K Number
K220277
Date Cleared
2022-02-17

(17 days)

Product Code
Regulation Number
872.1800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The digital intraoral X-ray sensors XVD2121 Plus, XVD2530 are intended for any dental practice that uses X-ray equipment for intraoral diagnostic purposes. Each can be used by trained dental professionals for patients receiving intraoral X-ray examinations and produces digital images that can be displayed and archived digitally.

Device Description

The subject device digital intraoral X-ray sensor, XVD2121 Plus or XVD2530, is used as X-ray receptor to acquire dental X-ray radiographic images in clinical entities. The subject device doesn't emit X-rays. The X-rays are to be generated from X-ray dental machine, a separate device. The subject device can be used by trained dental professionals for patients receiving intraoral X-ray imaging examinations and produces digital images that can be displayed, and archived digitally. To acquire X-ray intraoral radiographic image, first the X-rays are to be generated from the X-ray dental machine and passing through human teeth and surrounding anatomy; then the attenuated X-ray beams are detected by the sensor and subsequently converted into digital data that are to be transmitted to the software (computer). The imaging data are then stored in DICOM format images that can be displayed on the monitor and viewed by dentists.

The subject device digital intraoral X-ray sensor consists of sensor, adaptor, cable, and software (computer or laptop NOT part of the device).

The intraoral sensor detects the incident X-rays and converts X-rays into digital signal or data. The sensor consists of tiny silicon chip-based pixels and their associated electronics encased in a plastic housing. The intraoral sensor is small, thin, flat, rigid rectangular box, usually black in color and similar in size to intraoral film packets. Sensors vary in thickness from about 5 to 7 mm. Sensor is cabled to allow data to be transferred directly from the mouth to the computer (software) via the USB port.

When used clinically the sensor shall be covered with a protective plastic barrier envelope for infection control purposes. New sheath is required for each new patient and must be disposed of after patient use. One usually slide the sensor into the sheath to provide a secure barrier around the sensor. The protective cover sheath is disposable and is a separate device, NOT a part of this device.

To acquire the intraoral X-ray radiographic images, the sensor (with sheath) is placed inside the mouth as X-ray receptor. First the X-rays are to be generated from the dental X-ray generating machine outside of the mouth. The X-rays pass through and are attenuated by tooth/teeth and surrounding anatomy; the attenuated X-rays are detected by the sensor and then converted into digital data that are to be transmitted to the software (computer).

The digital data are then stored in DICOM format image that can be displayed on the monitor and viewed by dentists. The key clinical feature of intraoral sensor is the rapid availability of the image after exposure. The sensor is utilized as X-ray receptor, and does NOT emit X-rays. The X-ray generating equipment is a separate device. NOT a part of subject device.

The physical size of model XVD2530 is 31.5mm * 38.5mm, while that of model XVD2121 Plus is 26.5mm * 32mm. The effective imaging area of model XVD2530 is 29.5mm * 24.6mm, while that of model XVD2121 Plus is 20.0mm * 19.8mm. The differences between model XVD2121 Plus and model XVD2530 are the physical sizes and effective imaging areas. Otherwise model XVD2121 Plus and model XVD2530 are exactly same device. Different size is offered to better fit with size of oral cavity of individual patient and different anatomic region, such as posterior teeth or anterior teeth.

AI/ML Overview

The provided text describes a 510(k) summary for a medical device cleared by the FDA, primarily focusing on proving its substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a study demonstrating the device meets those criteria in a typical clinical trial sense.

This document is a premarket notification (510(k)) to the FDA, which generally focuses on demonstrating that a new device is as safe and effective as a legally marketed predicate device. This is different from a Post-Market Approval (PMA) application, which typically requires rigorous clinical trials to demonstrate safety and effectiveness for a novel device.

Therefore, the information regarding specific acceptance criteria and detailed study results that "prove the device meets acceptance criteria" for clinical performance, as one might expect from a full clinical trial, is limited or not present in the provided text. The testing described is primarily non-clinical performance testing to verify functionality and safety, and bench testing for image quality and electrical safety, to support substantial equivalence.

