(57 days)
The CLAROX is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors.
CLAROX is a battery-operated, portable dental X-ray source designed for handheld operation. It is designed to produce diagnostic quality X-ray images. The CLAROX is designed for use in a dental office. It can also be used in other similar environments (orthodontic office, general practitioner's office, hospital ward, etc.) where appropriate safeguards are implemented. The device uses a rechargeable battery to allow for the use of the CLAROX where transportation or use of other x-ray devices might be prohibitive due to ther device's size and/or lack of mobility. The CLAROX is an X-ray device with a AC/DC adaptor. The handheld device features a main body (tube head), Cone for sensor, Acryl shielding, cradle, and AC/DC adaptor.
The provided document is a 510(k) summary for the CLAROX dental X-ray system, which is an X-ray generator. It focuses on demonstrating substantial equivalence to a predicate device based on technical specifications and adherence to recognized standards, rather than clinical performance (e.g., diagnostic accuracy). Therefore, many of the requested sections related to clinical studies, ground truth, expert review, and MRMC studies are not applicable or not detailed in this document.
Here's an analysis of the provided text in relation to your request:
Acceptance Criteria and Study to Prove Device Meets Criteria:
The document does not specify "acceptance criteria" in terms of diagnostic performance metrics (e.g., sensitivity, specificity, AUC) for the CLAROX device itself. Instead, it demonstrates substantial equivalence to a predicate device (KaVo NOMAD Pro2 Handheld X-ray System) by comparing technological characteristics and showing compliance with international safety and performance standards.
The study that "proves the device meets acceptance criteria" is described as non-clinical testing and bench testing to demonstrate compliance with these standards and similar technological characteristics to the predicate device.
1. Table of acceptance criteria and the reported device performance
Since the "acceptance criteria" are not explicit performance metrics for diagnostic capability, but rather conformance to standards and comparable technical specifications, the table below will reflect the comparison between the subject device (CLAROX) and the predicate device (KaVo NOMAD Pro2) as presented in the document. The "reported device performance" refers to the specifications of the CLAROX device.
| Characteristic | Acceptance Criteria (Predicate Device Specification) | Reported Device Performance (CLAROX) |
|---|---|---|
| Indications for Use | "The KaVo NOMAD Pro 2 Handheld X-ray System is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors." | "The CLAROX is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors." |
| Body Size | 279mm(L) × 140mm(W) × 267mm(H) | 300mm(W) × 100mm(D) × 252mm(H) |
| Weight | 2.73kg | 1.5kg |
| Source to skin distance | 210mm | 200mm |
| Cone diameter | 60mm | 60mm |
| User Interface | Control buttons, display, and a trigger | control panel, control buttons, display, and an exposure button |
| Energy Source | Rechargeable 22.2 V DC Li-polymer battery core pack | Rechargeable 22.2 V DC Li-polymer battery core pack |
| Battery Capacity | 1.7 A-hr | 1.46 A-hr |
| Recharge capability | 70% remaining capacity after 300 cycles | 80% or above remaining capacity after 300 cycles |
| Exposure time | 0.02 - 1.0 seconds | 0.12 - 1.2 seconds |
| Timer Accuracy | $\pm$ (10% + 1ms) | Less than 5% or 20 ms |
| mA | 2.5 mA | 2.0 mA |
| kVp | 60 kVp | 70 kVp |
| Waveform | Constant Potential (DC) | Constant Potential (DC) |
| Compliance with Standards | - (Implicitly compliant with relevant standards) | AAMI ES60601-1; IEC 60601-1-2 (Ed. 4); IEC 60601-1-3; IEC 60601-2-65; ANSI AAMI IEC 62304:2006; EN ISO 14971:2019; ANSI AAMI IEC 62366-1:2015+AMD1:2020; IEC 60601-1-6 Ed. 3.1; IEC/EN 62133-2 Ed. 1.0 |
Note on Differences: While some specifications (e.g., weight, battery capacity, exposure time range, mA, kVp) differ, the document concludes that these differences do not raise new questions of safety or effectiveness and are within a range considered substantially equivalent for an extraoral X-ray source. For instance, the subject device has a lower weight and higher kVp compared to the predicate.
