K Number
K220105
Device Name
Saige-Dx
Manufacturer
Date Cleared
2022-05-12

(120 days)

Product Code
Regulation Number
892.2090
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Saige-Dx analyzes digital breast tomosynthesis (DBT) mammograms to identify the presence of soft tissue lesions and calcifications that may be indicative of cancer. For a given DBT mammogram, Saige-Dx analyzes the DBT image stacks and the accompanying 2D images, including full field digital mammography and/or synthetic images. The system assigns a Suspicion Level, indicating the strength of suspicion that cancer may be present, for each detected finding and for the entire case. The outputs of Saige-Dx are intended to be used as a concurrent reading aid for interpreting physicians on screening mammograms with compatible DBT hardware.

Device Description

Saige-Dx is a software device that processes screening mammograms using artificial intelligence to aid interpreting radiologists. By automatically detecting the presence or absence of soft tissue lesions and calcifications in mammography images, Saige-Dx can help improve reader performance, while also reducing time. The software takes as input a set of x-ray mammogram DICOM files from a single digital breast tomosynthesis (DBT) study and generates finding-level outputs for each image analyzed, as well as an aggregate case-level assessment. Saige-Dx processes both the DBT image stacks and the associated 2D images (full-field digital mammography (FFDM) and/or synthetic 2D images) in a DBT study. For each image, Saige-Dx outputs bounding boxes circumscribing any detected findings and assigns a Finding Suspicion Level to each finding, indicating the degree of suspicion that the finding is malignant. Saige-Dx uses the results of the finding-level analysis to generate a Case Suspicion Level, indicating the degree of suspicion for malignancy across the case. Saige-Dx encapsulates the finding and caselevel results into a DICOM Structured Report (SR) object containing markings that can be overlaid on the original mammogram images using a viewing workstation and a DICOM Secondary Capture (SC) object containing a summary report of the Saige-Dx results.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implicit)Reported Device Performance
Reader Performance Improvement (MRMC Study)
- Increase in Radiologist AUC when aided by Saige-Dx.The average AUC of radiologists increased from 0.865 (unaided) to 0.925 (aided), a difference of 0.06 (95% CI: 0.041, 0.079, p < 0.00001). All 18 readers showed an increase.
- Increase in Radiologist Sensitivity when aided by Saige-Dx.Average reader sensitivity increased by 8.8% (95% CI: 7.0%, 10.6%).
- Stability/Improvement in Radiologist Specificity when aided by Saige-Dx.Average reader specificity increased by 0.9% (95% CI: -0.9%, 2.7%).
- Consistent performance across various subgroups (breast densities, ages, race/ethnicities, lesion types/sizes, radiologist specialization).Similar trends observed:
- Lesion type: AUC increased from 0.866 to 0.918 for soft tissue, and 0.795 to 0.899 for calcifications.
- Radiologist specialization: AUC for breast imaging specialists increased from 0.885 to 0.931; for generalists, from 0.826 to 0.911.
Standalone Performance (Algorithm Only)
- Demonstrate strong standalone performance (e.g., high AUC).Saige-Dx exhibited an AUC of 0.930 (95% CI: 0.902, 0.958) on the dataset, demonstrating strong performance relative to the unaided reader performance in the reader study.
- Consistent standalone performance across various subgroups.Similar standalone performance trends were observed across breast densities, ages, race/ethnicities, and lesion types and sizes. Assessed on recalled/non-recalled and visible/non-visible cancers.
Safety and EffectivenessNon-clinical and clinical testing confirmed that Saige-Dx is safe and effective. Minor differences from predicate do not alter intended use or affect safety/effectiveness.
Substantial EquivalenceInformation presented in the 510(k) submission demonstrates Saige-Dx is substantially equivalent to the predicate device, with similar indications for use, patient population, technical characteristics, and principles of operation. Differences do not alter suitability for intended use or safety/effectiveness.

Study Details

Here's a breakdown of the studies conducted:

