(547 days)
The HeartBuds Stethoscope is an electronic stethoscope that enables transmission of auscultation sound data, whereby a clinician at one location via the HeartBuds mobile application can listen to the auscultation sounds of a patient at a different location. The HeartBuds Stethoscope is intended to be used by professional users in a clinical environment or by non medical professional adult users in a non clinical environment. The device is for medical diagnostics purposes only. The device is not intended for self-diagnosis.
HeartBuds is an electronic stethoscope intended to auscultate heart and lung sounds. HeartBuds Stethoscope can be used on any patient undergoing a physical assessment. This stethoscope consists of a hand-held device with integral plastic sound diaphragm, that must be used together with a smartphone equipped with the HeartBuds application software. HeartBuds Stethoscope allows users to digitally record heart and lung internal auscultation and transmit the sound data to a medical professional in another location. The HeartBuds system includes two (2) main components: The HeartBuds Electronic Stethoscope and the HeartBuds mobile app.
The HeartBuds Electronic Stethoscope connects to mobile devices (Apple and Android) by means of an audio cable. The HeartBuds app with the integrated HeartBuds software provides users a platform to listen, display, and record sounds simultaneously using an external microphone. The users can share their recordings with their healthcare provider within the HeartBuds app for remote examination.
The provided text describes the HeartBuds Electronic Stethoscope, which is an electronic stethoscope enabling the transmission of auscultation sound data. It specifically states that "Clinical data was not required for this type of device" (page 11), indicating that no clinical study was conducted to prove the device meets acceptance criteria related to its diagnostic performance.
Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (Tyto Stethoscope K181612) through verification and validation activities against recognized standards. These activities are designed to ensure the device's safety and effectiveness in its stated use, rather than clinical diagnostic accuracy.
Therefore, the following information cannot be extracted from the provided text as it pertains to clinical studies that were not performed:
- A table of acceptance criteria and the reported device performance regarding diagnostic accuracy.
- Sample size used for the test set and data provenance (e.g., country of origin of the data, retrospective or prospective).
- Number of experts used to establish the ground truth for the test set and their qualifications.
- Adjudication method for the test set.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of how much human readers improve with AI vs. without AI assistance.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done (as this device transmits sound data for a clinician to interpret, it's inherently human-in-the-loop).
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for diagnostic performance.
- The sample size for the training set (as no AI algorithm for diagnostic inference is mentioned).
- How the ground truth for the training set was established.
However, the document does provide information on performance testing related to safety and usability. These are not diagnostic performance metrics, but rather engineering and operational validations.
Here's a summary of the acceptance criteria and performance as described for the device's overall design verification and validation:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Standard/Requirement | Reported Device Performance |
|---|---|---|
| Risk Management | ISO 14971: Application of risk management to medical devices | Potential risks identified, analyzed, and mitigated. All risk mitigations were satisfactorily verified and validated. |
| Electrical & Mechanical Safety | ANSI/AAMI 60601-1: Medical electrical equipment - General requirements for basic safety and essential performance | Compliant with the standard. Successful bench testing results included in submission. |
| Electromagnetic Disturbances | IEC 60601-1-2: Electromagnetic disturbances | Compliant with the standard. Successful bench testing results included in submission. |
| Usability Engineering | IEC 62366-1: Application of usability engineering to medical devices | Compliant with the standard. Human factors testing demonstrated usability in a simulated use environment for both healthcare professionals and lay users. |
| Biocompatibility | ISO 10993 (5th ed, 2018): Biological evaluation of medical devices - Part 1 Evaluation and testing within a risk management process | Compliant with the standard for patient contacting materials. |
| Software Testing | Company's Design Control process (21 CFR Part 820.30) | Verification and validation testing performed, including test cases for off-the-shelf software and cybersecurity features. Results were successful. |
| Cybersecurity | Not explicitly stated as a standard, but a specific mention | Cybersecurity controls incorporated, and supporting data provided in software information sections. |
2. Sample size used for the test set and the data provenance: Not applicable as no clinical diagnostic performance study was conducted.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no clinical diagnostic performance study was conducted.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable as no clinical diagnostic performance study was conducted.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device transmits raw sound data; it does not include AI for diagnostic assistance. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device's function is to transmit sound data for human interpretation; it does not perform standalone diagnostic analysis.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the verification and validation listed (e.g., electrical safety, usability), the "ground truth" would be compliance with the respective standards and design requirements, as demonstrated by successful test results. For instance, for usability, the ground truth was that users could effectively operate the device in a simulated environment. For biocompatibility, the ground truth was meeting ISO 10993 requirements.
