K Number
K220506
Manufacturer
Date Cleared
2022-03-07

(13 days)

Product Code
Regulation Number
878.4430
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SkinPen® Precision system is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older.

Device Description

The SkinPen® Precision System consists of a microneedling pen handpiece (SkinPen® Precision) and a sterile needle cartridge (SkinPen® Precision Cartridge). The accessories are a charging base and a BioSheath. A SkinPen® Precision System treatment kit is provided separately and contains the following: SkinPen® Precision Cartridge: sterile, disposable needle cartridge. Not to be resterilized or reused. (K202243, Class 2, Regulation 878.4430, Product Code: QAI.); Lift HG: hydrogel wound dressing (without drugs and/or biologics), to protect against abrasion and friction during the microneedling procedure. May be applied to prevent skin from drying out post procedure. (Class I, 510(k) Exempt, Regulation 878.4022, Product code: NAE.); SkinPen® BioSheath: nonsterile, disposable cover for the microneedling pen handpiece to avoid contamination of the SkinPen® Precision.; Sani-Cloth AF3: Sanitizing cloth available for purchase along with device to sanitize between uses.

AI/ML Overview

The provided FDA 510(k) clearance letter (K220506) for the SkinPen® Precision System does not contain information about acceptance criteria or a study proving the device meets acceptance criteria for its clinical indications (improving the appearance of wrinkles of the neck and facial acne scars).

The letter states, "No Clinical testing was conducted as part of this submission." This submission is a "Special 510(k)" for a modification to an already cleared device (K202243 SkinPen Precision System). The "acceptance criteria" and "study that proves the device meets the acceptance criteria" would have been part of the original 510(k) for the predicate device, K202243, or a prior submission that established the clinical effectiveness for the indications for use.

Instead, this submission focuses on a non-clinical performance testing related to a change in the disinfecting cloth used with the device.

Here's a breakdown of the information that is present, in relation to your request, and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

  • No clinical acceptance criteria or performance data are provided in this document for the indications of use (wrinkles of the neck, facial acne scars).
  • Non-clinical (cleaning validation) acceptance criteria and performance:
    • Acceptance Criteria for Cleaning Validation: The goal was to confirm that the Sani-Cloth AF3 is sufficient for bacterial removal from the device even under worst-case contamination scenarios. The implicit acceptance criterion is that the bacteria are effectively removed.
    • Reported Device Performance (Cleaning Validation): "it was confirmed that the bacteria were removed from the device when cleaned with Sani-Cloth AF3... The new Sani-Cloth AF3 shows the same or improved results than the previously used Sani-Cloth HB."

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Test Set: Not applicable, as no clinical testing was performed for this submission.
  • Non-clinical (Cleaning Validation) Test Set: The document describes "application of bacteria on the SkinPen device surface." The exact sample size (number of devices, number of wipes, etc.) is not explicitly stated, but it refers to "worst-case conditions."
  • Data Provenance: Not explicitly stated, but likely conducted in a laboratory setting as part of the manufacturer's testing for regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Clinical Test Set: Not applicable.
  • Non-clinical (Cleaning Validation) Test Set: The "ground truth" here is the presence or absence of bacteria (before and after cleaning). This would be established by standard microbiological laboratory techniques, not by human experts in the context of clinical interpretation. The qualifications of the microbiologists or lab technicians are not provided.

4. Adjudication Method for the Test Set

  • Clinical Test Set: Not applicable.
  • Non-clinical (Cleaning Validation) Test Set: Not applicable in the sense of clinical adjudication. The result (bacteria removed or not) would be determined by laboratory assays.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Readers Improve with AI vs Without AI Assistance

  • This device is a microneedling system, not an AI diagnostic or assistive device. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable and was not performed.

6. If a Standalone (Algorithm Only Without Human-in-the Loop Performance) Was Done

  • This device is a microneedling system, not an algorithm. Therefore, standalone algorithm performance is not applicable.

7. The Type of Ground Truth Used

  • Clinical Ground Truth: For the clinical indications (wrinkles and acne scars), the ground truth for the original predicate device clearance would likely have been based on clinician assessments (e.g., blinded expert photographic evaluations, severity scales) and/or patient-reported outcomes. This document does not provide that information.
  • Non-clinical (Cleaning Validation) Ground Truth: The ground truth for the cleaning validation was the microbiological assessment of bacterial presence/absence on the device surface.

8. The Sample Size for the Training Set

  • Clinical Training Set: Not applicable, as no clinical studies were performed for this submission, and it's not an AI/model development submission.
  • Non-clinical (Cleaning Validation) Training Set: Not applicable. This was a validation study, not a model training process.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable for the reasons stated above.