1. Multi-Reader Multi-Case (MRMC) Reader Study (Performance Testing: Reader Study)

  • Sample Size for Test Set: 240 cases (100 pathology-proven cancer cases, 140 confirmed non-cancer cases).
    • Data Provenance: Retrospectively collected DBT mammogram exams from unique female patients 35 years of age or older, acquired from Hologic equipment. Data was from different clinical sites than those used for AI algorithm training. Patients represented a racially and ethnically diverse population in the US.
  • Number of Experts for Ground Truth: Two MQSA qualified, highly experienced (>10 years in practice) breast imaging specialists, plus a third as an adjudicator.
  • Qualifications of Experts for Ground Truth: MQSA qualified, highly experienced (>10 years in practice) breast imaging specialists.
  • Adjudication Method: For exams with discrepancies between the two truthers' assessment of density, lesion type, and/or lesion location, a third truther served as the adjudicator.
  • MRMC Comparative Effectiveness Study: Yes.
    • Effect Size (Human Reader Improvement with AI vs. without AI):
      • Average AUC increased by 0.06 (from 0.865 unaided to 0.925 aided).
      • Average reader sensitivity increased by 8.8%.
      • Average reader specificity increased by 0.9%.
  • Standalone Performance: No, this specific study was for human reader performance with and without AI.
  • Type of Ground Truth: Expert consensus with pathology confirmation for cancer cases. Each mammogram had a ground truth status of "cancer" or "non-cancer." For cancer exams, malignant lesions were annotated based on the biopsied location that led to malignant pathology.
  • Sample Size for Training Set: Not explicitly stated, but the text mentions "six datasets across various geographic locations in the US and the UK," indicating a large, diverse dataset.
  • How Ground Truth for Training Set was Established: Not explicitly detailed for the training set, but it is stated that "DeepHealth ensured that there was no overlap between the data used to train and test the Saige-Dx Al algorithm." It can be inferred that similar robust methods (likely expert review and pathology confirmation) were used, given the thoroughness described for the test set.

2. Standalone Study (Performance Testing: Standalone Study)

  • Sample Size for Test Set: 1304 cases (136 cancer, 1168 non-cancer).
    • Data Provenance: Retrospective, blinded, multi-center study. Collected from 9 clinical sites in the United States. All data came from clinical sites that had never been used previously for training or testing of the Saige-Dx AI algorithm.
  • Number of Experts for Ground Truth: "Truthed using similar procedures to those used for the reader study," which implies two highly experienced breast imaging specialists and a third adjudicator.
  • Qualifications of Experts for Ground Truth: Implied to be MQSA qualified, highly experienced (>10 years in practice) breast imaging specialists, consistent with the reader study.
  • Adjudication Method: Implied to be consistent with the reader study (third truther for discrepancies).
  • MRMC Comparative Effectiveness Study: No, this was a standalone performance study of the algorithm only.
  • Standalone Performance: Yes. Saige-Dx exhibited an AUC of 0.930 (95% CI: 0.902, 0.958).
  • Type of Ground Truth: Implied to be expert consensus with pathology confirmation, consistent with the reader study, as data was "collected and truthed using similar procedures."
  • Sample Size for Training Set: Not explicitly stated, but the data used was specifically excluded from the test set for this study, confirming separation.
  • How Ground Truth for Training Set was Established: Implied to be through expert review and pathology confirmation, given the "similar procedures" used for test set truthing and the isolation of training data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

DeepHealth, Inc. % B. Nathan Hunt VP, Quality Assurance and Regulatory Affairs 1000 Massachusetts Avenue CAMBRIDGE MA 01238

May 12, 2022

Re: K220105

Trade/Device Name: Saige-Dx Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological computer assisted detection and diagnosis software Regulatory Class: Class II Product Code: QDQ Dated: May 2, 2022 Received: May 4, 2022

Dear B. Nathan Hunt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

2

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Michael D. O'Hara. Ph.D. Deputy Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220105

Device Name Saige-Dx

Indications for Use (Describe)

Saige-Dx analyzes digital breast tomosynthesis (DBT) mammograms to identify the presence of soft tissue lesions and calcifications that may be indicative of cancer. For a given DBT mammogram, Saige-Dx analyzes the DBT image stacks and the accompanying 2D images, including full field digital mammography and/or synthetic images. The system assigns a Suspicion Level, indicating the strength of suspicion that cancer may be present, for each detected finding and for the entire case. The outputs of Saige-Dx are intended to be used as a concurrent reading aid for interpreting physicians on screening mammograms with compatible DBT hardware.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for DeepHealth. On the left side of the logo, there is a stack of three squares, each slightly offset from the one below it. To the right of the squares, the word "DeepHealth" is written in a combination of red and gray. The word "Deep" is in red, while the word "Health" is in gray.

