(110 days)
The single use andorate® Suction Valve is used to control the suction of an endoscope during GL endoscopic procedures.
The single use andorate® Air/Water Valve is used to control the air / water function of an endoscope during GI endoscopic procedures.
The subject devices are intended for single-use and are supplied sterile. The suction valve is designed to be attached to the suction port of the endoscope and the air/water valve is designed to be attached to the air/water port of the endoscope. The activation of the suction valve allows the user to aspirate excess fluids or other debris obscuring the endoscope image, while the activation of the air/water valve allows the user to control air and water flow to assist in cleansing the lens during procedures.
This is a 510(k) premarket notification for andorate Suction Valve and andorate Air/Water Valve. The document states that performance tests were conducted to support substantial equivalence to a predicate device. However, the document does not provide specific acceptance criteria or detailed results of these tests in a quantitative manner that would allow for a table of acceptance criteria and reported performance. It only lists the types of tests performed.
Therefore, many of the requested details cannot be extracted from the provided text.
Here is the information that can be extracted:
1. A table of acceptance criteria and the reported device performance
The provided document lists the types of performance tests conducted but does not specify the numerical acceptance criteria or the quantitative results of these tests.
| Performance Test | Acceptance Criteria (Not explicitly stated in document) | Reported Device Performance (Not explicitly stated in document with quantitative values) |
|---|---|---|
| Suction Valve | ||
| Endoscope Compatibility Test | (e.g., Compatible with specified endoscopes) | Met established specifications |
| Suction Flow Test | (e.g., Flow rate within X-Y range) | Met established specifications |
| Vacuum Leak Test | (e.g., Leakage below Z level) | Met established specifications |
| Pressing Force Test | (e.g., Force within A-B range) | Met established specifications |
| Air/water Valve | ||
| Endoscope Compatibility Test | (e.g., Compatible with specified endoscopes) | Met established specifications |
| Air Flow Test | (e.g., Flow rate within X-Y range) | Met established specifications |
| Water Flow Test | (e.g., Flow rate within P-Q range) | Met established specifications |
| Water Leakage Test | (e.g., Leakage below R level) | Met established specifications |
| Pressing Force Test | (e.g., Force within C-D range) | Met established specifications |
| Backflow Prevention Test | (e.g., No backflow detected) | Met established specifications |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not specify the sample size for the test set or the data provenance. It only states that "bench testing was performed."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not applicable as the study involved bench testing (performance tests) of a medical device (suction and air/water valves), not data analysis requiring expert interpretation or ground truth establishment.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not applicable for bench testing of a medical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable as the study involved bench testing of a physical medical device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable for bench testing of a physical medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For bench testing, the "ground truth" would be the engineering specifications and performance standards that the device is designed to meet. The document states that the performance data "demonstrated that the subject devices met established specifications."
8. The sample size for the training set
This information is not applicable as the study involved bench testing of a physical medical device, not an AI model that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable as there was no training set for an AI model. For the device itself, the "ground truth" or acceptable performance standards for the bench tests would have been established based on engineering specifications and relevant regulatory guidelines.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 12, 2022
GA Health Company Limited Cindy Ye Chief Executive Officer Unit 18, 21/F, Metropole Square, 2 On Yiu Street, Shatin Hong Kong, CHINA
Re: K213095
Trade/Device Name: andorate Suction Valve, andorate Air/Water Valve Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: ODC Dated: December 13, 2021 Received: December 16, 2021
Dear Cindy Ye:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213095
Device Name andorate® Suction Valve andorate® Air/Water Valve
Indications for Use (Describe)
The single use andorate® Suction Valve is used to control the suction of an endoscope during GL endoscopic procedures.
The single use andorate® Air/Water Valve is used to control the air / water function of an endoscope during GI endoscopic procedures.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K213095 Page 1 of 4
510(k) Premarket Notification: Traditional Suction Valve and Air/Water Valve GA Health Company Limited CONFIDENTIAL
510(k) Summary
The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.
1. Submission Sponsor
Submitter's Name: Submitter's Address:
Establishment Registration No.:
2. Sponsor Contact
2 On Yiu Street, Shatin, N.T, Hong Kong. CHINA 3014749926
GA Health Company Limited Unit 18, 21/F, Metropole Square
Cindy Ye Chief Executive Officer Telephone: +852 2833 9010 Email: cindy@gahealth.net
Lam Wing Yu Senior Requlatory Affairs Specialist Telephone: +852 2833 9010 Email: rainy@gahealth.net
3. Date Prepared
16 November 2021
4. Device Identification
Device Name: Common Name: Classification Number: Classification Name: Product Code: Product Code Name: Regulation Class: Device Panel:
Device Name: Common Name: Classification Number: Classification Name: Product Code: Product Code Name: Regulation Class: Device Panel:
andorate® Suction Valve Suction valve for endoscope 21 CFR 876.1500 Endoscope and accessories ODC Endoscope channel accessory 2 Gastroenterology/Urology
andorate® Air/Water Valve Air/Water valve for endoscope 21 CFR 876.1500 Endoscope and accessories ODC Endoscope channel accessory 2 Gastroenterology/Urology
{4}------------------------------------------------
5. Predicate Device Identification
Predicate Device 510(k) No .: Predicate Device Trade Name: Predicate Device Product Code: K180341 FUJIFILM 600 Series Endoscope EG-600WR v2 FDS – Endoscope and Accessories
6. Device Description:
The subject devices are intended for single-use and are supplied sterile. Table 1 shows the components included in the submission.
