(373 days)
The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semiquantitative determination of the presence of tricyclic antidepressants (TCAs) in human serum, plasma, or urine of patients at a cutoff concentration of 300 ng/mL in patients suspected of drug overdose.
The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical test result. A more specific alternative chemical must be used to obtain a confirmed analytical result. Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method.
Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
For In Vitro Diagnostic Use Only.
The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay using ready to-use liquid reagents. Specific tricyclic antibodies were used to detect most tricyclic antidepressants in serum, plasma, or urine. The test is based on the competition of an enzyme, glucose-6-phosphate dehydrogenase (G6PDH), labeled-drug and the drug from the sample for a fixed amount of specific antibody binding sites. In the absence of the drug from the sample, the specific antibody binds the enzyme-labeled drug and the enzyme activity is inhibited. This phenomenon creates a direct relationship between drug concentration in the sample and the enzyme activity. The enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.
The DRI™ Tricyclics Serum Tox assay is supplied as a two liquid reagent kit (Reagent A and Reagent E). Thev are bottled separately within the same kit, see details below:
- Antibody/Substrate Reagent (Reagent A): Contains polyclonal anti-tricyclics . antibodies (sheep), glucose-6- phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in Tris buffer with sodium azide as a preservative.
- Enzyme Conjugate Reagent (Reagent E): Contains glucose-6-phosphate ● dehydrogenase (G6PDH) labeled with nortriptyline in Tris buffer with sodium azide as a preservative.
The provided document describes the analytical performance of the DRI™ Tricyclics Serum Tox Assay. It details various studies conducted to demonstrate its performance characteristics, but it does not explicitly state specific pass/fail acceptance criteria for each test. Instead, it presents the results of each study, implying that these results met the internal criteria used by the manufacturer for demonstrating sufficient performance and substantial equivalence to the predicate device.
Here's an analysis of the provided information, addressing your points where possible, and noting when the information is unavailable:
1. A table of acceptance criteria and the reported device performance
As mentioned, explicit acceptance criteria (e.g., "Accuracy must be >95%") are not stated in this document. The tables below summarize the reported device performance for several key analytical studies. The implication is that these reported performances were deemed acceptable for the 510(k) clearance.
| Study Type | Reported Device Performance (Serum) | Reported Device Performance (Urine) |
|---|---|---|
| Precision (Repeatability) | Qualitative Mode (n=80 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 80/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 10/70 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/80 (Negative/Positive)Semi-Quantitative Mode (n=80 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 80/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 19/61 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/80 (Negative/Positive) | Qualitative Mode (n=80 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 80/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 68/12 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/80 (Negative/Positive)Semi-Quantitative Mode (n=80 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 80/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 73/7 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/80 (Negative/Positive) |
| Precision (Reproducibility) | Qualitative Mode (n=75 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 75/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 30/45 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/75 (Negative/Positive)Semi-Quantitative Mode (n=75 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 75/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 29/46 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/75 (Negative/Positive) | Qualitative Mode (n=75 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 75/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 61/14 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/75 (Negative/Positive)Semi-Quantitative Mode (n=75 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 75/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 66/9 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/75 (Negative/Positive) |
| Spike Recovery | Low Control (225 ng/mL, n=20): All 20 Negative (Below Cutoff)High Control (375 ng/mL, n=20): All 20 Positive (Above Cutoff) | Low Control (225 ng/mL, n=20): All 20 Negative (Below Cutoff)High Control (375 ng/mL, n=20): All 20 Positive (Above Cutoff) |
| Dilution Linearity (Recovery) | Range 96-115% for expected values 125 ng/mL to 1000 ng/mL (average of 5 replicates) | Range 94-118% for expected values 125 ng/mL to 1000 ng/mL (average of 5 replicates) |
| Method Comparison (Accuracy) | Qualitative Mode (n=117):- % Negative Sample Agreement: 98% (60/61)- % Positive Sample Agreement: 100% (56/56)- % Total Sample Agreement: 99% (116/117)Semi-Quantitative Mode (n=117):- % Negative Sample Agreement: 97% (59/61)- % Positive Sample Agreement: 100% (56/56)- % Total Sample Agreement: 98% (115/117) | Qualitative Mode (n=100):- % Negative Sample Agreement: 96% (48/50)- % Positive Sample Agreement: 98% (49/50)- % Total Sample Agreement: 97% (97/100)Semi-Quantitative Mode (n=100):- % Negative Sample Agreement: 96% (48/50)- % Positive Sample Agreement: 98% (49/50)- % Total Sample Agreement: 97% (97/100) |
| Matrix Equivalency | Serum: 50 patient samples total (25 positive, 25 negative). Full agreement for K2 EDTA Plasma, K3 EDTA Plasma, Lithium Heparin Plasma, Sodium Citrate Plasma, Potassium Oxalate Plasma across qualitative and semi-quantitative modes.Sodium Heparin Plasma: 45 patient samples (25 positive, 20 negative). Full agreement across qualitative and semi-quantitative modes. | Not applicable (serum comparison) |
| Interference | Spiking endogenous/exogenous/physiological substances at high concentrations into low and high controls (225 ng/mL and 375 ng/mL) showed accurate detection of controls (negative for low, positive for high), indicating no interference. (See Table 20 for list of compounds and concentrations tested). | Spiking endogenous/exogenous/physiological substances and pH variations (pH 3-11) at high concentrations into low and high controls (225 ng/mL and 375 ng/mL) showed accurate detection of controls (negative for low, positive for high), indicating no interference. (See Table 21 for list of compounds and concentrations tested). Specific gravity (1.004-1.030) also showed no interference. |
| Specificity (Cross-Reactivity) | Extensive tables (Table 16, 18, 20) are provided listing cross-reactivity percentages for numerous structurally related and unrelated compounds in serum. The study demonstrated cross-reactivity for various TCA drugs and metabolites (e.g., Amitriptyline 100%, Desipramine 120%, Imipramine 157.9%). Minimal to no cross-reactivity (<0.3% to low single digits) was observed for most structurally unrelated compounds at high concentrations, except for a few exceptions like Chlorpromazine (57.1%), Cyclobenzaprine (66.7%), and Perphenazine (66.7%). | Extensive tables (Table 17, 19, 21) are provided listing cross-reactivity percentages for numerous structurally related and unrelated compounds in urine. Similar to serum, cross-reactivity was observed for various TCA drugs and metabolites (e.g., Amitriptyline 100%, Desipramine 120%, Imipramine 136.4%). Minimal to no cross-reactivity (<0.3% to low single digits) was observed for most structurally unrelated compounds at high concentrations, except for a few exceptions like Chlorpromazine (50%), Cyclobenzaprine (60%), and Perphenazine (46.2%). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Precision (Repeatability): 80 replicates per concentration level for serum and urine.
- Precision (Reproducibility): 75 replicates per concentration level for serum and urine (across 3 different instruments, 2 reagent lots).
- Spike Recovery: 20 replicates for low and high controls for both serum and urine.
- Dilution Linearity: 9 levels, each with 5 replicates, for both serum and urine.
- Method Comparison and Accuracy:
- Serum: 61 negative patient samples + 56 positive patient samples = 117 total patient samples.
- Urine: 50 negative patient samples + 50 positive patient samples = 100 total patient samples.
