K Number
K220280

Validate with FDA (Live)

Date Cleared
2022-08-30

(210 days)

Product Code
Regulation Number
870.1380
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arrow Stiffening Stylet is intended to be placed inside a catheter to stiffen the catheter for placement.

Device Description

The Arrow Stiffening Stylet is a sterile, single use, percutaneous peripheral vasculature catheter stylet intended for transient use (up to 24 hours) to be placed within a catheter to render it stiff to aid in catheter placement. The proposed device will be delivered as a sterile accessory within a convenience kit for the Arrow Peripherally Inserted Central Catheter (PICC)s. The Stylet is a PTFE coated solid nitinol core wire with straight tip. The stylet is available in 0.015-0.017-inch diameters and in 29.5-inch length with distance markings along the stylet body.

The Arrow Stiffening Stylet is provided with an existing, commercially available flushable hub with side arm and slide clamp cleared under K790408. The flushable hub maintains the stylet position within the catheter during catheter placement, is intended to minimize blood loss during catheter introduction, and allows the clinician to flush through the catheter to aid in catheter placement, if needed. There is no change to the flushable hub cleared under K790408 introduced by this submission.

AI/ML Overview

This document describes a 510(k) Pre-Market Notification for the Arrow Stiffening Stylet. The primary purpose of this submission is to demonstrate the substantial equivalence of the Arrow Stiffening Stylet to a legally marketed predicate device, the Galt Medical Corp's GaltTWS Stylet (K182660).

Acceptance Criteria and Device Performance (as derived from the provided document):

The document does not explicitly state "acceptance criteria" as a pass/fail threshold but rather lists the tests performed and implies that the device "passed all acceptance criteria" during non-clinical performance testing. The "reported device performance" is essentially that the device met the requirements of these tests.

Acceptance Criteria (Implied by Testing)Reported Device Performance (Implied by Conclusion)
Mechanical/Physical testing in accordance with ISO 11070, Guidance Documents, and Arrow International Internal Requirements:Passed
Surface: Extraneous Matter and DefectsPassed
Corrosion ResistancePassed
Radio-detectabilityPassed
Guide wire FracturePassed
Guide wire FlexurePassed
Guide wire TensilePassed
Retainer Leak (Flushable Hub Interface)Passed
Retainer Grip (Flushable Hub Interface)Passed
Coating IntegrityPassed
ParticulatePassed
Torque TestPassed
StiffnessPassed
Removal Force from CatheterPassed
Human Factors/Usability TestingPassed
Biocompatibility testing in accordance with ISO 10993-1:Passed
Cytotoxicity (MEM Elution)Passed
Sensitization Kligman Maximization Test (polar and non-polar extracts)Passed
Intracutaneous Injection (polar and non-polar extracts)Passed
Acute Systemic Toxicity (polar and non-polar extracts)Passed
Materials Mediated PyrogenicityPassed
Hemolysis (direct and indirect contact)Passed

Study Details:

  1. Sample Size and Data Provenance:

    • Test Set Sample Size: The document does not specify the exact sample sizes for each test. It broadly states "testing conducted" without providing numerical details for the number of units or distinct test conditions evaluated.
    • Data Provenance: The data appears to be entirely from non-clinical performance testing conducted by Arrow International LLC. There is no indication of patient data or clinical trials. The country of origin for the data is not explicitly stated, but as Arrow International LLC is based in Morrisville, North Carolina, USA, the testing was likely performed there or in accordance with US regulatory standards. The data is not retrospective or prospective clinical data as no human clinical data was used.
  2. Number of Experts and Qualifications for Ground Truth:

    • Not Applicable. This submission is for a medical device (stylet) that relies on physical and mechanical testing, not interpretation of data like images or patient records that would require expert consensus for ground truth. Therefore, there's no mention of experts establishing ground truth in the context of interpretation. Internal engineering and quality control experts at Arrow International LLC would have developed the test protocols and evaluated the results against internal requirements and relevant ISO standards.
  3. Adjudication Method for Test Set:

    • Not Applicable. As this is non-clinical physical/mechanical testing, there isn't an "adjudication method" in the sense of resolving disagreements between human readers or experts. Test results would be objectively measured against predefined specifications.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. An MRMC study was not conducted. This type of study is typically performed for AI or diagnostic imaging devices where human interpretation is involved. This submission is for a physical medical device.
  5. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

