K Number
K220736
Date Cleared
2022-08-01

(140 days)

Product Code
Regulation Number
880.6880
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Consolidated HC Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities using saturated steam as the sterilizing agent. They are equipped with factory programmed cycles. The factory programmed cycles, the chamber and overall dimensions, and the number of standard test items per AAMI ST8:2013 for all models of the HC Series steam sterilizers are provided below.

Device Description

The steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam (reference 21 CFR Part 880.6880). Three models have been selected for the Consolidated HC Sterilizer product line: Model 3AV-HC, Model 26AV-HC, and Model 26BV-HC. The sterilizer will be targeted for sale into Ambulatory Surgery Centers (ASCs), Clinics, and Hospitals.

The Consolidated HC Steam Sterilizers are for general purpose gravity, vacuum, or liquid steam sterilization of heat and moisture-stabile medical goods, surgical instruments, and supplies. All models utilize both gravity/downward air displacement with positive-pressure pulse conditioning and pressure/vacuum pulsing for dynamic air removal. The programmed cycles are easily accessed and are password protected. All cycle phases are sequenced and monitored by a PLC-based control system, providing both audible and visual notification of deviation from critical operating parameters. These sterilizers can either be supplied for connection to direct building steam supply of 50-80 PSI of steam pressure or come equipped with an integral, electrically heated steam generator.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Consolidated HC Steam Sterilizer, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
PerformanceMeets requirements of AAMI ST8: 2013/(R)2018Pass
General Electrical SafetyMeets requirements of UL 61010-1, 3rd EditionPass
Sterilizer Electrical SafetyMeets requirements of UL 61010-2-040Pass
Electromagnetic CompatibilityMeets requirements of IEC 61326-1:2020, FCC Part 15 subpart b, and Table 9 Per IEC 60601-1-2:2014Pass
Pressure Vessel SafetyMeets requirements of Section VIII, ASME Boiler & Pressure Vessel Code, 2019Pass
Generator SafetyMeets requirements of Section I, ASME Boiler & Pressure Vessel Code, 2019Pass
Sterility Assurance Level (SAL)10^-610^-6 (stated under Technological Characteristics)

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes non-clinical tests against various standards. It does not provide specific sample sizes for these tests in terms of the number of devices tested or the number of sterilization cycles performed for each criterion. The provenance of this data is internal testing conducted by Consolidated Machine Corp. (the manufacturer) to demonstrate compliance with recognized industry standards. The data is reported as part of a premarket notification to the FDA. It is prospective data generated specifically for regulatory submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the document. For non-clinical tests based on engineering and performance standards (like AAMI, UL, IEC, ASME), the "ground truth" is typically established by fulfilling the defined criteria and measurement methodologies specified within those standards. This would involve calibrated equipment and trained technicians/engineers, rather than a panel of clinical experts.

4. Adjudication Method for the Test Set

This information is not applicable in the context of the provided non-clinical testing. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation or subjective assessment (e.g., image reading), not for objective performance tests against engineering standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No clinical testing was used in support of this submission." This device is a medical sterilizer, and its effectiveness is determined by its ability to achieve a Sterility Assurance Level (SAL) through validated physical and biological parameters, not by human interpretation of results.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable as the device is a steam sterilizer, not an AI algorithm. Its performance is inherent to its mechanical and control systems.

7. The Type of Ground Truth Used

The ground truth for the non-clinical tests is based on established industry standards and regulatory requirements. For example:

  • Performance: AAMI ST8:2013 (R)2018 (a standard for steam sterilization in healthcare facilities). This standard outlines the microbiological and physical requirements for effective sterilization, often involving the use of biological indicators that demonstrate a defined kill rate.
  • Safety (Electrical, EMC, Pressure Vessel, Generator): UL, IEC, and ASME standards, which define objective measurement criteria for electrical safety, electromagnetic compatibility, and structural integrity under pressure.
  • Sterility Assurance Level (SAL): Defined as 10^-6, meaning there is a 1 in 1,000,000 probability of a viable microorganism being present on a sterilized item. This is a universally accepted microbiological standard for sterilization.