- Matrix Equivalency: 50 patient samples for each of 5 plasma types (K2 EDTA, K3 EDTA, Lithium Heparin, Sodium Citrate, Potassium Oxalate), and 45 patient samples for Sodium Heparin plasma. Each run with 2 replicates.
- Specificity (Cross-Reactivity) & Interference: Tested with known spiked concentrations in drug-free serum/urine and low/high controls. The exact number of individual samples/replicates for each substance is not explicitly stated beyond what is in the tables (e.g., "known amount," "single determinations" implied for cross-reactivity at a given concentration).
- Data Provenance: The document does not specify the country of origin of the data or whether the patient samples were collected retrospectively or prospectively.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This device is an in vitro diagnostic (IVD) assay for detecting tricyclic antidepressants. The ground truth for the test set (method comparison study) was established by comparison to Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) lab testing. This is a highly accurate analytical chemical method, considered the "preferred confirmatory method" by the device's indications for use. Therefore, human experts were not used to establish the primary ground truth in the way they might be for image-based diagnostics. The LC-MS/MS results serve as the objective reference standard.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as the ground truth was based on a reference chemical analytical method (LC-MS/MS), not on human expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an IVD assay, not an AI-based diagnostic tool requiring human interpretation or a multi-reader study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the studies presented are standalone performance studies of the device (the assay) itself. There is no human-in-the-loop component for the analytical performance evaluations described. The device provides a preliminary analytical test result independently.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for all analytical performance studies (especially method comparison) was Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS), which is a gold-standard chemical analytical method.
8. The sample size for the training set
This document describes the analytical validation of an enzyme immunoassay kit, not a machine learning algorithm. Therefore, there is no "training set" in the context of AI/ML. The samples used for various analytical performance studies (precision, spike recovery, linearity, method comparison, etc.) are implicitly part of the validation process, not for "training" an algorithm.
9. How the ground truth for the training set was established
As there is no training set for a machine learning algorithm, this question is not applicable. The assay's performance is based on its biochemical interaction, not on learned patterns from a training dataset.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
December 21, 2022
Microgenics Corporation Pranjali Shinde Senior Regulatory Affairs Specialist 46500 Kato Road Fremont, CA 94538
Re: K213875
Trade/Device Name: DRI™ Tricyclics Serum Tox Assay Regulation Number: 21 CFR 862.3910 Regulation Name: Tricyclic antidepressant drugs test system Regulatory Class: Class II Product Code: LFH Dated: August 19, 2022 Received: August 23, 2022
Dear Pranjali Shinde:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
- for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Paula --Digitally signed by Caposino - Date: 2022.12.21 Paula Caposino -S 11:51:25 -05'00'
Paula Caposino, Ph.D. Acting Deputy Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213875
Device Name DRI™ Tricyclics Serum Tox Assay
Indications for Use (Describe)
The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semiquantitative determination of the presence of tricyclic antidepressants (TCAs) in human serum, plasma, or urine of patients at a cutoff concentration of 300 ng/mL in patients suspected of drug overdose.
The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical test result. A more specific alternative chemical must be used to obtain a confirmed analytical result. Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method.
Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
For In Vitro Diagnostic Use Only.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
510(k) Number: K213875
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of Safe Medical Device Act of 1990 and 21 CFR 807.92.
A. Device Information
| Sponsor: | Microgenics CorporationThermo Fisher Scientific46500 Kato RoadFremont, CA 94538Phone: 510-979-5000FAX: 510-979-5002 |
|---|---|
| Correspondent ContactInformation: | Pranjali ShindeMicrogenics CorporationThermo Fisher Scientific46500 Kato RoadFremont, CA-94538Email: Pranjali.Shinde@Thermofisher.comPhone: 213-344-9952Fax: 510-979-5002 |
| Device Common Name: | Tricyclics Serum Tox Assay |
| Trade or Proprietary Name | DRI™ Tricyclics Serum Tox Assay |
| Brand Name | DRI™ Tricyclics Serum Tox Assay |
| Candidate Device ProductCode, Classification,Classification Name & Panel | LFH, Class II,21 CFR 862. 3910 – Tricyclic antidepressant drugs testsystem, 91 - Toxicology |
Predicate Device Information
| Predicate Device: | Tricyclic Serum Tox EIA Assay |
|---|---|
| Predicate Device Manufacturer: | Microgenics Corporation |
| Predicate Device Common Name | Tricyclic Serum Tox Assay |
| Predicate Device Premarket Notification #: | K983268 |
| Predicate Device Product Code, Classification, Classification Name & Panel | LFH, Class II,21 CFR 862. 3910 – Tricyclic antidepressant drugs testsystem, 91 - Toxicology |
B. Date Summary Prepared
November 14, 2022
{4}------------------------------------------------
C. Description of Device
The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay using ready to-use liquid reagents. Specific tricyclic antibodies were used to detect most tricyclic antidepressants in serum, plasma, or urine. The test is based on the competition of an enzyme, glucose-6-phosphate dehydrogenase (G6PDH), labeled-drug and the drug from the sample for a fixed amount of specific antibody binding sites. In the absence of the drug from the sample, the specific antibody binds the enzyme-labeled drug and the enzyme activity is inhibited. This phenomenon creates a direct relationship between drug concentration in the sample and the enzyme activity. The enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.
The DRI™ Tricyclics Serum Tox assay is supplied as a two liquid reagent kit (Reagent A and Reagent E). Thev are bottled separately within the same kit, see details below:
- Antibody/Substrate Reagent (Reagent A): Contains polyclonal anti-tricyclics . antibodies (sheep), glucose-6- phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in Tris buffer with sodium azide as a preservative.
- Enzyme Conjugate Reagent (Reagent E): Contains glucose-6-phosphate ● dehydrogenase (G6PDH) labeled with nortriptyline in Tris buffer with sodium azide as a preservative.
CI. Intended Use
The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semiquantitative determination of the presence of tricyclic antidepressants (TCAs) in human serum, plasma, or urine of patients at a cutoff concentration of 300 ng/mL in patients suspected of drug overdose.
The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method.
Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
For In Vitro Diagnostic Use Only.
