(63 days)
Dental: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the maxillofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients.
Neuro: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the cranial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients.
The Stryker Resorbable Fixation System is a cranio-maxillofacial plating system intended for use in the fixation of bones of the craniofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients but is not intended for use in the mandible and/or full load bearing procedures. It consists of resorbable bone fixation plates, meshes, and screws made of a copolymer of poly lactide and poly glycolide.
The provided text describes a 510(k) premarket notification for the Stryker Resorbable Fixation System. This submission focuses on design modifications related to the sterilization process and packaging, not on validating the core performance of a new device or an AI/ML algorithm. Therefore, the information needed to answer the prompt, particularly regarding acceptance criteria and studies proving the device meets those criteria in the context of an AI/ML device, is not present in the provided document.
The document states:
- "Performance bench testing was not required as a basis for substantial equivalence."
- "Animal testing was not required as a basis for substantial equivalence."
- "Clinical testing was not required as a basis for substantial equivalence."
The "studies" mentioned are limited to:
- Sterilization Validation: Product Adoption and Sterility, EO Residual, Endotoxin Testing. These confirm the new sterilization process achieves sterility and meets safety limits for ETO residuals and endotoxins.
- Summative Usability Validation of Packaging: A study to confirm correct aseptic transfer despite a packaging change.
Since the prompt specifically asks about acceptance criteria and studies for a device (implying a functional, potentially AI/ML, device performance) and the document indicates these types of performance tests were not required for this particular 510(k) submission (which focused on demonstrating substantial equivalence for manufacturing/packaging changes), I cannot fill out the requested table or answer the questions as they pertain to the performance validation of an AI/ML device.
However, I can extract the information provided about the limited validation studies that were performed for the changes described in the document:
1. Table of acceptance criteria and the reported device performance (for the changes evaluated):
| Characteristic | Test | Acceptance Criteria (Implied by "Passed") | Reported Performance | Standards Used |
|---|---|---|---|---|
| Sterilization Validation / Sterility | Product Adoption and Sterility | Passed | Achieved sterility | ISO 11135:2014 +AMD1:2018 |
| Sterilization / Biocompatibility | EO Residual | Passed | Below limits | ISO 10993-7:2008 +AMD1:2019 |
| Sterilization / Biocompatibility | Endotoxin Testing | Passed | Below limits | ISO 10993-1 +ANSI/AAMI ST72:2019 |
| Packaging Usability / Aseptic Transfer | Aseptic transfer performance | All test participants perform correctly | Passed | (Implicitly, no use errors or difficulties observed) |
Regarding the other questions, the document does not contain the required information as it is not an AI/ML device validation study:
-
Sample sizes used for the test set and the data provenance:
- Packaging Usability: Sample size of 19 representative users.
- Data Provenance: Not specified, but generally, usability studies are prospective.
- For sterilization tests, "product adoption" would involve a specific number of units, but this detail is not provided.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for sterility is laboratory measurement against standards. For packaging usability, "correct performance" by representative users serves as the "ground truth."
-
Adjudication method for the test set: Not applicable. For usability, direct observation of performance. For biochemical tests, measurement against established limits.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Sterilization: Laboratory measurements and standardized biological indicators (implicit in ISO standards).
- Packaging Usability: Observational performance of users against defined correct steps.
-
The sample size for the training set: Not applicable. This is not an AI/ML device.
-
How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
2/4/2022
Stryker Leibinger Micro Implants Gregory Gohl Senior Regulatory Affairs Specialist 750 Trade Centre Way -Suite 200 Portage, Michigan 49002
Re: K213777
Trade/Device Name: Stryker Resorbable Fixation System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY, GXR, HBW Dated: January 21, 2022 Received: January 26, 2022
Dear Gregory Gohl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K213777
Device Name Stryker Delta Resorbable Fixation System
Indications for Use (Describe)
Dental: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the maxillofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients.
