(56 days)
No
The device description and performance studies focus on mechanical and fluid dynamics properties, with no mention of AI or ML.
Yes
The device is indicated for the "removal of fresh, soft emboli and thrombi in the peripheral arterial system," which describes a therapeutic action to treat a medical condition.
No.
The device is indicated for the removal of emboli and thrombi, not for diagnosing conditions.
No
The device description clearly outlines physical components (catheter, tubing) and mentions bench and animal testing of these physical attributes, indicating it is a hardware device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are devices used to examine specimens taken from the human body (like blood, urine, or tissue) to provide information about a person's health.
- Device Description and Intended Use: The Merlin Aspiration System is a medical device used within the body (in the peripheral arterial system) to physically remove blood clots (emboli and thrombi). It does not analyze samples taken from the body.
- Lack of IVD Characteristics: The description focuses on the physical components, their function in a surgical procedure, and performance testing related to mechanical properties and aspiration effectiveness. There is no mention of analyzing biological samples or providing diagnostic information based on such analysis.
Therefore, the Merlin Aspiration System is an invasive medical device used for therapeutic purposes, not an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
· Merlin Aspiration Catheter
As a part of the Merlin Aspiration System, the Merlin Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi in the peripheral arterial system.
· High Flow Tubing
As a part of the Merlin Aspiration System, the High Flow Tubing is indicated to connect the guide catheter to an aspiration pump.
Product codes
QEZ
Device Description
The Merlin Aspiration System is comprised of two devices:
- Merlin Aspiration Catheter
- High Flow Tubing
The Merlin Aspiration Catheter is introduced through a guide catheter or long sheath into the peripheral vasculature and guided over a guidewire to the site of primary occlusion. The Merlin Aspiration Catheter is used in conjunction with an aspiration pump, which is connected to High Flow Tubing. These devices are visible under fluoroscopy.
Merlin Aspiration Catheter
The Merlin Aspiration Catheter consists of single lumen, coiled construction, variable stiffness shaft with radiopaque markers on distal and proximal end of the catheter portion for angiographic visualization. The Merlin Aspiration Catheter is identical in design to the MIVI Q Distal Access Catheter (K192558).
High Flow Tubing
The MIVI High Flow Tubing is a sterilized and disposable (onetime use) product consisting of high-pressure PVC tubing with end fittings and an on/off switch. The MIVI High Flow Tubing connects an aspiration pump to a reperfusion catheter, providing a means for introducing vacuum during procedures. The MIVI High Flow Tubing has a flow valve that allows the physician to start and stop the flow of aspiration.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
fluoroscopy
Anatomical Site
peripheral arterial system
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Bench Testing
The Merlin Aspiration System is identical to the MIVI Q Distal Access Catheter (K192558), which has been evaluated through risk analysis and design verification testing, which confirmed the device met design specifications.
Testing includes:
Coating Particulate Testing - Met established criteria
Coating Adhesion - Met established criteria
Coating Uniformity - Met established criteria
Surface Integrity - Met established criteria
Tip Inspection - Met established criteria
Liquid Leakage under Pressure - Met established criteria
Dimensional Verification - Met established criteria
Kink Resistance - Met established criteria
Push/Track - Met established criteria
Suction Flow Rate - Met established criteria
System Introduction / Aspiration of Thromboemboli - Met established criteria
Tensile Strength-Push Wire - Met established criteria
Tensile Strength-Distal Section - Met established criteria
Catheter Torque Testing - Met established criteria
Catheter Burst Testing - Met established criteria
High Flow Tubing
Dimensional Verification - Met established criteria
Vacuum Pressure - Met established criteria
Flow Rate - Met established criteria
Degree of Collapse - Met established criteria
Simulated Use - Met established criteria
Leak Test - Met established criteria
Distal Male Luer Tubing Tensile Strength - Met established criteria
Female Luer Tubing Tensile Strength - Met established criteria
Proximal Male to Tubing Tensile Strength - Met established criteria
Connector to Tubing Tensile Strength - Met established criteria
Animal Testing
The Animal Testing data was performed in the porcine carotid, renal, and subclavian vasculature of 8 animals to ensure that there are no new questions of safety and effectiveness.
Key Metrics
Not Found
Predicate Device(s)
Penumbra INDIGOTM Aspiration System (K161523)
Reference Device(s)
MIVI Q Distal Access Catheter (K192558)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.5150 Embolectomy catheter.
