(188 days)
The product is intended to be used in a hospital or at home to obtain capillary blood samples from the fingertip for tests using small amounts of blood. The lancet is intended to be assembled with a lancing device, such that once the lancing device is launched, the needle of the lancet can prick the skin.
The Disposable Blood Lancets consists of a needle, a body and a cap. The models of the Disposable Blood Lancets are 18G; 21G; 23G; 26G; 28G; 30G; 32G; 33G. Disposable Blood Lancets are used to collect blood samples for hospital clinic test. It is sterilized by Irradiation and is a single-use product.
The provided document (K220387 510(k) summary for Disposable Blood Lancets) details acceptance criteria and the results of a study to demonstrate substantial equivalence to a predicate device, rather than a study proving significant improvement over human readers or standalone performance. The document focuses on performance testing against defined specifications for the device itself.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
No. | Testing Item | Acceptance Criteria (Specification) | Reported Device Performance (Result) |
---|---|---|---|
01 | Appearance | Lancet should have the same color, no bubble, no flash, no slip. | Pass |
02 | Launch Length | The length of the needles in the Disposable Blood Lancets is different in different gauge. The general depth is 3.2mm. | Pass |
03 | Sharpness | Penetration force ≤1.00N. | Pass |
04 | Exterior | The connection between needle and needle body should be firm. | Pass |
05 | Cap twist | Cap twist should be smooth. | Pass |
06 | Initial bioburden | Initial bioburden of the device shall be less than 100CFU/g. | Pass |
07 | Sterile | The sterile blood lancet shall be sterile. | Pass |
In addition to these direct performance tests, biocompatibility, sterilization validation, and shelf life testing were also performed, with implied "Pass" results for substantial equivalence:
Biocompatibility Testing (Evaluated in accordance with ISO 10993-1:2018):
- Cytotoxic test (ISO 10993-5:2009)
- Skin sensitization test (ISO 10993-10:2010)
- Intracutaneous test (ISO 10993-10:2010)
- Acute systemic toxicity test (ISO 10993-11:2017)
- Hemolysis test (ISO 10993-4:2017)
- Pyrogen Test (USP )
Sterilization and Shelf Life Testing:
- Irradiation sterilization validation per ISO 11173-1 and ISO 11173-3.
- Pyrogen testing per USP .
- Simulated shipping per ASTM D4169.
- Stability study (including aging test) for a 5-year shelf life.
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes used for each individual test (e.g., how many lancets were tested for sharpness or appearance). It only reports "Pass" for each criterion.
The data provenance is from non-clinical testing performed by the manufacturer, Suzhou Kyuan Medical Apparatus Co., Ltd., in China. The study is a retrospective evaluation of the device's adherence to established performance and safety standards, typical for a 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This type of information is not applicable to the provided document. The "ground truth" for this device's performance is based on objective, quantifiable engineering and biological test standards (e.g., penetration force of 1.00N, bioburden
§ 878.4850 Blood lancets.
(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.