1000 Massachusetts Ave Cambridge. MA 01238 Phone: 424-832-1480 www.deep.health

K220105 510(k) Summary DeepHealth, Inc. Saige-Dx

In accordance with 21 CFR 807.92 the following summary of information is provided, on this date, May 10, 2022:

1. 510(k) SUBMITTER

DeepHealth, Inc. 1000 Massachusetts Avenue Cambridge, MA 02138 Tel: 424-832-1480

Contact Person:

B. Nathan Hunt Vice President, Quality Assurance and Requlatory Affairs DeepHealth, Inc. 1000 Massachusetts Avenue Cambridge, MA 02138 Tel: 424-832-1480

Date Prepared:

May 10, 2022

2. DEVICE

Trade Name of Device: Saige-Dx

Common or Usual Name: Medical Image Software

Regulation Name and Number: Radiological Computer Assisted Detection and Diagnosis Software (21 CFR 892.2090)

Regulation Class: II

Product Code: QDQ

3. PREDICATE DEVICE

Trade Name: Transpara™

Common Name or Usual Name: Medical Image Software

Regulation Name and Number: Radiological Computer Assisted Detection and Diagnosis Software (21 CFR 892.2090)

Regulation Class: II

Product Code: QDQ

510(K) No .: K181704

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DeepHealth. The logo consists of a stack of three squares on the left, with the word "DeepHealth" to the right. The word "Deep" is in a dark red color, while the word "Health" is in gray.

1000 Massachusetts Ave. Cambridge. MA 01238 Phone: 424-832-1480 www.deep.health

4. DEVICE DESCRIPTION

Saige-Dx is a software device that processes screening mammograms using artificial intelligence to aid interpreting radiologists. By automatically detecting the presence or absence of soft tissue lesions and calcifications in mammography images, Saige-Dx can help improve reader performance, while also reducing time. The software takes as input a set of x-ray mammogram DICOM files from a single digital breast tomosynthesis (DBT) study and generates finding-level outputs for each image analyzed, as well as an aggregate case-level assessment. Saige-Dx processes both the DBT image stacks and the associated 2D images (full-field digital mammography (FFDM) and/or synthetic 2D images) in a DBT study. For each image, Saige-Dx outputs bounding boxes circumscribing any detected findings and assigns a Finding Suspicion Level to each finding, indicating the degree of suspicion that the finding is malignant. Saige-Dx uses the results of the finding-level analysis to generate a Case Suspicion Level, indicating the degree of suspicion for malignancy across the case. Saige-Dx encapsulates the finding and caselevel results into a DICOM Structured Report (SR) object containing markings that can be overlaid on the original mammogram images using a viewing workstation and a DICOM Secondary Capture (SC) object containing a summary report of the Saige-Dx results.

5. INDICATIONS FOR USE

Saige-Dx analyzes digital breast tomosynthesis (DBT) mammograms to identify the presence or absence of soft tissue lesions and calcifications that may be indicative of cancer. For a given DBT mammogram, Saige-Dx analyses the DBT image stacks and the accompanying 2D images, including full field digital mammography and/or synthetic images. The system assigns a Suspicion Level, indicating the strength of suspicion that cancer may be present, for each detected finding and for the entire case. The outputs of Saige-Dx are intended to be used as a concurrent reading aid for interpreting physicians on screening mammograms with compatible DBT hardware.

Intended User Population

The intended users of Saige-Dx are interpreting physicians qualified to read screening mammography exams.

Intended Patient Populations

The device is intended to be used on women thirty-five (35) years of age or older undergoing screening mammography.

Warnings and Precautions

Saige-Dx is an adjunct tool and is not intended to replace a physician's own review of a mammogram. Decisions should not be made solely based on analysis by Saige-Dx.

6. PREDICATE DEVICE COMPARISON

Saige-Dx and the predicate device have similar indications for use, patient population, technical characteristics, and principles of operation. The differences between Saige-Dx and the predicate device do not alter the suitability or of the subject device for its intended use.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for DeepHealth. The logo consists of a stack of three squares on the left, with the word "DeepHealth" on the right. The word "Deep" is in red, and the word "Health" is in gray. The squares are also red.

1000 Massachusetts Ave. Cambridge. MA 01238 Phone: 424-832-1480 www.deep.health

The devices are intended to be used by physicians to aid in the interpretation of screening mammograms. The devices are not intended to be used as a replacement for a full physician review or their own clinical judgement.

The design of Saige-Dx is similar to that of the predicate device. Both devices detect and characterize findings in radiological breast images and provide information regarding the presence and location of the findings to the user. As both devices use proprietary algorithms, there are assumed differences in the algorithmic components, as well as minor differences in the specific formats of the outputs provided to users.

Non-clinical and clinical testing has been completed ensuring that the differences do not affect the safety and effectiveness of the proposed subject device.

7. PERFORMANCE DATA

Saige-Dx is a software device and has been determined of Moderate Level of Concern. Verification testing included software unit testing, software integration testing, and system testing. Testing confirmed that the software, as designed and implemented, satisfies the software requirements.

