K Number
K220469
Date Cleared
2022-06-16

(118 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DISPOSABLE VINYL EXAMINATION GLOVE is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Device Description

The subject device is powder free vinyl examination gloves. The subject device's color is clear. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

AI/ML Overview

The provided text describes the acceptance criteria and the study conducted for a Disposable Vinyl Examination Glove (K220469).

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D5250Physical Dimensions TestLength (mm): S/M/L/XL: ≥230; Width (mm): S: 85±5; M: 95±5; L: 105±5; XL: 115±5; Thickness (mm): Finger: ≥0.08, Palm: ≥0.08Length (mm): > 230/Pass; Width (mm): S: 85-89 /Pass, M: 95-97/ Pass, L: 104-107/ Pass, XL: 116-118/ Pass; Thickness (mm): Finger: 0.11-0.17/Pass, Palm: 0.10-0.13/Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125/Pass
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.06-0.09mg/Pass
ASTM D412Physical properties (Before Aging)Tensile Strength ≥11MPa; Ultimate Elongation ≥300%Tensile Strength: 11-19MPa/Pass; Ultimate Elongation: 320-369%/Pass
ASTM D412Physical properties (After Aging)Tensile Strength ≥11MPa; Ultimate Elongation ≥300%Tensile Strength: 11-18MPa/Pass; Ultimate Elongation: 310-352%/Pass
ISO 10993-5CytotoxicityIn Vitro Cytotoxicity: Not cytotoxicUnder conditions of the study, device extract is not cytotoxic./Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant/ Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer./Pass

2. Sample size used for the test set and the data provenance

The document does not explicitly state the full sample size for all tests. However, for the Watertightness Test (ASTM D5151), a sample size of 125 was used (0/125/Pass).

The data provenance is not explicitly mentioned as "country of origin" or "retrospective/prospective" studies. However, the tests are non-clinical, meaning they are laboratory-based and not derived from patient data. The submitter is CHIFENG HUAWEI MEDICAL SCIENCE & TECHNOLOGY CO.,LTD. (China), suggesting the non-clinical testing likely took place in China or a contracted lab.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the described tests are non-clinical (material and performance testing of a physical product) and do not involve expert interpretation or ground truth establishment in a clinical sense. The "ground truth" for these tests are the established standards (e.g., ASTM, ISO) and their defined acceptance criteria.

4. Adjudication method for the test set

This information is not applicable for these non-clinical, objective tests. Adjudication methods are typically used in studies involving subjective interpretation, often in clinical imaging or diagnostic contexts, where multiple experts might disagree on an assessment.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not performed. This type of study is relevant for AI-powered diagnostic tools or imaging analysis where human readers are involved. The device in question is a disposable vinyl examination glove, which does not involve human reader interpretation or AI assistance in its core function.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

A standalone performance study in the context of an algorithm or AI was not done. The tests performed are for the physical and biological properties of the glove itself, not for an algorithm.

7. The type of ground truth used

The "ground truth" for these tests is based on established standard specifications and analytical measurements. For example:

  • Physical dimensions are measured against defined ranges.
  • Tensile strength and ultimate elongation are measured against minimum values.
  • Powder content is measured against a maximum value.
  • Biocompatibility (cytotoxicity, irritation, sensitization) is assessed by laboratory tests based on ISO standards to determine if the device elicits a specific biological response.

8. The sample size for the training set

This information is not applicable. The tests described are for a physical product, a disposable glove. There is no AI algorithm being "trained" in this context, so there is no training set in the conventional machine learning sense.

9. How the ground truth for the training set was established

This information is not applicable as there is no training set mentioned for this medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 16, 2022

Chifeng Huawei Medical Science&Technology Co.,Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K220469

Trade/Device Name: Disposable Vinyl Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: June 8, 2022 Received: June 16, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220469

Device Name DISPOSABLE VINYL EXAMINATION GLOVE

Indications for Use (Describe)

The DISPOSABLE VINYL EXAMINATION GLOVE is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K220469

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

CHIFENG HUAWEI MEDICAL SCIENCE & TECHNOLOGY CO.,LTD. Name: Address: No. 2-4, Second-stage Standardized Plant, Songshan (Anqing) Industrial Park, Chifeng, China. Phone Number: +86-18248056569 Contact: Li Xiaohong Date of Preparation: Jan.7, 2022

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: DISPOSABLE VINYL EXAMINATION GLOVE Common name: Vinyl Patient Examination Glove Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

3.0 Classification

Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: ZHICHENG TRADING CO., LTD. Device: Vinyl Examination Glove (Clear, Non-Colored) 510(k) number: K180861