8. The sample size for the training set: Not applicable as there is no mention of a machine learning algorithm or AI model requiring a training set for diagnostic classification.
9. How the ground truth for the training set was established: Not applicable for the same reason as above.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 14, 2023
Heartbuds LLC % Michelle Lott Principal and Founder leanRAQA 2081 Longden Circle Los Altos, California 94024
Re: K212906
Trade/Device Name: HeartBuds Electronic Stethoscope Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DOD Dated: March 7, 2023 Received: March 8, 2023
Dear Sheila Pickering:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert T. Kazmierski -S
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K212906
Device Name HeartBuds Stethoscope
Indications for Use (Describe)
The HeartBuds Stethoscope is an electronic stethoscope that enables transmission of auscultation sound data, whereby a clinician at one location via the HeartBuds mobile application can listen to the auscultation sounds of a patient at a different location. The HeartBuds Stethoscope is intended to be used by professional users in a clinical environment or by non medical professional adult users in a non clinical environment. The device is for medical diagnostics purposes only. The device is not intended for self-diagnosis.
Type of Use (Select one or both, as applicable)
| Deposition Use (Part 21 CFR 321 Subpart B) | |
|---|---|
| Over-The-Counter Use (21 CFR 331 Subpart C) |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
HeartBuds Electronic Stethoscope K212906 510(k) Summary Prepared August 11 2022
| Sponsor: | HeartBuds LLC711 London RoadWinter Park, Florida 32792 |
|---|---|
| Contact Person: | Seth Ellis |
| Telephone: | 407 906 2484 |
| Submission Date: | August 10, 2022 |
| Device Name: | HeartBuds Electronic Stethoscope |
| Common Name: | Electronic Stethoscope |
| Trade Name: | HeartBuds Model 1 |
| Classification: | |
| Regulatory Class: | II |
| Review Category: | 21 CFR 870.1875 (DQD) |
| Classification Panel: | Cardiovascular diagnostics |
A. Legally Marketed Predicate Devices
The predicate device is the Tyto Stethoscope (K181612).
B. Device Description:
HeartBuds is an electronic stethoscope intended to auscultate heart and lung sounds. HeartBuds Stethoscope can be used on any patient undergoing a physical assessment. This stethoscope consists of a hand-held device with integral plastic sound diaphragm, that must be used together with a smartphone equipped with the HeartBuds application software. HeartBuds Stethoscope allows users to digitally record heart and lung internal auscultation and transmit the sound data to a medical professional in another location. The HeartBuds system includes two (2) main components: The HeartBuds Electronic Stethoscope and the HeartBuds mobile app.
The HeartBuds Electronic Stethoscope connects to mobile devices (Apple and Android) by means of an audio cable. The HeartBuds app with the integrated HeartBuds software provides users a platform to listen, display, and record sounds simultaneously using an external microphone. The users can share their recordings with their healthcare provider within the HeartBuds app for remote examination
{4}------------------------------------------------
C. Intended Use / Indications for Use
The HeartBuds Stethoscope is an electronic stethoscope that enables transmission of auscultation sound data, whereby a clinician at one location via the HeartBuds mobile application can listen to the auscultation sounds of a patient at a different location. The HeartBuds Stethoscope is intended to be used by professional users in a clinical environment or by non medical professional adult users in a non clinical environment. The device is for medical diagnostics purposes only. The device is not intended for selfdiagnosis
D. Substantial Equivalence
The HeartBuds electronic stethoscope (HeartBuds) is substantially equivalent to the Tyto Stethoscope manufactured by Tyto Care Ltd. and cleared pursuant to 510(k) premarket notification K181612. Both are portable hand-held devices intended to enable transmission of auscultation sound data. A detailed comparison of the HeartBuds device and the predicate is presented in the Table below.
