AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diamondback 360® Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The OAS supports removal of stenotic from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

The Stealth 360® Peripheral Orbital Atherectorny System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

The Diamondback 360® Peripheral Orbital Atherectomy System Exchangeable Series is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The Exchangeable Series OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

Device Description

The DIAMONDBACK 360 Peripheral OAS, Stealth 360 Peripheral OAS, and DIAMONDBACK 360 OAS Exchangeable Series are designed to remove or reduce occlusive material and restore luminal patency by using an orbiting, diamond-coated, eccentrically mounted crown. Each OAS consists of the following main components:

  1. Reusable Saline Pump (provided non-sterile)
  2. Single-use Orbital Atherectomy Device (OAD) (provided sterile). The Exchangeable Series OAD consists of a physician-operated handle and an interchangeable crown cartridge.
  3. Single-use Atherectomy lubricant (provided sterile)
  4. Single-use Atherectomy guide wire (provided sterile)
AI/ML Overview

The provided text describes a 510(k) summary for the DIAMONDBACK 360® Peripheral Orbital Atherectomy System, DIAMONDBACK 360® Peripheral Orbital Atherectomy System, Exchangeable Series, and Stealth 360® Peripheral Orbital Atherectomy System.

Here's an analysis of the acceptance criteria and study information:

1. A table of acceptance criteria and the reported device performance:

The document states:

  • "Functional and Safety Testing: Only software verification and associated testing was required as the software updates do not affect the device performance and hardware."
  • "The modified devices are different from the predicate device in that the software has been updated to include additional cybersecurity protection. Also, a hypotube was added to the Exchangeable driveshaft and a coupler was updated to accommodate the change. The hypotube and coupler are located within a fully enclosed, non-patient contacting portion of the device inside the cartridge."
  • "The hypotube/coupler update does not impact the functionality of the device to ablate the lesion nor its intended use."
  • "The subject devices are substantially equivalent to the predicate devices and the devices continue to perform as intended."

Given this, the acceptance criteria are implicitly that the software updates for cybersecurity and the mechanical changes (hypotube/coupler) do not adversely affect the device's original performance specifications and intended use. The performance reported is that the device continues to perform as intended and its functionality for lesion ablation and intended use are not impacted.

Acceptance CriteriaReported Device Performance
Software updates for cybersecurity do not affect device performance.Software updates do not affect device performance.
Hardware modifications (hypotube/coupler) do not affect device performance, functionality, or intended use.Does not impact the functionality of the device to ablate the lesion nor its intended use.
Device maintains substantial equivalence to predicate devices.Subject devices are substantially equivalent to predicate devices and continue to perform as intended.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

The document does not specify a separate "test set" in the context of clinical or performance data for the device's atherectomy function. The primary testing mentioned is "software verification and associated testing" and evaluation of the impact of mechanical changes. No sample size for patients or clinical data is provided, nor is data provenance mentioned, as this submission focuses on modifications rather than new clinical claims.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

Not applicable. The submission does not detail an expert-reviewed ground truth for a clinical test set. The evaluation seems to be based on engineering and functional assessments rather than clinical opinion.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

Not applicable. There is no mention of an adjudication method for a clinical test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is an atherectomy system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is not an AI algorithm. The device has software for cybersecurity, but its primary function is mechanical.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" used for this 510(k) submission appears to be regulatory and engineering standards, and the performance characteristics of the previously cleared predicate devices. The modifications were assessed against the existing performance specifications and intended use of the predicate devices.

8. The sample size for the training set:

Not applicable. There is no mention of a training set as this is not an AI model.

