Contacting Material:Ultem
Spot Separation:12mm
Energy Range:0.5 to 0.8 Joule (in 0.1J steps)

The main practical differences between the NeoGen 25mm nozzle and the Rhytec solution are:

  • . The NeoGen PSR automatically detects the presence of the 25mm nozzle and adjusts the on-screen energy settings to reflect the actual energy delivered to the skin surface. This reduces the burden on the operator, and the possibility of errors;
  • It is not possible to exceed the maximum ratings of the nozzle in respect of enerqy level or ● rated lifetime;
  • The possibility of the standoff being incorrectly fitted, or becoming detached, is reduced;
  • Available energy levels differ. At such low levels the clinical differences are minimal however.

The main similarities are:

  • Same spot size treatment techniques therefore remain the same; ●
  • Same material used for standoff biocompatibility and thermal stability considerations ● therefore remain the same.

12

CONCLUSION

The comparison of Technological Characteristics and the results of the performance and compliance testing conducted demonstrate that the subject device is Substantially Equivalent to the predicate device as cleared in K132754, and that the 25mm nozzle is Substantially Equivalent to that of the Rhytec nozzle/25mm standoff combination as cleared under K072394.