(55 days)
The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Surgical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided nonsterile.
The Surgical Face Masks are blue color, three-layer, flat-pleaded masks with nose piece and ear loops, which are composed of outer facing layer and inner facing layer. The colorant is blue polypropylene (PP) master batch.
The outer facing layer and inner facing layer of the mask are made of spun-bond polypropylene, the middle layer is made of melt blown polypropylene. The ear loop of the subject mask is held in place over the users' mouth and nose by two ear loops welded to the face mask. The ear loop is made with nylon and spandex. The nose piece in the layers of face mask is to allow the user to fit the mask around their nose, which is made of polypropylene wrapped aluminium.
The Surgical Face Mask is sold non-sterile and are intended to be single use, disposable devices.
The mask is designed and manufactured in accordance with ASTM F2100-19 Standard Specification for Performance of Materials Used in Surgical Face Masks.
Please find below the acceptance criteria and study information based on the provided text.
Note: The provided document is a 510(k) Premarket Notification for a Surgical Face Mask. It details performance testing against established standards for physical properties and biocompatibility, not for a software-based AI/ML medical device. Therefore, many of the requested categories (e.g., sample size for test set with data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance with humans, training set details) are not applicable to this type of device and the information is not present in the document. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance and biocompatibility testing.
Acceptance Criteria and Reported Device Performance
1. Table of acceptance criteria and the reported device performance
| Test Methodology | Purpose | Acceptance Criteria for Level 1 Barrier | Result |
|---|---|---|---|
| Bacterial Filtration Efficiency (ASTM F2101) | Measure bacterial filtration efficiency | $\ge$ 95% | Passed |
| Differential Pressure (mmH₂O/cm²) (EN 14683:2019 Annex C) | Determine breathability of the mask | <5.0 mmH₂O/cm² | Passed |
| Sub-micron Particulate Filtration Efficiency (ASTM F2299-17) | Measure initial particle filtration efficiency | $\ge$ 95% | Passed |
| Resistance to Penetration by Synthetic Blood (ASTM F1862-17) | Evaluate the resistance to penetration by impact of small volume of synthetic blood | 29 out of 32 pass at 80 mmHg | Passed |
| Flammability (16 CFR 1610-2008 Part) | Response of materials to heat and flame | Class I | Passed |
Biocompatibility Testing:
| Biocompatibility Evaluation | Biological Effect | Standard | Acceptance Criteria (Implied by standard) | Result |
|---|---|---|---|---|
| Cytotoxicity | Non-cytotoxic | ISO 10993-5 | Non-cytotoxic | Non-cytotoxic Passed |
| Sensitization | Non-sensitizing | ISO 10993-10 | Non-sensitizing | Non-sensitizing Passed |
| Irritation | Negligibly irritating | ISO 10993-10 | Negligibly irritating | Negligibly irritating Passed |
Study Details (Applicable sections based on a non-AI/ML medical device)
2. Sample size used for the test set and the data provenance:
- Sample Size: The document provides acceptance criteria for tests like "29 out of 32 pass at 80 mmHg" for synthetic blood penetration, indicating a sample size of 32 for that specific test. For other tests (Bacterial Filtration Efficiency, Differential Pressure, Particulate Filtration Efficiency, Flammability, and Biocompatibility tests), the specific sample sizes used in testing are not explicitly stated within the public 510(k) summary, only the "Passed" result against the criteria.
- Data Provenance: Not specified, but generally, such testing is performed in laboratories according to the cited ASTM, EN, and ISO standards. The tests were performed as "Non-clinical Test Conclusion" and "Non-clinical Testing."
- Retrospective or Prospective: Not applicable as this refers to device performance testing against physical and chemical properties as per standards, not clinical data collection.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. This is a physical device subject to standardized performance testing, not an AI/ML device requiring human expert ground truth for image interpretation or similar. The "truth" is established by the test methods and their defined pass/fail criteria.
4. Adjudication method for the test set:
- Not Applicable. See point 3.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is not an AI/ML medical device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is not an AI/ML medical device.