{5}------------------------------------------------
E. Comparison to Predicate Device
| Characteristics | Candidate Device:DRI™ Tricyclics Serum Tox Assay(K213875) | Predicate Device:Tricyclic Serum Tox EIA Assay(K983268) |
|---|---|---|
| Intended Use | The DRI™ Tricyclics Serum ToxAssay is a homogeneous enzymeimmunoassay intended for thequalitative and/orsemiquantitative determination ofthe presence of tricyclicantidepressants (TCAs) in humanserum, plasma, or urine of patientsat a cutoff concentration of 300ng/mL in patients suspected ofdrug overdose.The semi-quantitative mode is forthe purpose of enablinglaboratories to determine anappropriate dilution of thespecimen for confirmation by aconfirmatory method such asLiquid Chromatography/TandemMass Spectrometry (LC-MS/MS)or permitting laboratories toestablish quality controlprocedures.The assay provides only apreliminary analytical test result.A more specific alternativechemical method must be used toobtain a confirmed analyticalresult. LiquidChromatography/Tandem MassSpectrometry (LC-MS/MS) is thepreferred confirmatory method.Clinical and professionaljudgment should be applied to anydrug of abuse test result | Same |
| particularly when preliminarypositive results are used. | ||
| For In Vitro Diagnostic Use Only. | ||
| OperatingPrinciple(Technology) | DRI | Same |
| Measured Analyte | Nortriptyline | Same |
| Test Matrix | Serum, Plasma, Urine | Same |
| Cut-off Levels | 300 ng/mL | Same |
| Methodology | Homogeneous EnzymeImmunoassay | Same |
| Reagents Form | Liquid ready-to-use. | Same |
| Antibody | Sheep polyclonal antibodies | Monoclonal antibodies |
| Storage | 2-8 °C until expiration date. | Same |
| Principal Operator | Trained professionals | Same |
| Calibrator Levelsfor Semi-Quant | 5-point Calibrator | 5-point Calibrator |
{6}------------------------------------------------
F. Test Principle
The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay using ready-to use liquid reagents. The assay uses specific tricyclic antibodies, which can detect most tricyclic antidepressants in serum, plasma, or urine. The DRI technology is based on the competition of an enzyme glucose-6-phosphate dehydrogenase (G6PDH) labeled-drug and the drug from the serum, plasma, or urine sample for a fixed amount of specific antibody binding sites. In the presence of free drug from the sample, the free drug occupies the antibody binding sites, allowing the drug-labeled G6PDH to interact with the substrate, resulting in enzyme activity. In the absence of drug from the sample, the specific antibody binds to the drug labeled with G6PDH and the enzyme activity is inhibited. This phenomenon creates a direct relationship between the drug concentration in the serum, plasma, or urine and the enzyme activity. The enzyme G6PDH activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.
{7}------------------------------------------------
G. Summary of Supporting Data
Analytical Performance:
The performance was evaluated on Architect c4000 clinical chemistry analyzer.
- a) Precision
Precision studies were performed in accordance with CLSI Guideline EP05-A3.
Samples were prepared by spiking Nortriptyline in both drug-free serum and drug-free urine at the cutoff, 25%, 50%, 75%, and 100% above and below the cutoff.
Repeatability:
The study was conducted for 20 days with 2 runs per day, 2 replicates per run in both qualitative and semi-quantitative modes with 1 lot of reagent, calibrators, and controls. There were 80 replicates for each level of precision samples. Serum and urine samples were run separately. The results of the precision study (Repeatability) are presented in the table below.
Table 1: Precision study data (Repeatability) in qualitative and semi-quantitative mode for 300 ng/mL cutoff - Serum
| SpikedConcentration(ng/mL) | % ofCutoff | # ofDeterminants | Total Precision (n=80) | |
|---|---|---|---|---|
| QualitativeImmunoassay Results(Negative/Positive) | Semi-QuantitativeImmunoassay Results(Negative/Positive) | |||
| 0 | -100 | 80 | 80/0 | 80/0 |
| 75 | -75 | 80 | 80/0 | 80/0 |
| 150 | -50 | 80 | 80/0 | 80/0 |
| 225 | -25 | 80 | 80/0 | 80/0 |
| 300 | 100 | 80 | 10/70 | 19/61 |
| 375 | +25 | 80 | 0/80 | 0/80 |
| 450 | +50 | 80 | 0/80 | 0/80 |
| 525 | +75 | 80 | 0/80 | 0/80 |
| 600 | +100 | 80 | 0/80 | 0/80 |
{8}------------------------------------------------
| SpikedConcentration(ng/mL) | % ofCutoff | # ofDeterminants | Total Precision (n=80) | |
|---|---|---|---|---|
| QualitativeImmunoassay Results(Negative/Positive) | Semi-QuantitativeImmunoassay Results(Negative/Positive) | |||
| 0 | -100 | 80 | 80/0 | 80/0 |
| 75 | -75 | 80 | 80/0 | 80/0 |
| 150 | -50 | 80 | 80/0 | 80/0 |
| 225 | -25 | 80 | 80/0 | 80/0 |
| 300 | 100 | 80 | 68/12 | 73/7 |
| 375 | +25 | 80 | 0/80 | 0/80 |
| 450 | +50 | 80 | 0/80 | 0/80 |
| 525 | +75 | 80 | 0/80 | 0/80 |
| 600 | +100 | 80 | 0/80 | 0/80 |
Table 2: Precision study data (Repeatability) in qualitative and semi-quantitative mode for 300 ng/mL cutoff = Urine
Reproducibility:
The study was conducted for 5 days with 1 run per day, 5 replicates per run in both qualitative and semi-quantitative modes with 2 lots of reagents on 3 different instruments. Serum and urine samples were run separately. The results of the precision study (Reproducibility) are presented in the table below.
Table 3: Precision study data (Reproducibility) in qualitative and semi-quantitative mode for 300 ng/mL cutoff - Serum
| SpikedConcentration(ng/mL) | % ofCutoff | # ofDeterminants | Total Precision (n=75) | |
|---|---|---|---|---|
| QualitativeImmunoassay Results(Negative/Positive) | Semi-quantitativeImmunoassay Results(Negative/Positive) | |||
| 0 | -100 | 75 | 75/0 | 75/0 |
| 75 | -75 | 75 | 75/0 | 75/0 |
| 150 | -50 | 75 | 75/0 | 75/0 |
| 225 | -25 | 75 | 75/0 | 75/0 |
| 300 | 100 | 75 | 30/45 | 29/46 |
| 375 | +25 | 75 | 0/75 | 0/75 |
| 450 | +50 | 75 | 0/75 | 0/75 |
| 525 | +75 | 75 | 0/75 | 0/75 |
| 600 | +100 | 75 | 0/75 | 0/75 |
{9}------------------------------------------------
Table 4: Precision study data (Reproducibility) in qualitative and semi-quantitative mode for 300 ng/mL cutoff - Urine
| SpikedConcentration(ng/mL) | % ofCutoff | # ofDeterminants | Total Precision (n=75) | |
|---|---|---|---|---|
| QualitativeImmunoassay Results(Negative/Positive) | Semi-quantitativeImmunoassay Results(Negative/Positive) | |||
| 0 | -100 | 75 | 75/0 | 75/0 |
| 75 | -75 | 75 | 75/0 | 75/0 |
| 150 | -50 | 75 | 75/0 | 75/0 |
| 225 | -25 | 75 | 75/0 | 75/0 |
| 300 | 100 | 75 | 61/14 | 66/9 |
| 375 | +25 | 75 | 0/75 | 0/75 |
| 450 | +50 | 75 | 0/75 | 0/75 |
| 525 | +75 | 75 | 0/75 | 0/75 |
| 600 | +100 | 75 | 0/75 | 0/75 |
{10}------------------------------------------------
- b) Spike Recovery
The spike recovery study was performed using 20 replicates. Samples were prepared by spiking Nortriptyline in both drug-free serum and drug-free urine at 225 ng/mL, 300 ng/mL, 375 ng/mL. The study demonstrated the accuracy of spike recovery at low control (225 ng/mL) and high control (375 ng/mL) in both qualitative and semiquantitative mode. Serum and urine samples were run separately. The results of the study are presented in the table below.
