K Number
K221750

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-06-30

(14 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CPS Aim™ Universal II slittable inner catheter (subselector/cannulator) is designed for intracardiac access of the coronary sinus and subselection of the venous system of the heart, and to serve as a conduit during implantation for delivery of contrast medium and Abbott Medical devices (such as guidewires and implantable left heart leads). In addition, the CPS Aim™ Universal II slittable inner catheter (subselector/cannulator) can work with outer guide catheters as a system.

Device Description

The CPS Aim™ Universal II Slittable Inner Catheter (subselector/cannulator) is used to facilitate left heart lead delivery procedures. It is an introducer that is used to cannulate the coronary venous system and act as a conduit for contrast medium, implantable coronary leads, or other devices.

The CPS Aim™ Universal II Slittable Inner Catheter design is based on the currently marketed CPS Aim™ Universal Slittable Inner Catheter design and components. Three modifications were made:

  1. Modified hydrophilic coating: the subject device uses a different hydrophilic coating material as compared to the currently marketed CPS Aim Universal catheters.
  2. Use of a modified braid design pattern to provide improved kink resistance: the predicate device braid is a single braided wire, and the subject braid contains paired braided wires.
  3. Reduction in shelf-life from three (3) years to two (2) years.

The CPS Aim™ Universal II Slittable Inner Catheter will be added as a line extension to the currently marketed family of CPS catheters. New. unique model numbers will distinguish the CPS Aim™ Universal II inner catheter models from existing CPS catheter models.

The CPS Aim™ Universal II Slittable Inner Catheters will be available in the same working lengths, 59 cm and 65 cm, as the predicate device, and will be available with the same shape of curves as the predicate device. The key design features of the CPS Aim™ Universal II Slittable Inner Catheters have only undergone a minimal change to support the coating and braid change. There is no change to the accessories, including the slitter and valve bypass tool (VBT).

The key design features CPS Aim Universal and CPS Aim Universal II Slittable Inner Catheters and their accessory, the CPS Valve Bypass Tools include:

  • Braid reinforced, varying durometer PEBAX catheter shaft with molded proximal hub.
  • Inner diameter of the catheter is PTFE lined.
  • Atraumatic distal soft tip.
  • Embedded marker band on soft tip for fluoroscopic visibility.
  • Outside surface of the catheter shaft is coated with a hydrophilic coating to provide lubricity during use.
  • Hub of the catheter is fitted with a retention cap and an integrated valve that provides hemostasis and facilitates contrast injection. The hub also includes a sideport assembly with 3-way stopcock.
  • The catheters are available in cannulator and sub-selector models.
  • Accessories such as valve bypass tool (VBT) used to assist the insertion of Abbott Medical devices (leads, guidewires, inner catheters, etc.).

Similar to the existing CPS catheters, the CPS Aim™ Universal II Slittable Inner Catheters are supplied sterile (via ethylene oxide) in one package and are intended for single procedure.

There is no change in accessory product or process for the different models as a result of this line extension. The proposed indications for use is consistent with the cleared indications.

AI/ML Overview

Based on the provided text, the device in question is a medical catheter, the CPS Aim™ Universal II Slittable Inner Catheter. The document describes a 510(k) premarket notification for this device, claiming substantial equivalence to a predicate device (CPS Aim™ Universal Slittable Inner Catheter, K130252). This type of submission generally focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate device, rather than proving novel clinical efficacy. Therefore, the information provided primarily pertains to design verification and biocompatibility testing against established standards and the predicate device.

Here's the breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly list quantified acceptance criteria for each test and then present the specific performance numbers. Instead, it states that "performance testing demonstrated that the CPS Aim™ Universal II Slittable Inner Catheter performs in a substantially equivalent manner to the currently marketed predicate device" and that the device "meets its design specifications and is as safe and effective as the predicate devices."

However, it does list the types of tests conducted, which imply the areas where performance was assessed:

Acceptance Criteria (Implied)Reported Device Performance
Biocompatibility (as per ISO 10993-1 and FDA GLP 21 CFR 58)Met standards; testing included Cytotoxicity, Sensitization, Irritation, Materials-Mediated Pyrogenicity, Acute Systemic Toxicity, Hemocompatibility (Direct Contact and Extract), Complement Activation Assay - SC5b-9, Partial Thromboplastin Time, Platelet & Leukocyte Counts, Particulate Matter per USP <788>, Surface Characterization.
Design Verification (Physical and Dimensional Characteristics)Met design specifications.
Design Verification (Functional Characteristics)Met design specifications.
Design Verification (Particulate Testing)Met design specifications.
Design Verification (Hydrophilic Coating Lubricity and Durability)Met design specifications.
Shelf-Life (Accelerated Aging)Demonstrated 24-month shelf life.