2. Sample size used for the test set and the data provenance
The document states, "Clinical testing is not necessary for the current submission, based on the device type (an x-ray generator) and based on the technological characteristics similar to the predicate system." Therefore, there is no test set of clinical images or patient data described in this submission. All testing was non-clinical (bench testing) to verify technical specifications and compliance with standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable, as no clinical test set with a ground truth established by experts was used.
4. Adjudication method for the test set
Not applicable, as no clinical test set was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The CLAROX device is an X-ray generator, not an AI-powered diagnostic tool. Its submission is based on substantial equivalence as an X-ray source, not on improving human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable, as the CLAROX is an X-ray generator, not an algorithm, and does not perform diagnostic interpretation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no clinical test set requiring ground truth for diagnostic accuracy was utilized. The "ground truth" in this context would implicitly be the validated specifications and performance of the predicate device and the requirements of the international standards.
8. The sample size for the training set
Not applicable. This device is an X-ray generator, not a machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable. There is no training set for this device.
{0}------------------------------------------------
June 29, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
VSI Co., Ltd. % Mr. Edward Park CEO LightenBridge LLC 4408 Tortuga Ln MCKINNEY TX 75070
Re: K221286
Trade/Device Name: CLAROX Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: EHD Dated: May 3, 2022 Received: May 3, 2022
Dear Mr. Park:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K221286
Device Name CLAROX
Indications for Use (Describe)
The CLAROX is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Value Service Innovation (VSI). The logo consists of a blue symbol on the left, followed by the text "Value Service Innovation" in a smaller font size above the letters "VSI" in a larger, bold font. The symbol on the left appears to be three circles connected in a triangular pattern.
VSI Co., Ltd. 151-33. Namseok-ro Nami-myeon, Seowon-gu Cheongju-si, Chungcheongbuk-do 28182 South Korea Tel. +82-43-267-9039 / Fax. +82-43-269-4039
510(k) Summary - Traditional 510(k)
K221286
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
Submitter Information
| Submitter Name: | VSI Co., Ltd. |
|---|---|
| Address: | 151-33, Namseok-ro Nami-myeon, Seowon-gu, Cheongju-si,Chungcheongbuk-do, 28182 Republic of Korea |
| Phone/Fax | +82-43-267-9039 / +82-43-269-4039 |
| Contact Person: | Edward Park, official correspondent of AssembleCircle Corp. |
| Date of submission: | Apr 15, 2022 |
| Device Information | |
| Proprietary Name(s): | CLAROX |
| Model Name: | VX-30 |
| Common Name: | Portable X-ray Equipment |
| Regulation Name: | Extraoral Source X-Ray System |
| Product Code: | EHD |
| Regulation Number: | 21 CFR 872.1800 |
| Classification Panel: | Radiology |
Device Class: II
Device Description
CLAROX is a battery-operated, portable dental X-ray source designed for handheld operation. It is designed to produce diagnostic quality X-ray images. The CLAROX is designed for use in a dental office. It can also be used in other similar environments (orthodontic office, general practitioner's office, hospital ward, etc.) where appropriate safeguards are implemented. The device uses a rechargeable battery to allow for the use of the CLAROX where transportation or use of other x-ray devices might be prohibitive due to ther device's size and/or lack of mobility. The CLAROX is an X-ray device with a AC/DC adaptor. The handheld device features a main body (tube head), Cone for sensor, Acryl shielding, cradle, and AC/DC adaptor.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Value Service Innovation (VSI). The logo consists of a blue symbol on the left, resembling three interconnected circles. To the right of the symbol, the text "Value Service Innovation" is written in a smaller font size above the letters "VSI", which are written in a larger, bold font.