9. How the ground truth for the training set was established:

Not applicable. There is no mention of a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 1, 2022

Cardiovascular Systems Inc. Nikita Basandra Regulatory Affairs Manager 1225 Old Highway 8 NW Saint Paul, Minnesota 55112

Re: K220109

Trade/Device Name: DIAMONDBACK 360® Peripheral Orbital Atherectomy System, DIAMONDBACK 360® Peripheral Orbital Atherectomy System, Exchangeable Series, Stealth 360® Peripheral Orbital Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal artery stripper Regulatory Class: Class II Product Code: MCW Dated: January 12, 2022 Received: January 13, 2022

Dear Nikita Basandra:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220109

Device Name

Diamondback 360® Peripheral Orbital Atherectomy System

Indications for Use (Describe)

The Diamondback 360® Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The OAS supports removal of stenotic from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K220109

Device Name Stealth 360® Peripheral Orbital Atherectomy System

Indications for Use (Describe)

The Stealth 360® Peripheral Orbital Atherectorny System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K220109

Device Name

Diamondback 360® Peripheral Orbital Atherectomy System Exchangeable Series

Indications for Use (Describe)

The Diamondback 360® Peripheral Orbital Atherectomy System Exchangeable Series is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The Exchangeable Series OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Cardiovascular Systems, Inc. The logo consists of the letters "CSI" in a bold, sans-serif font, with a green arc underneath. To the right of the letters, separated by a vertical line, is the company name "CARDIOVASCULAR SYSTEMS, INC." in a smaller, sans-serif font.

l 510(K) SUMMARY (K220109)