In summary, this document primarily outlines the regulatory clearance (510(k)) for a minor modification (change in disinfecting wipe) to an already cleared microneedling device. It does not provide the details of clinical acceptance criteria or studies proving clinical effectiveness, as those would have been part of the original clearance for the predicate device (K202243). The only "study" mentioned is a non-clinical cleaning validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 7, 2022

Crown Aesthetics Marie Fogartie Director of Regulatory Affairs 5005 Lyndon B. Johnson Freeway, Suite 370 Dallas. Texas 75244

Re: K220506

Trade/Device Name: SkinPen®Precision System Regulation Number: 21 CFR 878.4430 Regulation Name: Microneedling Device for Aesthetic Use Regulatory Class: Class II Product Code: QAI Dated: February 18, 2022 Received: February 22, 2022

Dear Marie Fogartie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements,

{1}------------------------------------------------

including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and. if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations. please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220506

Device Name SkinPen® Precision System

Indications for Use (Describe)

The SkinPen® Precision system is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over The Counter Use (21 CFR 801 Subpart C)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Date Prepared: March 2, 2022

    1. Submitter and Owner:
      Crown Aesthetics 5005 Lyndon B. JohnsonFreeway Suite 370 Dallas, Texas 75244, USA
    1. Official Correspondent:
      Marie Fogartie, Director of Regulatory Affairs Phone: 423-630-2296 Email: mfogartie@crownlaboratories.com
    1. Submission Type: Special 510(k)
  • Proprietary Name: SkinPen® Precision System 4.

    1. Device Classification Information:

Classification Name: Microneedling Device for Aesthetic Use Requlation Number: 878.4430 Product Code: QAI Device Class: 2

    1. Predicate Device: K202243 SkinPen Precision System
    1. Device Description:

The SkinPen® Precision System consists of a microneedling pen handpiece (SkinPen® Precision) and a sterile needle cartridge (SkinPen® Precision Cartridge).

The accessories are a charging base and a BioSheath. A SkinPen® Precision System treatment kit is provided separately and contains the following:

  • SkinPen® Precision Cartridge: sterile, disposable needle cartridge. Not ● to be resterilized or reused. (K202243, Class 2, Regulation 878.4430, Product Code: QAI.)
  • Lift HG: hydrogel wound dressing (without drugs and/or biologics), to ● protect against abrasion and friction during the microneedling procedure. May be applied to prevent skin from drying out post procedure. (Class I, 510(k) Exempt, Regulation 878.4022, Product code: NAE.)
  • SkinPen® BioSheath: nonsterile, disposable cover for the ● microneedling pen handpiece to avoid contamination of the SkinPen® Precision.
  • . Sani-Cloth AF3: Sanitizing cloth available for purchase along with device to sanitize between uses.

{4}------------------------------------------------

    1. Indications/Intended use:
      The SkinPen® Precision system is a microneedling device and accessories intendedto be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older.
    1. Indications for Use Comparison:
      The subject and the predicate device (K202243) have identical indications for use statements.
CharacteristicSubject DevicePredicate DeviceK202243Comparison
Trade NameSkinPen® Precision SystemSkinPen® Precision SystemSame
ClassificationNameMicroneedling deviceforaesthetic useMicroneedling deviceforaesthetic useSame
Classification22Same
Product CodeQAIQAISame
Requlation No.878.4430878.4430Same
PrescriptionUseYesYesSame
TargetPopulationAdults age 22 and overAdults age 22 and overSame
Target UserHealthcare professionalstrained in use of the deviceHealthcare professionalstrained in use of the deviceSame
EnvironmentClinicalClinicalSame
Treatment AreaFaceFaceSame
Shape of needlecartridgeRoundRoundSame
Needles14 total solid medical gradestainless steel14 total solid medical gradestainless steelSame
NeedleProtrusionsettings0–2.5 mm0–2.5 mmSame
Max. NeedleLength used intheclinical study2.5mm2.5mmSame
Frequency6300 RPM to 7700 RPM(105-128.3 Hz)6300 RPM to 7700 RPM(105-128.3 Hz)Same
    1. Technological characteristics comparison:
      The subject device is identical to the predicate device, except the use of a different disinfecting cloth. The cleaning validation described below demonstrated the device modification raises no safety or effectiveness concerns.

11. Non-clinical performance testing

Cleaning validation testing was conducted for the disinfecting cloth that is the subject of this submission. The cleaning test applied bacteria to the device after being wiped clean with a Sani-Cloth AF3. After applying the bacteria, the devicewas cleaned again with the Sani-Cloth AF3. After application of bacteria on the SkinPen device surface under worst-case conditions of no bio-sheath,

{5}------------------------------------------------

and a minimal cleaning time of 1 minute, it was confirmed that the bacteria were removedfrom the device when cleaned with Sani-Cloth AF3. In conclusion, even if the device is contaminated by bacteria, Sani-Cloth AF3 is considered sufficient for bacterial removal from the device, as the number of bacteria applied to the device was sufficiently high enough to simulate a worst-case scenario of possible microbial contamination. The new Sani-Cloth AF3 shows the same or improved results than the previously used Sani-Cloth HB.

    1. Clinical Testing Summary
      No Clinical testing was conducted as part of this submission.
    1. Statement of Substantial Equivalence:
      Crown Aesthetics has demonstrated that the SkinPen® Precision System has the same device classification, same intended use and technological characteristics as the predicate device. The only change in this submission is the cloth used for disinfection is shown effective in a validation report provided with this application.

§ 878.4430 Microneedling device for aesthetic use.

(a)
Identification. A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.