K Number
K221648
Date Cleared
2022-08-25

(79 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

The subject device in this 510(k) Notification is Iconic Latex Glove - Powder Free Latex Examination Glove. The subject device is a patient examination glove made from Latex compound, Natural White color, powder free and non-sterile (Per 21 CFR 880.6250 - class I). The device meets the specifications in ASTM D3578-19 Standard specification for Rubber Examination Gloves. The available sizes of the subject devices are Small, Medium, Large, X-Larqe.

AI/ML Overview

The information provided describes the acceptance criteria and the study conducted for the Iconic Latex Glove, a non-powdered patient examination glove.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

Characteristics / Test MethodologyAcceptance Criteria (from Section 4.0 and 7.0)Reported Device Performance (Average Results from Section 8.0)Final Results (from Section 8.0)
Product CodeLYYLYYsame
Intended UseA powder-free patient examination glove, disposable, for medical purposes, worn on examiner's hand to prevent contamination between patient and examiner. For over-the-counter use.This powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The device is for over-the-counter use.same
Material UsedLatex - Natural RubberLatex - Natural Rubbersame
ColorN/A (Predicate: Natural White)Natural Whitesame
SterilitySterile/Non-sterileNon sterile- (Not explicitly "same", but matches expectation)
Single UseSingle useSingle usesame
Freedom from pinholes (ASTM D5151-19)AQL 2.5 Inspection Level G-1Pass (across all sizes)Pass
Overall Length (ASTM D3578-19)Min 230mmSmall: 230.8mm, Medium: 240.5mm, Large: 237.5mm, X-Large: 236.8mmPass
Width (ASTM D3578-19)Small - 80mm (±10mm); Medium - 95mm (±10mm); Large - 110mm (±10mm); X-Large - 120mm (±10mm) (Note: Table in 4.0 has typo for Large - 110 vs table in 7.0/8.0 as 111)Small: 83mm, Medium: 96mm, Large: 108.4mm, X-Large: 118.7mmPass
Thickness at Palm (ASTM D3578-19)Min 0.05 mm (from Section 4.0) Min: 0.08 mm (from Section 7.0/8.0)0.14mm (Small), 0.13mm (Medium), 0.12mm (Large), 0.14mm (X-Large)Pass
Thickness at Fingertip (ASTM D3578-19)Min 0.05 mm (from Section 4.0) Min: 0.08 mm (from Section 7.0/8.0)0.10mm (Small), 0.10mm (Medium), 0.09mm (Large), 0.14mm (X-Large)Pass
Physical Properties Before Aging (ASTM D3578-19)Tensile Strength = 18 MPa, min.; Ultimate Elongation = 650 % minTensile Strength: 20.8 (S), 23.4 (M), 21.4 (L), 20.7 (XL); Ultimate Elongation: 1018.7 (S), 878.9 (M), 1127.9 (L), 1097.6 (XL)Pass
Stress at 500% Elongation (Before Aging)N/A (Standard specifies min tensile and ultimate elongation)3.3 (S), 3.9 (M), 3.7 (L), 3.3 (XL)Pass
Physical Properties After Accelerated Aging (ASTM D3578-19)Tensile Strength = 14 MPa, min.; Ultimate Elongation = 500 % minTensile Strength: 15.3 (S), 17.3 (M), 16.9 (L), 17.4 (XL); Ultimate Elongation: 856.4 (S), 890.8 (M), 1077.3 (L), 956.8 (XL)Pass
Residual Powder (ASTM D6124-06)< 2.0 mg/pc0.12 (S), 0.26 (M), 0.38 (L), 0.18 (XL)Pass
Extractable Protein (ASTM D5712-15)Maximum 50 µg/dm²Small = 10.3 µg/dm², Medium = 8.7 µg/dm², Large = 9.6 µg/dm², X-Large = 8.8 µg/dm²Pass
In vitro cytotoxicity (ISO 10993-5)Complies with the requirement of this standardComplies with the requirement of this standardPass
Skin sensitization (ISO 10993-10)Non-SensitizerUnder the conditions of this study, the test article was a non-sensitizer (also stated as "Non - Skin Sensitized")Pass
Systemic toxicity (ISO 10993-11)Not induce systemic toxicityNot induce systemic toxicityPass
Irritation (ISO 10993-23)Non-irritantUnder the conditions of this study, the test article was a non-irritant (also stated as "Non-Irritant")Pass

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes used for each individual test (e.g., number of gloves tested for pinholes, physical properties, etc.). It only provides the average results for different glove sizes (Small, Medium, Large, X-Large). The data provenance is not explicitly stated beyond the fact that the tests were conducted by a company in Malaysia (Iconic Medicare Sdn Bhd is in Malaysia, and the general testing information implies this is where their reports originated). These appear to be retrospective tests conducted on manufactured batches of the device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This is not applicable as the device is a medical glove, and the "ground truth" is established through standardized laboratory testing against international and national standards (ASTM, ISO). There are no "experts" in the sense of clinicians or radiologists establishing ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable. Adjudication methods like "2+1" or "3+1" are typically used for interpreting results (e.g., medical images) where human judgment is involved in establishing a diagnosis or reading. The tests performed for this device are objective measurements against defined acceptance criteria in laboratory settings.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a medical glove, not an AI-powered diagnostic or assistive tool. MRMC studies are used to evaluate diagnostic performance with and without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not applicable. The device is a medical glove, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is established by objective measurements and results from standardized laboratory tests according to recognized national and international standards (ASTM D3578-19, ASTM D5151-19, ASTM D6124-06, ASTM D5712 -15, ISO 10993-5:2009, ISO 10993-10:2021, ISO 10993-11:2017, ISO 10993-23:2021).

