(230 days)
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci System, Model IS4000) and the Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System, Model 154200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
The da Vinci XXi (IS4200/IS4000) 8mm EndoWrist Instruments have a unique articulating design at their distal tips the human wrist. While seated at the Surgeon Console of the Surgical System, the surgeon can precisely control movements of the end effectors/instrument tips to perform one or more specific surgical tasks e.g., grasping, suturing, cutting, or tissue manipulation.
This document describes the 510(k) premarket notification for the da Vinci Xi/X (IS4200/IS4000) 8mm Reusable Instruments. The submission focuses on changes to the labeling regarding an increased number of lives (uses) and reprocessing cycles for these instruments.
Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state numerical acceptance criteria for each test in a table format, nor does it provide quantitative performance data for the subject device beyond stating that it "successfully met the acceptance criteria" or "meet the requirements." However, the general types of tests and their objectives serve as implicit acceptance criteria.
Type of Test | Implicit Acceptance Criteria (Objective) | Reported Device Performance |
---|---|---|
Cleaning Validation | Efficacy of manual and automated cleaning processes, meeting standards and guidance (FDA Guidance: "Reprocessing Medical Devices in Health Care Settings...", AAMI TIR 12:2020, AAMI TIR 30:2011/(R)2016, ANSI/AAMI ST15883-1:2009/(R)2014) | "The da Vinci X/Xi 8mm Reusable Instruments successfully met the acceptance criteria for all markers. The test results demonstrate that the da Vinci X/Xi 8mm Reusable Instruments can be cleaned using the following cleaning methods: Automated cleaning process using a compatible automated washer/disinfector. Manual cleaning process using an ultrasonic bath." |
Reliability/Life Testing | Instruments are not adversely affected by the increased number of lives (uses) and reprocessing cycles specified in Table 1. | "Reliability/Life Testing was performed to ensure that da Vinci X/Xi 8mm Reusable Instruments are not adversely affected by the increased number of lives (uses) and reprocessing cycles for these as listed in Table 1." (It implicitly states they did ensure this by performance of the test, and the conclusion states substantial equivalence, meaning criteria were met). |
Electrical Performance Testing | Meet requirements within the FDA guidance, "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery," issued March 9, 2020, after multiple reuse and reprocessing cycles. | "This testing was performed to ensure the da Vinci X/Xi 8mm Extended Lives Instruments family (subject devices as listed in Table 1) meet the requirements within the FDA guidance, Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, Document Issued on March 9, 2020." (Implies requirements were met). |
Thermal Effects Testing | Thermal effects on tissue are comparable between the subject and predicate devices, in accordance with the FDA guidance, "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery," issued March 9, 2020. | "Thermal Effects testing was performed to confirm that thermal effects on tissue are comparable between the subject and predicate devices. The testing was performed in accordance with the FDA guidance, 'Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery', Document Issued on March 9, 2020." (Implies comparability was confirmed and guidance met). |
2. Sample sizes used for the test set and the data provenance
The document does not specify exact sample sizes (e.g., number of instruments tested) for any of the performance tests (Cleaning Validation, Reliability/Life Testing, Electrical Performance Testing, Thermal Effects Testing). The data provenance (e.g., country of origin of the data, retrospective or prospective) is also not mentioned, but the tests described are typically prospective, laboratory-based engineering and validation tests conducted by the manufacturer.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided. The tests described are engineering and laboratory performance tests, not clinical studies involving expert interpretation of medical data (like imaging). The "ground truth" for these types of tests is generally defined by calibrated measurements against established standards and specifications, not expert consensus.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. The described tests are objective performance evaluations against predefined standards, not a review of clinical outcomes or imaging by experts.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document pertains to the reusability and safety of surgical instruments, not an AI-assisted diagnostic or interpretative device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This document is about surgical instruments, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the performance tests:
- Cleaning Validation: The ground truth is the absence of biological contaminants and residues, measured against specific acceptance criteria defined by industry standards (AAMI TIR 12, AAMI TIR 30, ANSI/AAMI ST15883-1) and FDA guidance for reprocessing.
- Reliability/Life Testing: The ground truth is the continued functionality and integrity of the instruments after the increased number of uses and reprocessing cycles, as compared to design specifications.
- Electrical Performance Testing: The ground truth is adherence to electrical safety and performance parameters outlined in FDA guidance for electrosurgical devices.
- Thermal Effects Testing: The ground truth is comparability of thermal effects to the predicate device, consistent with FDA guidance for electrosurgical devices.
8. The sample size for the training set
Not applicable. There is no mention of a "training set" as this device is not an AI/ML algorithm.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for an AI/ML algorithm.
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.