Based on the provided text, here's an attempt to answer your questions:

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table of "acceptance criteria" for clinical performance. Instead, it lists technical specifications and performance characteristics compared to a predicate device. The general "acceptance criteria" are implied to be that the device performs functionally and safely at a level comparable to the predicate device and meets relevant industry standards.

Here's a table primarily extracting the technical specifications and reported characteristics that would serve as a proxy for performance criteria in this context of a 510(k) submission, comparing the subject device to the predicate:

DescriptionAcceptance Criteria (from Predicate Device / Standards)Reported Device Performance (Subject Device)
General
Device/Trade NameComputed Oral Radiology SystemXVD2121 Plus, XVD2530
Classification NameExtraoral Source X-ray SystemSame
Product CodeMUHSame
Regulation Number21 CFR 872.1800Same
Device ClassClass IISame
Indications for UseFor intra-oral dental x-ray examination; produces instant, digital, intra-oral images while reducing necessary x-ray dosage.For intraoral diagnostic purposes; produces digital images that can be displayed and archived digitally.
Intended User GroupTrained dental professionalsSame
Deployment MethodsCovered with a protective sheath then placed inside the mouth.Same
Contact Body SiteDevice does not directly contact human body or organ (covered by protective plastic barrier envelope).Same (covered by protective plastic barrier envelope).
Patient PopulationsGeneral population (excluding pregnant women) needing intraoral X-ray imaging.Same
Installation TypePortableSame
Image Quality & Functionality
Sensor StructureScintillator + CMOSSemiconductor + CMOS ASIC
Actual Spatial Resolution7 lp/mmSame (7 lp/mm)
Low Contrast ResolutionCan distinguish 1mm-diameter hole on aluminum platesSame (Can distinguish 1mm-diameter hole on aluminum plates)
Image Non-uniformity≤ 2%Same (≤ 2%)
Ghost and ArtifactNoSame (No)
Gray Scale12 bit16 bit
Power Consumption5V DC, 250mA5V DC, 100mA
CommunicationsUSB 2.0Same
CoolingAmbient air coolingSame
Safety & Environment
Protection against Matter/WaterIP68IP67
Protection against ShockType BF applied partSame
Operation EnvironmentTemp: 10-40°C, Humidity: ≤ 75% (Non-Condensing), Atmos. Pressure: 700-1060hPaTemp: 10-40°C, Humidity: ≤ 80% (Non-Condensing), Atmos. Pressure: 700-1060hPa
Storage & Transportation EnvironmentTemp: -40-70°C, Humidity: 10-100% (Non-Condensing), Atmos. Pressure: 500-1060hPaTemp: -20-55°C, Humidity: ≤ 93% (No condensation), Atmos. Pressure: 700-1060hPa
Software
SoftwareZoom WindowXVDental
Standards ComplianceIEC 60601-1:2005+A1:2012, IEC 60601-1-2:2014IEC 60601-1:2005+A1:2012, IEC 60601-1-2:2014, IEC 60601-2-65:2017

The Study Proving Acceptance Criteria:

The document explicitly states: "Clinical data is NOT required for a finding of substantial equivalence." (Section 13). Therefore, the "study" proving the device meets acceptance criteria regarding clinical performance is not a full-scale clinical trial.