    • Not Applicable. This is a physical medical device, not a software algorithm or AI.
  6. Type of Ground Truth Used:

    • The "ground truth" for this device's performance is established by objective measurements against predefined engineering specifications and relevant international standards (e.g., ISO 11070, ISO 10993-1). It is based on physical and mechanical properties, not expert consensus, pathology, or outcomes data from a clinical setting.
  7. Training Set Sample Size:

    • Not Applicable. There is no "training set" in the context of this device. This is not an AI/machine learning device. The testing described is verification testing against design specifications.
  8. How Ground Truth for Training Set Was Established:

    • Not Applicable. As there is no training set, there is no ground truth established for one.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

August 31, 2022

Arrow International, LLC (a subsidiary of Teleflex, Inc.) Fallon Young Regulatory Affairs Specialist 3015 Carrington Mill Blvd Morrisville, North Carolina 27560

Re: K220280

Trade/Device Name: Arrow Stiffening Stylet Regulation Number: 21 CFR 870.1380 Regulation Name: Catheter Stylet Regulatory Class: Class II Product Code: DRB Dated: July 25, 2022 Received: July 28, 2022

Dear Fallon Young:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Hetal Odobasic Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationIndications for UseForm Approved: OMB No. 0910-0120Expiration Date: 06/30/2023See PRA Statement below.
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510(k) Number (if known)K220280
Device NameArrow Stiffening Stylet

Indications for Use (Describe)

The Arrow Stiffening Stylet is intended to be placed inside a catheter to stiffen the catheter for placement.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review
instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of
information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it
displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Arrow International. The word "ARROW" is in large, white, sans-serif font. Below it, in a smaller font, is the word "INTERNATIONAL". The background is a solid blue color.

510(k) SUMMARY

(as required by the Safe Medical Devices Act of 1990 and in accordance with 21 CFR 807.92)

Arrow Stiffening Stylet

1. Applicant Information:

Name:Arrow International LLC (subsidiary of Teleflex Incorporated)
Address:3015 Carrington Mill BlvdMorrisville, NC 27560 USA
Contact Person:Fallon Young
Email:fallon.young@teleflex.com
Telephone Number:(610) 984-7188
Fax Number:(610) 478-3128
Date Prepared:August 30, 2022
Device Name:
• Trade/Proprietary Name:Arrow Stiffening Stylet
• Classification Name:Catheter, Stylet
• CFR Number:21 CFR 870.1380
  • . Device Class: II
  • DRB Product Code: .

3. Predicate Device(s):

  • . K182660: Galt Medical Corp's GaltTWS Stylet

{4}------------------------------------------------

4. Description of Device:

The Arrow Stiffening Stylet is a sterile, single use, percutaneous peripheral vasculature catheter stylet intended for transient use (up to 24 hours) to be placed within a catheter to render it stiff to aid in catheter placement. The proposed device will be delivered as a sterile accessory within a convenience kit for the Arrow Peripherally Inserted Central Catheter (PICC)s. The Stylet is a PTFE coated solid nitinol core wire with straight tip. The stylet is available in 0.015-0.017-inch diameters and in 29.5-inch length with distance markings along the stylet body.

The Arrow Stiffening Stylet is provided with an existing, commercially available flushable hub with side arm and slide clamp cleared under K790408. The flushable hub maintains the stylet position within the catheter during catheter placement, is intended to minimize blood loss during catheter introduction, and allows the clinician to flush through the catheter to aid in catheter placement, if needed. There is no change to the flushable hub cleared under K790408 introduced by this submission.

ನ. Indications for Use:

The Arrow Stiffening Stylet is intended to be placed inside a catheter to stiffen the catheter for placement.

Substantial Equivalence 6.

Intended Use / Indications for Use:

The intended use of the subject Arrow Stiffening Stylet is equivalent to the intended use of the predicate stylet. The subject and predicate devices are both intended to be placed within a catheter to stiffen the catheter for placement.

The indications for use of the Arrow Stiffening Stylet and the predicate stylet include the placement within a percutaneous catheter to stiffen the device for placement. The indications for use of the subject Arrow Stiffening Stylet are equivalent to the indications for use of the predicate stylet.