| # | Parameters | Subject Device | Predicate Device TytoStethoscope(K181612) | SubstantiallyEquivalent (SE) |
|---|---|---|---|---|
| 1 | Trade Name | HeartBuds | Tyto Stethoscope | N/A |
| 2 | Manufacturer | HeartBuds LLC | Tyto Care Ltd. | N/A |
| 3 | Classification | Class II | Class II | Same |
| 4 | Regulation No. | 21 CFR 870.1875 | 21 CFR 870.1875 | Same |
| 5 | ProductCode(s) | DQD | DQD | Same |
| 6 | Indicationsfor use | The HeartBuds Stethoscope isan electronic stethoscope thatenables transmission ofauscultation sound data,whereby a clinician at onelocation via the HeartBudsmobile application can listento the auscultation sounds of apatient at a different location.The HeartBuds Stethoscope isintended to be used byprofessional users in a clinicalenvironment or by lay users ina nonclinical environment.The device is for medicaldiagnostics purposes only. Thedevice is not intended for self-diagnosis. | The Tyto Stethoscopeis an electronicstethoscope thatenables transmission ofauscultation sounddata, whereby aclinician at onelocation on an IPnetwork can listen tothe auscultation soundsof a patient on site orat a different locationon the IP network withthe signal carried on anIP connection betweenthe two locations. TheTyto Stethoscope isintended to be used byprofessional users in a | SE |
| clinical environment orby lay users in anonclinicalenvironment. Thedevice is for medicaldiagnostics purposesonly. The device is notintended for self-diagnosis. | ||||
| 7 | DeviceDescription | The HeartBuds Stethoscopewhen connected to theHeartBuds App, capturessounds by the HeartBudsStethoscope which can berecorded by the App andstored to the cloud for futureuse. The HeartBuds mobileapplication on the smartphonerecords those sounds as anaudio recording to be storedand if desired by the user to beshared with a healthcareprovider. The HeartBuds Apprequires a user to register andcreate a secure login in orderto provide a means to associatethe recordings to eachindividual user and to bestored in the cloud (Googleapplication Firebase). TheHeartBuds App provides thecapability to start and stoprecording at the user'sdiscretion. The HeartBudsApp also provides thecapability for the user tosample the sounds receivedfrom the HeartBudsStethoscope before starting arecording. This provides ameans for the user to eitherchange the position of theStethoscope to ensure therecording includes sounds ofthe heart or lung. The userstops the recording, theHeartBuds App names the fileto facilitate futureidentification (both the userand a specific anatomicallocation) and loads it to the | The Tyto Stethoscope(OTC) is designed foruse by professional aswell as lay users inclinical or non-clinicalenvironments. Itenables four types ofstethoscope exams:Heart, Lungs, HeartRate and Audio (Audiois for clinician only).The operation processof the TytoStethoscope uses four(4) primary functionalelements: (1) The TytoStethoscope(composed of aStethoscope Tip and aBase Unit - TytoDevice and supportedwith proprietarysoftware). (2) Amobile device (e.g., asmartphone, not part ofthe Tyto Stethoscopedevice, not supplied byTytoCare, on whichthe proprietaryTytoCare App isrunning), (3) The TytoServer platform(composed of serverhardware not part ofthe Tyto Stethoscopedevice, not supplied byTytoCare, on whichthe proprietary serversoftware is running).(4) A clinicianreceiving platformlocated in a clinical | SE |
| cloud for storage. The | environment (e.g., a | |||
| HeartBuds App provides the | PC at the clinic, not | |||
| user a list of all saved | part of the Tyto | |||
| recordings. From the list of | Stethoscope device, | |||
| recordings the user can select | not supplied by | |||
| one for replaying or to share | TytoCare, on which | |||
| via email. | the proprietary | |||
| Clinician App is | ||||
| running). | ||||
| Two operational flows | ||||
| are optional for using | ||||
| the Tyto Stethoscope: | ||||
| store-and forward flow | ||||
| and on-line exam flow. | ||||
| Both flows are | ||||
| essentially similar and | ||||
| share the same | ||||
| fundamental steps: | ||||
| performing one or | ||||
| more measurements | ||||
| using the Tyto | ||||
| Stethoscope, recording | ||||
| the data and sending to | ||||
| a clinician, review of | ||||
| the recorded | ||||
| measurements by the | ||||
| clinician, and user | ||||
| receiving a written | ||||
| summary from the | ||||
| clinician presenting | ||||
| his/her assessment | ||||
| and/or | ||||
| recommendations. | ||||
| While in the store and- | ||||
| forward flow the user | ||||
| can perform the | ||||
| measurements and | ||||
| send the recorded data | ||||
| to the clinician | ||||
| whenever convenient | ||||
| for him/her, an on-line | ||||
| flow may be executed | ||||
| only when also the | ||||
| clinician is available | ||||
| on-line | ||||
