K Number
K213643
Date Cleared
2022-01-14

(57 days)

Product Code
Regulation Number
872.3770
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

UltraPrint-Dental Temp C&B UV is indication of temporary dental restorations in conjunction with extra-oral curing light equipment. Duration is not more than 28 days in oral environment.

Device Description

The Subject device is a light-cured methacrylate-based resin used in 3D printers for the production of temporary dental restorations. The Subject device is used by a dentist or dental technician for the CAD/CAM manufacturing of temporary dental restorations. Restorations fabricated using the Subject device are one-time use, temporary, prescription-only devices. The Subject device is a viscous solution consisting of methacrylate-based resins, photo initiators and pigments. Commonly used dental CAD software is used by dental professionals to virtually design a fixed indirect restoration and generate an industry-standard "STL" 3D dataset which reflects the intended shape and contour. The Subject resin is used within a validated manufacturing workflow to create the intended restoration. The Subject device is available in a variety of optional shades to reproduce the intended tooth shade of the restoration. Methacrylates are known materials, commonly used in the dental industry for fixed and removable prosthetic devices due to their physical-chemical, mechanical and biocompatible properties. The Subject device is intended to be sold by the bottle and used with compatible hardware 3D printers and their nesting software, and post-curing devices.

AI/ML Overview

The provided text is a 510(k) summary for the UltraPrint-Dental Temp C&B UV device and does not describe AI/ML performance. Instead, it focuses on demonstrating substantial equivalence to a predicate device (FREEPRINT temp) through non-clinical performance testing against established standards for dental materials. Therefore, much of the requested information regarding AI/ML models, such as sample size for test sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and training set details, is not applicable to this document.

Here's an analysis based on the provided text, focusing on the acceptance criteria and the study that proves the device meets them, as much as possible for a non-AI/ML device:

1. A table of acceptance criteria and the reported device performance

The document does not present a formal table of acceptance criteria with reported device performance in the way one would for an AI/ML diagnostic device with metrics like sensitivity, specificity, or AUC. Instead, it refers to compliance with ISO standards for material properties and biocompatibility as its "acceptance criteria," and the "reported performance" is that the device meets these requirements.

Acceptance Criteria (Standards Met)Reported Device Performance
Material Properties:
ISO 10477:2018 (Dentistry - Polymer-based crown and veneering materials):Meets the property requirements for: Depth of Cure, Flexural Strength, Water Solubility, Surface Finish, Shade Consistency, and Color Stability.
Biocompatibility:
ISO 10993-1:2018 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process)Biocompatibility testing was performed.
ISO 7405:2014 or 2018 (Dentistry - Evaluation of biocompatibility of medical devices used in dentistry)Biocompatibility testing was performed.
ISO 10993-3:2014 (Tests for genotoxicity, carcinogenicity and reproductive toxicity)Biocompatibility testing was performed (for genotoxicity).
ISO 10993-5:2009 (Tests for in vitro cytotoxicity)Biocompatibility testing was performed (for cytotoxicity).
ISO 10993-6:2016 (Tests for local effects after implantation)Biocompatibility testing was performed (for implantation).
ISO 10993-10:2013 (Tests for irritation and skin sensitization)Biocompatibility testing was performed (for irritation and sensitization).
ISO 10993-11:2017 (Tests for systemic toxicity)Biocompatibility testing was performed (for acute systemic toxicity, material-mediated pyrogenicity, subacute toxicity).
Shelf Life:Validated device packaging and material stability during storage.

Study Proving Acceptance Criteria Met:

The study proving the device meets these acceptance criteria is described as Non-Clinical Performance Testing.

2. Sample size used for the test set and the data provenance

For material property and biocompatibility testing, "test set" and "data provenance" don't apply in the same way they would for an AI/ML model. The "samples" would be physical specimens of the UltraPrint-Dental Temp C&B UV resin, prepared and cured according to the manufacturing workflow, and subjected to various laboratory tests as prescribed by the ISO standards. The specific number of physical samples used for each test (e.g., how many specimens for flexural strength, how many cell cultures for cytotoxicity) is not detailed in this document. The provenance is implicitly the manufacturer's testing facilities (Guangzhou Heygears Imc. Inc, China) or contracted labs, as these are non-clinical studies. These are prospective tests conducted on manufactured samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This concept is not applicable here. "Ground truth" in the context of this device refers to objective measurements against established ISO standards rather than expert interpretation of data. The "experts" involved would be technicians and scientists conducting the laboratory tests and interpreting the results against the numerical criteria defined in the ISO standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This concept is not applicable as there is no "adjudication" of expert opinions for ground truth establishment. The results are objective measurements from laboratory tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This document is for a dental resin, not an AI-powered diagnostic tool. Clinical data was specifically stated as not included in the submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not Applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is compliance with established international standards (ISO standards) for physical material properties and biocompatibility. This involves objective measurements obtained in laboratory settings.

