(57 days)
MR Elastography is an optional package for Canon Medical Systems magnetic resonance imaging (MRI) systems. This option allows the user to obtain an image that reflects the stiffness, in kPa, of body tissue such as liver and muscle.
MR Elastography is an optional package for Canon Medical Systems magnetic resonance imaging (MRI) system. This option allows the user to obtain an image that reflects the stiffness of body tissue.
MR Elastography consists of Acoustic Driver System including Active Driver, Passive Driver, QA Phantom and Other parts, TTL trigger converter includes AC adaptor, Manuals and Software (License).
By MRI scanning while applying vibration from the Acoustic Driver System, it is possible to obtain an image that reflects the stiffness of the body tissue in kPa. The sequence to be used with Acoustic Driver System is FE2D or SE-EPI2D. The MRI scanner outputs the following images from the acquired phase image and magnitude image.
- · Wave image
- · Stiffness image (capable of registration and display in color image format)
- · Confidence image (capable of superimposing confidence image on the stiffness image)
The provided text describes the 510(k) submission for Canon Medical Systems' MR Elastography (MZEK-001A) device. Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance (MZEK-001A) |
|---|---|
| Slope of Measured vs. Reference Stiffness Value (by linear regression analysis) | Within 1.0 ± 0.1 |
| 95% Confidence Limits of Measured vs. Reference Value (by Bland Altman analysis) | Within ± 10% |
2. Sample Size Used for the Test Set and Data Provenance
The text states: "The phantoms were scanned 5 times per day, multiple days per scanner, on 2 different 1.5T and 2 different 3T scanners by FE2D and SE-EPI2D techniques."
- Test Set Sample Size: While an exact number of "scans" isn't explicitly given beyond "5 times per day, multiple days per scanner," it indicates a sufficient number of measurements across different scanner types (2x 1.5T, 2x 3T) and scanning techniques (FE2D, SE-EPI2D). The total number of individual phantom measurements would be 5 (scans/day) * X (days) * 4 (scanners) * 2 (techniques). The document implies a robust bench testing setup rather than patient data.
- Data Provenance: The data appears to be from prospective bench testing conducted by Canon Medical Systems, likely in a controlled laboratory environment. The country of origin of the data is not specified, but given the manufacturer is Japan and the US agent is in the US, the testing could have occurred in either location or a combination. The data is synthetic (phantom-based) rather than clinical patient data.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Number of Experts: Not applicable/not mentioned. The ground truth was established using "Resoundant-provided calibration phantoms having known stiffness values." This indicates a physical phantom with pre-calibrated, known stiffness properties, rather than expert interpretation of images.
- Qualifications of Experts: Not applicable, as expert interpretation was not used for ground truth.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable/not mentioned. Since the ground truth was based on physical phantoms with known stiffness values, there was no need for expert adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported. The study focused on the device's ability to accurately measure stiffness against known phantom values, not on human reader performance or improvement with AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Standalone Performance: Yes, the described testing is akin to a standalone (algorithm only) performance evaluation. The device (hardware and software) outputs stiffness values based on the acquired data, and these values are compared directly against the known stiffness of the phantoms. There is no mention of human interpretation being part of this core performance evaluation. The output stiffness image and confidence image are for user display, but the performance metrics are system-derived.
7. The Type of Ground Truth Used
- Ground Truth Type: The ground truth used was phantom-based with known stiffness values. Specifically, "three Resoundant-provided calibration phantoms having known stiffness values representing the expected clinical range."
8. The Sample Size for the Training Set
- Training Set Sample Size: The document does not provide details on the sample size of any training set. It describes the performance testing of the final device. The software was "based on an estimation algorithm provided by the Mayo Clinic and Resoundant, Inc." and "developed in conjunction with The Mayo Clinic and Resoundant, Inc.," suggesting external development and/or collaboration, but no specifics on training data are given.
9. How the Ground Truth for the Training Set Was Established
- Training Set Ground Truth Establishment: The document does not provide details on how the ground truth for any training set was established. Given the software's origin from Mayo Clinic and Resoundant, Inc., it's plausible they used a combination of phantom data and potentially de-identified clinical data with histopathological correlation for development and training, but this is speculative based on common practice for such devices and not explicitly stated in the provided text.