VSI Co., Ltd. 151-33. Namseok-ro Nami-mveon. Seowon-gu Cheongju-si, Chungcheongbuk-do 28182 South Korea Tel. +82-43-267-9039 / Fax. +82-43-269-4039
The power is supplied by a rechargeable Lithium Ion battery core pack built into a main body. This facilitates portability of the device. A beam-limiting cone is incorporated within the device. The acryl shielding provide sufficient radiation protection to allow the clinician to remain in the operatory with the patient. To make the system as simple as possible for the operator, CLAROX uses a fixed tube voltage of 70kV and a fixed tube current of 2.0 mA. The only operator-adjustable parameter is the exposure time. This adjustment can be quickly accomplished through the user-friendly control panel. control buttons, display, and an exposure button provide the primary operator interface. Exposures settings can be selected and displayed. Voltage (70 kV) and current (2.0 mA) are fixed with the exposure time varying based on patient type, detector type, and anatomical feature. Exposures can be completed using the exposure button. The CLAROX should be used with an X-ray detector, and this detector has digital type and analog film type. But the X-ray detector is not included in the device package. The functions of the CLAROX x-ray system are controlled by software (firmware). The device software is not based on the predicate and it is of Moderate level of concern.
Principle of Operation / Mechanism of Action:
The CLAROX is used like any other extraoral dental Xray source for intraoral application. An image receptor, such as film, is placed in the patient's oral cavity behind the teeth. The device is powered on, and the appropriate exposure time is set by the operator. The operator should follow appropriate instructions to ensure proper alignment of the X-ray beam to the receptor, and proper positioning of a detector in the patient's mouth. The irradiation time is set manually. To prevent inadvertent exposure to X-rays, the value adjusted by the MCU electronically controls the semiconductor switching element. Using this X-ray, it passes through the body part to be diagnosed to obtain an image with a detector so that it can be used for diagnosis. The irradiation time setting method sets the irradiation time by the user and can select a value from 0.12 seconds to 1.2 seconds by turning the dial manually. After setting, press the investigation button to perform investigation. The system has numerous beep sound alerts to communicate with the operator.
Predicate Device
- . KaVo NOMAD Pro2 Handheld X-ray System (Aribex K173319, 10/20/2017)
- Common Name: Extraoral Source X-ray System
- Regulation Name: Extraoral Source X-Ray System
- Device Class: II:
- Product Code: EHD
- Regulation Number: 21 CFR 872.1800
- Classification Panel: Radiology
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo with a blue circular design on the left. To the right of the design, the text "Valas Service Innovation" is stacked on top of the letters "VSI" in a larger, bold font. The text is black and appears to be a company logo.
VSI Co., Ltd. 151-33. Namseok-ro Nami-myeon, Seowon-gu Cheongju-si, Chungcheongbuk-do 28182 South Korea Tel. +82-43-267-9039 / Fax. +82-43-269-4039
Indications for Use
The CLAROX is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors.
Technological Characteristics
The CLAROX retains the same basic design components and operating features as the predicate device, KaVo NOMAD Pro 2 Handheld X-ray System (K173319). The handheld device features a main body, Acryl shielding, and cradle for charging. The X-ray tube (70 kV, 2.0 mA) and battery pack are integrated in the main body. The functionality of the user interface is also similar to the previous product, including the pre-programmed exposure times for adult and pediatric patients. The power is supplied by a rechargeable Lithium-Ion battery core pack built into a handset. Both devices have 22.2 VDC but CLAROX has lower battery capacity, but recharge capacity of the subject device is better than the predicate device. The design of the subject device also looks similar to the predicate device. The battery core packs in both devices are compliant with IEC 62133. Testing has been completed on basic safety and essential performance and the device complies with AAMI ES60601-1; IEC 60601-1-2 (Ed. 4); IEC 60601-1-3, and IEC 60601-2-65.