7. The type of ground truth used:
- For performance criteria (e.g., filtration efficiency, breathability, synthetic blood penetration, flammability), the "ground truth" is established by the specified international and national standards (ASTM, EN, CFR) and their associated validated test methodologies.
- For biocompatibility, the "ground truth" for safety is established by the ISO 10993 series of standards (ISO 10993-5 for Cytotoxicity, ISO 10993-10 for Sensitization and Irritation), specifically evaluating the biological response of the materials.
8. The sample size for the training set:
- Not Applicable. This is not an AI/ML medical device that uses a "training set."
9. How the ground truth for the training set was established:
- Not Applicable. See point 8.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 17, 2022
Shanghai Hua En Industrial CO LTD % Ms. Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai. 200122 China
Re: K221838
Trade/Device Name: Surgical Face Mask (HNFM0103) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: June 23, 2022 Received: June 23, 2022
Dear Ms. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K221838
Device Name Surgical Face Mask (HNFM0103)
Indications for Use (Describe)
The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Surgical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided nonsterile.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary
(As requirement by 21 CFR 807.92)
Date prepared: 16th, June, 2022
A. Applicant:
Name: SHANGHAI HUA EN INDUSTRIAL CO LTD Address: #30, NO988 HUI JI ROAD, XUANQIAO TOWN, PUDONG,SHANGHAI CHINA 2013 14 Contact: Lisa Zhao Title: General Manager Tel: +86- 15618875177 Email: lisa.zhao@hnind.com
Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@@gmail.com
B. Device:
Trade Name: Surgical Face Mask Model: HNFM0103
Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 21 CFR 878.4040 Review Panel: Surgical Apparel
C. Predicate device:
K212471 Medical Face Mask Haian Medigauze Co., Ltd
Regulatory Information Classification Name: Surgical Face Mask
{4}------------------------------------------------
Classification: Class II Product code: FXX Regulation Number: 21 CFR 878.4040 Review Panel: Surgical Apparel
D. Indications for use of the device:
The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Surgical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.
E. Device Description:
The Surgical Face Masks are blue color, three-layer, flat-pleaded masks with nose piece and ear loops, which are composed of outer facing layer and inner facing layer. The colorant is blue polypropylene (PP) master batch.
The outer facing layer and inner facing layer of the mask are made of spun-bond polypropylene, the middle layer is made of melt blown polypropylene. The ear loop of the subject mask is held in place over the users' mouth and nose by two ear loops welded to the face mask. The ear loop is made with nylon and spandex. The nose piece in the layers of face mask is to allow the user to fit the mask around their nose, which is made of polypropylene wrapped aluminium.
The Surgical Face Mask is sold non-sterile and are intended to be single use, disposable devices.
The mask is designed and manufactured in accordance with ASTM F2100-19 Standard Specification for Performance of Materials Used in Surgical Face Masks.
F. Non-clinical Test Conclusion
The Surgical Face Mask was tested in accordance with the tests recommended in the FDA guidance document, Guidance for Industry and FDA Staff Surgical Masks – Premarket Notification [510(k)] Submission issued March of 2004. Based upon the guidance document the following testing has been performed.
| Test Methodology | Purpose | Acceptance Criteria forLevel 1 Barrier | Result |
|---|---|---|---|
| Bacterial FiltrationEfficiencyASTM F2101 | Measure bacterial filtrationefficiency | $\ge$ 95% | Passed |
| Differential Pressure(mmH2O/cm²)EN 14683:2019Annex C | Determine breathability ofthe mask | <5.0 mmH2O/cm² | Passed |
| Sub-micron | Measure initial particle | $\ge$ 95% | Passed |
{5}------------------------------------------------
| ParticulateFiltrationEfficiencyF2299-17 | filtration efficiencyASTM | |||
|---|---|---|---|---|
| ResistancePenetrationSynthetic BloodASTM F1862-17 | toby | Evaluate the resistance topenetration by impact ofsmall volume of syntheticblood | 29 out of 32 pass at 80 mmHg | Passed |
| Flammability16 CFR1610-2008 | Part | Response of materials to heatand flame | Class I | Passed |
Biocompatibility Testing
The biocompatibility evaluation for the Surgical Face Mask was conducted in accordance with ISO 10993-1:2018 Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing within a Risk Management Process, as recognized by FDA. The Surgical Face Mask is classified as a surface contacting device. Specific biocompatibility tests were selected under the guidance of ISO 10993-1:2018 Annex A.