Neuro: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the cranial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
- SUBMITTER [§807.92(a)(1)] I.
| 510(k) Owner: | Stryker Leibinger GmbH& Co. KGBoetzinger Strasse 41D-79111 Freiburg, Germany |
|---|---|
| Submitter/ ContactPerson: | Gregory GohlSr. Regulatory Affairs SpecialistStryker Craniomaxillofacial (CMF)1941 Stryker WayPortage, MI 49002 |
| Phone: 269-370-1476Fax: 877-648-7114 | |
| Date prepared: | February 4, 2022 |
- DEVICE [§807.92(a)(2)] K213777 II.
| Trade Name: | Stryker Resorbable Fixation System |
|---|---|
| Abbreviated Name: | Delta System / Delta Resorbable Fixation System |
| Common or Usual Name: | Resorbable Bone Plating System |
| Device: | Stryker Resorbable Fixation System |
| Classification Name &Regulation Description: | Plate, Fixation, Bone; per 21 CFR §888.3030 |
| Regulation MedicalSpecialty & Review Panel: | Dental (OHT1/DHT1B – Office of Ophthalmic,Anesthesia, Respiratory, ENT, & Dental); Neurology(OHT5/DHT5A – Office of Neurological and PhysicalMedicine Devices) |
| Primary Product Code: | JEY |
| Subsequent Prod. Codes: | GXR, HBW |
| Regulatory Device Class: | Class II |
| *Note the company Stryker or legacy name Stryker Leibinger precedes theproduct/trade name and predicate device in some documentation. |
{4}------------------------------------------------
III. PREDICATE DEVICE [§807.92(a)(3)]
- A. Submission Branch of Predicate Device: Division of Surgical, Orthopedic, and Restorative Devices; Office of Device Evaluation Center for Devices and Radiological Health
- B. Predicate Device: Stryker Resorbable Fixation System K113109
IV. DEVICE DESCRIPTION [§807.92(a)(4)]
- A. Submission Branch of Subject Device: Dental (OHT1/DHT1B-Office of Ophthalmic, Anesthesia, Respiratory, ENT, & Dental); Neurology (OHT5/DHT5A - Office of Neurological and Physical Medicine Devices)
- B. Subject Device: Stryker Resorbable Fixation System (also referred to as Stryker Delta Resorbable Fixation System or Delta System
The Stryker Resorbable Fixation System is a cranio-maxillofacial plating system intended for use in the fixation of bones of the craniofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients but is not intended for use in the mandible and/or full load bearing procedures. It consists of resorbable bone fixation plates, meshes, and screws made of a copolymer of poly lactide and poly glycolide.
- V. INDICATIONS FOR USE [§807.92(a)(5)]
| Predicate Device - K113109 | Subject Device | |
|---|---|---|
| Indications forUse | The Delta Resorbable Fixation System isintended for use in the fixation of bonesof the craniofacial and midfacialskeleton affected by trauma or forreconstruction. The system can be usedin both adult and pediatric patients but isnot intended for use in the mandibleand/or full load bearing procedures. | Dental: The Delta Resorbable FixationSystem is intended for use in the fixationof bones of the maxillofacial andmidfacial skeleton affected by trauma orfor reconstruction. The system can beused in both adult and pediatric patients.Neuro: The Delta Resorbable FixationSystem is intended for use in the fixationof bones of the cranial skeleton affectedby trauma or for reconstruction. Thesystem can be used in both adult andpediatric patients. |
TABLE 5-1: COMPARISON OF INDICATIONS FOR USE
{5}------------------------------------------------
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE [§807.92(a)(6)]
The subject device is compared to its predicate device for substantial equivalence of technological characteristics based on the following criteria:
- A. Principles of Operation
- B. Technological Characteristics
C. Principles of Operation / Operating Principle
The basic operational principle of the Stryker Resorbable Fixation System remains the same as for the predicate: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the craniofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients but is not intended for use in the mandible and/or full load bearing procedures. This submission does not propose a change in the operating principle as there is no change to the implant design. interface with accessories or implant materials.
D. Technological Characteristics
The technological characteristics remain the same as the predicate:
- -Same Intended Use: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the craniofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients but is not intended for use in the mandible and/or full load bearing procedures.
- -Same Mode of Fixation: fixated to the native bone with Stryker Delta Resorbable screws.