| Device Name | CLAROX | KaVo NOMAD Pro2Handheld X-ray System |
|---|---|---|
| 510k number | K173319 | |
| Manufacturer | VSI Co., Ltd. | Aribex |
| Indicationsfor Use | The CLAROX is indicated for use only by atrained and qualified dentist or dentaltechnician for both adult and pediatricsubjects as an extraoral diagnostic dental X-ray source to produce E-ray images usingintraoral image receptors. | The KaVo NOMAD Pro 2 Handheld X-raySystem is indicated for use only by a trainedand qualified dentist or dental technician forboth adult and pediatric subjects as anextraoral diagnostic dental X-ray source toproduce X-ray images using intraoral imagereceptors. |
| Body Size | 300mm(W) × 100mm(D) × 252mm(H) | 279mm(L) × 140mm(W) × 267mm(H) |
| Weight | 1.5kg | 2.73kg |
| Source to skindistance | 200mm | 210mm |
| Cone diameter | 60mm | 60mm |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a blue abstract symbol on the left and the text "Value Service Innovation VSI" on the right. The abstract symbol consists of three interconnected circles, forming a triangular shape. The text "Value Service Innovation" is smaller and positioned above the larger letters "VSI", which are in a bold, serif font.
VSI Co., Ltd. 151-33. Namseok-ro Nami-myeon, Seowon-gu Cheongju-si, Chungcheongbuk-do 28182 South Korea Tel. +82-43-267-9039 / Fax. +82-43-269-4039
| Device Name | CLAROX | KaVo NOMAD Pro2Handheld X-ray System |
|---|---|---|
| 510k number | K173319 | |
| Manufacturer | VSI Co., Ltd. | Aribex |
| User Interface | control panel. control buttons, display,and an exposure button | Control buttons, display, and a trigger |
| Energy Source | Rechargeable 22.2 V DC Li-polymerbattery core pack | Rechargeable 22.2 V DC Li-polymerbattery core pack |
| BatteryCapacity | 1.46 A-hr | 1.7 A-hr |
| Rechargecapability | 80% or above remaining capacity after300 cycles | 70% remaining capacity after 300cycles |
| Exposure time | 0.12 - 1.2 seconds in | 0.02 - 1.0 seconds |
| TimerAccuracy | Less than 5% or 20 ms | $\pm$ (10% + 1ms) |
| mA | 2.0 mA | 2.5 mA |
| kVp | 70 kVp | 60 kVp |
| Waveform | Constant Potential (DC) | Constant Potential (DC) |
Performance Data
Non-clinical test was performed in accordance with the following international standards,
- ANSI AAMI IEC 62304:2006 Medical Device Software Software Life Cycle Processes -
- EN ISO 14971:2019 Medical devices Applications of risk management to medical devices
- ANSI AAMI IEC 62366-1:2015+AMD1:2020 Medical devices Part 1: Application of usability engineering to medical devices
- -AAMI ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1; General Requirements for Basic Safety and Essential Performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-3 Edition 2.1 2013-04 Medical Electrical Equipment Part 1-3: General -Requirements for Basic Safety And Essential Performance - Collateral Standard: Radiation Protection in Diagnostic X-ray Equipment
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for VSI, which stands for Value Service Innovation. The logo consists of a blue symbol on the left, followed by the text "Value Service Innovation" in a smaller font size above the letters "VSI" in a larger, bold font. The symbol on the left appears to be three interconnected circles, forming a unique design element for the logo.
- IEC 60601-1-6 Edition 3.1 2013-10 Medical Electrical Equipment Part 1-6: General -Requirements for Basic Safety And Essential Performance - Collateral Standard: Usability
- IEC 60601-1-2 Edition 4.0 2014-02 Medical Electrical Equipment Part 1-2: General -Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
- IEC/EN 62133-2 Edition1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- -IEC 60601-2-65 Edition 1.0 2012-09 Medical Electrical Equipment - Part 2-65: Particular Requirements for the Basic Safety And Essential Performance of Dental Intra-Oral X-Ray Equipment
Complete device testing has been performed for the CLAROX handheld x-ray system and successful test results indicate device safety and effectiveness.
Clinical testing is not necessary for the current submission, based on the device type (an x-ray generator) and based on the technological characteristics similar to the predicate system. Successful results of bench testing should be sufficient evidence of substantial equivalence.
Conclusion
The subject device is substantially equivalent in the areas of indications for use, general functions & features, principle of operation, and technological characteristics. The new device does not introduce a fundamentally new scientific technology. Therefore, we conclude that the subject device described in this submission is substantially equivalent to the predicate device.
§ 872.1800 Extraoral source x-ray system.
(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.