| Table 5: Spike Recovery in qualitative and semi-quantitative mode for 300 ng/mL cutoff - | |||
|---|---|---|---|
| Serum |
| Replicates | Low Control: 225 ng/mL(-25% of cutoff) (n=20) | High Control: 375 ng/mL(+25% of cutoff)(n=20) |
|---|---|---|
| 1 | Negative | Positive |
| 2 | Negative | Positive |
| 3 | Negative | Positive |
| 4 | Negative | Positive |
| 5 | Negative | Positive |
| 6 | Negative | Positive |
| 7 | Negative | Positive |
| 8 | Negative | Positive |
| 9 | Negative | Positive |
| 10 | Negative | Positive |
| 11 | Negative | Positive |
| 12 | Negative | Positive |
| 13 | Negative | Positive |
| 14 | Negative | Positive |
| 15 | Negative | Positive |
| 16 | Negative | Positive |
| 17 | Negative | Positive |
| 18 | Negative | Positive |
| 19 | Negative | Positive |
| 20 | Negative | Positive |
| Overlap | No | No |
| Relative to C/O | All 20 below C/O | All 20 above C/O |
Table 6: Recovery in qualitative and semi-quantitative mode for 300 ng/mL cutoff -Urine
| Replicates | Low Control: 225 ng/mL-25% of cutoff (n=20) | High Control: 375 ng/mL+25% of cutoff (n=20) |
|---|---|---|
| 1 | Negative | Positive |
| 2 | Negative | Positive |
| 3 | Negative | Positive |
| 4 | Negative | Positive |
{11}------------------------------------------------
| Replicates | Low Control: 225 ng/mL-25% of cutoff (n=20) | High Control: 375 ng/mL+25% of cutoff (n=20) |
|---|---|---|
| 5 | Negative | Positive |
| 6 | Negative | Positive |
| 7 | Negative | Positive |
| 8 | Negative | Positive |
| 9 | Negative | Positive |
| 10 | Negative | Positive |
| 11 | Negative | Positive |
| 12 | Negative | Positive |
| 13 | Negative | Positive |
| 14 | Negative | Positive |
| 15 | Negative | Positive |
| 16 | Negative | Positive |
| 17 | Negative | Positive |
| 18 | Negative | Positive |
| 19 | Negative | Positive |
| 20 | Negative | Positive |
| Overlap | No | No |
| Relative to C/O | All 20 below C/O | All 20 above C/O |
{12}------------------------------------------------
c) Dilution Linearity
The Dilution Linearity study is performed using the CLSI guideline CLSI EP06-A.
To demonstrate the dilution linearity for purposes of sample dilution and quality control of the entire assay range, drug-free serum or drug-free urine were spiked using Nortriptyline and diluted with drug-free serum or drug-free urine, respectively to generate 9 levels between 0 and 1000 ng/mL. Each sample was run in replicates of 5 in semi-quantitative mode and the average was used to determine percent recovery compared to the expected target value. Serum and urine samples were run separately. The study results are presented in the table below.
| Level# | Expected NortriptylineValues (ng/mL) | Serum | |
|---|---|---|---|
| Observed Values (ng/mL) | Recovery (%) | ||
| 1 | 0 | 7 | N/A |
| 2 | 125 | 121 | 97 |
| 3 | 250 | 242 | 97 |
| 4 | 375 | 385 | 103 |
| 5 | 500 | 479 | 96 |
| 6 | 625 | 720 | 115 |
| 7 | 750 | 839 | 112 |
| 8 | 875 | 913 | 104 |
| 9 | 1000 | 1046 | 105 |
Table 7: Dilution linearity data - Serum
| Level# | Expected NortriptylineValues (ng/mL) | Urine | |
|---|---|---|---|
| Observed Values (ng/mL) | Recovery (%) | ||
| 1 | 0 | 8 | N/A |
| 2 | 125 | 148 | 118 |
| 3 | 250 | 271 | 108 |
| 4 | 375 | 393 | 105 |
| 5 | 500 | 470 | 94 |
| 6 | 625 | 620 | 99 |
| 7 | 750 | 741 | 99 |
| 8 | 875 | 838 | 96 |
| 9 | 1000 | 946 | 95 |
Table 8: Dilution linearity data – Urine
{13}------------------------------------------------
- d) Method Comparison and Accuracy
The method comparison study was performed in accordance with CLSI guideline CLSI EP09c-A3.
One (1) replicate of each of the 61 negative patient samples for serum and 50 negative samples for urine and 56 positive patient samples for serum and 50 positive samples for urine were analyzed in both qualitative and semi-quantitative modes. Serum and urine samples were run separately. The results were compared to LC-MS/MS lab testing.
Serum and urine samples were run separately. The results of the study are presented in the tables below.
| Table 9: Stratified data comparing immunoassay in qualitative mode with LC-MS/MS – | ||||||
|---|---|---|---|---|---|---|
| Serum | ||||||
| Near Cutoff | ||||||
| DRITCA STAssayResults | Negativeby LC-MS/MS | < 50% ofCutoffconcentrationby LC-MS/MS (<150 ng/mL) | Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS)(150-299 ng/mL) | Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450ng/mL) | HighPositives(Greater than50% abovecutoffconcentration(> 450ng/mL) | |
| Positive | 0 | 11 | 0 | 31 | 25 | |
| Negative | 1 | 35 | 24 | 0 | 0 | |
| % Negative Sample Agreement | 98% or (60/61) | |||||
| % Positive Sample Agreement | 100% or (56/56) | |||||
| % Total Sample Agreement | 99% or (116/117) |
1 Refer to Table 11 for discordant samples of serum.
Table 10: Stratified data comparing immunoassay in semi-quantitative mode with LC MS/MS -
| Serum | |||||
|---|---|---|---|---|---|
| DRITCA STAssayResults | Negativeby LC-MS/MS | < 50% ofCutoffconcentrationby LC-MS/MS (<150 ng/mL) | Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS)(150-299ng/mL) | Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450ng/mL) | HighPositives(Greater than50% abovecutoffconcentration(> 450ng/mL) |
| Positive | 0 | 21 | 0 | 31 | 25 |
| Negative | 1 | 34 | 24 | 0 | 0 |
| % Negative Sample Agreement | 97% or (59/61) |
{14}------------------------------------------------
| % Positive Sample Agreement | 100% or (56/56) |
|---|---|
| % Total Sample Agreement | 98% or (115/117) |
1 Refer to Table 11 for discordant samples of serum.
| Sample ID | Qualitative(mA/min)Result | Semi-QuantitativeObserved concentration(ng/mL) | Final LC-MS/MSValue(ng/mL) |
|---|---|---|---|
| APP9489-22 | Negative | 354 (Positive) | 110 |
| APP9474-23 | Positive | 418 (Positive) | 120 |
| 1Table 11: Discordant Samples - Serum |
|---|
| --------------------------------------- |
2 Sample APP9489-2 contains Amitriptyline at 8.08 ng/mL, Imipramine at 18.35 ng/mL, Desipramine at 46.35 ng/mL, and Nortriptyline at 17 ng/mL by LC-MS/MS and cross-reacts at 100%, 158%, 120%, and 100% by immunoassay, respectively.
3 Sample APP9474-2 contains Amitriptyline at 9.08 ng/mL, Imipramine at 19.9 ng/mL, Desipramine at 49.8 ng/mL, and Nortriptyline at 19.65 ng/mL by LC-MS/MS and cross-reacts at 100%, 158%, 120%, and 100% by immunoassay, respectively.