Submitter:Cardiovascular Systems, Inc.1225 Old Highway 8 NWSaint Paul, MN 55112
ContactPerson:Nikita BasandraRegulatory Affairs ManagerCardiovascular Systems, Inc.1225 Old Highway 8 NWSaint Paul, MN 55112Ph: 651-259-8206nbasandra@csi360.com
DatePrepared:January 20, 2022
TradeName:• DIAMONDBACK 360 Peripheral Orbital Atherectomy System• DIAMONDBACK 360 Peripheral Orbital Atherectomy SystemExchangeable Series• Stealth 360 Peripheral Orbital Atherectomy System
CommonName:Intraluminal Artery Stripper
RegulationNumber:21 CFR 870.4875
Classification:Class II
ProductCode:MCW
PredicateDevice(s):K190634 - DIAMONDBACK 360® and Stealth 360® Peripheral OrbitalAtherectomy Systems (Cardiovascular Systems, Inc.)• K182397 - DIAMONDBACK 360® Peripheral Orbital AtherectomySystem Exchangeable Series
DeviceDescription:The DIAMONDBACK 360 Peripheral OAS, Stealth 360 Peripheral OAS, andDIAMONDBACK 360 OAS Exchangeable Series are designed to remove orreduce occlusive material and restore luminal patency by using an orbiting,diamond-coated, eccentrically mounted crown. Each OAS consists of thefollowing main components:1. Reusable Saline Pump (provided non-sterile)2. Single-use Orbital Atherectomy Device (OAD)(provided sterile). The Exchangeable Series OAD
consists of a physician-operated handle and aninterchangeable crown cartridge.3. Single-use Atherectomy lubricant (provided sterile)4. Single-use Atherectomy guide wire (provided sterile)
Mechanism of ActionThe Diamondback, Stealth, and Exchangeable Series OAS mechanism ofaction is identical to the predicate device and is defined by:
• Centrifugal force• Orbital rotation• Differential sanding• Bi-directional sanding
The rapidly rotating eccentric crown creates a centrifugal force that presses thediamond-coated crown against the calcified plaque. With each pass of thecrown, plaque is reduced and the diameter of the orbit increases.
IntendedUse:The Diamondback 360® Peripheral Orbital Atherectomy SystemThe Diamondback 360® Peripheral Orbital Atherectomy System is apercutaneous orbital atherectomy system indicated for use as therapy inpatients with occlusive atherosclerotic disease in peripheral arteries andwho are acceptable candidates for percutaneous transluminalatherectomy.
The OAS supports removal of stenotic material from artificial arteriovenousdialysis fistulae (AV shunt). The system is a percutaneous orbitalatherectomy system indicated as a therapy in patients with occludedhemodialysis grafts who are acceptable candidates for percutaneoustransluminal angioplasty.
Stealth 360® Peripheral Orbital Atherectomy SystemThe Stealth 360® Peripheral Orbital Atherectomy System is apercutaneous orbital atherectomy system indicated for use as therapy inpatients with occlusive atherosclerotic disease in peripheral arteries andwho are acceptable candidates for percutaneous transluminalatherectomy.
The OAS supports removal of stenotic material from artificial arteriovenousdialysis fistulae (AV shunt). The system is a percutaneous orbitalatherectomy system indicated as a therapy in patients with occluded
hemodialysis grafts who are acceptable candidates for percutaneoustransluminal angioplasty.
Diamondback 360® Peripheral Orbital Atherectomy System ExchangeableSeries
The Diamondback 360® Peripheral Orbital Atherectomy SystemExchangeable Series is a percutaneous orbital atherectomy systemindicated for use as therapy in patients with occlusive atheroscleroticdisease in peripheral arteries and who are acceptable candidates forpercutaneous transluminal atherectomy.
The Exchangeable Series OAS supports removal of stenotic material fromartificial arteriovenous dialysis fistulae (AV shunt). The system is apercutaneous orbital atherectomy system indicated as a therapy in patientswith occluded hemodialysis grafts who are acceptable candidates forpercutaneous transluminal angioplasty.
Comparisonto PredicateDevice:The modified Diamondback 360® [Stealth 360®] Peripheral OrbitalAtherectomy System [Exchangeable Series] devices are identical to thepredicate devices as follows:Same regulation number, product code and classification Same intended use and indications for use Same vessel diameter range and anatomic location of use Same design and materials Same hardware Same principles of operation Same sterilization method and SAL Same number of uses per device (single use) Same performance specifications The modified devices are different from the predicate device in that thesoftware has been updated to include additional cybersecurity protection.Also, a hypotube was added to the Exchangeable driveshaft and a coupler wasupdated to accommodate the change. The hypotube and coupler are locatedwithin a fully enclosed, non-patient contacting portion of the device inside thecartridge. The driveshaft continues to be identical in length, material, filarcount, tip, and crown to the predicate driveshaft.
Functionaland SafetyTesting:Only software verification and associated testing was required as the softwareupdates do not affect the device performance and hardware.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Cardiovascular Systems, Inc. The logo consists of the letters "CSI" in a bold, sans-serif font, with a green arc underneath the "S" and "I". To the right of the letters, there is a vertical line separating the letters from the words "CARDIOVASCULAR SYSTEMS, INC.", which are stacked on top of each other in a smaller, sans-serif font.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Cardiovascular Systems, Inc. The logo consists of the letters "CSI" in a stylized font, with a green arc underneath. To the right of the letters is a vertical line, followed by the words "CARDIOVASCULAR SYSTEMS, INC." in a smaller, sans-serif font. The logo is simple and modern, and the use of green suggests a focus on health and wellness.

This confidential document is the property of Cardiovascular Systems, Inc. and shall not be reproduced, disclosed or used without the express written consent of Cardiovascular Systems, Inc.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Cardiovascular Systems, Inc. The logo consists of the letters "CSI" in a gray, sans-serif font, with a green arc underneath the letters. To the right of the letters is a vertical line, followed by the words "CARDIOVASCULAR SYSTEMS, INC." in a smaller, gray, sans-serif font. The logo is simple and modern, and it is likely used to represent the company's brand.

The hypotube/coupler update does not impact the functionality of the device to ablate the lesion nor its intended use. The subject devices are substantially equivalent to the predicate devices and Conclusion: the devices continue to perform as intended.

§ 870.4875 Intraluminal artery stripper.

(a)
Identification. An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)(b)
Classification. Class II (performance standards).