Instead, the "study" is a series of non-clinical performance testing (bench testing, phantom tests, system validation/verification), electrical safety, EMC, and software testing, as described in sections 8, 9, 10. The goal of these tests is to demonstrate that the device performs functionally and safely in a manner comparable to the predicate device and meets recognized consensus standards.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not explicitly stated as a number of patient cases or images for clinical evaluation, because no clinical study was performed. The "test set" pertains to the non-clinical and bench testing. For phantom tests and system validation, the "sample size" would refer to the number of tests performed on the physical devices/software, but these quantities are not specified in the summary.
  • Data Provenance: The document does not specify the country of origin for the non-clinical test data. It is implied to be from the manufacturer's own testing facilities.
  • Retrospective or Prospective: Not applicable, as there was no clinical data collection of patient images described. The non-clinical testing would be prospective (tests conducted at the time of device development).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not applicable, as there was no clinical study involving expert readers to establish ground truth for a test set of patient images.
  • Qualifications of Experts: N/A.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable, as there was no clinical study involving multiple readers and adjudication of findings.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. The device described is an intraoral X-ray sensor (hardware), not an AI-powered diagnostic tool for image interpretation. Therefore, the concept of "human readers improving with AI assistance" is not relevant to this device's submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable in the context of an "algorithm only" performance, as this device is a hardware sensor, not a standalone AI algorithm. The image quality and technical specifications (e.g., spatial resolution, low contrast resolution, gray scale) can be considered "standalone" performance characteristics of the hardware and its associated software for image acquisition and display. These were evaluated through non-clinical bench testing.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth: For the non-clinical performance testing, the "ground truth" would be established by physical measurements against known standards (e.g., phantom test results, resolution charts, electrical safety parameters) and engineering specifications. There is no mention of clinical ground truth (e.g., expert consensus, pathology) for patient data, as no clinical study was conducted.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable. This device is a hardware sensor. While it has embedded software, the document describes it as a "hardware-based device that incorporates software" (Section 10). There is no indication of a machine learning or AI component that would require a "training set" of data in the typical sense for a clinical imaging algorithm.

9. How the ground truth for the training set was established

  • How Ground Truth for Training Set was Established: Not applicable, as there is no mention of an algorithm requiring a training set in the document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol is a stylized representation of a human figure, and the text is written in a clear, sans-serif font. The FDA logo is commonly used on products and materials related to food and drug regulation in the United States.

February 17, 2022

Shenzhen Xpectvision Technology Co., Ltd. % Mengsi Peng Regulatory Affairs Manager B507, Block A and B. Nanshan Medical Device Industrial Park, Nanhai Avenue 1019, Nanshan District Shenzhen, Guangdong 518067 CHINA

Re: K220277

Trade/Device Name: XVD2121 Plus, XVD2530 Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: January 17, 2022 Received: January 31, 2022

Dear Mengsi Peng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220277

Device Name XVD2121 Plus, XVD2530

Indications for Use (Describe)

The digital intraoral X-ray sensors XVD2121 Plus, XVD2530 are intended for any dental practice that uses X-ray equipment for intraoral diagnostic purposes. Each can be used by trained dental professionals for patients receiving intraoral X-ray examinations and produces digital images that can be displayed and archived digitally.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 510(k) Summary K220277

{4}------------------------------------------------

510 (k) Summary

(As Required by 21 CFR 807.92)

1. Date Prepared

January 17, 2022

2. Submitter's Information

Company Name:Shenzhen Xpectvision Technology Co., Ltd.
Company Address:B507, Block A and B,Nanshan Medical Device Industrial Park,Nanhai Avenue 1019, Nanshan District,Shenzhen City, Guangdong Province, China.
Contact Person:Mengsi Peng
Phone:(+86) 0755-26897621
Email:msp1022@xpectvision.com

3. Device/Trade Name, Common Name, and Classification Name

Device/Trade Name:XVD2121 Plus, XVD2530
Common Name:intraoral X-ray sensor
Classification Name:Extraoral Source X-Ray System
Regulation Number:21 CFR 872.1800
Product Code:MUH
Device Class:Class II

{5}------------------------------------------------

4. Identification of Predicate Device(s)

Relevant information of predicate device within this submission are as follows:

Manufacturer:Schick Technologies Inc
Device/Trade Name:Computed Oral Radiology System
Regulation Number:21 CFR 872.1800
Product Code:MUH
Classification Name:Extraoral source x-ray system
FDA 510 (k) #:K072134
Device Class:Class II

5. Indications for Use

The digital intraoral X-ray sensors XVD2121 Plus, XVD2530 are intended for any dental practice that uses X-ray equipment for intraoral diagnostic purposes. Each can be used by trained dental professionals for patients receiving intraoral X-ray examinations and produces digital images that can be displayed and archived digitally.