{0}------------------------------------------------
March 3, 2022
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Canon Medical Systems Corporation % Mr. Paul Biggins Sr. Dir. Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive TUSTIN CA 92780
Re: K220018
Trade/Device Name: MR Elastography; MZEK-001A Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: December 29, 2021 Received: January 5, 2022
Dear Mr. Biggins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
- for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
MR Elastography; MZEK-001A
Indications for Use (Describe)
MR Elastography is an optional package for Canon Medical Systems magnetic resonance imaging (MRI) systems. This option allows the user to obtain an image that reflects the stiffness, in kPa, of body tissue such as liver and muscle.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92
1. CLASSIFICATION and DEVICE NAME
| Classification Name: | Magnetic Resonance Diagnostic Device |
|---|---|
| Regulation Number: | 90-LNH (Per 21 CFR § 892.1000) |
| Trade Proprietary Name: | MR Elastography |
| Model Number: | MZEK-001A |
2. SUBMITTER'S NAME
Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
3. OFFICIAL CORRESPONDENT
Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance Canon Medical Systems Corporation
4. CONTACT PERSON, U.S. AGENT and ADDRESS
- Contact Person/ Official Correspondent/U.S. Agent Paul Biggins Senior Director, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-7808 Fax: (714) 730-1310 E-mail: pbiggins@us.medical.canon
5. MANUFACTURING SITE
Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan
6. ESTABLISHMENT REGISTRATION 9614698
{4}------------------------------------------------
DICAL SYSTEMS USA, INC.
Made For life
7. DATE PREPARED
December 29, 2021
8. DEVICE NAME
MR Elastography, MZEK-001A
9. TRADE NAME
MR Elastography, MZEK-001A
10. Classification Name
Magnetic Resonance Diagnostic Device (MRDD)
11. CLASSIFICATION PANEL
Radiology
12. DEVICE CLASSIFICATION
Class II (per 21 CFR 892.1000, Magnetic Resonance Diagnostic Device)
13. PRODUCT CODE
90-LNH
14. PREDICATE DEVICE
Predicate Device:
- . MR Elastography (K140666)
Reference Devices:
- . Achieva, Intera, Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K193215)
- . MR Elastography, Resoundant (K201389)
| System | Subject Device | Predicate Device | Reference Device | Reference Device |
|---|---|---|---|---|
| MR Elastography | MR Elastography | Achieva, Intera, Ingenia,Ingenia CX, Ingenia Elition,and Ingenia Ambition MRSystems | MR Elastography | |
| Marketed By | Canon Medical Systems USA, Inc. | Philips | Philips | Resoundant |
| 510(k) Number | This Submission | K140666 | K193215(Referred to MR Elastographyextension) | K201389 |
| Clearance Date | June 13, 2014 | April 10, 2020 | July 10, 2020 |
15. REASON FOR SUBMISSION
Modification of a cleared device
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Canon Medical Systems USA, Inc. The word "Canon" is written in red, with the rest of the text in black. The text is simple and easy to read, with a focus on the company name.
Made For life
16. SUBMISSION TYPE
Traditional 510(k) Premarket Notification
17. DEVICE DESCRIPTION
MR Elastography is an optional package for Canon Medical Systems magnetic resonance imaging (MRI) system. This option allows the user to obtain an image that reflects the stiffness of body tissue.
MR Elastography consists of Acoustic Driver System including Active Driver, Passive Driver, QA Phantom and Other parts, TTL trigger converter includes AC adaptor, Manuals and Software (License).
By MRI scanning while applying vibration from the Acoustic Driver System, it is possible to obtain an image that reflects the stiffness of the body tissue in kPa. The sequence to be used with Acoustic Driver System is FE2D or SE-EPI2D. The MRI scanner outputs the following images from the acquired phase image and magnitude image.
- · Wave image
- · Stiffness image (capable of registration and display in color image format)
- · Confidence image (capable of superimposing confidence image on the stiffness image)
18. SUMMARY OF CHANGE(S)
This submission is to report the following functionalities have been added:
Summary of Hardware Changes:
- . MR Elastography:
MR Elastography consists of Acoustic Driver System including Active Driver, Passive Driver, QA Phantom and Other parts, TTL converter includes AC adaptor.
Summary of Software Changes:
● MR Elastography:
MR Elastography is a newly added optional package that allows the user to obtain an image that reflects the stiffness of body tissue.