8. The Sample Size for the Training Set

This question is not applicable. This device is a physical sterilizer, not an AI/machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

August 1, 2022

Consolidated Machine Corp. Arthur Trapotsis CEO 3 Enterprise Rd, Suite C Billerica, Massachusetts 01821

Re: K220736

Trade/Device Name: Consolidated HC Steam Sterilizer Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: June 28, 2022 Received: June 30, 2022

Dear Arthur Trapotsis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220736

Device Name Consolidated HC Steam Sterilizer

Indications for Use (Describe)

The Consolidated HC Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities using saturated steam as the sterilizing agent. They are equipped with factory programmed cycles. The factory programmed cycles, the chamber and overall dimensions, and the number of standard test items per AAMI ST8:2013 for all models of the HC Series steam sterilizers are provided below.

Table 1: Validated Sterilization Cycles

Cycle TypeSterilizationTemperatureSterilizationTimeDry TimeLoad Type
Gravity250°F (121°C)30 minutes30 minutesFabric Packs
Gravity250°F (121°C)30 minutes30 minutesDouble Wrapped InstrumentTrays Maximum weight pertray: 25lbs (11.3 kg)
Gravity - IUSS270°F (132°C)3 minutes1 minutesSingle Unwrapped TrayMaximum weight per tray: 25lbs(11.3 kg)
Gravity270°F (132°C)15 minutes30 minutesDouble Wrapped InstrumentTrays Maximum weight pertray: 25lbs (11.3 kg)
Pre-Vacuum -IUSS270°F (132°C)4 minutes1 minutesSingle Unwrapped TrayMaximum weight per tray: 25lbs(11.3 kg)
Pre-Vacuum270°F (132°C)4 minutes20 minutesFabric Packs
Pre-Vacuum270°F (132°C)4 minutes30 minutesDouble Wrapped InstrumentTrays Maximum weight pertray: 25lbs (11.3 kg)
Pre-Vacuum275°F (135°C)3 minutes30 minutesDouble Wrapped InstrumentTrays Maximum weight pertray: 25lbs (11.3 kg)
Pre-Vacuum270°F (132°C)4 minutes5 minutesSingle Fabric Pack and SingleUnwrapped Tray Maximumweight per tray: 25lbs (11.3 kg)

{3}------------------------------------------------

Table 1 (cont.): Validated Sterilization Cycles

Liquids250°F (121°C)30 minutesN/A1L Liquid Volumes
Bowie Dick Test273°F (134°C)3.5 minutes0 minutesBowie Dick Test Pack
Vacuum LeakTest270°F (132°C)N/AN/AEmpty Chamber
Warm Up270°F (132°C)3 minutes1 minutesEmpty Chamber

Table 2: Load Size

ModelSterilizerChamber SizeDoubleWrappedInstrumentTrays1FabricPacks2Number ofLiquidContainers3
26AV-HC26" X 26" X39"9, maximumweight 25lbeach1270, 1 LiterVolume PerContainer
26BV-HC26" X 26" X49"12, maximumweight 25lbeach1690, 1 LiterVolume PerContainer
3AV-HC20" X 20" X38"3, maximumweight 25lbeach928, 1 LiterVolume PerContainer

1 For instrument trays, the maximum weight per tray is 25lbs (11.3 kg).

2 The weight of each fabric pack should be limited to aproximately 3.2 lbs (1.4 kg) with a density of 11.3 lb/ft3 (181 kg/m3) or less. 3 For liquid containers, the maximum volume of liquid per container is 1,000 ml. Please note, liquid loads processed in the sterlizer are inappropriate for direct patient contact.

{4}------------------------------------------------

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

| X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary for Consolidated HC Steam Sterilizer

Consolidated Sterilizer Systems 3 Enterprise Road, Suite C Billerica, MA 01821

Contact: Arthur Trapotsis Phone: 617-782-6072 ext2200 Email: Arthur@consteril.com

Premarket Notification Number: K220736 Date of Preparation: July 27, 2022

{6}------------------------------------------------

Device Name

Trade Name:Consolidated HC Steam Sterilizer
Device Class:Class II
Common Name:Steam Sterilizer/Autoclave
Classification Name:Sterilizer, Steam
Classification Number:21 CFR 880.6880
Product Code:FLE

Predicate Device

K183410 AMSCO 600 Steam Sterilizer
Device Class:Class II
Common Name:Steam Sterilizer/Autoclave
Classification Name:Sterilizer, Steam
Classification Number:21 CFR 880.6880
Product Code:FLE

Description of Device

The steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam (reference 21 CFR Part 880.6880). Three models have been selected for the Consolidated HC Sterilizer product line: Model 3AV-HC, Model 26AV-HC, and Model 26BV-HC. The sterilizer will be targeted for sale into Ambulatory Surgery Centers (ASCs), Clinics, and Hospitals.