Note: In addition, both samples APP9489-2 and APP9474-2 were confirmed by LC-MS/MS to contain Carbamazepine and Carbamazepine Epoxide with concentrations (4275 ng/mL & 5285 ng/mL) and (595 ng/mL & 955 ng/mL), respectively.
| Urine | |||||
|---|---|---|---|---|---|
| DRITCA STAssayResults | Negativeby LC-MS/MS | < 50% ofCutoffconcentrationby LC-MS/MS (<150 ng/mL) | Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS)(150-299 ng/mL) | Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450ng/mL) | HighPositives(Greater than50% abovecutoffconcentration(> 450ng/mL) |
| Positive | 0 | 0 | 24 | 7 | 42 |
| Negative | 22 | 11 | 15 | 14 | 0 |
| % Negative Sample Agreement | 96% or (48/50) | ||||
| % Positive Sample Agreement | 98% or (49/50) | ||||
| % Total Sample Agreement | 97% or (97/100) |
Table 12: Stratified data comparing immunoassay in qualitative mode with LC-MS/MS -I Irine
4 Refer to Table 14 for discordant samples of urine.
{15}------------------------------------------------
| DRITCA STAssayResults | Negativeby LC-MS/MS | < 50% ofCutoffconcentrationby LC-MS/MS (<150 ng/mL) | Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS) (150-299 ng/mL) | Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450ng/mL) | HighPositives(Greater than50% abovecutoffconcentration(> 450ng/mL) |
|---|---|---|---|---|---|
| Positive | 0 | 0 | 24 | 7 | 42 |
| Negative | 22 | 11 | 15 | 14 | 0 |
| % Negative Sample Agreement | 96% or (48/50) | ||||
| % Positive Sample Agreement | 98% or (49/50) | ||||
| % Total Sample Agreement | 97% or (97/100) |
Table 13: Stratified data comparing immunoassay in semi-quantitative mode with LC-MS/MS - Urine
4 Refer to Table 14 for discordant samples of urine.
4Table 14: Discordant Samples - Urine
| Sample ID | Qualitative (mA/min)Result | Semi-QuantitativeObserved concentration(ng/mL) | LC-MS/MS Value(ng/mL) |
|---|---|---|---|
| APP6058-25 | Positive | 347 (Positive) | 285 |
| APP6060-26 | Positive | 375 (Positive) | 230 |
| APP6056-27 | Negative | 267 (Negative) | 372 |
5 Sample APP6058-2 contains Amitriptyline at 146 ng/mL and Nortriptyline at 139 ng/mL by LC-MS/MS and cross-reacts at 100% and 100% by immunoassay, respectively.
6 Sample APP6060-2 contains Amitriptyline at 114 ng/mL, Nordoxepin at 3.05 ng/mL, and Nortriptyline at 115 ng/mL by LC-MS/MS and cross-reacts at 100%, 17.1% and 100% by immunoassay, respectively.
7 Sample APP6056-2 contains Amitriptyline at 186 ng/ mL and Nortriptyline at 186 ng/mL by LC-MS/MS and cross-reacts at 100% and 100% by immunoassay, respectively.
{16}------------------------------------------------
- e) Matrix equivalency
Matrix Equivalency is performed in accordance with CLSI Guideline CLSI-EP35.
Two (2) replicates of 50 patient samples of K2 EDTA Plasma, K3 EDTA plasma, Lithium Heparin Plasma, Sodium Citrate Plasma, Potassium Oxalate Plasma and 45 patient samples of Sodium Heparin Plasma were run in both qualitative and semiquantitative modes to demonstrate Nortriptyline concentrations obtained in different test plasma matrices with different anticoagulants are equivalent to those measured in the primary or control matrix, serum. The results of the matrix equivalency study are presented in the table below.
| Qualitative | Semi-quantitative | ||||
|---|---|---|---|---|---|
| Serum Matrix | Positive | Negative | Positive | Negative | |
| K2 EDTA Plasma | Positive | 25 | 0 | 25 | 0 |
| Negative | 0 | 25 | 0 | 25 | |
| K3 EDTA plasma | Positive | 25 | 0 | 25 | 0 |
| Negative | 0 | 25 | 0 | 25 | |
| Lithium Heparin Plasma | Positive | 25 | 0 | 25 | 0 |
| Negative | 0 | 25 | 0 | 25 | |
| Sodium Citrate Plasma | Positive | 25 | 0 | 25 | 0 |
| Negative | 0 | 25 | 0 | 25 | |
| Potassium Oxalate Plasma | Positive | 25 | 0 | 25 | 0 |
| Negative | 0 | 25 | 0 | 25 | |
| Sodium Heparin Plasma | Positive | 25 | 0 | 25 | 0 |
| Negative | 0 | 20 | 0 | 20 |
{17}------------------------------------------------
- f) Specificity
Cross-Reactivity to structurally related compounds:
The cross-reactivity of Tricyclic compounds and other structurally related compounds were evaluated by adding known amount of each compound into drug-free serum and drug-free urine at concentrations indicated. Serum and urine samples were run separately.