- Same Materials of Construction: Implants are made of a copolymer of poly lactide and poly glycolide. No change to the formulation of the polymer.
- -Same Design: The change of the sterilization supplier and location does not change the implant design by any means.
VII. PERFORMANCE DATA [§807.92(b)(7)]
The modifications described in this submission are the change to a new Ethylene Oxide (EO) sterilization supplier and location and a modification to the packaging. The new EO sterilization cycle is equivalent to the predicate device sterilization cycle operating at lower EO gas concentrations. There is no change to the sterilization method regarding the final end-product. The new supplier location and cycle were proven to be as substantially equivalent for sterilization as the predicate location.
Packaging components, design, and materials remain the same. Labeling requirements also remain unchanged. The modification of the packaging is an update of the sterile barrier layout of the double pouch packaging system (i.e., only the inner pouch is the sterile barrier instead of the outer and inner pouch as for the predicate devices). This change is
{6}------------------------------------------------
communicated to the customer through an inclusion of a warning sticker in regard. Useability validation confirms the warning sticker label mitigates the risk associated with the change in the sterile barrier packaging.
A risk analysis was performed, and sterilization testing was performed in support of the substantial equivalence determination.
There is no change in the design, material, manufacturing process, or duration/location of contact. The tests necessary to evaluate the change regarding the new sterilization location and supplier are listed below. The following tests were performed to adopt the Stryker Resorbable Fixation System plates and meshes into the existing EO Cycle.
| Characteristic | Test | Result | Standards |
|---|---|---|---|
| SterilizationValidation | Product Adoption andSterility | Passed | ISO 11135:2014+AMD1:2018 |
| Sterilization /Biocompatibility | EO Residual | Passed | ISO 10993-7:2008+AMD1:2019 |
| Sterilization /Biocompatibility | Endotoxin Testing | Passed | ISO 10993-1+ANSI/AAMI ST72:2019 |
TABLE 5-2: STERILIZATION AND BIOCOMPATIBILITY TESTING
The aseptic transfer (i.e., the transfer of the devices from the non-sterile field into the sterile field during surgery) is modified by the change in the sterile barrier declaration. The aseptic transfer has been identified as the only critical task to be subjected to a summative usability validation of packaging. The human factors testing/study was designed to evaluate in particular possible adverse effects caused by negative transfer. The study was conducted with a sample size of 19 representative users including experienced as well as new users. The full packaging configuration was subjected to test. All test participants were able to perform the aseptic transfer correctly. No use errors or use difficulties were observed. The packaging change was evaluated to be non-critical.
TABLE 5-3: SUMMATIVE USABILITY VALIDATION OF PACKAGING
| Characteristic | Test | Result |
|---|---|---|
| Summative Validation | Aseptic transferperformance | Passed |
Performance Bench Testing
Performance bench testing was not required as a basis for substantial equivalence.
{7}------------------------------------------------
Animal Testing
Animal testing was not required as a basis for substantial equivalence.
Clinical Testing [§807.92(b)(2)]
Clinical testing was not required as a basis for substantial equivalence.
VIII. CONCLUSION [§807.92(b)(3)]
In summary, the Stryker Resorbable Fixation System is substantially equivalent to its predicate device. The design of implants has not changed, nor has the manufacturing of them changed. There is no change to the sterilization method, and the sterilization cycles of the subject and predicate device are evaluated to be equivalent throughout this submission. The fundamental scientific technology has not changed from the predicate device since the design of the implants or subcomponents has not changed. The intended use, principle of operation, mechanism of action, technological characteristics, and materials of construction also have not changed. In reference to packaging and labeling, specifically, the packaging components, design, and materials remain the same. The moisture barrier change in packaging of the subject devices led to change in labeling. Human factors data supports that the modified packaging, including the change in labeling, is substantially equivalent to the predicate. Overall, the modifications do not raise new questions of safety or effectiveness. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein. it is concluded that the information included in this submission supports substantial equivalence to the predicate device.
§ 872.4760 Bone plate.
(a)
Identification. A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.(b)
Classification. Class II.