8. The sample size for the training set

Not Applicable. This is not an AI/ML device, so there is no training set.

9. How the ground truth for the training set was established

Not Applicable. As there is no training set for an AI/ML model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 14, 2022

Guangzhou Heygears Imc. Inc Chris Brown Manager Aclivi, LLC 3250 Brackley Drive Ann Arbor, Michigan 48105

Re: K213643

Trade/Device Name: UltraPrint-Dental Temp C&B UV Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown And Bridge Resin Regulatory Class: Class II Product Code: EBG Dated: November 16, 2021 Received: November 18, 2021

Dear Chris Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

4. Indications for Use Statement

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K213643

Device Name UltraPrint-Dental Temp C&B UV

Indications for Use (Describe)

UltraPrint-Dental Temp C&B UV is indication of temporary dental restorations in conjunction with extra-oral curing light equipment. Duration is not more than 28 days in oral environment.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K213643 510(k) Summary GUANGZHOU HEYGEARS IMC.INC UltraPrint-Dental Temp C&B UV 12/30/2021

ADMINISTRATIVE INFORMATION

Manufacturer NameGUANGZHOU HEYGEARS IMC.INC
Block B2, 501, 601
Enterprise Accelerator, Kaifa District
Guangzhou, Guangdong, 510700, China
Telephone: +86-17612078809
Official ContactLin Liu - Registration Supervisor
Emailhzhang5@heygears.com

DEVICE NAME AND CLASSIFICATION

Proprietary/Device Name:UltraPrint-Dental Temp C&B UV
Common Name:Crown and Bridge, Temporary, Resin
Regulation Description:Temporary Crown And Bridge Resin
Regulation Number:21 CFR 872.3770
Device Class:Class II
Product Code:EBG
Review Panel:Dental
Reviewing Branch:Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)Dental Devices (DHT1B)

PREDICATE DEVICE INFORMATION

The Subject device in this submission is substantially equivalent in indications, use and design principles to the following Predicate device.

510(k)Predicate Device NameCompany Name
K200273FREEPRINT tempDETAX GmbH & Co.KG

INDICATIONS FOR USE

UltraPrint-Dental Temp C&B UV is indication of temporary dental restorations in conjunction with extra-oral curing light equipment. Duration is not more than 28 days in oral environment.

DEVICE DESCRIPTION

The Subject device is a light-cured methacrylate-based resin used in 3D printers for the production of temporary dental restorations. The Subject device is used by a dentist or dental technician for the CAD/CAM manufacturing of temporary dental restorations.

Restorations fabricated using the Subject device are one-time use, temporary, prescription-only devices. The Subject device is a viscous solution consisting of methacrylate-based resins, photo initiators and pigments.

Commonly used dental CAD software is used by dental professionals to virtually design a fixed indirect restoration and generate an industry-standard "STL" 3D dataset which reflects the intended shape and contour. The Subject

{5}------------------------------------------------

resin is used within a validated manufacturing workflow to create the intended restoration. The Subject device is available in a variety of optional shades to reproduce the intended tooth shade of the restoration. Methacrylates are known materials, commonly used in the dental industry for fixed and removable prosthetic devices due to their physical-chemical, mechanical and biocompatible properties.

The Subject device is intended to be sold by the bottle and used with compatible hardware 3D printers and their nesting software, and post-curing devices.

EQUIVALENCE TO MARKETED DEVICE

The Subject device is highly similar to the Primary Predicate device with respect to Indications for Use and technological principles. The comparison tables below compare the Indications for Use and Technological Characteristics of the Subject and Predicate devices.

Indications For Use

DeviceIndications for Use Statement
Subject DeviceUltraPrint-Dental Temp C&B UVGUANGZHOU HEYGEARS IMC.INCUltraPrint-Dental Temp C&B UV is indicated for the fabrication of temporary dental restorations inconjunction with extra-oral curing light equipment. Duration is not more than 28 days in oralenvironment.
Predicate DeviceFREEPRINT temp (K200273)DETAX GmbH & Co.KGFREEPRINT temp is indicated for the fabrication of temporary dental restorations in conjunctionwith extra-oral curing light equipment. Duration is less than 30 days in oral environment.

The Subject and Predicate Indications for Use Statement (IFUS) are highly similar, differing only by the device name and two days difference in duration. Slight differences in the Indications for Use Statements does not change the intended use of the Subject and Reference device to fabricate temporary dental restorations.