6. Description of the Device

The subject device digital intraoral X-ray sensor, XVD2121 Plus or XVD2530, is used as X-ray receptor to acquire dental X-ray radiographic images in clinical entities. The subject device doesn't emit X-rays. The X-rays are to be generated from X-ray dental machine, a separate device. The subject device can be used by trained dental professionals for patients receiving intraoral X-ray imaging examinations and produces digital images that can be displayed, and archived digitally. To acquire X-ray intraoral radiographic image, first the X-rays are to be generated from the X-ray dental machine and passing through human teeth and surrounding anatomy; then the attenuated X-ray beams are detected by the sensor and subsequently converted into digital data that are to be transmitted to the software (computer). The imaging data are then stored in DICOM format images that can be displayed on the monitor and viewed by dentists.

{6}------------------------------------------------

The subject device digital intraoral X-ray sensor consists of sensor, adaptor, cable, and software (computer or laptop NOT part of the device), as shown in Fig.1.

Image /page/6/Figure/1 description: The image shows a device setup with four labeled components. Component 1 is a black square device connected by a cable to component 2, a small, rounded device with a light. Component 2 is connected via a cable to component 3, a USB connector. Component 3 is plugged into component 4, a laptop displaying an image processing application.

Figure 1. Components of the device

The intraoral sensor detects the incident X-rays and converts X-rays into digital signal or data. The sensor consists of tiny silicon chip-based pixels and their associated electronics encased in a plastic housing. The intraoral sensor is small, thin, flat, rigid rectangular box (as denoted (1) in Fig.1 ), usually black in color and similar in size to intraoral film packets. Sensors vary in thickness from about 5 to 7 mm. Sensor is cabled to allow data to be transferred directly from the mouth to the computer (software) via the USB port.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a dental X-ray sensor in a clear plastic bag. The sensor is black and rectangular, with a cable extending from one side. The cable is also black and appears to be flexible. The plastic bag is sealed around the sensor and cable, protecting it from contamination.

Figure 2. Sensor inside a protective plastic sheath

When used clinically the sensor shall be covered with a protective plastic barrier envelope for infection control purposes. New sheath is required for each new patient and must be disposed of after patient use. One usually slide the sensor into the sheath to provide a secure barrier around the sensor (as shown in Fig. 2). The protective cover sheath is disposable and is a separate device, NOT a part of this device.

Image /page/7/Picture/3 description: The image shows a person getting an x-ray of their mouth. The person is facing to the right, and their mouth is open. A device is pointed at their mouth, and a beam of light is shining into their mouth. The x-ray is being taken to check for any problems with the person's teeth or gums.

Figure 3. Placement of sensor (with sheath) during intraoral X-ray imaging

To acquire the intraoral X-ray radiographic images, the sensor (with sheath) is placed inside the mouth as X-ray receptor. First the X-rays are to be generated from the dental X-ray generating machine outside of the mouth (as shown in Fig. 3). The X-rays pass through and are attenuated by tooth/teeth and surrounding anatomy; the attenuated X-rays are detected by the sensor and then converted into digital data that are to be transmitted to the software (computer).

{8}------------------------------------------------

The digital data are then stored in DICOM format image that can be displayed on the monitor and viewed by dentists. The key clinical feature of intraoral sensor is the rapid availability of the image after exposure. The sensor is utilized as X-ray receptor, and does NOT emit X-rays. The X-ray generating equipment is a separate device. NOT a part of subject device.

Image /page/8/Figure/1 description: The image shows two sets of technical drawings of an object, each set containing a side view and a front view. The first set of drawings shows the object with dimensions of 5.8 ± 0.1 mm for the side view and 26.5 ± 0.1 mm in width and 32.0 ± 0.1 mm in height for the front view. The second set of drawings shows the object with dimensions of 6.0 ± 0.1 mm for the side view and 31.5 ± 0.1 mm in width and 38.5 ± 0.1 mm in height for the front view.

