K Number
K221433
Date Cleared
2022-07-15

(59 days)

Product Code
Regulation Number
878.4850
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Facet 28G Universal Lancet is a sterile, disposable single used to obtain a droplet of capillary blood for subsequent diagnostic testing from the finger or an alternative site, such as the palm, upper arm, or forearm. The Lancet is to be properly disposed of after a single use on an individual child, adolescent, or adult patient in a home setting.

Device Description

The Facet 28G Universal Lancet) is a sterile, single use, blood sampling device used to obtain a sample of capillary blood for diagnostic purposes, primarily for blood glucose monitoring in diabetic patients. The lancets consist of a 28G stainless steel needle overmolded with low density polyethylene (LDPE) and integral sterility cap. The lancet is single use, disposable and is sterilized by gamma radiation to a SAL of 10-9.

AI/ML Overview

This document is a 510(k) summary for the Facet 28G Universal Lancet, a medical device. It does not describe an AI/ML-enabled device, nor does it contain information about clinical studies with human participants, expert ground truth establishment, or multi-reader multi-case (MRMC) studies. Therefore, many of the requested points regarding acceptance criteria and study details for AI/ML devices cannot be answered from this document.

However, I can extract the acceptance criteria and performance related to the non-clinical bench testing for this physical device.


Device: Facet 28G Universal Lancet

Nature of the Device: This is a physical medical device (blood lancet), not an AI/ML software device. The acceptance criteria and performance data presented are for physical and material properties, and compatibility, not for diagnostic accuracy or algorithmic performance.


1. Table of Acceptance Criteria and the Reported Device Performance

The document describes non-clinical bench testing, which serves as the "study" to prove the device meets pre-determined criteria. The reported performance is not explicitly given in numerical terms for all tests, but rather stated that "predetermined criteria were met" and "risks were appropriately managed." However, the acceptance criteria (requirements) are provided.

Property/CharacteristicAcceptance Criteria (Requirement)Reported Device Performance (Implied)
Bond Strength of lancet blade (needle) to lancet body$\geq$ 13.23 NMet the criteria (stated "predetermined criteria were met")
Sterility Cap Torque$\leq$ 4.9 N-cmMet the criteria (stated "predetermined criteria were met")
Compatibility with lancet bases (lancing devices)Ensure usability with commercially available lancet basesMet the criteria (stated "predetermined criteria were met")
BiocompatibilityMeet ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11 requirements (Cytotoxicity, Sensitization, Irritation or Intracutaneous reactivity, Acute Systemic toxicity)Met the criteria (stated "predetermined criteria were met")
SterilitySterilized by Gamma Radiation SAL = $10^{-6}$Met the criteria (Same as predicate)

Note: The document only states that the criteria were "met" or that the performance was "same" as the predicate. Specific numerical results of the tests (e.g., the exact bond strength measured) are not provided in this summary.


2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: The document does not specify the numerical sample sizes for each non-clinical test (e.g., how many lancets were tested for bond strength or cap torque). It broadly states that "non-clinical bench testing was performed."
  • Data Provenance: This is not applicable in the context of clinical data for an AI/ML device. For this physical device, the data provenance is from internal non-clinical bench testing performed by Facet Technologies, LLC. The testing ensures the device meets specified engineering and biocompatibility requirements.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Not applicable. This device is a physical lancet, not an AI/ML system requiring expert interpretation of data. The "ground truth" here is objective measurement of physical properties (e.g., force, torque) and standard biocompatibility testing.

4. Adjudication Method for the Test Set

  • Not applicable. There is no human interpretation or adjudication required for the physical property tests or biocompatibility assessments described.

5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI-enabled device and no MRMC studies were performed or are relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is not an AI/ML algorithm.

7. The type of ground truth used

  • For physical properties (Bond Strength, Cap Torque): The ground truth is established by objective physical measurements against engineering specifications.
  • For Biocompatibility: The ground truth is established by standardized international (ISO 10993 series) and FDA-recognized test methods performed in a laboratory setting.
  • For Compatibility: The ground truth is established via simulated use with commercially available lancing devices.

8. The sample size for the training set

  • Not applicable. This is not an AI/ML device, so there is no "training set." The focus is on verifying compliance with design specifications and regulatory standards.

9. How the ground truth for the training set was established

  • Not applicable. There is no training set as this is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it, also in blue.

July 15, 2022

Facet Technologies LLC James Bonds Director Regulatory Affairs 3900 N. Commerce Dr. Atlanta, Georgia 30344

Re: K221433

Trade/Device Name: Facet 28G Universal Lancet Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: QRK, QRL Dated: May 16, 2022 Received: May 17, 2022

Dear James Bonds:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221433

Device Name Facet 28G Universal Lancet

Indications for Use (Describe)

The Facet 28G Universal Lancet is a sterile, disposable single used to obtain a droplet of capillary blood for subsequent diagnostic testing from the finger or an alternative site, such as the palm, upper arm, or forearm. The Lancet is to be properly disposed of after a single use on an individual child, adolescent, or adult patient in a home setting.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Facet 28G Universal Lancet

510(k) Summary

(Reference 21 CFR 807.92)

Submitted by:Facet Technologies, LLC3900 North Commerce DriveAtlanta, GA 30344-8149Phone Number: (770) 590-6462Fax Number: (770) 590-6412
Contact:James R. Bonds
Date of Preparation:July 13, 2022
Device Trade Name:Facet 28G Universal Lancet
Common Name:Blood Lancet
Classification Name:Single use only blood lancet without an integral sharps injuryprevention feature
Regulation:878.4850
Product codes:QRK (Single use only blood lancet without an integral sharpsinjury prevention feature)QRL (Multiple use blood lancet for single patient use only)
Product Classification:II
Panel:General & Plastic Surgery
Predicate Device:Facet 28G Universal Lancet (Product Code FMK, 510(k)Exempt)

{4}------------------------------------------------

Device Description

The Facet 28G Universal Lancet) is a sterile, single use, blood sampling device used to obtain a sample of capillary blood for diagnostic purposes, primarily for blood glucose monitoring in diabetic patients.