Validation of the software was performed using two retrospective studies as described below. The data used in the validation testing was obtained from different clinical sites than that used for Saige-Dx Al algorithm training, DeepHealth ensured that there was no overlap between the data used to train and test the Saige-Dx Al algorithm. The data used to train the Saige-Dx algorithm consisted of six datasets across various geographic locations in the US and the UK, including diverse regions such as New York City.

Performance Testing: Reader Study

A fully balanced, multi-reader multi-center (MRMC) reader study was conducted to evaluate the performance of 18 MQSA qualified radiologists when reading a set of retrospectively collected DBT mammogram exams with and without the aid of Saige-Dx. The primary objective was to compare breast cancer detection performance of radiologists reading with the aid of Saige-Dx versus without Saige-Dx. Each radiologist read 240 cases twice in two reading sessions, once with and once without Saige-Dx, with a washout period of at least 4 weeks in between the two sessions. The mammograms were collected from unique female patients 35 years of age or older according to an IRB approved protocol and were acquired from Hologic equipment. The cases in the study included 100 pathology-proven cancer cases and 140 confirmed non-cancer cases. To provide a mix of cancer cases that might be encountered in clinical practice, 67 of the cancer cases were recalled in clinical practice, indicating the interpreting radiologist detected a suspicion lesion and 33 of the cancer cases were not recalled in clinical practice, indicating the interpreting radiologist did not detect anything suspicious and the cancer was diagnosed at a later time. The patients in the study represented a racially and ethnically diverse population. Testing was also performed on a subgroup representative of the racial distribution of a United States screening population, which also met performance goals.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for DeepHealth. The logo consists of a stack of three squares on the left, with the word "DeepHealth" on the right. The word "Deep" is in red, and the word "Health" is in gray. The squares are also red.

1000 Massachusetts Ave. Cambridge. MA 01238 Phone: 424-832-1480 www.deep.health

Each mammogram included in the study had one ground truth status: cancer or non-cancer, Two MQSA qualified, highly experienced (>10 years in practice) breast imaging specialists participated in establishing the reference standard for cancer and non-cancer exams. For each exam, the truthers confirmed and recorded the cancer status, as well as an interpretation of breast density. For cancer exams, the truthers annotated each malignant lesion on all views where the lesion was visible (based on the biopsied location that led to the malignant pathology) and indicated the lesion type. For exams where there were discrepancies between the two truther's assessment of density, lesion type, and/or lesion location, a third truther served as the adjudicator.

The reader performance increased with the aid of Saige-Dx from an average AUC of 0.865 when unaided to 0.925 when aided (difference of 0.06; 95% Cl: 0.041, 0.079, p < 0.00001). All 18 readers demonstrated an increase in AUC performance when using Saige-Dx. The average reader sensitivity increased by 8.8% (95% Cl: 7.0%, 10.6%). The average reader specificity increased by 0.9% (95% CI: -0.9%, 2.7%).

Similar reader performance trends were observed across breast densities, ages, racelethnicities, lesion types and sizes, and radiologist specialization. For lesion type, the average AUC increased from 0.866 to 0.918 for soft tissue densities and from 0.795 to 0.899 for calcifications. For radiologist specialization, the average AUC for breast imaging specialists increased from 0.885 to 0.931 and from 0.826 to 0.911 for generalists.

Performance Testing: Standalone Study

A retrospective, blinded, multi-center study was conducted to evaluate the standalone performance of Saige-Dx on DBT mammograms. A total of 1304 cases were collected from 9 clinical sites in the United States, consisting of 136 cancer and 1168 non-cancer cases. The data was collected and truthed using similar procedures to those used for the reader study. All data came from clinical sites that had never been used previously for training or testing of the Saige-Dx AI algorithm. Saige-Dx exhibited an AUC of 0.930 (95% CI: 0.902, 0.958) on the dataset, demonstrating strong performance relative to the unaided reader performance in the reader study. Similar standalone performance trends were observed across breast densities, ages, race/ethnicities, and lesion types and sizes. Saige-Dx was assessed on both recalled and nonrecalled cancers, as well as visible and non-visible cancers.

8. CONCLUSION

The non-clinical and clinical testing conducted to support this submission confirm that Saige-Dx is safe and effective. The minor differences, including technological differences, between Saige-Dx and the predicate do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. Therefore, the information presented in this 510(k) submission demonstrates that Saige-Dx is substantially equivalent to the predicate device.

§ 892.2090 Radiological computer-assisted detection and diagnosis software.

(a)
Identification. A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.