2. Sample Sizes Used for the Test Set and Data Provenance

The document does not specify the sample sizes (number of units or batches) used for each individual test (biocompatibility, design verification, shelf-life). It also does not mention the country of origin of the data or whether it was retrospective or prospective, as these types of studies typically involve laboratory bench testing and animal model testing (where applicable for biocompatibility).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of information, relating to expert consensus or clinical adjudication for establishing ground truth, is not applicable or provided in this 510(k) summary. The studies described are bench tests and biocompatibility tests, not clinical performance evaluations that would require human expert interpretation of device output.

4. Adjudication Method for the Test Set

Not applicable for the types of tests described (bench and biocompatibility testing).

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence through bench and biocompatibility testing, not on comparative clinical effectiveness with human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This device is a physical medical catheter, not an AI algorithm. Therefore, "standalone" performance in the context of an algorithm is not applicable.

7. The Type of Ground Truth Used

For biocompatibility testing, the "ground truth" is adherence to established biological safety standards (ISO 10993-1, FDA GLP 21 CFR 58), and for specific tests, it would be the absence of toxic reactions, irritation, etc., as measured by standardized assays. For design verification, the "ground truth" is adherence to predefined engineering specifications and functional requirements. For shelf-life, it's the maintenance of performance characteristics over the specified period.

8. The Sample Size for the Training Set

This device is not an AI algorithm, and therefore there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, while the text on the right is written in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.

June 30, 2022

Abbott Medical Indira Verdekel Regulatory Affairs Project Manager 15900 Valley View Court Sylmar, California 91342

Re: K221750

Trade/Device Name: CPS AIM Universal II Slittable Inner Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DQY Dated: June 15, 2022 Received: June 16, 2022

Dear Ms. Verdekel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K221750

Device Name CPS AIM™ Universal II Slittable Inner Catheter

Indications for Use (Describe)

The CPS Aim™ Universal II slittable inner catheter (subselector/cannulator) is designed for intracardiac access of the coronary sinus and subselection of the venous system of the heart, and to serve as a conduit during implantation for delivery of contrast medium and Abbott Medical devices (such as guidewires and implantable left heart leads). In addition, the CPS Aim™ Universal II slittable inner catheter (subselector/cannulator) can work with outer guide catheters as a system.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date Prepared

June 15, 2022

Submitter

Abbott Medical (formerly St. Jude Medical) 15900 Valley View Court Sylmar, CA 91342 USA Phone: (818) 362-6822

Establishment Registration: 2017865

Contact Person

Indira Verdekel Regulatory Affairs Project Manager Phone: (818) 493-3360 Indira.Valdivia@abbott.com

Natasha Loghmanpour Manager Regulatory Affairs Phone: (818) 493-3015 Natasha.Loghmanpour(@abbott.com

Device

Name of Device:CPS Aim™ Universal II Slittable Inner Catheter
Model Numbers:DS2N100-59, DS2N100-65, DS2N101-59, DS2N101-65, DS2N102-59, DS2N102-65, DS2N103-65, DS2N104-65
Common or Usual Name:Percutaneous catheter
Classification Name:Catheter Introducer (21 CFR 870.1250)
Regulatory Class:Class II
Code:DQY

Predicate Device

The line extension. CPS Aim™ Universal II Slittable Inner Catheter, is substantially equivalent in intended use and method of operation to the following predicate manufactured by Abbott Medical.

  • CPS Aim™ Universal Slittable Inner Catheter (K130252, cleared on March 20, 2013) -

{4}------------------------------------------------

Device Description

The CPS Aim™ Universal II Slittable Inner Catheter (subselector/cannulator) is used to facilitate left heart lead delivery procedures. It is an introducer that is used to cannulate the coronary venous system and act as a conduit for contrast medium, implantable coronary leads, or other devices.

The CPS Aim™ Universal II Slittable Inner Catheter design is based on the currently marketed CPS Aim™ Universal Slittable Inner Catheter design and components. Three modifications were made:

    1. Modified hydrophilic coating: the subject device uses a different hydrophilic coating material as compared to the currently marketed CPS Aim Universal catheters.
    1. Use of a modified braid design pattern to provide improved kink resistance: the predicate device braid is a single braided wire, and the subject braid contains paired braided wires.
    1. Reduction in shelf-life from three (3) years to two (2) years.

The CPS Aim™ Universal II Slittable Inner Catheter will be added as a line extension to the currently marketed family of CPS catheters. New. unique model numbers will distinguish the CPS Aim™ Universal II inner catheter models from existing CPS catheter models.