| Serum | |||
|---|---|---|---|
| Structurally Related Compounds(TCA drugs and Metabolites) | TestedConcentrations(ng/mL) | Positive/Negative | Cross-Reactivity (%) |
| 2-Hydroxy Imipramine | 1,300 | Positive | 23.1 |
| 7-Hydroxy Quetiapine | 100,000 | Negative | < 0.3 |
| 7-Hydroxy Amoxapine | 100,000 | Negative | < 0.3 |
| 8-Hydroxy Amoxapine | 100,000 | Negative | < 0.3 |
| Amitriptyline | 300 | Positive | 100 |
| Amoxapine | 125,000 | Positive | 0.2 |
| 10-Hydroxyamitriptyline | 1,000 | Positive | 30 |
| Clomipramine | 300 | Positive | 100 |
| Desipramine | 250 | Positive | 120 |
| Dosulepin | 475 | Positive | 63.2 |
| Doxepin | 600 | Positive | 50 |
| Imipramine | 190 | Positive | 157.9 |
| Lofepramine | 430 | Positive | 69.8 |
| N-Desmethyltrimipramine | 350 | Positive | 85.7 |
| Norclomipramine | 375 | Positive | 80 |
| Nordoxepin | 1,750 | Positive | 17.1 |
| Nortriptyline | 300 | Positive | 100 |
| Opipramol | 350 | Positive | 85.7 |
| Protriptyline | 450 | Positive | 66.7 |
| Quetiapine Fumarate | 50,000 | Positive | 0.6 |
| Trimipramine | 390 | Positive | 76.9 |
Table 16: Cross Reactivity to structurally related compounds in Serum (TCA drugs and Metabolites)
Table 17: Cross Reactivity to structurally related compounds in Urine (TCA drugs and Metabolites)
| Urine | |||
|---|---|---|---|
| Structurally Related Compounds(TCA drugs and Metabolites) | TestedConcentrations(ng/mL) | Positive/Negative | Cross-Reactivity (%) |
| 2-Hydroxy Imipramine | 1,000 | Positive | 30 |
| 7-Hydroxy Quetiapine | 100,000 | Negative | < 0.3 |
| 7-Hydroxy Amoxapine | 100,000 | Negative | < 0.3 |
| 8-Hydroxy Amoxapine | 100,000 | Negative | < 0.3 |
| Amitriptyline | 300 | Positive | 100 |
| Amoxapine | 100,000 | Positive | 0.3 |
{18}------------------------------------------------
| Structurally Related Compounds(TCA drugs and Metabolites) | Urine | ||
|---|---|---|---|
| TestedConcentrations(ng/mL) | Positive/Negative | Cross-Reactivity (%) | |
| 10-Hydroxyamitriptyline | 700 | Positive | 42.9 |
| Clomipramine | 350 | Positive | 85.7 |
| Desipramine | 250 | Positive | 120 |
| Dosulepin | 425 | Positive | 70.6 |
| Doxepin | 550 | Positive | 54.5 |
| Imipramine | 220 | Positive | 136.4 |
| Lofepramine | 460 | Positive | 65.2 |
| N-Desmethyltrimipramine | 325 | Positive | 92.3 |
| Norclomipramine | 450 | Positive | 66.7 |
| Nordoxepin | 1750 | Positive | 17.1 |
| Nortriptyline | 300 | Positive | 100 |
| Opipramol | 350 | Positive | 85.7 |
| Protriptyline | 400 | Positive | 75 |
| Quetiapine Fumarate | 45,000 | Positive | 0.7 |
| Trimipramine | 400 | Positive | 75 |
Table 18: Cross Reactivity to structurally related compounds in Serum (Other compounds)
| Serum | |||
|---|---|---|---|
| Other Structurally RelatedCompounds | TestedConcentrations(ng/mL) | Positive/Negative | Cross-Reactivity (%) |
| Alimemazine | 12,000 | Positive | 2.5 |
| Blonanserin | 100,000 | Negative | < 0.3 |
| Chlorpromazine | 525 | Positive | 57.1 |
| Clozapine | 100,000 | Negative | < 0.3 |
| Cyclobenzaprine | 450 | Positive | 66.7 |
| Desloratadine | 100,000 | Negative | < 0.3 |
| Diphenhydramine | 60,000 | Positive | 0.5 |
| Fluoxetine | 100,000 | Negative | < 0.3 |
| Fluphenazine | 2,000 | Positive | 15 |
| Haloperidol | 100,000 | Negative | < 0.3 |
| Loratadine | 100,000 | Negative | < 0.3 |
| Loxapine | 100,000 | Negative | < 0.3 |
| Maprotiline | 100,000 | Positive | 0.3 |
| Mianserin | 100,000 | Negative | < 0.3 |
| Mirtazapine | 100,000 | Negative | < 0.3 |
| N-Desmethylclozapine | 100,000 | Negative | < 0.3 |
| Nefazodone | 100,000 | Negative | < 0.3 |
| N-Desmethylmirtazapine | 100,000 | Negative | < 0.3 |
| Normaprotiline | 100,000 | Positive | 0.3 |
| Olanzapine | 100,000 | Negative | < 0.3 |
| Paroxetine | 100,000 | Negative | < 0.3 |
{19}------------------------------------------------
| Perphenazine | 450 | Positive | 66.7 |
|---|---|---|---|
| Phenoxybenzamine | 100,000 | Negative | < 0.3 |
| Promazine | 400 | Positive | 75 |
| Promethazine | 20,000 | Positive | 1.5 |
| Risperidone | 100,000 | Negative | < 0.3 |
| Sertraline | 100,000 | Negative | < 0.3 |
| Thioridazine | 6,000 | Positive | 5 |
| Thiothixene | 100,000 | Negative | < 0.3 |
| Tianeptine | 100,000 | Negative | < 0.3 |
| Trazodone | 100,000 | Negative | < 0.3 |
| Ziprasidone | 100,000 | Negative | < 0.3 |
Table 19: Cross Reactivity to structurally related compounds in Urine
(Other compounds)
| Urine | |||
|---|---|---|---|
| Other Structurally RelatedCompounds | TestedConcentrations(ng/mL) | Positive/Negative | Cross-Reactivity (%) |
| Alimemazine | 12,000 | Positive | 2.5 |
| Blonanserin | 100,000 | Negative | < 0.3 |
| Chlorpromazine | 600 | Positive | 50 |
| Clozapine | 100,000 | Negative | < 0.3 |
| Cyclobenzaprine | 500 | Positive | 60 |
| Desloratadine | 100,000 | Negative | < 0.3 |
| Diphenhydramine | 30,000 | Positive | 1 |
| Fluoxetine | 100,000 | Negative | < 0.3 |
| Fluphenazine | 2,000 | Positive | 15 |
| Haloperidol | 100,000 | Negative | < 0.3 |
| Loratadine | 100,000 | Negative | < 0.3 |
| Loxapine | 100,000 | Positive | 0.3 |
| Maprotiline | 100,000 | Positive | 0.3 |
| Mianserin | 100,000 | Negative | < 0.3 |
| Mirtazapine | 100,000 | Negative | < 0.3 |
| N-Desmethylclozapine | 100,000 | Negative | < 0.3 |
| Nefazodone | 100,000 | Negative | < 0.3 |
| N-Desmethylmirtazapine | 100,000 | Negative | < 0.3 |
| Normaprotiline | 100,000 | Positive | 0.3 |
| Olanzapine | 100,000 | Negative | < 0.3 |
| Paroxetine | 100,000 | Negative | < 0.3 |
| Perphenazine | 650 | Positive | 46.2 |
| Phenoxybenzamine | 100,000 | Negative | < 0.3 |
| Promazine | 410 | Positive | 73.2 |
| Promethazine | 15,000 | Positive | 2 |
| Risperidone | 100,000 | Negative | < 0.3 |
| Sertraline | 100,000 | Negative | < 0.3 |
{20}------------------------------------------------
| Urine | |||
|---|---|---|---|
| Other Structurally RelatedCompounds | TestedConcentrations(ng/mL) | Positive/Negative | Cross-Reactivity (%) |
| Thioridazine | 4,000 | Positive | 7.5 |
| Thiothixene | 100,000 | Negative | < 0.3 |
| Tianeptine | 90,000 | Positive | 0.33 |
| Trazodone | 100,000 | Negative | < 0.3 |
| Ziprasidone | 100,000 | Negative | < 0.3 |
Cross-reactivity to structurally unrelated compounds:
Structurally unrelated compounds and/or concurrently used drugs were spiked at the concentration listed below into low and high controls (225 and 375 ng/mL) in serum and urine. Low control as negative and high control as positive indicate that all the compounds evaluated exhibited minimal cross-reactivity at the concentration tested. Serum and urine samples were run separately. The study results are presented in the table below:
| Serum | |||
|---|---|---|---|
| Structurally Unrelated Compounds | TestedConcentrations(ng/mL) | Low Control-25% of cutoff(225 ng/mL) | High Control+25% of cutoff(375 ng/mL) |
| 11-nor- $\Delta$ 9-THC-COOH | 100,000 | Negative | Positive |
| 6-Acetyl morphine | 75,000 | Negative | Positive |
| Acetaminophen | 100,000 | Negative | Positive |
| Acetylsalicylic acid | 100,000 | Negative | Positive |
| Amisulpride | 100,000 | Negative | Positive |
| Amoxicillin | 100,000 | Negative | Positive |
| Amphetamine | 100,000 | Negative | Positive |
| Benztropine Methane Sulfonate | 3,000 | Negative | Positive |
| Benzoylecgonine | 100,000 | Negative | Positive |
| Brompheniramine | 3,000 | Negative | Positive |
| Caffeine | 100,000 | Negative | Positive |
| Carbamazepine | 3,000 | Negative | Positive |
| Carbamazepine Epoxide | 10,000 | Negative | Positive |
| Chloroquine phosphate | 100,000 | Negative | Positive |
| Cimetidine | 100,000 | Negative | Positive |
| Cocaine | 75,000 | Negative | Positive |
| Codeine | 100,000 | Negative | Positive |
| Dextromethorphan | 100,000 | Negative | Positive |
| Diacetylmorphine (Heroin) | 100,000 | Negative | Positive |
| Diazepam | 100,000 | Negative | Positive |
| Digoxin | 100,000 | Negative | Positive |
| Dihydrocodeine | 100,000 | Negative | Positive |
| EDDP Perchlorate | 100,000 | Negative | Positive |
| Structurally Unrelated Compounds | Serum | ||
| TestedConcentrations(ng/mL) | Low Control-25% of cutoff(225 ng/mL) | High Control+25% of cutoff(375 ng/mL) | |
| EMDP-HCL | 100,000 | Negative | Positive |
| Fentanyl | 25,000 | Negative | Positive |
| Glutethimide | 100,000 | Negative | Positive |
| Hydrocodone | 100,000 | Negative | Positive |
| Hydrocortisone | 100,000 | Negative | Positive |
| Hydromorphone | 100,000 | Negative | Positive |
| Hydroxyzine | 3,000 | Negative | Positive |
| Ibuprofen | 100,000 | Negative | Positive |
| Levorphanol-D3 | 100,000 | Negative | Positive |
| Levothyroxine | 100,000 | Negative | Positive |
| Meperidine | 25,000 | Negative | Positive |
| Methadone | 75,000 | Negative | Positive |
| Methamphetamine | 100,000 | Negative | Positive |
| Methaqualone | 500 | Negative | Positive |
| Methsuximide | 75,000 | Negative | Positive |
| Methylphenidate | 100,000 | Negative | Positive |
| Morphine | 100,000 | Negative | Positive |
| Morphine-3β-glucuronide | 100,000 | Negative | Positive |
| Morphine-6β-glucuronide | 100,000 | Negative | Positive |
| Nalbuphine | 100,000 | Negative | Positive |
| Nalorphine | 100,000 | Negative | Positive |
| Naloxone | 100,000 | Negative | Positive |
| Naltrexone | 100,000 | Negative | Positive |
| Naproxen | 100,000 | Negative | Positive |
| Norcodeine | 100,000 | Negative | Positive |
| Nordiazepam | 100,000 | Negative | Positive |
| Norethindrone | 100,000 | Negative | Positive |
| Norhydrocodone | 100,000 | Negative | Positive |
| Noroxycodone | 100,000 | Negative | Positive |
| Noroxymorphone | 100,000 | Negative | Positive |
| Norpropoxyphene | 75,000 | Negative | Positive |
| Oxaprozin | 100,000 | Negative | Positive |
| Oxazepam | 100,000 | Negative | Positive |
| Oxycodone | 100,000 | Negative | Positive |
| Oxymorphone | 100,000 | Negative | Positive |
| PCP | 50,000 | Negative | Positive |
| Phenobarbital | 100,000 | Negative | Positive |
| Phenytoin | 100,000 | Negative | Positive |
| Primidone | 100,000 | Negative | Positive |
| Procyclidine | 100,000 | Negative | Positive |
| Propoxyphene | 100,000 | Negative | Positive |
| Secobarbital | 100,000 | Negative | Positive |
| Serum | |||
| Structurally Unrelated Compounds | TestedConcentrations(ng/mL) | Low Control-25% of cutoff(225 ng/mL) | High Control+25% of cutoff(375 ng/mL) |
| Tapentadol | 100,000 | Negative | Positive |
| Temazepam | 100,000 | Negative | Positive |
| Triprolidine | 100,000 | Negative | Positive |
| Valproic Acid | 100,000 | Negative | Positive |
| Venlafaxine | 100,000 | Negative | Positive |
| Verapamil | 100,000 | Negative | Positive |
Table 20: Cross Reactivity to structurally unrelated compounds in Serum
{21}------------------------------------------------
{22}------------------------------------------------
Table 21: Cross Reactivity to structurally unrelated compounds in Urine
| Structurally Unrelated Compounds | TestedConcentrations(ng/mL) | Urine | |
|---|---|---|---|
| Low Control-25% of cutoff(225 ng/mL) | High Control+25% of cutoff(375 ng/mL) | ||
| 11-nor-Δ9-THC-COOH | 100,000 | Negative | Positive |
| 6-Acetyl morphine | 100,000 | Negative | Positive |
| Acetaminophen | 100,000 | Negative | Positive |
| Acetylsalicylic acid | 100,000 | Negative | Positive |
| Amisulpride | 100,000 | Negative | Positive |
| Amoxicillin | 100,000 | Negative | Positive |
| Amphetamine | 100,000 | Negative | Positive |
| Benztropine Methane Sulfonate | 7,000 | Negative | Positive |
| Benzoylecgonine | 100,000 | Negative | Positive |
| Brompheniramine | 5,000 | Negative | Positive |
| Caffeine | 100,000 | Negative | Positive |
| Carbamazepine | 5,000 | Negative | Positive |
| Carbamazepine Epoxide | 30,000 | Negative | Positive |
| Chloroquine phosphate | 100,000 | Negative | Positive |
| Cimetidine | 100,000 | Negative | Positive |
| Cocaine | 100,000 | Negative | Positive |
| Codeine | 100,000 | Negative | Positive |
| Dextromethorphan | 100,000 | Negative | Positive |
| Diacetylmorphine (Heroin) | 100,000 | Negative | Positive |
| Diazepam | 100,000 | Negative | Positive |
| Digoxin | 100,000 | Negative | Positive |
| Dihydrocodeine | 100,000 | Negative | Positive |
| EDDP Perchlorate | 100,000 | Negative | Positive |
| EMDP-HCL | 100,000 | Negative | Positive |
| Fentanyl | 25,000 | Negative | Positive |
| Glutethimide | 100,000 | Negative | Positive |
| Hydrocodone | 100,000 | Negative | Positive |
| Hydrocortisone | 100,000 | Negative | Positive |
| Hydromorphone | 100,000 | Negative | Positive |
| Hydroxyzine | 5,000 | Negative | Positive |
| Ibuprofen | 100,000 | Negative | Positive |
| Levorphanol-D3 | 100,000 | Negative | Positive |
| Urine | |||
| Structurally Unrelated Compounds | TestedConcentrations(ng/mL) | Low Control-25% of cutoff(225 ng/mL) | High Control+25% of cutoff(375 ng/mL) |
| Levothyroxine | 100,000 | Negative | Positive |
| Meperidine | 25,000 | Negative | Positive |
| Methadone | 75,000 | Negative | Positive |
| Methamphetamine | 100,000 | Negative | Positive |
| Methaqualone | 1,000 | Negative | Positive |
| Methsuximide | 75,000 | Negative | Positive |
| Methylphenidate | 100,000 | Negative | Positive |