{6}------------------------------------------------

Technological Characteristics

ParameterSubject DevicePredicate DeviceComparison withPredicate Device
UltraPrint-Dental Temp C&B UVGUANGZHOU HEYGEARS IMC.INCFREEPRINT temp (K200273)DETAX GmbH & Co.KG
Product CodeEBGEBGSame
Regulation Number872.3770872.3770Same
Regulatory ClassClass IIClass IISame
Intended UseUltraPrint-Dental Temp C&B UV isindicated for the fabrication oftemporary dental restorations inconjunction with extra-oral curing lightequipment. Duration is not more than28 days in oral environment.FREEPRINT temp is indicated for thefabrication of temporary dentalrestorations in conjunction with extra-oral curing light equipment. Duration isless than 30 days in oral environment.Highly Similar
Technology3D liquid (light-cured) print resin fordental CAD/CAM3D liquid (light-cured) print resin fordental CAD/CAMSame
MaterialMethacrylate polymer resin(includes dimethacrylate)Methacrylate polymer resin(dimethacrylate)Highly Similar
Material ShadesCommon VITA-shadesCommon VITA-shadesSame
BiocompatibleYesYesSame
OTC or RxRxRxSame
SterileNon-sterileNon-sterileSame
Chemical CompositionMethacrylate polymer resin with photoinitiator, and pigments(Meth)acrylate-based resin with photoinitiator and pigmentsHighly Similar
Polymerization (Curing)MethodVisible light, 385 nmw/post curingVisible light, 385 nmw/post curingSame
EquipmentValidated 3D-Printer and post curingdevicesValidated 3D-Printer and post curingdevicesSame
Performance TestingISO 10477:2018ISO 10477:2018Same
Biocompatibility TestingISO 7405:2018ISO 10993-1:2018ISO 10993-3:2014ISO 10993-5:2009ISO 10993-6:2016ISO 10993-10:2013ISO 10993-11:2017ISO 7405:2014ISO 10993-1:2018ISO 10993-3:2014ISO 10993-5:2009ISO 10993-10:2013ISO 10993-11:2017Highly Similar

The Technological Characteristics of the Subject and Predicate devices are the same or Highly Similar.

Material/Chemical Composition - The Subject and Predicate devices are same Highly Similar in they are both methacrylate polymer resins. Slight differences in chemical composition do not change the intended use of the Subject and Predicate devices to be used in the fabrication of temporary dental prostheses. The Subject device has demonstrated suitability for intended use through material non-clinical performance testing.

Technological differences between the Subject and Predicate devices have been evaluated through non-clinical performance testing. The results of the tests performed show that Subject device meets the requirements mentioned in the applicable standards and confirm that the device performs similarly to Predicate device.

CLINICAL TESTING

The performance of methacrylate-based polymer resins in the clinical environment has been well established. No clinical data is included in this submission.

NON-CLINICAL PERFORMANCE TESTING

Validation of the manufacturing process was performed demonstrating consistency of the process output with that of the process input and confirm compatibility with 3D Printers and Post-curing equipment. Assessment of printer compatibility was performed through evaluation of light output, printer resolution, and printer accuracy testing to internal protocols and confirmatory biocompatibility testing.

Physical material property testing was performed on the Subject device to ISO 10477:2018, Dentistry - Polymerbased crown and veneering materials. Results demonstrated the Subject device meets the property requirements

{7}------------------------------------------------

of the referenced standard for; Depth of Cure, Flexural Strength, Water Solubility, Surface Finish, Shade Consistency and Color Stability.

A biological evaluation was performed on the Subject device. Biocompatibility testing was performed on the Subject device according to ISO 10993-1:2018 and ISO 7405:2014. Test included: Cytotoxicity, Irritation, Sensitization, Acute Systemic Toxicity, Material-mediated Pyrogenicity, Subacute Toxicity, Genotoxicity and Implantation.

Accelerated and Real-time Shelf life testing was performed to validate device packaging and material stability during storage.

No animal testing data is included in this submission.

Non-clinical performance testing of the Subject device met the acceptance criteria for each validation and test described above. This non-clinical performance testing demonstrates that the Subject device is suitable for intended use.

CONCLUSION

Overall, the Indications for Use statements for the Subject and Predicate devices are highly similar differing only in device name and slightly in use duration. Overall, the Technological Characteristics of the Subject device are the same or highly similar to the Predicate device with any differences mitigated through nonclinical performance testing.

Overall, these similarities between the Subject and Predicate devices, support a determination of substantial equivalence.

§ 872.3770 Temporary crown and bridge resin.

(a)
Identification. A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.(b)
Classification. Class II.