8. The sample size for the training set

This is not applicable. The device is a physical product (a glove), not a machine learning model that requires a training set.

9. How the ground truth for the training set was established

This is not applicable (see point 8).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 25, 2022

Iconic medicare sdn bhd % A.C. Thirumaran Official Correspondent Integrated Assessment Services Pvt Ltd 1495, Manasarovar, 16th Main road, Anna Nagar West Chennai, Tamil Nadu 600040 India

Re: K221648

Trade/Device Name: Iconic Latex Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LYY Dated: May 20, 2022 Received: June 7, 2022

Dear A.C. Thirumaran:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221648

Device Name Iconic Latex Glove

Indications for Use (Describe)

This non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1.0 Submitter:

Mr. Tan Cho Chia Managing Director Company Name: Iconic Medicare Sdn Bhd. Company Address: PMT 798, LINGKARAN CASSIA SELATAN. TAMAN PERINDUSTRIAN BATU KAWAN, 14110 BANDAR CASSIA, PULAU PINANG. MALAYSIA Email: cctan@iconic.com.my Telephone: 60 4 504159 Date of Summary Prepared: 15th March 2022

2.0 Subject Device Identification:

Trade Name / Proprietary Name: Iconic Latex Glove Device Common Name: Non-powder Patient Examination gloves. Device Classification Name: Natural Rubber Latex Patient Examination gloves Device Classification: 1 Regulation Number: 21 CFR 880.6250 Product Code: LYY

3.0 Official Correspondent

Mr. A.C. Thirumaran Integrated Assessment Services Private Limited No.1495, Manasarovar, 16th Main road, Anna Nagar west, Chennai- 600040, India. Telephone: 91-44-26162670 Email: iasfda16@gmail.com

4.0 Identification of the Legally Marketed Device:

Predicate Device: Careglove Global SDN BHD 510k Number: - K161833 Device Name: Latex Examination Gloves Powder Free Classification Name: Natural Rubber Latex Patient Examination Gloves Device Classification: 1 Regulation Number: 21 CFR 880.6250

5.0 Device Description

The subject device in this 510(k) Notification is Iconic Latex Glove - Powder Free Latex Examination Glove. The subject device is a patient examination glove made from Latex compound, Natural White color, powder free and non-sterile (Per 21 CFR 880.6250 - class I). The device meets the specifications in ASTM D3578-19 Standard specification for Rubber Examination Gloves. The available sizes of the subject devices are Small, Medium, Large, X-Larqe.

6.0 Indications for Use

{4}------------------------------------------------

lconic Latex Glove. This non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

7.0 Technological characteristics Comparison for the proposed and predicate devices

The Iconic Latex Glove is a Powder Free Latex Examination Glove, Non-sterile, with the following summarized technological characteristics, in comparison to the Predicate device and specifications in ASTM D3578-19.

