K Number
K213006
Date Cleared
2021-12-21

(92 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vinyl Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Device Description

The subject device is powder free vinyl examination gloves. The subject device is clear. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for Vinyl Examination Gloves, not an AI/ML medical device. Therefore, the information requested in the prompt regarding AI/ML device acceptance criteria and study details (such as sample sizes for test and training sets, expert ground truth, multi-reader multi-case studies, etc.) is not applicable and cannot be extracted from this document.

The document focuses on demonstrating substantial equivalence of the Vinyl Examination Gloves to a predicate device through non-clinical performance testing.

Here's a breakdown of the non-clinical acceptance criteria and reported performance, as directly extractable from the provided text, adapted to the table format requested:

1. A table of acceptance criteria and the reported device performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6319Physical Dimensions Test (Length)Length(mm): S/M/L/XL: ≥ 230Length(mm): > 230/Pass
ASTM D6319Physical Dimensions Test (Width)Width(mm): S: 85±5; M: 95±5; L: 105±5; XL: 115±5Width(mm): S: 80-89 /Pass; M: 90-98/ Pass; L: 100-107/ Pass; XL:111-118/ Pass
ASTM D6319Physical Dimensions Test (Thickness)Thickness (mm): Finger: ≥ 0.08; Palm: ≥ 0.08Thickness (mm): Finger: 0.08-0.17/Pass; Palm: 0.08/Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125/Pass
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.16-0.20mg/Pass
ASTM D412Physical properties (Tensile Strength - Before Aging)≥11MPa11-27MPa/Pass
ASTM D412Physical properties (Ultimate Elongation - Before Aging)≥300%304-607%/Pass
ASTM D412Physical properties (Tensile Strength - After Aging)≥11MPa11-24MPa/Pass
ASTM D412Physical properties (Ultimate Elongation - After Aging)≥300%300-499%/Pass
ISO 10993-5CytotoxicityIn Vitro Cytotoxicity (Pass, implied by "Under conditions of the study, device extract is cytotoxic" for predicate but not for subject device - clarification needed for subject device's specific result from the provided table)[Result not explicitly stated as "Pass" or "Fail" in the table for the main device, but a separate analysis states "Under conditions of the study, device extract is cytotoxic" which would imply fail for the subject device - this seems contradictory to implied substantial equivalence and needs careful interpretation of the table. Assuming a pass based on the conclusion.]
ISO 10993-11Acute Systemic ToxicityNon-acute systemic toxicityUnder the condition of acute systemic toxicity test, the test article did not show acute systemic toxicity in vivo / Pass
ISO 10993-10IrritationNon-irritating[Result not explicitly stated as "Pass" or "Fail" in the table for the main device, but a separate analysis states "Under the conditions of the study, not an irritant or a sensitizer" for subject device which would imply pass.]
ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer / Pass

Regarding the other points, as this is a traditional medical device (gloves) and not an AI/ML algorithm, the following points are not applicable:

  1. Sample sized used for the test set and the data provenance: Not an AI/ML device. Test "set" refers to the specific gloves tested according to standards, not a dataset.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for gloves is defined by physical/chemical standards.
  3. Adjudication method: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done: No, this is for AI/ML devices assisting human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, this is for AI/ML algorithms.
  6. The type of ground truth used: Not applicable in the AI/ML sense. Ground truth is established by standardized physical and chemical testing methods (e.g., ASTM, ISO).
  7. The sample size for the training set: Not applicable; there is no training set for gloves.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 21, 2021

Taian Hengchang Medical Technology Co.,Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 1801. No. 161 Luiiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K213006

Trade/Device Name: Vinyl Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: September 6, 2021 Received: September 20, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213006

Device Name Vinyl Examination Gloves

Indications for Use (Describe)

The Vinyl Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K213006

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Taian Hengchang Medical Technology Co.,Ltd. Address: No. 88 Central Road, Sunbo Town Industrial Park in Feicheng, Taian City, Shandong Province, China. Phone Number: +86 15963319855 Contact: Jing Li Date of Preparation: Sept 6, 2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Vinyl Examination Gloves Common name: Vinyl Patient Examination Glove Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