(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
January 27, 2022
MIVI Neurovascular, Inc. Janel Hurtado Regulatory Affairs Director 6545 City West Parkway Eden Prairie, Minnesota 55443
Re: K213771
Trade/Device Name: Merlin Aspiration System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEZ Dated: November 24, 2021 Received: December 2, 2021
Dear Janel Hurtado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known)
Device Name Merlin Aspiration System
Indications for Use (Describe)
· Merlin Aspiration Catheter
As a part of the Merlin Aspiration System, the Merlin Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi in the peripheral arterial system.
· High Flow Tubing
As a part of the Merlin Aspiration System, the High Flow Tubing is indicated to connect the guide catheter to an aspiration pump.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(K) SUMMARY
Date Prepared: January 24, 2022 | ||||
---|---|---|---|---|
Table 1. General Information | ||||
510(k) Submitter | Contact | |||
MIVI Neuroscience, Inc. | Janel Hurtado | |||
6545 City West Parkway | Regulatory Affairs Director | |||
Eden Prairie, MN 55344 | Email: jhurtado@mivineuro.com | |||
General Information | ||||
Trade Name | Merlin Aspiration System | |||
Common Name | Catheter, Embolectomy | |||
Classification | Percutaneous catheter; 21 CFR 870.5150 (Class II) | |||
Information | ProCode: QEZ | |||
Panel: Cardiovascular | ||||
Predicate | Penumbra INDIGOTM Aspiration System (K161523) | |||
Device | ||||
Reference | MIVI Q Distal Access Catheter (K192558) | |||
Device |
Device Description
The Merlin Aspiration System is comprised of two devices:
- Merlin Aspiration Catheter
- High Flow Tubing
The Merlin Aspiration Catheter is introduced through a guide catheter or long sheath into the peripheral vasculature and guided over a guidewire to the site of primary occlusion. The Merlin Aspiration Catheter is used in conjunction with an aspiration pump, which is connected to High Flow Tubing. These devices are visible under fluoroscopy.
Merlin Aspiration Catheter
The Merlin Aspiration Catheter consists of single lumen, coiled construction, variable stiffness shaft with radiopaque markers on distal and proximal end of the catheter portion for angiographic visualization. The Merlin Aspiration Catheter is identical in design to the MIVI Q Distal Access Catheter (K192558).
Table 2: Device Information | |
---|---|
Device Description | Model Number |
Merlin Aspiration Catheter, 3F | M3-36 |
Merlin Aspiration Catheter, 4F | M4-43 |
4
Merlin Aspiration Catheter, 5F | M5-57 |
---|---|
Merlin Aspiration Catheter, 6F | M6-69 |
High Flow Tubing | HFT 110 |
High Flow Tubing
The MIVI High Flow Tubing is a sterilized and disposable (onetime use) product consisting of high-pressure PVC tubing with end fittings and an on/off switch. The MIVI High Flow Tubing connects an aspiration pump to a reperfusion catheter, providing a means for introducing vacuum during procedures. The MIVI High Flow Tubing has a flow valve that allows the physician to start and stop the flow of aspiration.
Intended Use / Indications
● Merlin Aspiration Catheter
As a part of the Merlin Aspiration System, the Merlin Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombus in the peripheral arterial system.
- High Flow Tubing ●
As a part of the Merlin Aspiration System, the High Flow Tubing is indicated to connect the guide catheter to an aspiration pump.
Substantial Equivalence Comparison
Table 3 compares the substantial equivalence of the subject, predicate and reference devices.
Table-3: Comparison with Predicate Device | |||
---|---|---|---|
Feature | Subject Device | ||
Merlin Aspiration | |||
System | Predicate Device | ||
INDIGO Aspiration | |||
System | Reference Device | ||
MIVI Q Distal Access | |||
Catheter | |||
510(k) Holder & | |||
Manufacturer | MIVI Neuroscience, Inc. | Penumbra, Inc. | MIVI Neuroscience, Inc. |
510(k)# | K213771 | K161523 | K192558 |
Product Code | QEZ | QEW | DQY |
Regulatory Class | Class II | ||
Regulatory Number | 21 CFR 870.5150 | ||
Regulatory Name | Percutaneous Catheter | ||
Indications / Intended Use / Principle of Operation | |||
Indications / Intended | |||
Use | Merlin Aspiration | ||
Catheter | |||
As a part of the Merlin | |||
Aspiration System, the | |||
Merlin Aspiration | INDIGO Aspiration | ||
Catheters | |||
and Separators | |||
As part of the Penumbra | |||
Embolectomy Aspiration | |||
System | Q Distal Access Catheter is | ||
indicated for use with | |||
compatible guide catheters in | |||
facilitating the insertion | |||
and guidance of | |||
microcatheters into a selected | |||
Table-3: Comparison with Predicate Device | |||
Feature | Subject Device | ||
Merlin Aspiration | |||
System | Predicate Device | ||
INDIGO Aspiration | |||
System | Reference Device | ||
MIVI Q Distal Access | |||
Catheter | |||
Catheter is indicated for | |||
the removal of fresh, soft | |||
emboli and thrombus in | |||
the peripheral arterial | |||
systems. |
High Flow Tubing
As a part of the Merlin
Aspiration System, the
HFT 110 High Flow
Tubing is indicated to
connect the guide
catheter to an aspiration
pump. | (INDIGOTM Aspiration
System),
the INDIGO Aspiration
Catheters
and Separators are
indicated for
the removal of fresh, soft
emboli
and thrombi from vessels
of the
peripheral arterial and
venous
systems.