| Morphine | 100,000 | Negative | Positive |
| Morphine-3β-glucuronide | 100,000 | Negative | Positive |
| Morphine-6β-glucuronide | 100,000 | Negative | Positive |
| Nalbuphine | 100,000 | Negative | Positive |
| Nalorphine | 100,000 | Negative | Positive |
| Naloxone | 100,000 | Negative | Positive |
| Naltrexone | 100,000 | Negative | Positive |
| Naproxen | 100,000 | Negative | Positive |
| Norcodeine | 100,000 | Negative | Positive |
| Nordiazepam | 100,000 | Negative | Positive |
| Norethindrone | 100,000 | Negative | Positive |
| Norhydrocodone | 75,000 | Negative | Positive |
| Noroxycodone | 100,000 | Negative | Positive |
| Noroxymorphone | 100,000 | Negative | Positive |
| Norpropoxyphene | 75,000 | Negative | Positive |
| Oxaprozin | 100,000 | Negative | Positive |
| Oxazepam | 100,000 | Negative | Positive |
| Oxycodone | 100,000 | Negative | Positive |
| Oxymorphone | 100,000 | Negative | Positive |
| PCP | 75,000 | Negative | Positive |
| Phenobarbital | 100,000 | Negative | Positive |
| Phenytoin | 100,000 | Negative | Positive |
| Primidone | 100,000 | Negative | Positive |
| Procyclidine | 100,000 | Negative | Positive |
| Propoxyphene | 100,000 | Negative | Positive |
| Secobarbital | 100,000 | Negative | Positive |
| Tapentadol | 100,000 | Negative | Positive |
| Temazepam | 100,000 | Negative | Positive |
| Triprolidine | 100,000 | Negative | Positive |
| Valproic Acid | 100,000 | Negative | Positive |
| Venlafaxine | 100,000 | Negative | Positive |
| Verapamil | 100,000 | Negative | Positive |
{23}------------------------------------------------
{24}------------------------------------------------
- g) Interference
The interference study is performed per the CLSI EP07. 3rd edition. The potential interference of endogenous, exogenous, physiological substances, and pH on the recovery of nortriptyline using DRI Tricyclics Serum Tox Assay was assessed. Potentially interfering substances were spiked into the low control, 225ng/mL (-25% of the cutoff concentration of 300 ng/mL) and high controls, 375 ng/mL (+25% of the cutoff concentration of 300 ng/mL). Serum and urine samples were run separately.
As shown in the tables below, the controls were detected accurately. In the presence of the compounds listed below, the controls were detected accurately, indicating that these compounds did not show interference in the assay.
| Compounds | TestedConcentrations(mg/dL) | Low Control-25% of cutoff(225 ng/mL) | High Control+25% of cutoff(375 ng/mL) |
|---|---|---|---|
| Bilirubin (Conjugated) | 40 | Negative | Positive |
| Bilirubin (Unconjugated) | 40 | Negative | Positive |
| Hemoglobin | 1000 | Negative | Positive |
| Albumin | 7500 | Negative | Positive |
| y-globulin | 5000 | Negative | Positive |
| Rh Factor | 1300 IU | Negative | Positive |
| Triglycerides | 1500 | Negative | Positive |
| Cholesterol | 1400 | Negative | Positive |
Table 22: Results of Interfering Substances Testing - Serum
Table 23: Results of Interfering Substances Testing - Urine
| Compounds | TestedConcentrations(mg/dL) | Low Control-25% of cutoff(225 ng/mL) | High Control+25% of cutoff(375 ng/mL) |
|---|---|---|---|
| Acetone | 500 | Negative | Positive |
| Ascorbic Acid | 150 | Negative | Positive |
| Caffeine | 5 | Negative | Positive |
| Creatinine | 400 | Negative | Positive |
| Ethanol | 1000 | Negative | Positive |
| Galactose | 5 | Negative | Positive |
| Glucose | 1000 | Negative | Positive |
| Hemoglobin | 150 | Negative | Positive |
| Human Serum Albumin(HSA) | 200 | Negative | Positive |
| Oxalic acid | 50 | Negative | Positive |
| Riboflavin | 3 | Negative | Positive |
| Sodium Chloride | 1000 | Negative | Positive |
| Urea | 1000 | Negative | Positive |
{25}------------------------------------------------
pH interference for urine only:
The intermediate drug stock at 10.000 ng/mL was used to spike the pH adjusted drug free urine samples to make low control, 225 ng/mL and high control, 375 ng/mL with pH range from 3 to 11. The results are presented in the table below:
| Table 24: pH interference study results for urine | ||
|---|---|---|
| pH | Low Control-25% of cutoff(225 ng/mL) | High Control+25% of cutoff(375 ng/mL) |
| 3 | Negative | Positive |
| 4 | Negative | Positive |
| 5 | Negative | Positive |
| 6 | Negative | Positive |
| 7 | Negative | Positive |
| 8 | Negative | Positive |
| 9 | Negative | Positive |
| 10 | Negative | Positive |
| 11 | Negative | Positive |
Table 24: pH interference study results for urine
There is no need to test pH as interfering factor for serum because blood pH range is very narrow. Any continuous pH < 7.3 and > 7.8 are lethal. Varied urine pH is to support in case of sample adulteration.
Specific gravity for urine only:
The specific gravity of multiple negative urine samples from different donors were tested on available clinical chemistry analyzer to select ten (10) negative urine samples with specific gravity range from 1.004 to 1.030.
The selected negative urine samples with a specific gravity range from 1.004 to 1.030 were spiked at control levels of 225 ng/mL and 375ng/mL using intermediate drug stock at 10,000 ng/mL.
| Specific Gravity | Low Control-25% of cutoff(225 ng/mL) | High Control+25% of cutoff(375 ng/mL) |
|---|---|---|
| 1.004 | Negative | Positive |
| 1.006 | Negative | Positive |
| 1.008 | Negative | Positive |
| 1.010 | Negative | Positive |
| 1.011 | Negative | Positive |
| 1.012 | Negative | Positive |
| 1.013 | Negative | Positive |
| 1.016 | Negative | Positive |
| 1.022 | Negative | Positive |
| 1.030 | Negative | Positive |
| Specific Gravity | Number of Patients |
|---|---|
| 1.001-1.010 | 10 |
| 1.011-1.020 | 25 |
| 1.021-1.030 | 15 |
| >1.030 | 5 |
{26}------------------------------------------------
H. Conclusion
The information supports a determination of substantial equivalence between DRI™ Tricyclics Serum Tox Assay (K213875) and the predicate device Tricyclic Serum Tox EIA Assay (K983268).
§ 862.3910 Tricyclic antidepressant drugs test system.
(a)
Identification. A tricyclic antidepressant drugs test system is a device intended to measure any of the tricyclic antidepressant drugs in serum. The tricyclic antidepressant drugs include imipramine, desipramine, amitriptyline, nortriptyline, protriptyline, and doxepin. Measurements obtained by this device are used in the diagnosis and treatment of chronic depression to ensure appropriate therapy.(b)
Classification. Class II (special controls). A tricyclic antidepressant drugs test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).