{4}------------------------------------------------

5.0 Indication for Use

The DISPOSABLE VINYL EXAMINATION GLOVE is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free vinyl examination gloves. The subject device's color is clear. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

ItemSubject Device(K220469)Predicated Device(K180861)Remark
Product CodeLYZLYZSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended Use /Indications for UseThe DISPOSABLEVINYL EXAMINATIONGLOVE is a disposabledevice intended formedical purposes that isworn on the examiner'shands to preventcontamination betweenpatient and examiner.The Vinyl ExaminationGlove (Clear,Non-Colored) is adisposable deviceintended for medicalpurposes that is worn onthe examiner's hands toprevent contaminationbetween patient andexaminer.Same
MaterialVinylVinylSame
Powdered orPowered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
ColorantClearClearSame
LabelingInformationSingle-use indication,powder free, devicecolor, device name,glove size and quantity,Non-SterileSingle-use indication,powder free, devicecolor, device name,glove size and quantity,Non-SterileSame
Dimensions(mm)Length:S/M/L/XL: ≥230;Width:S: 85±5;Length:S≥230;M≥235;L≥245;SimilarAnalysis 1

Table1-General Comparison

{5}------------------------------------------------

M: 95±5;L: 105±5;XL: 115±5XL≥245;Width:S: 85±5;M: 95±5;L: 105±5;XL: 115±5
Thickness(mm)Finger: ≥0.08;Palm: ≥0.08Finger: ≥0.05;Palm: ≥0.08Similar Analysis 1
PhysicalPropertiesBeforeAgingTensile Strength11MPa, minUltimate Elongation300% minTensile Strength15MPa, minUltimate Elongation380% minSimilar Analysis 2
AfterAgingTensile Strength11MPa, minUltimate Elongation300%minTensile Strength15MPa, minUltimate Elongation380%minSimilar Analysis 2
Freedom fromHolesBe free from holes whentested in accordance withASTMD5151AQL=2.5Be free from holes whentested in accordance withASTMD5151AQL=1.5Similar Analysis 3
Powder ContentMeet the requirements ofASTM D6124Meet the requirements ofASTM D6124Same
BiocompatibilityISO 10993-10;Under the conditions ofthe study, not an irritantor a sensitizerComply withISO10993-10Same
Under conditions of thestudy, did not showpotential toxicity to L-929cells. Complies with ISO10993-5Under conditions of thestudy, did not showpotential toxicity to L-929cells. Complies with ISO10993-5Same

Analysis 1: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250, so the differences do not raise any new safety or performance questions.

Analysis 2: The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250, so the differences do not raise any new safety or performance questions.

Analysis 3: Freedom from holes of subject device is similar with that of the predicate, but they all meet the requirements of ASTMD5151, so the differences do not raise any new safety or performance questions.

{6}------------------------------------------------

8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

TestMethodPurposeAcceptance CriteriaResults
ASTMD5250PhysicalDimensionsTestLength(mm):S/M/L/XL: ≥230;Width(mm):S: 85±5;M: 95±5;L: 105±5;XL: 115±5Thickness (mm):Finger: ≥0.08Palm: ≥0.08Length(mm):> 230/Pass;Width(mm):S: 85-89 /PassM: 95-97/ PassL: 104-107/ PassXL:116-118/ PassThickness (mm):Finger: 0.11-0.17/PassPalm: 0.10-0.13/Pass
ASTMD5151WatertightnessTestforDetectionofHolesMeet the requirements of ASTM D5151AQL 2.50/125/Pass
ASTMD6124PowderContentMeet the requirements of ASTM D6124< 2.0mg0.06-0.09mg/Pass;
ASTMD412PhysicalpropertiesBefore AgingTensile Strength≥11MPa11-19MPa/Pass;
Before AgingUltimate≥300%320-369%/Pass;

Table 2 - Summary of non-clinical performance testing

{7}------------------------------------------------

Elongation
AfterAgingTensileStrength≥11MPa11-18MPa/Pass;
UltimateElongation≥300%310-352%/Pass;
ISO10993-5CytotoxicityIn Vitro CytotoxicityUnder conditions of thestudy, device extract isnot cytotoxic./Pass
ISO10993-10IrritationNon-irritatingUnder the conditions ofthe study, not anirritant/ Pass
ISO10993-10SensitizationNon-sensitizingUnder conditions of thestudy, not a sensitizer./Pass

9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device DISPOSABLE VINYL EXAMINATION GLOVE is as safe, as effective, and performs as well as or better than the legally marketed predicated device K180861.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.