| 8 | Intended users | Health care professionals and | professional as well as | SE |
| adult lay users (i.e., users who | lay users in clinical or | |||
| are not health careprofessionals) | nonclinical | |||
| environments | ||||
| 9 | Intended UseEnvironments | clinical or non-clinicalenvironments | clinical or non-clinicalenvironments | Same |
| 10 | Types ofstethoscopeexams | Heart, Lungs, and Audio | Heart, Lungs, HeartRate and Audio | SE |
| 11 | Functionalelements | stethoscopeA mobile device (e.g., asmartphone, not part of theHeartBuds Stethoscope device.not supplied by HeartBuds, on which the proprietaryHeartBuds App is running), | stethoscopea mobile device (e.g., asmartphone, not part ofthe Tyto Stethoscopedevice, not supplied byTytoCare, on which theproprietary TytoCareApp is running),The Tyto Server platform(composed of serverhardware not part of theTyto Stethoscope device,not supplied byTytoCare, on which theproprietary serversoftware is running).A clinician receivingplatform located in aclinical environment(e.g., a PC at the clinic,not part of the TytoStethoscope device, notsupplied by TytoCare, onwhich the proprietaryClinician App isrunning). | SE |
| 12 | Hardware andsoftwareplatforms | Android and iOS | Android and iOS | Same |
| 13 | Dimensions | 2.0 x 1.7 x 1.375 inchesHeartBuds Stethoscope only | 3.35 x 2.87 x 1.85 inch(85 x 73 x 47mm) | SE |
| 14 | Weight | 1.44 ounces ( 41 grams)HeartBuds Stethoscope only | 0.33 lbs. (0.15kg) | SE |
| 15 | Display | No Display | 2.4" LCD touch screen | SE |
| 16 | Audio OutputPort | USB-C connection with audiojack adaptor provided by thephone manufacturer or only anaudio jack if the smart device. | Standard 3.5mmheadphone connector | SE |
| 17 | Audio outputmethod | Headphone or audio recording | Audio recording | SE |
| 18 | Power Output | No power output | Proprietary plug,5Vdc, 2.0A | SE |
| 19 | Frequencyrange | 50 – 3000 Hz | 20 - 3,500 Hz | SE |
| 20 | Audio Output | A recording playback on thesmartphone. | 3.5mm standardheadphone connector | SE |
| 21 | Battery | No battery provided | Li-ion, built-in,rechargeable | SE |
| 22 | Typical Batterylife | No battery provided. | Up to 400 cycles ofcharge/discharge | SE |
| 23 | Dimensions | 2.0 x 1.7 x 1.375 inchesHeartBuds Stethoscope only | 1.57 x 1.53 inch (40 x39 mm) | SE |
| 24 | Weight | 1.44 ounces (41 grams)HeartBuds Stethoscope only | 0.13 lbs. (0.06kg) | SE |
| 25 | Operationalflows/Methodof transmittingrecording tophysician | Cloud stored and forward bythe user by email. | store-and forward flowand on-line exam flow | SE |
| 26 | OperatingEnvironment | Temp range: 5°C - 40°CHumidity range: 15%-90%(non-condensing)Pressure: 700 hPa to 1060 hPa | 5° - 40°C (41° - 104°F)15 - 70% (non-condensing)700hPa to 1060hPa | SE |
| 27 | Communication | Wired | Wireless | SE |
| 28 | Connections | A wire connected to thesmartphone audio jack | Information NotAvailable in productlabeling | SE |
| 29 | Permits datatransfer ofstored digitalsignals | Yes | Yes | Same |
| 30 | Signal storage | Yes, virtually in the cloud | Yes | Same |
| 31 | Diaphragmdiameter | 1.45 inch diameter | Information NotAvailable in productlabeling | Same |
| 32 | User interface | Smartphone device display | Smartphone devicedisplay | Same |
| 33 | Amplification | Amplification by the smartphone volume control | Information NotAvailable in productlabeling | Same |
| 34 | Waterproof | No | No | Same |
| 35 | Patientcontactingmaterials | Compliant with ISO 10993 | Materials informationnot available;Compliant with ISO10993 | Same |
| 36 | ElectricalSafety | Compliant with ANSI/AMMI60601-1 and IEC 60601-1-2 | Compliant withANSI/AMMI 60601-1and 60601-1-2 | Same |
| 37 | Usability | Compliant with ISO 62366 | Compliant with ISO62366 | Same |
Substantial Equivalence Table
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Discussion of Similarities and Differences
Summary of Similar Technological Characteristics.
The HeartBuds electronic stethoscope and the predicate are both electronic stethoscopes intended for projecting the sounds associated with vibrations or sounds of the human body. Therefore, they have the same intended use and core technology,
Both the Tyto Stethoscope and the HeartBuds Stethoscope are intended to transmit the recorded auscultation sounds to a remote location where a clinician can listen to them. In addition, both these devices are indicated for use by adult lay users (i.e., adults who are not health care professionals).
As shown in the table above, the HeartBuds Stethoscope shares with its predicate device similar structural design and similar principles and mode of operation as follows:
- The indications for use is similar with the subject device having fewer capabilities for data 1. collection and processing.
-
- The core technology is the same, although the predicate utilizes wireless communications
- The classification code is the same 3.