(a) Model XVD2121 Plus (b) Model XVD2530 Figure 4. Lateral and frontal view of the sensor (the shaded region denotes effective imaging area)

The physical size of model XVD2530 is 31.5mm * 38.5mm, while that of model XVD2121 Plus is 26.5mm * 32mm. The effective imaging area of model XVD2530 is 29.5mm * 24.6mm, while that of model XVD2121 Plus is 20.0mm * 19.8mm. The differences between model XVD2121 Plus and model XVD2530 are the physical sizes and effective imaging areas. Otherwise model XVD2121 Plus and model XVD2530 are exactly same device. Different size is offered to better fit with size of oral cavity of individual patient and different anatomic region, such as posterior teeth or anterior teeth.

{9}------------------------------------------------

7. Comparable Technological Characteristics

DescriptionPredicate DeviceSubject Device
Device/TradenameComputed Oral RadiologySystemXVD2121 Plus, XVD2530
510(K) numberK072134K220277
ClassificationnameExtraoral Source X-raySystemSame
Product codeMUHSame
Regulationnumber21 CFR 872.1800Same
PanelRadiologySame
ClassificationIISame
Indications for useThe Computed OralRadiology System isindicated for patientsundergoing an intra-oraldental x-ray examination.It produces instant,digital, intra-oral imagesof a patient's mouth whilereducing the necessary x-ray dosage.The digital intraoral X-raysensors XVD2121 Plus,XVD2530 are intendedfor any dental practicethat uses X-ray equipmentfor intraoral diagnosticpurposes. Each can beused by trained dentalprofessionals for patientsreceiving intraoral X-rayexaminations andproduces digital imagesthat can be displayed andarchived digitally.
Intended usergroupTrained dentalprofessionalsSame
DescriptionPredicate DeviceSubject Device
DevicecomponentsImage: sensor, adaptor, cable, and USB connector.XVD2121 PlusImage: XVD2121 Plussame
XVD2530Image: XVD2530same
Sensor areaSize 1: 23.6mm x 31.9mmSize 2: 25.4mm x 38.3mmSize 3: 31.2mm x 43.0mmXVD2121 Plus:26.5mm x 32mmXVD253031.5mm x 38.5mm
Effective imagingareaSize 1: 18mm x 24mmSize 2: 20mm x 30mmSize 3: 25.6mm x 36mmXVD2121 Plus:20.0mm x 19.8mmXVD2530:24.6mm x 29.5mm
Deploymentmethods whererelevantCovered with a protectivesheath then placed inside themouth.Same
DescriptionPredicate DeviceSubject Device
Contact body siteThe device does notdirectly contact withhuman body or organ.Note: When usedclinically the sensorshall be covered with aprotective plastic barrierenvelope for infectioncontrol purposes andthen placed inside themouth during imagingprocedure. Theprotective cover sheathis disposable and is aseparate device, NOT apart of this device.Same
PatientpopulationsGeneral population(excluding pregnantwomen) who areevaluated by the dentist toneed intraoral X-rayimaging examination.Same
Installation typePortableSame
Sensor structureScintillator+ CMOSSemiconductor+ CMOS ASIC
Actual spatialresolution7 lp/mmSame
Low contrastresolutionThe imaging candistinguish 1mm -diameterhole on aluminum platesSame
Image non-uniformity$≤$ 2%Same
DescriptionPredicate DeviceSubject Device
Ghost and artifactNoSame
Gray scale12 bit16 bit
Powerconsumption5V DC, 250mA5V DC, 100mA
CommunicationsUSB 2.0Same
CoolingAmbient air coolingSame
Protection againstmatter/waterIP68IP67
Protection againstshockType BF applied partSame
OperationenvironmentTemperature: 10 to 40°CHumidity: ≤ 75% (Non-Condensing)Atmospheric pressure:700hPa ~ 1060hPaTemperature: 10 to 40°CHumidity: ≤ 80% (Non-Condensing)Atmospheric pressure:700hPa ~ 1060hPa
Storage andTransportationenvironmentTemperature: -40 to 70°CHumidity: 10 to 100%(Non - Condensing)Atmospheric pressure:500hPa ~ 1060hPaTemperature: -20 to 55°CHumidity: ≤ 93% (Nocondensation)Atmospheric pressure:700hPa ~ 1060hPa
SoftwareZoom WindowXVDental
StandardsComplianceIEC 60601-1:2005+A1:2012IEC 60601-1-2:2014IEC 60601-1:2005+A1:2012IEC 60601-1-2:2014IEC 60601-2-65:2017

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

8. Non-clinical Performance Testing

Non-clinical testing, (integration and functional) including phantom tests were conducted for the subject device during product development. The general purpose of each tests is to verify and validate the performance and functionality of the subject device.

