K Number
K213351
Date Cleared
2021-10-29

(21 days)

Product Code
Regulation Number
892.1000
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The dS Head 32ch 3.0T MR Coil is intended to be used in conjunction with a Philips 3.0T Magnetic Resonance Scanner to produce diagnostic images of the head on adult and pediatric patients that can be interpreted by a trained physician.

Device Description

The subject dS Head 32ch 3.0T Coil is a receive-only, phased-array radiofrequency (RF) coil designed for high-resolution head examinations on adult and pediatric patients. The coil is intended for use with Philips magnetic resonance (MR) scanners with a 3.0 tesla magnetic field strength.

AI/ML Overview

The provided text describes the 510(k) summary for the Philips dS Head 32ch 3.0T MR Coil. This document focuses on the safety and performance criteria for the device itself (a medical imaging coil), not on the performance of an AI algorithm interpreting images produced by the device.

Therefore, the information typically requested for AI/algorithm performance (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) is not applicable or not provided in this document as it's not a submission for an AI/ML powered device.

However, I can extract the relevant information regarding the device's acceptance criteria and the study proving it meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria / Standards AppliedReported Device Performance
Image QualityImage Signal to Noise and Image Uniformity characterization (NEMA MS 1, 3, 9 and IEC 62464-1)Met all criteria outlined
Safety - HeatingSurface heating (ANSI/AAMI ES 60601-1 and NEMA MS 14)Met all criteria outlined
Diagnostic UtilityAcquired Image quality assessed by a U.S. Board Certified radiologist to confirm images are sufficient for diagnostic use on both adult and pediatric patient populationsConfirmed sufficient quality for diagnostic use
Safety - DecouplingPresence of decoupling mechanismsMet all criteria outlined (decoupling mechanisms are present)
Safety - EMCEMC – Immunity, electrostatic discharge (IEC 60601-1-2)Met all criteria outlined
Safety - Electrical/MechanicalGeneral electrical/mechanical safety (IEC 60601-2-33 and AAMI/ANSI ES 60601-1)Met all criteria outlined
BiocompatibilityBiocompatibility evaluation (ISO 10993 series)Met all criteria outlined

The study proving the device meets the acceptance criteria is detailed under "Summary of Non-Clinical and Clinical Performance Data":

The subject dS Head 32ch 3.0T Coil met all safety and performance criteria outlined in the FDA guidance "Magnetic Resonance (MR) Receive-only Coil – Performance Criteria for Safety and Performance Based Pathway" issued December 11, 2020.

The specific tests performed include:

  • Image Signal to Noise and Image Uniformity characterization (NEMA MS 1, 3, 9 and IEC 62464-1)
  • Surface heating (ANSI/AAMI ES 60601-1 and NEMA MS 14)
  • Acquired Image quality assessment by a U.S. Board Certified radiologist
  • Confirmation of the presence of decoupling mechanisms
  • EMC – Immunity, electrostatic discharge (IEC 60601-1-2)
  • General electrical/mechanical safety (IEC 60601-2-33 and AAMI/ANSI ES 60601-1)
  • Biocompatibility evaluation (ISO 10993 series)

Regarding the specific questions about an AI study (which are not directly applicable to this device submission as it's not an AI device):

2. Sample size used for the test set and the data provenance: Not applicable. This submission is for an MR coil and focuses on its physical and imaging performance, not an AI algorithm's diagnostic performance on a test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for an AI algorithm test set. For the device's image quality assessment, it states "Acquired Image quality was assessed by a U.S. Board Certified radiologist." It does not specify the number beyond "a" radiologist.

4. Adjudication method for the test set: Not applicable for an AI algorithm test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.

7. The type of ground truth used: For the device's performance, physical measurements and adherence to engineering standards (NEMA, IEC, ANSI/AAMI, ISO) served as ground truth for technical performance aspects. For diagnostic utility, the "U.S. Board Certified radiologist" assessment of image quality serves as the ground truth (expert opinion) that the images produced are suitable for diagnostic use.

8. The sample size for the training set: Not applicable. This device does not use an AI algorithm that requires a training set.

9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Invivo Corporation (Business Trade Name: Philips) % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL, Minnesota 55114

Re: K213351

Trade/Device Name: ds Head 32ch 3.0T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: October 7, 2021 Received: October 8, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213351

Device Name dS Head 32ch 3.0T

Indications for Use (Describe)

The dS Head 32ch 3.0T MR Coil is intended to be used in conjunction with a Philips 3.0T Magnetic Resonance Scanner to produce diagnostic images of the head on adult and pediatric patients that can be interpreted by a trained physician.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The word is written in a bright blue color. The background is white, which makes the blue color of the word stand out. The image is a clear and simple representation of the Philips logo.