Technological Characteristics:

The subject Arrow Stiffening Stylet incorporates the same fundamental technology as the predicate stylet. The subject Arrow Stiffening Stylet and the predicate stylet are both metal body stylets of similar configurations with a flushable hub.

{5}------------------------------------------------

Table 2 summarizes the substantial equivalence comparison of the subject Arrow Stiffening Stylet with the predicate stylet.

Proposed DeviceArrow Stiffening StyletPredicate Device – K182660GaltTWS Stylet
Indications forUseThe Arrow Stiffening Stylet is intendedto be placed inside a catheter to stiffenthe catheter for placement.The GaltTWS Stylet is a percutaneous catheterstylet placed inside a catheter, lead, or cannulato stiffen the device for placement.
Device Design Feature
DesignA solid nitinol wireA stainless steel twisted/braided wire
AccessoriesExisting, commercially availableflushable hub with side arm (K790408)Existing, commercially availableflushable hub with side arm
Tip TypeStraight on both endsStraight on both ends
CoatingPTFEN/A
Core MaterialNitinolStainless Steel
Coil MaterialN/A - no coilN/A - no coil
DistalAtraumatic TipYesYes
DiameterDiameter: 0.015-0.017 inch (0.39 mm-0.44 mm)Diameter: 0.012-0.015 inch (0.30 mm-0.39 mm)
LengthLength: 29.5 inch (75 cm)Length: 30-72 inches (76 cm-182 cm)
Depth ControlDistance markings every 5 cm starting at20 cm from distal end of styletNo distance markings
Sterilization
SterilityEthylene OxideEthylene Oxide
Biocompatibility
BiocompatibilityBiocompatible materials used(per ISO 10993-1)Biocompatible materials used(per ISO 10993-1)

Table 2 - Substantial Equivalence Comparison Summary

7. Non-Clinical Performance Data.

The technological differences between the subject Arrow Stiffening Stylet and the predicate stylet have been evaluated through verification testing. The testing conducted that supports substantial equivalence of the subject Arrow Stiffening Stylet is listed below.

Testing Included:

  • Mechanical/Physical testing in accordance with ISO 11070, Guidance ●

{6}------------------------------------------------

Documents, and Arrow International Internal Requirements:

  • Surface: Extraneous Matter and Defects o
  • Corrosion Resistance o
  • Radio-detectability O
  • Guide wire Fracture O
  • Guide wire Flexure O
  • Guide wire Tensile O
  • Retainer Leak (Flushable Hub Interface) O
  • Retainer Grip (Flushable Hub Interface) O
  • Coating Integrity O
  • Particulate O
  • Torque Test O
  • Stiffness O
  • Removal Force from Catheter O
  • Human Factors/Usability Testing ●
  • Biocompatibility testing in accordance with ISO 10993-1:
    • Cytotoxicity (MEM Elution) o
    • Sensitization Kligman Maximization Test (polar and non-polar o extracts)
    • Intracutaneous Injection (polar and non-polar extracts) o
    • Acute Systemic Toxicity (polar and non-polar extracts) O
    • Materials Mediated Pyrogenicity о
    • Hemolysis (direct and indirect contact) O
  • The Arrow Stiffening Stylet will be packaged as a convenience kit ● component with Arrow Peripherally Inserted Central Catheter (PICC) products. There are no changes to the packaging introduced by this submission.

8. Clinical Performance Data.

No human clinical data was required to support substantial equivalence.

9. Conclusion Regarding Substantial Equivalence

The information included in this premarket notification supports the substantial equivalence of the subject Arrow Stiffening Stylet to the

{7}------------------------------------------------

predicate stylet. The subject device has a substantially equivalent intended use, indications for use, and incorporates the same fundamental technology as the legally marketed predicate device to which it is compared.

Performance, human factors/usability, and biocompatibility testing has been conducted and has passed all acceptance criteria to verify that the proposed Arrow Stiffening Stylet meets its design, physical integrity, functional, and safety requirements and that any dimensional or material differences between it and the predicate stylet do not raise new issues of safety and effectiveness. The results of the testing included in this premarket notification support a determination of substantial equivalence.

§ 870.1380 Catheter stylet.

(a)
Identification. A catheter stylet is a wire that is run through a catheter or cannula to render it stiff.(b)
Classification. Class II (performance standards).