INDIGO Aspiration
Tubing
As part of the Penumbra
Embolectomy Aspiration
System
(INDIGOTM Aspiration
System),
the INDIGO Sterile
Aspiration
Tubing is indicated to
connect the
INDIGO Aspiration
Catheters to
the Penumbra Pump
MAX.
Penumbra Pump MAX
The Penumbra Pump
MAX is
indicated as a vacuum
source for
the Penumbra Aspiration
System. | blood vessel in the
peripheral, coronary and
neuro vascular systems. |
| Aspiration Catheter | | | |
| Tip | Straight | Straight | Straight |
| Marker band | Radiopaque | Radiopaque | Radiopaque |
| Catheter Coating | Hydrophilic | Hydrophilic | Hydrophilic |
| Guidewire
compatibility | Yes | Yes | Yes |
| Catheter Sizes | 3F, 4F, 5F, 6F | 3.4F, 5F, 6F, 8F | 3F, 4F, 5F, 6F |
| Table-3: Comparison with Predicate Device | | | |
| Feature | Subject Device
Merlin Aspiration
System | Predicate Device
INDIGO Aspiration
System | Reference Device
MIVI Q Distal Access
Catheter |
| Total Length (cm) | 148cm,135cm,130cm,130cm | 150cm, 132cm, 135cm, 85cm | 148cm,135cm,130cm,130cm |
| Catheter Biomaterials | Patient contacting
materials are listed in
Table 11-3 in Section
11.7 | Biocompatible,
commonly utilized for
interventional devices | Same as Merlin Aspiration
Catheter |
| Principle of Operation | The principle of operation
is for removal of fresh,
soft emboli and thrombi
in the peripheral arterial
system, using an
endovascular
approach, under
fluoroscopy.
Contrast is often used
during catheter
placement. | The principle of operation
is for removal of fresh,
soft emboli and thrombi
in the peripheral arterial
and venous system, using
an endovascular
approach, under
fluoroscopy. Contrast is
often used during catheter
placement. | Used to endovascularly insert
and guide microcatheters
under fluoroscopy
during diagnostic and/or
therapeutic procedures for
patients with arterial
disease or damage |
| Aspiration Tubing | | | |
| Materials | The materials list is
provided in listed in
Table 11-3 in Section
11.7 | Biocompatible,
commonly utilized for
interventional devices | N/A |
| Inner Diameter (ID) | 0.110 +/- .003 inches | 0.071"- 0.110" [1.8mm –
2.8mm] | N/A |
| Length | 112 +/- 3 inches | 112.0 [284.5cm] | N/A |
| Principle of Operation | The High Flow Tubing is
a suction tube used to
pass fluids from a
catheter to a retention
canister based on vacuum
applied by a pump. It is
connected between a
commercially available
aspiration catheter and a
chosen source of
vacuum. | Tubing consists of a short
tubing segment with an
inline filter, and
connectors on each end to
facilitate attachment to
the pump's vacuum port.
The tubing is provided
pre-attached to the
canister reservoir lid. | N/A |
| Accessories / Packaging / Sterilization / Shelf Life / Labeling | | | |
| Provided Accessories | Not provided with any
accessories. | Torque Device and
Introducer Sheath | Not provided with any
accessories |
| Table-3: Comparison with Predicate Device | | | |
| Feature | Subject Device
Merlin Aspiration
System | Predicate Device
INDIGO Aspiration
System | Reference Device
MIVI Q Distal Access
Catheter |
| Required Accessories | 9F Rotating
hemostasis
valve Y-
Connector 8F guide
catheter / 6F
guide sheath 90-
95 cm in length
(Required
Accessory) | None | 9F Rotating
hemostasis valve Y-
Connector 8F guide catheter /
6F guide sheath 90-
95 cm in length
(Required
Accessory) |
| Package Configuration | Catheter inserted in a
plastic tube, mounted on
an insert card, and sealed
in a pouch. Sealed pouch
packaged in carton along
with Instructions for Use.