510(k) Summary K213351

prepared in accordance with 21 CFR §807.92

Date Prepared:October 26, 2021
Manufacturer:Invivo Corporation(Business Trade Name: Philips)3545 SW 47th AveGainesville, FL 32608Establishment Registration #1056069
Primary ContactPerson:Sarah PleaughRegulatory Affairs Specialistsarah.pleaugh@philips.com
Secondary ContactPerson:Ann LebarHead of Regulatory Affairsann.lebar@philips.com
Device Name:dS Head 32ch 3.0T
Classification:Classification Name:Coil, Magnetic Resonance, Specialty
Classification Regulation:21 CFR 892.1000
Classification Panel:Radiology
Device Class:Class II
Primary Product Code:MOS
Primary PredicateDevice:Trade name:Models HRB-127-32 High ResolutionBrain Coil
Manufacturer:Invivo Corporation
510(k) Clearance:K082916
Classification Name:Coil, Magnetic Resonance, Specialty
Classification Regulation:21 CFR 892.1000
Classification Panel:Radiology
Device Class:Class II
Primary Product Code:MOS
Reference Device:Trade name:Achieva, Intera, Ingenia, Ingenia CX,Ingenia Elition, and Ingenia Ambition MRSystems
Manufacturer:Philips Medical Systems Nederland B.V.
510(k) Clearance:K110151/K193215
Classification Name:Magnetic Resonance Diagnostic Device
Classification Regulation:21CFR 892.1000
Classification Panel:Radiology
Device Class:Class II
Product Codes:LNH, LNI, MOS
DeviceDescription:The subject dS Head 32ch 3.0T Coil is a receive-only, phased-arrayradiofrequency (RF) coil designed for high-resolution head examinationson adult and pediatric patients. The coil is intended for use with Philipsmagnetic resonance (MR) scanners with a 3.0 tesla magnetic fieldstrength.
Indications forUse:The dS Head 32ch 3.0T MR Coil is intended to be used in conjunctionwith a Philips 3.0T Magnetic Resonance Scanner to produce diagnosticimages of the head on adult and pediatric patients that can beinterpreted by a trained physician.
FundamentalScientificTechnology:The subject dS Head 32ch 3.0T MR Coil is similar in design, material, andenergy source to the legally marketed predicate device, Models HRB-127-32 High Resolution Brain Coil (K082916). The subject and predicatedevices are based on the following same technological elements:· Prescription use· Coil designs are 32-channel, receive-only, phased array RF coils· Compatible field strength (3.0T)· Decoupling methodology· Patient contacting materials are known materials that have beenassessed for compliance with recognized biocompatibility standards· Energy source for the coils is the MRI scanner· Coils designed for imaging the head anatomy· Coil mechanical design is a split housing to facilitate patientpositioning· Manufactured for use with Philips MRI scannersThe following technological differences exist between the subject andpredicate devices:· Predicate device is an analog coil whereas the subject dS Head 32ch3.0T MR coil is a digital coil.
· Different system cable connectors to interface with the digital MRI
scanners.
These differences in technology are supported by the MR system anddigital coil designs cleared in the reference device Philips Ingenia MRSystems (K110151/K193215), as well as the safety and performancetesting provided in this submission. The subject coil performs asintended and does not raise new issues of safety or effectiveness.
Summary of Non-Clinical and ClinicalPerformance Data:The subject dS Head 32ch 3.0T Coil met all safety and performancecriteria outlined in the FDA guidance "Magnetic Resonance (MR) Receive-only Coil – Performance Criteria for Safety and Performance BasedPathway" issued December 11, 2020:
- Image Signal to Noise and Image Uniformity characterization(NEMA MS 1, 3, 9 and IEC 62464-1)
- Surface heating (ANSI/AAMI ES 60601-1 and NEMA MS 14)
- Acquired Image quality was assessed by a U.S. Board Certifiedradiologist to confirm images produced on the subject coil aresufficient quality for diagnostic use on both adult and pediatricpatient populations
- Presence of decoupling mechanisms
- EMC – Immunity, electrostatic discharge (IEC 60601-1-2)
- General electrical/mechanical safety (IEC 60601-2-33 andAAMI/ANSI ES 60601-1)
- Biocompatibility evaluation (ISO 10993 series)
SubstantialEquivalenceConclusion:Substantial equivalence of the dS Head 32ch 3.0T MR Coil isdemonstrated through the Safety and Performance Based Pathway formagnetic resonance (MR) receive-only coils. The subject device has thesame indications for use and technological characteristics as thepredicate and reference devices. Substantially equivalent performance isdemonstrated by meeting all criterion in the guidance "MagneticResonance (MR) Receive-only Coil – Performance Criteria for Safety andPerformance Based Pathway" issued on December 11, 2020.

{4}------------------------------------------------

PHILIPS

{5}------------------------------------------------

PHILIPS

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.