| Biocompatibility Evaluation | |||
|---|---|---|---|
| Biological Effect | Standard | Result | |
| 1 | Cytotoxicity | ISO 10993-5 | Non-cytotoxic Passed |
| 2 | Sensitization | ISO 10993-10 | Non-sensitizing Passed |
| 3 | Irritation | ISO 10993-10 | Negligibly irritating Passed |
G. Summary of Technological Characteristics
Table 1 Comparison of Proposed and Predicate Devices
| Device | Proposed Device | Predicate Device | Result |
|---|---|---|---|
| 510K # | - | K212471 | - |
| Manufacturer | SHANGHAI HUA EN INDUSTRIAL CO LTD | Haian Medigauze Co., Ltd | - |
| Product Name | Surgical Face Mask | Medical face mask | Similar |
| Level | Level 1 | Level 1 | Same |
| Product Code | FXX | FXX | Same |
| Regulation Number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Indications for use | The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Surgical Face Mask is intended for use in infection | The medical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate materials in infection control practices to reduce the potential exposure | Same |
{6}------------------------------------------------
| control practices to reduce the potentialexposure to blood and body fluids. This asingle use, disposable device(s), providednon-sterile. | to blood and body fluids. It is forsingle-use and provided non-sterile. | ||
|---|---|---|---|
| Mask Style | Flat pleated | Flat pleated | Same |
| Configurations | Ear loop; | Ear loop; | Same |
| Color | Blue | Blue | Same |
| Dimension | $180mm×95mm$ | $17.5 cm×9.5cm$ | Similar |
| Sterility | Non-sterile | Non-sterile | Same |
| Use | Single Use, Disposable | Single Use, Disposable | Same |
| Material | |||
| Outerfacinglayer | Spun-bond Polypropylene | Polypropylene | Same |
| Middle layer | Meltblown Polypropylene | Meltblown Polypropylene | Same |
| Innerfacinglayer | Spun-bond Polypropylene | Polypropylene | Same |
| Nose clip | Polypropylene wrapped aluminium | Polypropylene coated steel wire | Different |
| Ear loops | Nylon, spandex | Nylon, spandex | Same |
| Biocompatibility | |||
| Cytotoxicity | Under the conditions of the study, theproposed device extract was determined tobe non-cytotoxic. | Under the conditions of the study, theproposed device extract was determinedto be non-cytotoxic. | Same |
| Irritation | Under the conditions of the study, theproposed device extract was determined tobe non-irritating. | Under the conditions of the study, theproposed device extract was determinedto be non-irritating. | Same |
| Sensitization | Under the conditions of the study, theproposed device extract was determined tobe non-sensitizing. | Under the conditions of the study, theproposed device extract was determinedto be non-sensitizing. | Same |
Difference Analysis:
The material of nose clip of the proposed device is different to the predicate device. However, biocompatibility test has been performed on the final form of the proposed device according to ISO 10993-5 and ISO 10993-10 and the results do not show any adverse effect. Thus, this difference will not affect the safety and effectiveness between the proposed device and the predicate device.
H. Summary of Non-Clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specification. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Mask - Premarket Notification [510(K)] Submission issued on March 5, 2004:
- A ISO 10993-05:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
- A ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin
{7}------------------------------------------------
Sensitization
-
ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks
-
ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
-
EN 14683, Medical Face Masks - Requirements and Test Methods
-
ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus Aureus
-
ASTM F2299, Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
-
16 CFR 1610, Standard for the Flammability of Clothing Textiles
I. Clinical Test Conclusion
No clinical study is included in this submission.
J. Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Surgical Face Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K212471.
§ 878.4040 Surgical apparel.
(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.