System Validation testings include:

  • Acceptance test (workflow and user manual test) ●
  • Legal and Regulatory test

System Verification testings include:

  • System Integration Test (functional) ●
  • Functionality verification
  • Image Quality (IQ) Evaluation ●

Testing covers all related aspects that contribute to the device performance and functions. The test specification and acceptance criteria are related to the corresponding requirements. The test results show that all of the performance specifications have met the acceptance criteria. Verification and validation testing of the device was found acceptable to support the claim of substantial equivalence.

{14}------------------------------------------------

A list of FDA-recognized consensus standards and general guidance documents considered for testing the subject device is listed as follows:

    1. FDA-recognized consensus standards
    • IEC 61223-3-4 First edition 2000-03 Evaluation and routine testing ● in medical imaging departments - Part 3-4: Acceptance tests -Imaging performance of dental X-ray.
    • IEC 60601-2-65 Edition 1.1 2017-05 CONSOLIDATED VERSION Medical electrical equipment - Part 2-65: Requirements for the basic safety and essential performance of dental intraoral X-ray equipment.
    1. FDA Guidance Documents
    • FDA Guidance for the Medical X-Ray Imaging Devices Conformance with IEC Standards.
    • O FDA Guidance for the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)].
    • FDA Guidance for the Device Labeling #G91.
    • FDA Guidance for the Information to Support a Claim of O Electromagnetic Compatibility (EMC) of Electrically - Powered Medical Devices.
    • FDA Guidance for the Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions.

{15}------------------------------------------------

9. Electrical Safety and EMC Testing

Electrical Safety and Electromagnetic Compatibility (EMC) testing were conducted on the subject device in accordance with the following standards or guide document(s):

  • IEC 60601-1:2005+CORR. 1:2006+CORR. 2:2007+A1:2012 Medical ● electrical equipment - Part 1: General requirements for basic safety and essential performance.
  • . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
  • . FDA Guidance for the Information to Support a a Claim of Electromagnetic Compatibility (EMC) of Electrically - Powered Medical Devices.

{16}------------------------------------------------

10. Software and Cybersecurity

The subject device is a hardware-based device that incorporates software. Tests are conducted for all software components developed in product development and for the complete product itself. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claims of substantial equivalence.

In accordance with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", the software documentation is also included in this submission that covers the following aspects:

  • Software description
  • Software hazard analysis ●
  • Software requirements specifications ●
  • Software architecture design chart ●
  • Software design specifications ●
  • Software traceability analysis ●
  • Software development environment description ●
  • Verification & Validation testing ●
  • Revision level history
  • Unresolved anomalies

The submitter conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from the subject device to an external recipient.

{17}------------------------------------------------

Cybersecurity information in accordance with guidance document "Content of Premarket Submissions for Management of Cybersecurity Medical Devices," is discussed in associated Cybersecurity documentation.

11. Biocompatibility Information

The subject device does NOT contact directly with human body or organ, hence no biocompatibility testing was performed.

12. Sterility Information

No sterility testing was accomplished, as this device is not delivered sterile, nor does it require sterility.

13. Clinical Testing

Clinical data is NOT required for a finding of substantial equivalence.

{18}------------------------------------------------

14. Conclusion

As discussed above, the subject device, either model XVD2121 Plus or model XVD2530, employs the same fundamental technology and has a set of same or similar technological characteristics as the predicate device. The non-clinical data supports the safety of the device. The hardware and software verification and validation demonstrates that the subject device should perform as intended in the specified use conditions. The data included in this submission demonstrates that the subject device performs comparably to the predicate device currently marketed for the same intended use. The differences between the subject device and predicate device do not raise different questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the predicate device.

§ 872.1800 Extraoral source x-ray system.

(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.