-
- Both devices are indicated for OTC devices that can be used in the home as well as in a clinical environment.
-
- Both devices are compliant with relevant safety standards (see performance data in this submission for ISO 10993, IEC 60601-1, IEC 60601-1-2 and ISO 62366)
- Both devices use downloadable Apps for Android and iOS 6.
- Both the HeartBuds Stethoscope and the Tyto Stethoscope record auscultation 7.
-
- Both the HeartBuds Stethoscope and the Tyto Stethoscope store the recording on smart phone storage
-
- Both the HeartBuds Stethoscope and the Tyto Stethoscope allow the user to share the recording
-
- Both the HeartBuds Stethoscope and the Tyto Stethoscope provide a transfer of the recording to on-line applications (via email, cloud storage)
Summary of Different Technological Characteristics.
As shown in the table above, the HeartBuds Stethoscope has some differences from the predicate device. Where these differences exist, performance data is provided in this submission to demonstrate that the differences do not raise any new issues of safety or effectiveness.
- The HeartBuds device has more limited capabilities than the Tyto device as listed in the table 1. above. Because there are fewer capabilities there are fewer risks associated with the correct delivery of information to the health care professional who will review the device output.
- Information on the specific materials used for the predicate are not available but are ISO 10993 2. compliant. Therefore, although the materials are not the same, both are substantially equivalent in that they are both compliant with ISO 10993.
{10}------------------------------------------------
K212906
-
- The HeartBuds device is wired whereas the Tyto device is wireless so that wireless transmission protocols were not required as part of the performance data provided in this submission.
-
- The software application is different in the HeartBuds and the Tyto device each using proprietary software algorithms. Details of the predicate software design are not available. However, the subject device software has been formally verified and validated according to the company's
-
- Design Control process (see performance data provided in the software information sections)
-
- There is no information available regarding cybersecurity protections in the predicate device. However, cybersecurity controls have been incorporated in the subject device and supporting data provided in the software information sections of this submission.
Discussion of Similarities and Differences
Verification and validation testing were completed in accordance with the company's Design Control process in compliance with 21 CFR Part 820.30. Successful results for the following tests are included in the submission as performance data supporting substantial equivalence:
-
- Bench testing for electrical and mechanical safety in compliance with applicable standards
- Software testing, consisted of verification and validation testing, including test cases related to off 2. the shelf software, as well as cybersecurity features.
- Human factors testing to demonstrate usability in a simulated use environment when used by 3. health care professionals and lay users.
Based on the comparison of indication for use and technological characteristics, the subject device is substantially equivalent to the predicate device. Based on the performance data provided in the submission the differences do not introduce new issues related to safety and efficacy.
E. Performance Data
Every specification of the HeartBuds Electronic Stethoscope has been verified and validated as required by the risk analysis. All design verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met.
| ISO 14971 | Application of risk management to medical devices |
|---|---|
| ANSI/AAMI 60601-1 | Medical electrical equipment- General requirements for basic safety andessential performance |
| IEC 60601-1-2 | Electromagnetic disturbances |
| IEC 62366-1 | Application of usability engineering to medical devices |
| ISO 10993 | ISO 10993 5 ed, 2018, Biological evaluation of medical devices- Part 1Evaluation and testing within a risk management process |
The verification and validation testing included testing to the following applicable standards:
Verification and validation testing were completed in accordance with the company's Design Control process in compliance with 21 CFR Part 820.30. Successful results for the following tests were included in the submission as performance data supporting substantial equivalence:
{11}------------------------------------------------
-
- Bench testing for electrical and mechanical safety in compliance with the standards cited above.
- Software testing, consisted of verification and validation testing, including test cases 2. related to off the shelf software, as well as cybersecurity features.
-
- Human factors testing to demonstrate usability in a simulated use environment when used by health care professionals and lay users.
Clinical data was not required for this type of device.
F. Conclusion
Potential risks were identified according to the ISO 62366 Standards. The risks were analyzed with regard to risk/benefit category and mitigations were implemented and tested as part of the performance testing described above. All risk mitigations were satisfactorily verified and validated. Where there were technological differences from the predicate, these were shown not to result in any new issues of safety or efficacy according to the performance data submitted.
Therefore, the HeartBuds Electronic Stethoscope is substantially equivalent to the predicate device with regards to intended use and technological characteristics. Results of performance testing demonstrated that the device met the design requirements and as well as the user needs.
§ 870.1875 Stethoscope.
(a)
Manual stethoscope —(1)Identification. A manual stethoscope is a mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.(b)
Electronic stethoscope —(1)Identification. An electronic stethoscope is an electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.