The predicate device has been in commercial distribution since 2006.

Facet lancets are intended to be used by diabetic patients to obtain a blood sample for assistance in self-monitoring of their blood glucose levels. The lancets consist of a 28G stainless steel needle overmolded with low density polyethylene (LDPE) and integral sterility cap. The lancet is single use, disposable and is sterilized by gamma radiation to a SAL of 10-9.

The predicate device is currently in commercial distribution in the United States, European, and other worldwide markets. Since commercial distribution of the current lancets began, there have been no design changes.

Indications for Use

The Facet 28G Universal Lancet is a sterile, disposable, single use device used to obtain a droplet of capillary blood for subsequent diagnostic testing from the finger or an alternative site, such as the palm, upper arm, or forearm. The Lancet is to be properly disposed of after a single use on an individual child, adolescent, or adult patient in a home setting.

Technoloqical Characteristics

The primary technological characteristics and intended use of the Facet 28G Universal Lancet are substantially equivalent to the predicate device.

As indicated in Table 1, the Facet 28G Universal Lancet is substantially equivalent to characteristics of the identified predicate device, the Facet 28 G Universal Lancet previously marketed as a 510(k) exempt device.

{5}------------------------------------------------

Subject Device

Predicate Device

Indication for UseThe Facet 28G UniversalLancet is a sterile,disposable single usedevice used to obtain adroplet of capillary bloodfor subsequent diagnostictesting from the finger oran alternative site, suchas the palm, upper arm,or forearm. The Lancet isto be properly disposed ofafter a single use on anindividual child,adolescent, or adultpatient in a home setting.Same
Intended UseThe Facet 28G UniversalLancet is intended toperform a skin puncture ofa finger or alternate site(palm of the hand, upperarm, or forearm) forcollection of a droplet ofcapillary blood forsubsequent diagnostictesting. The Lancet maybe used in combinationwith a reusable lancetbase (lancing device) thataccepts a universal-typelancet to perform alancing event. The Facet28G Universal Lancet isfor single use only on anindividual patient.Same
ManufacturerFacet Technologies, LLCSame
510(k) NumberN/AK221433
Product CodeFMKQRK, QRL
Tip configurationBevelBevel
Needle Length3.1±0.45 mmSame
Needle Gauge28GSame
Body ColorLight blueSame

Table 1: Comparison of Subject Device and Predicate Device

Characteristic

{6}------------------------------------------------

CharacteristicPredicate DeviceSubject Device
Needle Material304 Stainless steelSame
Body MaterialLow density polyethylene(LDPE)Same
Body/needle bondstrength$\geq$ 13.23 NSame
Cap Twist-off torque$\leq$ 4.9 N-cmSame
SterilitySterilized by Gamma RadiationSAL = $10^{-6}$Same

Non-clinical Testing Summary and Conclusions

Non-clinical bench testing was performed to ensure predetermined criteria were met and the special controls (21 CFR 878.4850) were satisfied. This includes mechanical design verification and validation testing in order to ensure the risks were appropriately managed in addition to verifying that the device continued to meet the specified requirements over the shelf life of the device. Physical testing included lancet body to needle bond strength, sterility cap twist-off torque, and compatibility with commercially available reusable lancet bases (lancing devices). Biocompatibility was evaluated through a battery of tests to meet ISO 10993-1 requirements.

Table 2:Summary of Nonclinical Tests Performed
------------------------------------------------------
Property/CharacteristicTest MethodImportanceReference
Bond Strength of lancetblade (needle) to lancetbodyTensile strengthtesterEnsure needle doesnot detach fromlancet body duringuseN/A
Sterility Cap TorqueTorque testerEnsure that the capcan be easilyremoved by userwhen desiredN/A
Compatibility with lancetbases (lancing devices)Simulated useEnsure usability withcommerciallyavailable lancetbasesN/A

{7}------------------------------------------------

Property/CharacteristicTest MethodImportanceReference
BiocompatibilityCytotoxicity,Sensitization,Irritation orIntracutaneousreactivity, AcuteSystemic toxicity(materials mediatedpyrogen)Ensure material ofconstruction arebiocompatible fortheir intended useFDA Guidance Useof InternationalStandard 10993-1,"Biological evaluationof medical devices -Part 1: Evaluationand testing within arisk managementprocess", Sept.2020,ISO 10993-1, ISO10993-5, ISO 0993-10, ISO 10993-11

In summary, the results of nonclinical testing demonstrate that the candidate device is substantially equivalent to the predicate device.

Conclusion

The intended use, technology, non-clinical testing, and functionality of the Facet 28G Universal Lancet demonstrate a substantially equivalent safety and effectiveness profile to the predicate device and should perform as well as the predicate in the intended use conditions.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.