The Consolidated HC Steam Sterilizers are for general purpose gravity, vacuum, or liquid steam sterilization of heat and moisture-stabile medical goods, surgical instruments, and supplies. All models utilize both gravity/downward air displacement with positive-pressure pulse conditioning and pressure/vacuum pulsing for dynamic air removal. The programmed cycles are easily accessed and are password protected. All cycle phases are sequenced and monitored by a PLC-based control system, providing both audible and visual notification of deviation from critical operating parameters. These sterilizers can either be supplied for connection to direct building steam supply of 50-80 PSI of steam pressure or come equipped with an integral, electrically heated steam generator.

{7}------------------------------------------------

Indications for Use

The Consolidated HC Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities using saturated steam as the sterilizing agent. They are equipped with factory programmed cycles. The factory programmed cycles, the chamber and overall dimensions, and the number of standard test items per AAMI ST8:2013 R2018 for all models of the HC Series steam sterilizers are provided below.

Cycle TypeSterilizationTemperatureSterilizationTimeDry TimeLoad Type
Gravity250°F (121°C)30 minutes30 minutesFabric Packs
Gravity250°F (121°C)30 minutes30 minutesDouble Wrapped Instrument TraysMaximum weight per tray: 25lbs(11.3 kg)
Gravity - IUSS270°F (132°C)3 minutes1 minutesSingle Unwrapped TrayMaximum weight per tray: 25lbs(11.3 kg)
Gravity270°F (132°C)15 minutes30 minutesDouble Wrapped Instrument TraysMaximum weight per tray: 25lbs(11.3 kg)
Pre-Vacuum -IUSS270°F (132°C)4 minutes1 minutesSingle Unwrapped Tray Maximumweight per tray: 25lbs (11.3 kg)
Pre-Vacuum270°F (132°C)4 minutes20 minutesFabric Packs
Pre-Vacuum270°F (132°C)4 minutes30 minutesDouble Wrapped Instrument TraysMaximum weight per tray: 25lbs(11.3 kg)
Pre-Vacuum275°F (135°C)3 minutes30 minutesDouble Wrapped Instrument TraysMaximum weight per tray: 25lbs(11.3 kg)
Pre-Vacuum270°F (132°C)4 minutes5 minutesSingle Fabric Pack and SingleUnwrapped Tray Maximum weightper tray: 25lbs (11.3 kg)
Liquids250°F (121°C)30 minutesN/A1L Liquid Volumes
Bowie Dick Test273°F (134°C)3.5 minutes0 minutesBowie Dick Test Pack
Vacuum Leak Test270°F (132°C)N/AN/AEmpty Chamber
Warm Up270°F (132°C)3 minutes1 minutesEmpty Chamber

Table 1: Validated Sterilization Cycles

{8}------------------------------------------------

Table 2: Load Size

ModelSterilizerChamber SizeDouble WrappedInstrument Trays1FabricPacks2Number of LiquidContainers3
26AV-HC26" X 26" X 39"9, maximumweight 25lb each1270, 1 Liter Volume PerContainer
26BV-HC26" X 26" X 49"12, maximumweight 25lb each1690, 1 Liter Volume PerContainer
3AV-HC20" X 20" X 38"3, maximumweight 25lb each928, 1 LiterVolume PerContainer

1 For instrument trays, the maximum weight per tray is 25lbs (11.3 kg).

2 The weight of each fabric pack should be limited to approximately 3.2 lbs (1.4 kg) with a density of 11.3 lb/ft3 (181 kg/m3) or less.

3 For liquid containers, the maximum volume of liquid per container is 1,000 ml. Please note, liquid loads processed in the sterilizer are inappropriate for direct patient contact.