The CPS Aim™ Universal II Slittable Inner Catheters will be available in the same working lengths, 59 cm and 65 cm, as the predicate device, and will be available with the same shape of curves as the predicate device. The key design features of the CPS Aim™ Universal II Slittable Inner Catheters have only undergone a minimal change to support the coating and braid change. There is no change to the accessories, including the slitter and valve bypass tool (VBT).

The key design features CPS Aim Universal and CPS Aim Universal II Slittable Inner Catheters and their accessory, the CPS Valve Bypass Tools include:

  • Braid reinforced, varying durometer PEBAX catheter shaft with molded proximal hub.
  • . Inner diameter of the catheter is PTFE lined.
  • . Atraumatic distal soft tip.
  • Embedded marker band on soft tip for fluoroscopic visibility. ●
  • . Outside surface of the catheter shaft is coated with a hydrophilic coating to provide lubricity during use.
  • . Hub of the catheter is fitted with a retention cap and an integrated valve that provides hemostasis and facilitates contrast injection. The hub also includes a sideport assembly with 3-way stopcock.
  • The catheters are available in cannulator and sub-selector models. .
  • Accessories such as valve bypass tool (VBT) used to assist the insertion of Abbott Medical . devices (leads, guidewires, inner catheters, etc.).

{5}------------------------------------------------

Similar to the existing CPS catheters, the CPS Aim™ Universal II Slittable Inner Catheters are supplied sterile (via ethylene oxide) in one package and are intended for single procedure.

There is no change in accessory product or process for the different models as a result of this line extension. The proposed indications for use is consistent with the cleared indications.

Pediatric Use

The CPS Aim™ Universal II Slittable Inner Catheter products have not been studied in pediatric patients.

Indications for Use

The are no changes to the Indications for Use as a result of this submission. The Indications for Use for the CPS Aim™ Universal II Slittable Inner Catheter is as follows:

The CPS Aim™ Universal II Slittable Inner Catheter (subselector/cannulator) is designed for intracardiac access of the coronary sinus and subselection of the venous system of the heart, and to serve as a conduit during implantation for delivery of contrast medium and Abbott Medical devices (such as guidewires and implantable left heart leads). In addition, the CPS Aim™ Universal II Slittable Inner Catheter (subselector/cannulator) can work with outer guide catheters as a system.

Comparison of Technological Characteristics with Predicate Device

The CPS Aim™ Universal II Slittable Inner Catheter has the same technological characteristics as the currently marketed CPS Aim™ Universal Slittable Inner Catheter with minor changes to the hydrophilic coating and braid. Where differences exist between the subject device and the predicate device, performance testing demonstrated that the CPS Aim™ Universal II Slittable Inner Catheter performs in a substantially equivalent manner to the currently marketed predicate device.

Performance Data

The CPS Aim™ Universal II Slittable Inner Catheter is substantially equivalent to the predicate devices based on comparisons of the intended use, device functionality, and technological characteristics. The following performance testing was conducted to demonstrate that the device meets its design specifications and is as safe and effective as the predicate devices:

{6}------------------------------------------------

Biocompatibility Testing

The biocompatibility evaluation for the CPS Aim™ Universal II Slittable Inner Catheter was conducted in accordance with ISO 10993-1 for biological evaluation and FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued September 4, 2020. For biological testing in animal models, testing was conducted in accordance with FDA Good Laboratory Practice (GLP) Regulations 21 CFR 58. Based on the ISO and FDA guidelines, the nature of patient contact for the CPS Aim™ Universal II Inner Catheter are externally communicating device with limited (224 hours) circulating blood contact. The Valve Bypass Tools are externally communicating, devices with limited (≤24 hour) indirect blood contact.

The testing included the following:

  • Cytotoxicity ●
  • Sensitization, Irritation ●
  • Materials-Mediated Pyrogenicity
  • Acute Systemic Toxicity
  • Hemocompatibility (Direct Contact and Extract)
  • Complement Activation Assay - SC5b-9
  • Partial Thromboplastin Time
  • . Platelet & Leukocyte Counts
  • Particulate Matter per USP <788> ●
  • Surface Characterization ●

Bench Testing

  • Design Verification ●
    • Physical and Dimensional Characteristics o
    • Functional Characteristics o
    • Particulate Testing o
    • Hydrophilic Coating Lubricity and Durability O
  • Shelf-Life Testing ●
    • o Accelerated Aging 24-month (device)

Conclusions

The resulting evidence obtained from the design verification testing and the risk analysis demonstrated that the subject device is as safe and effective as the predicate device, the CPS Aim™ Universal Slittable Inner Catheter (K130252).

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).