K Number
K220329
Device Name
G-CAM
Date Cleared
2022-08-16

(193 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

G-CAM discs are intended to be used for the manufacture of full and partial removable dentures as well as permanent and temporary restorations such as anterior or posterior crowns and bridges, inlays, veneers, copings and substructures.

Device Description

G-CAM is a thermoplastic acrylic disc made by a principal base of polymethyl methacrylate (PMMA) resin doped with graphene (allotropic form of carbon), suitable for the creation of dental prostheses using CAD/CAM technology.

G-CAM discs are intended to be used for the manufacture of full and partial removable dentures, implant overdentures as well as permanent and temporary restorations such as anterior or posterior crowns and bridges, inlays, onlays, veneers, copings and substructures.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a dental device called G-CAM. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel clinical effectiveness through extensive human studies. Therefore, the information regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" will largely relate to non-clinical performance testing against recognized standards and comparison with predicate devices, as clinical human-in-the-loop studies were explicitly not performed.

Here's a breakdown of the requested information based on the provided document:


Acceptance Criteria and Device Performance for G-CAM

The acceptance criteria for G-CAM are primarily based on demonstrating substantial equivalence to predicate devices through conformity with recognized international standards for dental materials.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a "table of acceptance criteria" in a traditional sense with numerical targets before the results. Instead, it provides a comparison table (Table 1) that lists the G-CAM's characteristics and its measured performance against the relevant ISO standards, alongside the characteristics of the predicate devices. The "acceptance" is implied by meeting or exceeding the minimum requirements of the specified ISO standards, and by showing substantial equivalence to the predicate devices.

Table.1 Comparison of G-CAM with Predicate and Reference Devices (with performance data)

FeatureAcceptance Criteria (from ISO Standard)Reported G-CAM PerformancePredicate (JUVORA Dental Disc) Performance (where stated)Reference (CAD/CAMouflage Milling Block) Performance (where stated)
Flexural ModulusISO20795 > 2000 MPa3200 MPa +/-7% (ISO 20795)3995 MPa (ISO 20795)Not stated
Flexural StrengthISO20795 > 65 MPa; ISO10477 ≥ 50 MPa140 MPa +/-7% (ISO 20795); 148 MPa (ISO 10477)165 MPa (ISO 20795); 192 MPa (ISO 10477)Not stated
Surface HardnessNot specified directly, but comparable to predicates88 ShoreD (ISO 48-4); 19.5 KHN (ASTM E384)Not statedNot stated
Water AbsorptionISO20795

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.