High Flow Tubing is
placed in a sealed,
labeled Tyvek pouch
packaged in a labeled
box. | Commonly utilized for
interventional devices | Catheter inserted in a plastic
tube, mounted on an insert
card, and sealed in a pouch.
Sealed pouch packaged in
carton along with Instructions
for Use. |
| Sterile & Non-
pyrogenic? | Yes | | |
| Sterilization Method | Ethylene Oxide (EO) | | |
| Shelf Life | 3 years | | |
| Single use | Yes | Yes | Yes |
| Labeling | Proposed IFU | Current IFU | Current IFU |
5
6
7
Performance Testing
Bench Testing
The Merlin Aspiration System is identical to the MIVI Q Distal Access Catheter (K192558), which has been evaluated through risk analysis and design verification testing, which confirmed the device met design specifications.
Testing includes:
Table 4: Design Verification Testing | |||
---|---|---|---|
Merlin Aspiration Catheter | |||
Test | Result | ||
Coating Particulate Testing | Met established criteria | ||
Coating Adhesion | Met established criteria |
8
Coating Uniformity | Met established criteria |
---|---|
Surface Integrity | Met established criteria |
Tip Inspection | Met established criteria |
Liquid Leakage under Pressure | Met established criteria |
Dimensional Verification | Met established criteria |
Kink Resistance | Met established criteria |
Push/Track | Met established criteria |
Suction Flow Rate | Met established criteria |
System Introduction / Aspiration of Thromboemboli | Met established criteria |
Tensile Strength-Push Wire | Met established criteria |
Tensile Strength-Distal Section | Met established criteria |
Catheter Torque Testing | Met established criteria |
Catheter Burst Testing | Met established criteria |
High Flow Tubing | |
Dimensional Verification | Met established criteria |
Vacuum Pressure | Met established criteria |
Flow Rate | Met established criteria |
Degree of Collapse | Met established criteria |
Simulated Use | Met established criteria |
Leak Test | Met established criteria |
Distal Male Luer Tubing Tensile Strength | Met established criteria |
Female Luer Tubing Tensile Strength | Met established criteria |
Proximal Male to Tubing Tensile Strength | Met established criteria |
Connector to Tubing Tensile Strength | Met established criteria |
Animal Testing
The Animal Testing data was performed in the porcine carotid, renal, and subclavian vasculature of 8 animals to ensure that there are no new questions of safety and effectiveness.
Biocompatibility
The MIVI Merlin Aspiration System patient contacting materials are the same materials used in the currently marketed reference device with an identical bio-contact. There is no change to colorants between the subject and reference devices.
Sterilization
Each Merlin Aspiration Catheter and High Flow Tubing device meets EO and ECH limits specified in the ISO 10933-7 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals (2008).
The sterilization validation for Merlin Aspiration Catheter and High Flow tubing was validated in accordance with ISO 11135-1: 2014 Sterilization of healthcare product- Ethylene oxide -Part 1: Requirements for development validation and routine control of a sterilization process for medical devices (2014).
Shelf-life and Expiration Dating
9
The Merlin Aspiration System will be labeled with an expiration date of 3 years from the date of sterilization, which is same as predicate and reference devices. There is no change in materials used and this does not impact the shelf-life of the product.
Substantial Equivalence Summary and Conclusion
The Merlin Aspiration System and the predicate device, the INDIGO Aspiration System, have identical indications, principles of operation, and a comparable design, but the Merlin catheter is indicated only for peripheral arterial use. While there are technological differences between the Merlin Aspiration System and the predicate device which are related to exclusion of separators, product labeling, and size matrix (3-6F and 130-148 cm lengths for the Merlin catheter, 3.4-8F and 85-150 cm lengths for the INDIGO aspiration catheter, 0.110+/- 0.003 inches ID and 112+/-3 inches length for High Flow Tubing, 0.071-0.110 inches ID and 112.0 inches length for INDIGO Aspiration Tubing), the Merlin Aspiration catheter is identical to currently marketed Q Distal Access catheter reference device in design and materials, and the scope of the change includes an indication change and rebranding. Therefore, the Merlin Aspiration Catheter has been evaluated through risk analysis and design verification testing (DVT). The testing verified the device met the design specifications and is sufficiently robust and suitable for its intended use as the predicate device. Based on the predicate device comparison, risk assessment, and DVT information provided in this 510(k), the subject device has been shown to be appropriate for its intended use and is therefore considered substantially equivalent to the predicate device.