CharacteristicsAcceptance CriteriaSubject device:Iconic Latex Glove(Small, Medium, Large,X-Large)Predicate Device LatexExamination GlovesPowder Free CaregloveGlobal SDN BHDComparisonAnalysis
510(k) Number-K221648K161833-
Product CodeLYYLYYLYYsame
Intended useA powder freepatient examinationglove is adisposable deviceintended formedicalpurposes that isworn on theexaminer's hand orfinger to preventcontaminationbetween patientand examiner. Thedevice is for over-the-counter use.This powder free patientexamination glove is adisposable deviceintended for medicalpurposes that is worn onthe examiner's hand topreventcontamination betweenpatient and examiner.The device is for over-the- counter use.A patient examinationglove is a disposabledeviceintended for medicalpurposes that is worn onthe examiner's hand orfinger to preventcontamination betweenpatient and examiner.same
Material usedLatex - NaturalRubberLatex - Natural RubberLatex - Natural Rubbersame
ColorN/ANatural WhiteNatural Whitesame
SterilitySterile/Non-sterileNon sterileNon sterilesame
Single useSingle useSingle useSingle usesame
Overall Length(mm) Min 230mmMeets ASTM D3578-19Meets ASTM D3578-19
DimensionsWidth (±10mm)Small - 80Medium- 95Large-110X-large-120Meets ASTM D3578-19Meets ASTM D3578-19same
CharacteristicsAcceptance CriteriaSubject device:Iconic Latex Glove(Small, Medium, Large,X-Large)Predicate Device LatexExamination GlovesPowder Free CaregloveGlobal SDN BHDComparisonAnalysis
PhysicalPropertiesThickness at Palm(mm) Min; 0.05 mmMeets ASTM D3578- 190.06-0.09mmsame
Thickness at FingerTip (mm) Min 0.05mmMeets ASTM D3578- 190.07-0.10mm
PhysicalPropertiesBefore AgingTensile Strength =18 MPa, min.UltimateElongation = 650% min Stress at500 % Elongation(MPa)After AcceleratedAgingTensile Strength=14 MPa, min.UltimateElongation = 500% minMeets ASTM D3578- 19Meets ASTM D3578- 19same
Freedom frompinholesAQL 2.5 InspectionLevel G-1Meets ASTM D5151- 19Meets ASTM D5151- 19same
ResidualPowder< 2.0 mg/pcMeets ASTM D6124- 06Meets ASTM D6124- 06same
Bio-CompatibilityISO 10993-23:2010Biologicalevaluation ofmedical devices:Tests for irritationUnder the conditions ofthis study, the test articlewas a non- irritant.N/Asame
ISO 10993-10:2021Biologicalevaluation ofmedical devices -Part10: Tests For skinsensitizationUnder the conditions ofthis study, the test articlewas a non- sensitizer.N/Asame
ISO 10993-5:2009Biologicalevaluation ofmedical devices -Part5: Tests for in vitrocytotoxicityComplies with therequirement of thisstandardN/Asame
CharacteristicsAcceptance CriteriaSubject device:Iconic Latex Glove(Small, Medium, Large,X-Large)Predicate Device LatexExamination GlovesPowder Free CaregloveGlobal SDN BHDComparisonAnalysis
ISO 10993-11:2017Biologicalevaluation ofmedical devices —Part11: Tests forsystemic toxicityNot induce systemictoxicityN/ASame
ExtractableProteinWater ExtractableProtein, Maximum50 µg/dm²ASTM D 5712-15Small = 10.3 µg/dm²Medium = 8.7 µg/dm²Large = 9.9 µg/dm²X Large = 8.8 µg/dm²< 50 µg/dm²Same

{5}------------------------------------------------

510(k) SUMMARY 510(k) No: K221648

{6}------------------------------------------------

510(k) SUMMARY 510(k) No: K221648

8.0 Summary of non-clinical testing results

lconic Nitrile Glove was tested and found in conformance with the following standards:

ASTM D3578-19Standard Specification for Rubber Examination Gloves for Medical Application
ASTM D5151-19Standard Test Method for detection of Holes in Medical Gloves
ASTM D6124-06Standard Test Method for Residual Powder on Medical Gloves
ASTM D5712 -15Standard Test Method for Analysis of Aqueous Extractable Protein in Latex,Natural Rubber, and Elastomeric Products Using the Modified Lowry Method
ISO 10993-5:2009Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021Biological evaluation on medical device Part 10: Test for Skin Sensitization
ISO 10993-11:2017Biological evaluation of medical devices Part 11: Tests for systemic toxicity
ISO 10993-23:2021Biological evaluation of medical devices Part 23: Tests for irritation
TestMethodologyPurposeAcceptanceCriteriaAverage ResultsFinalResults
SMALLMediumLargeX Large
Sterility-Non sterileNon sterileNon sterileNon sterile-
ASTM D3578-19Freedom fromhole - ASTMD5151-19AQL 2.5PassPassPassPassPass
Overall Length(mm) Min230mm.230.8240.5237.5236.8Pass
Dimension - width,Length,ThicknessWidth (±10mm)Small - 80Medium - 95Large -111X-large - 1208396108.4118.7Pass
Thickness at Palm & fingertip Min: 0.08 mm

{7}------------------------------------------------

510(k) SUMMARY 510(k) No: K221648

TestMethodologyPurposeAcceptanceCriteriaAverage ResultsFinal
SMALLMediumLargeX LargeResults
Palm0.140.130.120.14Pass
Fingertip0.100.100.090.14Pass
a. Before Aging
TensileStrength=18MPa, min.20.823.421.420.7Pass
UltimateElongation=650% min1018.7878.91127.91097.6Pass
Physicalproperties beforeaging, afteraccelerated agingStress at 500%Elongation3.33.93.73.3Pass
b. After Accelerated Aging
TensileStrength=14MPa, min.15.317.316.917.4Pass
UltimateElongation= 500% min856.4890.81077.3956.8Pass
Powder-freeResidue exceedsmaximum limit -ASTM D6124-06< 2.0 mg perglove0.120.260.380.18Pass
ExtractableProtein content$50 µg/dm2$10.38.79.68.8Pass
ISO 10993-5Test for InvitrocytotoxicityCytotoxicCharacteristicsPass
ISO 10993-10Test for irritationand SkinSensitizationNon - SkinSensitizedPass
ISO 10993-11Tests for systemictoxicityNot inducesystemic toxicityPass
ISO 10993-23Tests for irritationNon-IrritantPass

9.0 Summary of clinical Performance data

Not applicable - Clinical data was not used to assess performance of the subject device.

10.0Conclusion

The Conclusion drawn from the Non-Clinical test demonstrates that the subject device- Iconic Latex Glove is as safe, as effective, and performs as well as or better than the legally marketed Predicate device cleared under K161833.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.