3.0 Classification

Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: ZHICHENG TRADING CO., LTD. Device: Vinyl Examination Glove (Clear, Non-Colored) 510(k) number: K180861

{4}------------------------------------------------

5.0 Indication for Use

The Vinyl Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free vinyl examination gloves. The subject device is clear. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

ItemSubject DevicePredicated DeviceRemark
510(k) numberK213006K180861/
Product CodeLYZLYZSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseThe Vinyl ExaminationGlove is a disposabledevice intended formedical purposes that isworn on the examiner'shands to preventcontamination betweenpatient and examiner.The Vinyl ExaminationGlove (Clear,Non-Colored) is adisposable deviceintended for medicalpurposes that is worn onthe examiner's hands toprevent contaminationbetween patient andexaminer.Same
MaterialVinylVinylSame
Powdered orPowered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
ColorantClearClearSame
LabelingInformationSingle-use indication,powder free, devicecolor, device name,glove size and quantity,Non-SterileSingle-use indication,powder free, devicecolor, device name,glove size and quantity,Non-SterileSame
Dimensions(mm)Length:S/M/L/XL: ≥230;Width:S: 85+5;Length:S≥230;M≥235;L≥245;Similar

Table1-General Comparison

{5}------------------------------------------------

M: 95±5;L: 105±5;XL: 115±5XL≥245;Width:S: 85±5;M: 95±5;L: 105±5;XL: 115±5
Thickness(mm)Finger: ≥0.08;Palm: ≥0.08Finger: ≥0.05;Palm: ≥0.08Similar
PhysicalPropertiesBeforeAgingTensile Strength11MPa, minTensile Strength15MPa, minDifferent
Ultimate Elongation300% minUltimate Elongation380% minDifferent
AfterAgingTensile Strength11MPa, minTensile Strength15MPa, minDifferent
Ultimate Elongation300%minUltimate Elongation380%minDifferent
Freedom fromHolesBe free from holes when tested in accordance with ASTMD5151AQL=2.5Be free from holes when tested in accordance with ASTMD5151AQL=1.5Different
Powder ContentMeet the requirements of ASTM D6124Meet the requirements of ASTM D6124Same
BiocompatibilityISO 10993-10;Under the conditions of the study, not an irritant or a sensitizerComply with ISO10993-10Same
ISO 10993-5Under conditions of the study, device extract is cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells. Complies with ISO 10993-5/
ISO 10993-11;Under the condition of acute systemic toxicity test, the test article did not show acute systemic toxicity in vivo/

Analysis: The physical dimensions are different with that of the predicate.

{6}------------------------------------------------

Analysis: The tensile strength and ultimate elongation are different with that of the predicate.

8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

TestMethodPurposeAcceptance CriteriaResults
ASTMD6319PhysicalDimensionsTestLength(mm):S/M/L/XL: $\ge$ 230;Width(mm):S: 85±5;M: 95±5;L: 105±5;XL: 115±5Thickness (mm):Finger: $\ge$ 0.08Palm: $\ge$ 0.08Length(mm):> 230/Pass;Width(mm):S: 80-89 /PassM: 90-98/ PassL: 100-107/ PassXL:111-118/ PassThickness (mm):Finger: 0.08-0.17/PassPalm: 0.08/Pass
ASTMD5151WatertightnessTest forDetection ofHolesMeet the requirements of ASTM D5151AQL 2.50/125/Pass

Table 2 - Summary of non-clinical performance testing

{7}------------------------------------------------

ASTMD6124PowderContentMeet the requirements of ASTM D6124 < 2.0mg0.16-0.20mg/Pass;
ASTMD412PhysicalpropertiesBeforeAgingTensileStrength≥11MPa11-27MPa/Pass;
UltimateElongation≥300%304-607%/Pass;
AfterAgingTensileStrength≥11MPa11-24MPa/Pass;
UltimateElongation≥300%300-499%/Pass;
ISO10993-5CytotoxicityIn Vitro Cytotoxicity
ISO10993-11Acute SystemicToxicityNon- acute systemictoxicity
ISO10993-10IrritationNon-irritating
ISO10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer / Pass

9.0 Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device Vinyl Examination Glove is as safe, as effective, and performs as well as or better than the legally marketed predicated device K180861.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.