Technological Characteristics

Table 3: Comparison of the Proposed Device and the Predicate Device

FeatureConsolidated HC Steam SterilizerAMSCO 600 Steam Sterilizer(K183410)Comparison
Intended UseThe sterilization of heat andmoisture-stable materials used inhealthcare facilities.The sterilization of heat andmoisture-stable materials used inhealthcare facilities.Same
Critical ProcessParametersTime, Chamber Temperature,PressureTime, Chamber Temperature,PressureSame
ControlProgrammable Logic ControllerEmbedded ControllerDifferent
SAL$10^{-6}$$10^{-6}$Same
SterilantSaturated SteamSaturated SteamSame
UtilitiesSteam, Water, Electricity, AirSteam, Water, Electricity, AirSame
ChamberMaterial316L Stainless Steel316L Stainless SteelSame
NominalChamber Size26" X 26" X 39"26" X 26" X 49"20" X 20" X 38"26" X 26" X 39"26" X 26" X 49"Similar
Door316L Stainless Steel26" X 26" Power Vertical Sliding20" X 20" Power Vertical Sliding304L Stainless Steel26" X 26" Power Vertical SlidingSimilar
ChamberPressure Rating45 psig, 300 °F45 psig, 300 °FSame
Door SealAir Activated Door SealSteam Activated Door SealSimilar
ExternalProcessMonitorsElectronic ControlPrinterElectronic ControlPrinterSame

{9}------------------------------------------------

Internal ProcessMonitorsRTD temperature sensor in chamberdrainRTD temperature sensor in jacketdrainRTD temperature sensor in wastedrainPressure transducer in chamberRTD temperature sensor in chamberdrainRTD temperature sensor in jacketdrainRTD temperature sensor in heatexchangerPressure transducer in chamberSame
PerformanceMeets ANSI/AAMI ST8:2013 (R)2018Meets ANSI/AAMI ST8:2013Similar
AccessoriesShelves, Loading EquipmentBI, CI, Pouches, Trays, Wraps, Tape,Containers, Shelves, LoadingEquipmentDifferent
Test CyclesWarm Up, Leak Test, Bowie Dick TestWarm Up, Leak Test, DART (BowieDick) TestSame
CyclesSee Table 1270°F, Prevac, 4 min, Full fabric pack270°F, Prevac, 4 min, Full tray270°F, Prevac, 4 min, one fabric pack270°F, Prevac, 4 min, IUSS275°F, Prevac, 4 min, Full fabric pack250°F, Gravity, 30 min, Full traySimilar
Full Loads26AV: 9, 25 lb double wrapped traysor 12 fabric packs26BV: 12, 25 lb double wrapped traysor 16 fabric packs3AV: 3, 25 lb double wrapped trays or6 fabric packs39″: 9, 25 lb double wrapped trays or12 fabric packs49″: 12, 25 lb double wrapped traysor 16 fabric packsSimilar

The proposed device has the same intended use, instructions for use, and meets the same performance specification as the predicate device. The proposed device has similar technological characteristics as the predicate device. The proposed device offers more validated sterilization cycles than the predicate device. The proposed device offers an additional chamber size.

{10}------------------------------------------------

Non-Clinical Tests

The Consolidated HC Steam Sterilizer non-clinical testing was performed according to the following standards, with passing results in all cases:

TestAcceptance CriteriaResult
PerformanceMeets requirements of AAMI ST8: 2013/(R)2018Pass
General ElectricalSafetyMeets requirements of UL 61010-1, 3rd EditionPass
Sterilizer ElectricalSafetyMeets requirements of UL 61010-2-040Pass
ElectromagneticCompatibilityMeets requirements of IEC 61326-1:2020, FCC Part 15subpart b, and Table 9 Per IEC 60601-1-2:2014Pass
Pressure Vessel SafetyMeets requirements of Section VIII, ASME Boiler &Pressure Vessel Code, 2019Pass
Generator SafetyMeets requirements of Section I, ASME Boiler & PressureVessel Code, 2019Pass

Table 4: Non-Clinical Test Results

The testing demonstrated that the subject devices, 3AV-HC, and 26BV-HC, met the acceptance criteria described in these standards.

Clinical Tests

No clinical testing was used in support of this submission.

Conclusion

The conclusions drawn from the nonclinical performance data demonstrate that the device, the Consolidated HC Steam Sterilizer, is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K183410.

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).