| Components | Qty | Product Code | RegulationNumber | RegulatoryClassification |
|---|---|---|---|---|
| andorate® Suction Valve | 1 | ODC – Endoscopechannel accessory | 21 CFR 876.1500 | 2 |
| andorate® Air/WaterValve | 1 | ODC – Endoscopechannel accessory | 21 CFR 876.1500 | 2 |
Table 1 - Components included in the Submission
The suction valve is designed to be attached to the suction port of the endoscope and the air/water valve is designed to be attached to the air/water port of the endoscope. The activation of the suction valve allows the user to aspirate excess fluids or other debris obscuring the endoscope image, while the activation of the air/water valve allows the user to control air and water flow to assist in cleansing the lens during procedures.
There were no prior submissions for the subject devices.
7. Intended Use:
The single use andorate® Suction Valve is used to control the suction function of an endoscope during GI endoscopic procedures.
The single use andorate® Air/Water Valve is used to control the air / water function of an endoscope during GI endoscopic procedures.
8. Technological Characteristics
Table 2 summarizes the suction valve and air/water valve technological characteristics as compared to the predicate devices.
Table 2 - Summary of design, features and principles of operation and technological characteristics between the subject device and predicate devices
{5}------------------------------------------------
K213095 Page 3 of 4
510(k) Premarket Notification: Traditional Suction Valve and Air/Water Valve GA Health Company Limited CONFIDENTIAL
| Specification | Predicate Device | Subject Device | SubstantialEquivalence |
|---|---|---|---|
| Product code | FDS | ODC | SubstantialEquivalent |
| RegulatoryClassification | 2 | 2 | Identical |
| Regulation No | 21 CFR 876.1500 | 21 CFR 876.1500 | Identical |
| RegulationDescription | Endoscope and accessories | Endoscope and accessories | Identical |
| Intended Use | This product is the suction buttonused in combination with ourendoscopes in medical facilities.Do not use this product for anyother purpose.This product is the air/waterbutton used in combination withour endoscopes in medicalfacilities. Do not use this productfor any other purpose. | The single use andorate®Suction Valve is used to controlthe suction function of anendoscope during GI endoscopicprocedures.The single use andorate®Air/Water Valve is used to controlthe air / water function of anendoscope during GI endoscopicprocedures. | SubstantialEquivalent |
| Environment ofUse | Hospital and or clinics | Hospital and or clinics | Identical |
| Single Use orReusable | Reusable | Single Use | SubstantialEquivalent |
| Material | Fluorine Resin, Stainless Steel,Silicone, Thermoplastics | Polycarbonate, AcrylonitrileButadiene Styrene,Thermoplastic Elastomer,Silicone Rubber, Stainless Steel | SubstantialEquivalent |
| Manufacturingmethod | Injection molding, overmolding | Injection molding, compressionmolding, overmolding | SubstantialEquivalent |
| Packaging | Packaged in a plastic bag | Packaged in a sealed pouch | SubstantialEquivalent |
| Sterilization | Supplied non-sterileSterilized after each use | Yes | SubstantialEquivalent |
| Shelf Life | 1 year | 1 year | SubstantialEquivalent |
9. Performance Test
The bench testing was performed to support substantial equivalence on both the subject device and the predicate device. The performance data demonstrated that the subject devices met established specifications in the following non-clinical tests.
Suction Valve
-
- Endoscope Compatibility Test
-
- Suction Flow Test
-
- Vacuum Leak Test
-
- Pressing Force Test
{6}------------------------------------------------
K213095 Page 4 of 4
510(k) Premarket Notification: Traditional Suction Valve and Air/Water Valve GA Health Company Limited CONFIDENTIAL
Air/water Valve
-
- Endoscope Compatibility Test
-
- Air Flow Test
-
- Water Flow Test
-
- Water Leakage Test
-
- Pressing Force Test
-
- Backflow Prevention Test
10. Biocompatibility
The biocompatibility of the subject devices was conducted in accordance with the FDA guideline "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Biocompatibility testing is conducted on subject device in accordance with the ISO 10993 standard. The test result shows that both the subject devices are biocompatible.
11. Conclusion
The subject devices have the same intended use as the predicate devices. Based on the technological characteristics and overall performance of the devices in bench testing, there are no significant differences exist between the subject devices and the predicate devices. The subject devices do not raise any new issues of safety and effectiveness. From a clinical perspective and comparing design specifications, the subject devices and the predicate device are substantially equivalent.
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.