By MRI scanning while applying vibration from the Acoustic Driver System, it is possible to obtain an image that reflects the stiffness of the body tissue in kPa. The sequence to be used with Acoustic Driver System is FE2D or SE-EPI2D. The MRI scanner outputs the following images from the acquired phase image and magnitude image.
- · Wave image
- · Stiffness image (capable of registration and display in color image format)
- · Confidence image (capable of superimposing confidence image on the stiffness image)
{6}------------------------------------------------
19. INDICATIONS FOR USE
MR Elastography is an optional package for Canon Medical Systems magnetic resonance imaging (MRI) systems. This option allows the user to obtain an image that reflects the stiffness, in kPa, of body tissue such as liver and muscle.
20. SUMMARY OF DESIGN CONTROL ACTIVITIES
Risk Management activities for new software functionality is included in this submission. A declaration of conformity with design controls is included in this submission.
21. SAFETY
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards.
This modification is based upon the same technologies, materials, and software as the target Canon MRI systems. Risk activities were conducted in concurrence with established medical device development standards and guidance.
22. TECHNOLOGICAL CHARACTERISTICS:
MR Elastography contains both hardware and software elements. The hardware elements are supplied by Resoundant, Inc. (K201389) and consist of an active driver positioned outside the magnet room and a passive driver positioned on the scan subject. The MR Elastography software postprocessing package was based on an estimation algorithm provided by the Mayo Clinic and Resoundant, Inc. and the final MR Elastography software was developed in conjunction with The Mayo Clinic and Resoundant, Inc. The Philips MR Elastography (K140666) and Achieva, Intera, Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K193215) contain the same hardware elements provided by Resoundant, Inc. and used the same estimation algorithm and worked in conjunction with the Mayo Clinic and Resoundant, Inc. to develop the final software. The GE IDEAL IQ SOFTWARE OPTION (K103411) and MR-TOUCH OPTION (K083421), and the SIEMENS SOFTWARE SYNGO MR D13A FOR THE MAGNETOM SYSTEMS AERA/SKYRA/AVANTO/VERIO (K121434) and MAGNETOM AERA WITH SOFTWARE SYNGO MR E11A, MAGNETOM SKYRA WITH SOFTWARE SYNGO MR E11A, MAGNETOM SKYRA WITH 24 RF CHANNEL (K141977) are as well.
The phase contrast imaging sequences to be used for the MR Elastography data process are substantially equivalent to the existing phase contrast imaging sequences which was previously cleared Vantage Orian 1.5T, MRT-1550, V7.0 (K211633, K203053) and Vantage Galan 3T, MRT-3020, V7.0 (K212056). The acquisition, processing and output images are substantially equivalent to the previously cleared predicate, Philips MR Elastography (K140666) and Achieva, Intera, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K193215). As with the above predicate, the MR Elastography option provides for a sensitive phase-contrast acquisition synchronized with the vibration supplied by the Resoundant hardware. This captures images of the tissue displacements, which are then input to an algorithm to estimate tissue stiffness.
{7}------------------------------------------------
23. TESTING
MR Elastography (MRE) performance testing (bench testing), using three Resoundant-provided calibration phantoms having known stiffness values representing the expected clinical range, was conducted to measure tissue stiffness, based on the 2D Multi-Modality Direct Inversion (MMDI) algorithm. The phantoms were scanned 5 times per day, multiple days per scanner, on 2 different 1.5T and 2 different 3T scanners by FE2D and SE-EPI2D techniques. Results of the study demonstrated that a calculated slope of measured stiffness mean value and reference stiffness value by linear regression analysis as well as a 95% confidence limit of measured vs. reference value, by Bland Altman analysis were within the acceptance criteria of 1.0±0.1 (slope) and 95% confidence limits within ±10%.
24. SUBSTANTIAL EQUIVALENCE
Canon Medical Systems Corporation believes that the MR Elastography is substantially equivalent to the previously cleared predicate device, MR Elastography (K140666), MR Elastography of Achieva, Intera, Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K193215), using with Canon MRI systems, referenced in this submission.
Canon Medical Systems Corporation believes that the MR Elastography option added to Canon MRI systems are substantially equivalent to the previously cleared predicate device.
25. CONCLUSION
The MR Elastography indications for use along with, technological characteristics, safety and performance testing, verifies that the subject device meets the requirements that are considered adequate for its intended use and is substantially equivalent in design, principles of operation and indications for use to the predicate and reference devices.
§ 892.1000 Magnetic resonance diagnostic device.
(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.